manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hanna Instruments
  6. •
  7. Test Equipment
  8. •
  9. Hanna Instruments HI 98204 User manual

Hanna Instruments HI 98204 User manual

Other Hanna Instruments Test Equipment manuals

Hanna Instruments HI 95707C User manual

Hanna Instruments

Hanna Instruments HI 95707C User manual

Hanna Instruments HI 38077 User manual

Hanna Instruments

Hanna Instruments HI 38077 User manual

Hanna Instruments Pool Line HI981204 User manual

Hanna Instruments

Hanna Instruments Pool Line HI981204 User manual

Hanna Instruments HI 98311 User manual

Hanna Instruments

Hanna Instruments HI 98311 User manual

Hanna Instruments HI 38039 User manual

Hanna Instruments

Hanna Instruments HI 38039 User manual

Hanna Instruments HALO2 User manual

Hanna Instruments

Hanna Instruments HALO2 User manual

Hanna Instruments Gro Line HALO 2 User manual

Hanna Instruments

Hanna Instruments Gro Line HALO 2 User manual

Hanna Instruments DiST 1 User manual

Hanna Instruments

Hanna Instruments DiST 1 User manual

Hanna Instruments HI98330 User manual

Hanna Instruments

Hanna Instruments HI98330 User manual

Hanna Instruments HI 98129 User manual

Hanna Instruments

Hanna Instruments HI 98129 User manual

Hanna Instruments Pool Line HI983044 User manual

Hanna Instruments

Hanna Instruments Pool Line HI983044 User manual

Hanna Instruments HALO 2 HI9810442 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810442 User manual

