Hanna Instruments HI 99191 User manual

Range (*) -2.00 to 16.00 pH
-5.0 to 105.0 °C / 23.0 to 221.0 °F
Resolution 0.01 pH
0.1 °C / 0.1 °F
Accuracy ±0.02 pH
(@25 °C/77 °F) ±0.5 °C up to 60 °C; ±1.0 °C outside
±1.0 °F up to 140 °F; ±2.0 °F outside
Temperature Compensation Automatic
pH Calibration Automatic, 1 or 2 point
choose between 2 sets of buffers
(standard: 4.01/7.01/10.01 or NIST: 4.01/6.86/9.18)
Probe (included) FC 215D
preamplified pH/temperature
probe with DIN connector and 1 m (3.3’) cable
Battery Type 3 x 1.5V AAA
Battery Life Approximately 1200 hours of continuous use
Auto-off After 8 minutes of non-use
Environment 0 to 50 °C (32 to 122 °F);
RH max. 100%
Dimensions 152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)
Weight 205 g (7.2 oz.)
(*) the temperature range is limited from 0 to 50 °C (32 to
122 °F) when using FC 215D sensor.
To clean the meter, use water only.
HI 99191 is guaranteed for two years against defects in
workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrodes
are guaranteed for a period of six months. This warranty is
limited to repair or replacement free of charge.
Damage due to
accidents, misuse, tampering or lack of prescribed maintenance
is not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the
model number, date of purchase, serial number and the
nature of the problem. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain
a Returned Goods Authorization number from the Technical Service
department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly
packaged for complete protection.
Remove the instrument from the packing material and examine
it carefully to make sure that no damage has occurred during
shipment. If noticeable damage is evident, notify your dealer.
Note:Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing together with the
supplied accessories.
HI 99191
is a portable, microprocessor-based pH and temperature
meter specifically designed for Drinking Water Analysis.
Main features include: waterproof and compact casing; large
dual-line display; low battery detection; automatic pH calibration
at one or two points within two memorized buffer sets
(standard or NIST); selectable temperature unit (°C or °F).
The pH electrode has a built-in temperature sensor for simultaneous
temperature compensated pH and temperature readings, and
also contains an integral pH sensor preamplifier to provide
measurements impervious to noise and electrical interferences.
FC 215D pH electrode with a built-in temperature sensor,
features a rugged, easy to clean, PVDF body and a special
sensitive glass membrane which allows a quick response time in
low conductivity samples as drinking water is.
Each meter is supplied with:
•FC 215D pH electrode with internal temperature sensor, for
Drinking Water Analysis, with DIN connector and 1 m (3.3’)
cable
•pH 4.01 & pH 7.01 buffer sachets
•HI 700661 electrode cleaning solution (2x20 mL)
•3 x 1.5V AAA batteries
•rugged carrying case
•instruction manual
Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance
notice.
FC 215D pH electrode with internal temperature sensor, for
Drinking Water Analysis, with DIN connector and 1 m
(3.3’) cable
HI 7004L pH 4.01 buffer solution, 500 mL
HI 7006L pH 6.86 buffer solution, 500 mL
HI 7007L pH 7.01 buffer solution, 500 mL
HI 7009L pH 9.18 buffer solution, 500 mL
HI 7010L pH 10.01 buffer solution, 500 mL
HI 70300L pH electrode Storage solution, 500 mL
HI 7061L pH electrode Cleaning solution, 500 mL
HI 70630L pH electrode Acid Cleaning solution 500 mL
HI 70631L pH electrode Alkaline Cleaning solution, 500 mL
HI 70632L pH electrode Cleaning and Disinfection solution, 500 mL
HI 7071 Refilling Solution
1. Automatic Temperature Compensation indicator
2. Battery percentage (visible at power up)
3. Secondary display
4. Low battery indicator
5. Stability indicator
6. pH calibration indicator
7. Selectable temperature unit
