Hanna Instruments Silica ISM 93705 User manual

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments edge DO User manual

Hanna Instruments

Hanna Instruments edge DO User manual

Hanna Instruments HI774 User manual

Hanna Instruments

Hanna Instruments HI774 User manual

Hanna Instruments HI 2300 User manual

Hanna Instruments

Hanna Instruments HI 2300 User manual

Hanna Instruments HI 8733 User manual

Hanna Instruments

Hanna Instruments HI 8733 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments pH 21 User manual

Hanna Instruments

Hanna Instruments pH 21 User manual

Hanna Instruments HI 96804 User manual

Hanna Instruments

Hanna Instruments HI 96804 User manual

Hanna Instruments HI 9812-0 User manual

Hanna Instruments

Hanna Instruments HI 9812-0 User manual

Hanna Instruments HI 84100 User manual

Hanna Instruments

Hanna Instruments HI 84100 User manual

Hanna Instruments HI 761 User manual

Hanna Instruments

Hanna Instruments HI 761 User manual

Hanna Instruments HI 7635 User manual

Hanna Instruments

Hanna Instruments HI 7635 User manual

Hanna Instruments HI83325 User manual

Hanna Instruments

Hanna Instruments HI83325 User manual

Hanna Instruments HI 9812-0 User manual

Hanna Instruments

Hanna Instruments HI 9812-0 User manual

Hanna Instruments HI 95762C User manual

Hanna Instruments

Hanna Instruments HI 95762C User manual

Hanna Instruments HI736 User manual

Hanna Instruments

Hanna Instruments HI736 User manual

Hanna Instruments HI 98127 User manual

Hanna Instruments

Hanna Instruments HI 98127 User manual

Hanna Instruments HI 221 User manual

Hanna Instruments

Hanna Instruments HI 221 User manual

Hanna Instruments HI 98128 User manual

Hanna Instruments

Hanna Instruments HI 98128 User manual

Hanna Instruments HI781 User manual

Hanna Instruments

Hanna Instruments HI781 User manual

Hanna Instruments HI 931400 User manual

Hanna Instruments

Hanna Instruments HI 931400 User manual

Hanna Instruments HI 99161 User manual

Hanna Instruments

Hanna Instruments HI 99161 User manual

Hanna Instruments HI 713 User manual

Hanna Instruments

Hanna Instruments HI 713 User manual

Hanna Instruments HI 4221 User manual

Hanna Instruments

Hanna Instruments HI 4221 User manual

Hanna Instruments HI 96761C User manual

Hanna Instruments

Hanna Instruments HI 96761C User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

IRIMO 984-6D-1 owner's manual

IRIMO

IRIMO 984-6D-1 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Instruction Manual
HI 93705
Silica ISM
www.hannainst.com ThisInstrumentisin
Compliancewith the CEDirectives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93705iswarranted fortwoyearsagainstdefectsinworkman-
shipandmaterialswhenusedforitsintendedpurposeandmaintained
accordingtoinstructions.
Thiswarrantyislimitedtorepairorreplacementfreeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenancearenotcovered.
Ifserviceisrequired,contactthedealerfromwhomyoupurchasedthe
instrument.Ifunderwarranty, reportthemodelnumber, dateofpur-
chase,serialnumberand thenatureofthe failure.Iftherepair isnot
coveredbythewarranty,youwillbenotifiedofthechargesincurred.If
theinstrumentis tobereturned toHannaInstruments, firstobtaina
ReturnedGoodsAuthorizationNumber fromtheCustomerService de-
partmentandthensenditwithshipmentcostsprepaid.Whenshipping
anyinstrument,makesureitisproperlypackagedforcompleteprotec-
tion.
Tovalidateyourwarranty,filloutandreturntheenclosedwarrantycard
within14daysfromthedateofpurchase.
Allrights arereserved. Reproductioninwhole orin partis prohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket,RhodeIsland, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93705 meter measures the silica (SiO2) content in
water and wastewater in the 0.00 to 2.00 mg/L (ppm)
range.
Themeterusesan exclusive positive-locking systemtoensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
TypicalEMC ±0.01 mg/L
Deviation
Light Source Light Emitting Diode @ 890 nm
Method Adaptation of the ASTM D859 method of
the heteropoly blue method. The reaction
between silica and reagents causes a blue
tint in the sample
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93705A-0 Molybdate 6 drops
HI 93705B-0 Citricacid 1 packet
HI 93705C-0 Aminoacid 1 packet
REAGENT SETS
HI93705-01 Reagents for 100 tests
HI93705-03 Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumreadableconcentration (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays “- - -”, it is
ready.
• Fill the cuvet with 10 mL of
unreacted sample up to the
mark.
• Add6dropsofHI93705A
molybdate reagent. Re-
place the cap and swirl
the solution.
• Wait for 4 minutes, add
the content of one packet
of HI 93705B citric acid
reagent and shake until it
iscompletelydissolved.
• Wait for 1 minute. This is the blank.
• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and “SIP” will appear on the display.
• Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
ON
OFF
12
3
6
9
1’
ZERO
READ
TIMED
12
3
6
9
1’
READ
DIRECT
• Remove the cuvet and add
the content of one packet of
HI 93705C amino acid re-
agent and shake until it has
dissolved.
• Reinsert the cuvet into the in-
strument.
• Press READ TIMED and the display will show the count-
down prior to the measurement. Alternatively, wait for
exactly 1 minute and press READ DIRECT. In both cases
“SIP” will appear during measurement.
• Theinstrument directly displaysconcentration in mg/Lof
silica (SiO2) on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Phosphate above 60 mg/L (causes a 2% reduction in
reading)
Phosphate above 75 mg/L (causes an 11% reduction in
reading)
Sulfide and high concentration of iron
Eliminate color and turbidity interferences by zeroing the
meter with the original water sample.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiringtheoperatortotake all necessarystepstocorrect
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93705-01 Reagents for 100 tests
HI93705-03 Reagents for 300 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93705R2 10/99
PRINTED IN PORTUGAL
10 mL
x 6