Hanna Instruments HI 93739 User manual

Hanna Instruments

Hanna Instruments HI 93739 User manual

Hanna Instruments HI9810342 User manual

Hanna Instruments

Hanna Instruments HI9810342 User manual

Hanna Instruments DiST 3 User manual

Hanna Instruments

Hanna Instruments DiST 3 User manual

Hanna Instruments HI 3831T User manual

Hanna Instruments

Hanna Instruments HI 3831T User manual

Hanna Instruments HI96107 User manual

Hanna Instruments

Hanna Instruments HI96107 User manual

Hanna Instruments HI 83749 User manual

Hanna Instruments

Hanna Instruments HI 83749 User manual

Hanna Instruments HI 95701C User manual

Hanna Instruments

Hanna Instruments HI 95701C User manual

Hanna Instruments HI 96724C User manual

Hanna Instruments

Hanna Instruments HI 96724C User manual

Hanna Instruments HI 38081 User manual

Hanna Instruments

Hanna Instruments HI 38081 User manual

Hanna Instruments Total Chlorine Checker User manual

Hanna Instruments

Hanna Instruments Total Chlorine Checker User manual

Hanna Instruments HI 99121 User manual

Hanna Instruments

Hanna Instruments HI 99121 User manual

Hanna Instruments HI 3830 User manual

Hanna Instruments

Hanna Instruments HI 3830 User manual

Popular Test Equipment manuals by other brands

Ambit P.F. Series instruction manual

Ambit

Ambit P.F. Series instruction manual

Megger BMM500 Series user guide

Megger

Megger BMM500 Series user guide

RJG Lynx Sensor Tester operating instructions

RJG

RJG Lynx Sensor Tester operating instructions

Mark-10 TT02 Series user guide

Mark-10

Mark-10 TT02 Series user guide

Martel 3001 Operator's manual

Martel

Martel 3001 Operator's manual

ElproSys DiagProg4 instruction manual

ElproSys

ElproSys DiagProg4 instruction manual

Redtech TRAILERteck T05 user manual

Redtech

Redtech TRAILERteck T05 user manual

Venmar AVS Constructo 1.0 HRV user guide

Venmar

Venmar AVS Constructo 1.0 HRV user guide

Test Instrument Solutions SafetyPAT operating manual

Test Instrument Solutions

Test Instrument Solutions SafetyPAT operating manual

Kistler 5495C Series instruction manual

Kistler

Kistler 5495C Series instruction manual

Waygate Technologies DM5E Basic quick start guide

Waygate Technologies

Waygate Technologies DM5E Basic quick start guide

StoneL DeviceNet CK464002A manual

StoneL

StoneL DeviceNet CK464002A manual

Seica RAPID 220 Site preparation guide

Seica

Seica RAPID 220 Site preparation guide

Kingfisher KI7400 Series Training manual

Kingfisher

Kingfisher KI7400 Series Training manual

Kurth Electronic CCTS-03 operating manual

Kurth Electronic

Kurth Electronic CCTS-03 operating manual

SMART KANAAD SBT XTREME 3G Series user manual

SMART

SMART KANAAD SBT XTREME 3G Series user manual

Agilent Technologies BERT Serial Getting started

Agilent Technologies

Agilent Technologies BERT Serial Getting started

Agilent Technologies N3280A user guide

Agilent Technologies

Agilent Technologies N3280A user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