8. Primary display
9. Measuring unit for primary display
IST99191 06/13
WARRANTY
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
ACCESSORIES
LCD DESCRIPTION SPECIFICATIONS
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using this
instrument. This manual will provide you with the necessary
information for correct use of these instruments, as well as a
precise idea of their versatility. If you need additional technical
information, do not hesitate to e-mail us at
[email protected] or view our worldwide contact list at
www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright owner.
Instruction Manual
HI 99191
Portable Waterproof pH / Temperature
Meter for Drinking Water Analysis
If a single-point calibration with buffers pH 7.01 or 6.86 is
desired, after recognition of the buffer press any key to return
to the pH measurement mode. After the button is pressed, the
meter displays “7.01“(or “6.86“) - “OK 1“, and an audible
signal is produced.
Note:It is always recommended to carry out a two-point calibration
for better accuracy.
•For a
two-point calibration
, place the probe in pH 7.01 (or pH
6.86) buffer first. After the calibration point has been accepted,
the “pH 4.01 USE” message appears. Place the sensor into the
second calibration buffer. The message is held for 12 seconds,
unless a valid buffer is recognized. If no valid buffer is
recognized, then the ''---WRNG'' message is displayed. If a
valid buffer (pH 4.01, pH 10.01 or pH 9.18) is detected,
then the meter completes the calibration procedure. When
the buffer is accepted, the LCD displays the accepted value
with the “OK 2” message on the secondary display. The meter
then returns to the normal measuring mode.
Note:When the calibration procedure is completed, the
“Calibrated“ tag is turned on.
To exit calibration and to reset to the default values
•After entering the calibration mode and before the first point is
accepted, it is possible to quit the procedure and return to the
last calibration data by pressing the ON/OFF/MODE button.
The secondary LCD displays “---ESC“ for 1 second and the
meter returns to normal mode.
•To reset to the default values and clear a previous calibration,
press the SET/HOLD button after entering the calibration
mode and before the first point is accepted.
The secondary LCD displays “---CLR“ for 1 second, the meter
resets to the default calibration and the “Calibrated” tag on
the LCD disappears.
First time you use the instrument open the battery compartment
and put the batteries inside observing the polarity.
To connect the probe
With the meter turned off, connect the FC 215D probe to the DIN
socket on the bottom of the meter by aligning the pins and
pushing in the plug. Tighten the nut to ensure a good connection.
Remove the protective cap from the probe before taking any
measurements.
To turn the meter ON and check the battery status
Press the ON/OFF/MODE button until the display lights up. At
start-up, all the LCD segments are displayed for 1 second, then
the percent indication of the remaining battery life is displayed for
another second (E.g. “100 % BATT“). The meter then enters the
normal measuring mode.
Note:If the display needs to be checked, keep the ON button
pressed while turning the meter on. The meter will display
all segments as long as the button is pressed.
To freeze the display
While in measurement mode, press the SET/HOLD button, “HOLD“
appears on the secondary display and the reading will be frozen
on the LCD (E.g. “pH 5.73 HOLD“). Press any button to return to
normal mode.
To turn the meter OFF
While in normal measurement mode, press the ON/OFF/MODE
button. “OFF“ will appear on the secondary display. Release the
button.
The meter is provided with an acoustic signal active when a key is
pressed.
Note:When the meter detects the absence of a temperature probe
at its input, the message “Probe not connected“ appears
