Specications
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy
(@25 °C/77 °F) ±0.1
Calibration manual, one-point
Bat tery Type 1.5 V (3 pcs.)
Battery Life approximately 150 hours of use
Environment RH 95% max;
0 to 50°C (32 to 122°F);
Dimensions 150 x 30 x 21 mm (5.9 x 1.2 x 8“)
Weight 70 g (2.5 oz.)
Thank You
Thank you for choosing a Hanna Instruments
product. Please read this instruction manual
carefully before using the instrument.
For more information about Hanna and our
products, visit www.hannainst.com.
For technical support, contact your local Hanna
oce or email us at tech@hannainst.com
Find your local Hanna oce at
www.hannainst.com
Warranty
The meter is warranted for a period of six
months against defects in workmanship and
materials when used for their intended purpose
and maintained according to instructions. This
warranty is limited to repair or replacement free
of charge. Damage due to accidents, misuse,
tampering or lack of prescribed maintenance is
not covered. If service is required, contact your
local Hanna oce. If under warranty, report the
model number, date of purchase, serial number
and the nature of the problem. If the repair is
not covered by the warranty, you will be notied
of the charges incurred. If the instrument is
to be returned to Hanna Instruments, rst
obtain a Returned Goods Authorization (RGA)
number from the Technical Service department
and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is
properly packaged for complete protection.
Operational Guide
• Remove the protective cap.
Do not be alarmed if white
crystals appear around the
cap. This is normal with pH
electrodes and they dissolve
when rinsed with water.
• Turn the meter on sliding
the switch on the top.
• Immerse into solution
up to the max
immersion level.
• Stir gently
and wait until
the display
stabilizes.
• After use, rinse
the electrode
with water to minimize
contamination and
store it with a few
drops of storage (HI70300)
or pH 7 (HI 7007) solution
in the protective cap.
DO NOT USE DISTILLED OR
DEIONIZED WATER FOR
STORAGE PURPOSES.
• Large uctuations in readings could
be due to lack of calibration, dry
electrode or rundown batteries.
Calibration
• Immerse the tester up to the
maximum immersion level
in pH 7 buer (HI 7007)
• Allow the reading to stabilize
and using a small screwdriver
adjust the calibration
trimmer to read 7.0.
Battery Replacement
When the meter cannot be switched
on or the display fades, pull out the
battery compartment and replace
all three 1.5V batteries, while
paying attention to their polarity.
Batteries should only be replaced in
a safe area using the battery type
specied in this instruction manual.
pH
pH
MAX
IMMERSION
LEVEL
MIN
IMMERSION
LEVEL
Meter Overview
The pHep (pH electronic paper) has an accuracy of
±0.1 pH and is calibrated manually to one point.