pH/ORP/EC/Temperature in One Unit
HI 98204 Water Test
www.hannainst.com
ISTWATERTSTR2 07/05
ON / OFF
SWITCH RANGE SWITCH
LIQUID
CRYSTAL
DISPLAY
MAXIMUM
IMMERSION
LEVEL
BLACK
COMPARTMENT
FRONT
BATTERY
COVER
OFFSET & SLOPE
CALIBRATION
TRIMMERS
COVER
BACK
Water Test
ON / OFF
4 pH 7
RANGE
IMMERSION LEVEL
Water Test
ON / OFF
4 pH 7
RANGE
BATTERY
COVER
IMMERSION LEVEL
Recommendations for Users
Before using this product, make sure that it is entirely suitable for
the environment in which it is used. Operation of this instrument
in residential area could cause unacceptable interferences to
radio and TV equipments, requiring the operator to take all
necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharge. Avoid touching this glass bulb at all times.
For calibration it is recommended to use an antistatic
screwdriver.
Any variation introduced by the user to the supplied equipment
may degrade the instrument's EMC performance. To avoid
electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in
microwaveovens.
MAINTENANCE
• Separate the compartment from the tester.
• Remove deposits formed on the liquid junction
and pH glass bulb by soaking in HI 7061 cleaning
solution for 30 minutes.
• Soak the platinum wire in acetone for 30 minutes.
• Rinse glass bulb, platinum wire and liquid junc-
tion thoroughly with distilled water
• Re-assemble the compartment and the tester.
• Fill the compartment with tap water.
BATTERY REPLACEMENT
When the Water Test cannot be switched on or the
display fades, pull out the battery cover and replace
all three 1.5V batteries, while paying attention to their
polarity. Battery replacement must only take place in
a safe area and using the battery types specified in
this instruction manual.
ACCESSORIES
HI 7004M pH 4.01 buffer solution, 230 mL
HI 7004L pH 4.01 buffer solution, 500 mL
HI 7007M pH 7.01 buffer solution, 230 mL
HI 7007L pH 7.01 buffer solution, 500 mL
HI 7010M pH 10.01 buffer solution, 230 mL
HI 7010L pH 10.01 buffer solution, 500 mL
HI 7031M 1413 µS/cm conductivity solution, 230 mL
HI 7031L 1413 µS/cm Conductivity solution, 500 mL
HI 7061M General cleaning solution, 230 mL
HI 7061L General cleaning solution, 500 mL
HI 731326 Calibration screwdriver (20 pcs)
GENERAL DESCRIPTION
The Water Test is a "four-in-one" portable meter designed for quick
and reliable measurement of the four most important variables in
water analysis: pH, ORP, conductivity and temperature.
Water Test is engineered with great simplicity, that even non-techni-
cal personnel can use it: just fill the base with the sample to be
measured, press the RANGE key to select the mode and read the
measurement.
Water Test is perfect for use in the fields of water treatment analysis,
wastewater test, ecological studies, aquaculture and hydroponic ap-
plications.
SPECIFICATIONS
pH ORP Conductivity Temperature
Range
0.0 to 14.0 pH ±1000 mV 0 to 1999 µS/cm 0.0 to 60.0°C
Resolution
0.1 pH 1 mV 1 µS/cm 0.1°C
Accuracy
±0.2 pH ±5 mV ±2% F.S. ±1°C
(@20°C/68°F)
Typical EMC
±0.4 pH ±2mV ±2% F.S. ±1°C
Deviation
Calibration
manual factory manual factory
2 point calibrated 1 point calibrated
Battery Type/Life
3 x 1.5V / approx. 200 hours of continuous use
Environment
0 to 50°C (32 to 122°F); RH max 95%
Dimensions / Weight
190 x 85 x 85 mm (7.4 x 3.3 x 3.3") / 260 g (9.2 oz.)
Water Test
ON/ OFF
4 pH 7
RANGE
HI 7004
µS
pH
mV
Water Test
ON/ OFF
4 pH 7
RANGE
Water Test
ON / OFF
4 pH 7
RANGE
pH
Water Test
ON/ OFF
4 pH 7
RANGE
HI 7004
pH
°C
Water Test
ON/ OFF
4 pH 7
RANGE
HI 7007
Water Test
ON/ OFF
4 pH 7
RANGE
HI 7031
ON / OFF
RIGHT
SIDE
RANGE
µS
°C
ON / OFF
RANGE
RANGE
Water Test
ON / OFF
4 pH 7
RANGE
RANGE
OPERATION
• Do not be alarmed if white crystals appear around
the cap. This is normal with pH electrodes and
they dissolve when rinsed with water.
• Take a sample of the water to be tested and pour
into the black compartment of the Water Test.
If you are scooping up the water with the black
compartment, do not immerse above the indicated
maximum immersion level.
• Turn the tester on by pressing the ON/OFF
key on the front panel.
• The tester will enter the temperature mea-
surement mode, indicated by the "°C" sym-
bol on the display.
• Press the RANGE key once to enter the
conductivity (µS/cm) mode.
• Press the RANGE key again to display pH.
• Press the RANGE key once again to dis-
play ORP (mV).
• If the tester compartment has been left dry
for more than 2 days, wait for about 5 min-
utes before taking pH measurements.
• For more accurate measurements, make sure that Water Test is
calibrated for pH and conductivity before use.
• After measurements, empty the compartment, rinse it with tap water
and switch the tester off by pressing the ON/OFF key.
• When the tester is not in use, fill the compartment
with tap water.
For long term storage, keep the tester in its origi-
nal box.
NEVER USE DISTILLED OR DEIONIZED WATER
FOR STORAGE PURPOSES.
pH CALIBRATION
• Pour a small quantity of distilled water into the
black compartment, rinse by shaking the compart-
ment and then discard the content.
• Pour a small quantity of pH 7 (HI7007) buffer solu-
tion into the compartment and switch the tester on
by pressing the ON/OFF key.
• Press the RANGE key until the display enters the
pH measurement mode.
• Adjust the right trimmer located behind
the front cover until the display shows "7.0
pH".
• Discard the pH 7 solution and rinse the
compartment with distilled water.
• Pour a small quantity of pH 4 (HI7004) or
pH 10 (HI7010) buffer solution into the com-
partment.
• Adjust the left trimmer located behind the
front cover until the display shows "4.0 pH"
or "10.0 pH".
• Discard the buffer solution and rinse the
compartment with distilled water.
• The unit is now calibrated for pH and ready
for use.
CONDUCTIVITY CALIBRATION
• Rinse the black compartment with distilled
water, then fill it with HI7031 conductivity so-
lution to overflow and switch the tester on by
pressing the ON/OFF key. The temperature
of the solution will be displayed.
• Note the temperature of the solution and read
the corresponding conductivity from the table
on the solution bottle label (e.g. 1413 µS/cm
@25°C).
• Remove the screw on the right side of the casing to access the
conductivity calibration trimmer. Press the RANGE key to display
the conductivity measurement mode and adjust the trimmer
until the display shows the conductivity reading at the noted tem-
perature (e.g. 1278 µS/cm @20°C). Replace the screw after calibra-
tion.
°C °F HI7031
(µS/cm)
0 32 776
5 41 896
10 50 1020
15 59 1147
16 60.8 1173
17 62.6 1199
18 64.4 1225
19 66.2 1251
20 68 1278
21 69.8 1305
°C °F HI7031
(µS/cm)
22 71.6 1332
23 73.4 1359
24 75.2 1386
25 77 1413
26 78.8 1440
27 80.6 1467
28 82.4 1494
29 84.2 1521
30 86 1548
31 87.8 1575
ALWAYS USE FRESH BUFFERS
FOR CALIBRATION