on LCD and the Automatic Temperature Compensation is
turned off along with the “ATC“ tag. In this condition, the
secondary LCD displays “---- °C “ (“---- °F“) blinking.
When the probe is reconnected, the “Probe not connected“
tag is turned off, the temperature is displayed on the
secondary display and the “ATC“ tag is turned on.
Setup mode allows the selection of the temperature unit and the
type of pH buffer set.
To enter the Setup mode, press and hold the ON/OFF/MODE
button until “CAL“ on the secondary display is replaced by “TEMP“
and the current temperature unit (E.g. “TEMP °C“). Then:
•
for °C/°F selection
, use the SET/HOLD button. After the
temperature unit has been selected, press ON/OFF/MODE to
enter the buffer set selection mode; press ON/OFF/MODE
twice to return to the normal measuring mode.
•
to change the type of calibration buffer set
, after setting the
temperature unit, the meter will show the current buffer set:
“pH 7.01 BUFF” (for standard buffer set: 4.01/7.01/10.01)
or “pH 6.86 BUFF” (for NIST buffer set: 4.01/6.86/9.18).
Change the set with the SET/HOLD button, then press ON/OFF/
MODE to return to normal mode.
The meter is supplied with batteries.
The meter displays the remaining battery percentage when turned
on. When the level is below 5%, the “
- +
“ symbol on the LCD
blinks to indicate a low battery condition. If the battery level is low
enough to cause erroneous readings, the Battery Error Prevention
System (BEPS) turns the meter
off. It is recommended to replace
the batteries as soon as the display will flash the battery symbol.
To replace the batteries, follow the next steps:
•Open the battery compartment cap (on the bottom of the
instrument).
•Remove old batteries.
•Replace new batteries observing the polarity on the rear of
the instrument.
•Close the battery compartment cap.
BATTERY REPLACEMENT
OPERATIONAL GUIDE pH MEASUREMENT & CALIBRATION METER SETUP
•Make sure the meter has been calibrated before use.
•If the probe is dry, soak it in HI 70300 storage solution for
30 minutes to reactivate it.
•Place the tip of the probe into the sample to be tested.
•Wait until the “Not Stable“ tag on the LCD is turned off.
•The LCD displays the pH value (automatically compensated for
temperature) on the primary LCD, while the secondary LCD
displays the temperature of the solution.
•If measurements are taken in different samples successively,
rinse the probe tip thoroughly to eliminate cross-contamination.
For cleaning procedure choose the suitable Hanna cleaning
solutions (see Accessories section).
After cleaning, rinse the probe tip with some deionized water
and some of the sample to be measured.
pH calibration
For better accuracy, frequent calibration of the pH sensor with
the meter is recommended. In addition, the meter must be
recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where high accuracy is required.
d) At least once a month.
•From normal mode, press and hold the ON/OFF/MODE button
until “OFF“ on the secondary display is replaced by “CAL“.
Release the button. Place the sensor into the first calibration buffer.
•The meter enters the calibration mode, displaying “pH 7.01
USE” (or “pH 6.86 USE” if the NIST buffer set was selected).
After 1 second the meter activates the automatic buffer recognition
feature. If a valid buffer is detected, then its value is displayed
on the primary display, and “REC“ appears on the secondary
LCD. If no valid buffer is detected, the meter keeps the “USE“
indication active for 12 seconds, and then replaces it with
“---WRNG“, indicating that the sample being measured is not
a valid buffer.
•For a
single-point calibration
with buffers pH 4.01, 9.18 or
10.01, the meter automatically accepts the calibration when
the reading is stable; the meter will show on the primary display
the accepted buffer, with the message “OK 1” on the secondary
display, and an audible signal is produced.

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 981404N User manual

Hanna Instruments

Hanna Instruments HI 981404N User manual

Hanna Instruments HI 22091 User manual

Hanna Instruments

Hanna Instruments HI 22091 User manual

Hanna Instruments HI 991003 User manual

Hanna Instruments

Hanna Instruments HI 991003 User manual

Hanna Instruments PRIMO 3 User manual

Hanna Instruments

Hanna Instruments PRIMO 3 User manual

Hanna Instruments HI 93711 User manual

Hanna Instruments

Hanna Instruments HI 93711 User manual

Hanna Instruments HI 758 User manual

Hanna Instruments

Hanna Instruments HI 758 User manual

Hanna Instruments C 99 series User manual

Hanna Instruments

Hanna Instruments C 99 series User manual

Hanna Instruments HI96704C User manual

Hanna Instruments

Hanna Instruments HI96704C User manual

Hanna Instruments HI 933300 User manual

Hanna Instruments

Hanna Instruments HI 933300 User manual

Hanna Instruments Edge HI 2020 User manual

Hanna Instruments

Hanna Instruments Edge HI 2020 User manual

Hanna Instruments HI 8564 User manual

Hanna Instruments

Hanna Instruments HI 8564 User manual

Hanna Instruments HI 96769C User manual

Hanna Instruments

Hanna Instruments HI 96769C User manual

Hanna Instruments HI 83414 User manual

Hanna Instruments

Hanna Instruments HI 83414 User manual

Hanna Instruments HI 99301 User manual

Hanna Instruments

Hanna Instruments HI 99301 User manual

Hanna Instruments pH 21 User manual

Hanna Instruments

Hanna Instruments pH 21 User manual

Hanna Instruments HI 8010 User manual

Hanna Instruments

Hanna Instruments HI 8010 User manual

Hanna Instruments HI 96729C User manual

Hanna Instruments

Hanna Instruments HI 96729C User manual

Hanna Instruments HI 93733 User manual

Hanna Instruments

Hanna Instruments HI 93733 User manual

Hanna Instruments HI 83206 User manual

Hanna Instruments

Hanna Instruments HI 83206 User manual

Hanna Instruments HI 504903 User manual

Hanna Instruments

Hanna Instruments HI 504903 User manual

Hanna Instruments HI 254 User manual

Hanna Instruments

Hanna Instruments HI 254 User manual

Hanna Instruments HI97714 User manual

Hanna Instruments

Hanna Instruments HI97714 User manual

Hanna Instruments HI 746 User manual

Hanna Instruments

Hanna Instruments HI 746 User manual

Hanna Instruments HI96715 User manual

Hanna Instruments

Hanna Instruments HI96715 User manual

Popular Measuring Instrument manuals by other brands

wtw Cond 3110 quick start guide

wtw

wtw Cond 3110 quick start guide

Benning MM 5-2 operating manual

Benning

Benning MM 5-2 operating manual

BIRD 7027 Series Operation manual

BIRD

BIRD 7027 Series Operation manual

Colibri SMARTHUB user manual

Colibri

Colibri SMARTHUB user manual

Ametek WDG-V user manual

Ametek

Ametek WDG-V user manual

Citec LOOP420 instruction manual

Citec

Citec LOOP420 instruction manual

Mezon DSTM500 user manual

Mezon

Mezon DSTM500 user manual

Flexim FLUXUS F704**-NN Series user manual

Flexim

Flexim FLUXUS F704**-NN Series user manual

pH ELECTRODE PE-03 Operation manual

pH ELECTRODE

pH ELECTRODE PE-03 Operation manual

Honeywell DVC100 (M) MK2 installation guide

Honeywell

Honeywell DVC100 (M) MK2 installation guide

Milwaukee MW101 user manual

Milwaukee

Milwaukee MW101 user manual

Decagon Devices KD2 Pro Operator's manual

Decagon Devices

Decagon Devices KD2 Pro Operator's manual

Doepke DCTR B-X Installation and operating instructions

Doepke

Doepke DCTR B-X Installation and operating instructions

Smart Storm Universal Smart Meter instruction manual

Smart Storm

Smart Storm Universal Smart Meter instruction manual

M-system AS4M instruction manual

M-system

M-system AS4M instruction manual

Rixen SR-496 instruction manual

Rixen

Rixen SR-496 instruction manual

Vetek MO.DTR.526.R6 operating manual

Vetek

Vetek MO.DTR.526.R6 operating manual

ThermoWorks ThermaQ 2 operating instructions

ThermoWorks

ThermoWorks ThermaQ 2 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.