To connect the probe
Withthemeterturnedoff,connecttheHI 1292DprobetotheDIN
socketonthetopofthemeterbyaligningthepinsandpushingin
theplug.Tightenthenuttoensureagoodconnection.Remove
the protective cap from the probe before taking any measure-
ments.
To turn the meter ON and check the battery status
Press the ON/OFF/MODE button until the display lights up.At
start-up,alltheLCDsegmentsaredisplayedfor1second,then
the percent indication of the remaining battery life is shown for
anothersecond (E.g. % 100BATT).Themeterthen enters the
normal measuring mode.
Note: If the display needs to be checked, keep the ON button
pressedwhileturningthemeteron.Themeterwilldisplayall
segments as long as the button is pressed.
To freeze the display
While in measurement mode, press the SET/HOLD button,
HOLD appears on the secondary display and the reading will
be frozen on the LCD (E.g. pH 5.73 HOLD). Press any button
to return to normal mode.
To turn the meter OFF
Whileinnormalmeasurementmode,presstheON/OFF/MODE
button.OFF will appear on the secondarydisplay.Releasethe
button.
Note:Themeterisprovidedwithanacousticsignalfeature,which
canbedisabledusingtheswitchlocatedinthebatterycom-
partment.
Note:Whenthemeterdetectstheabsenceofatemperatureprobe
atitsinput,theAutomaticTemperatureCompensationisturned
off,andthemeterusesadefaultvalueof25°C(77°F)forthe
temperaturemeasurementandcompensation.Inthiscondi-
tion,thesecondaryLCDshows25.0°C(77.0°F)blinking.
Whenaprobeisconnected,themeterautomaticallyreturns
totheATCmode,theATCtagisturnedon,andthetempera-
tureisshownonthesecondarydisplay.
• Makesurethemeterhasbeencalibratedbeforeuse.
• If the probe is dry, soak it in HI 70300 storage solution for
one hour to reactivate it.
• Placethetipoftheprobeintothesampletobetested,stirbriefly
andwaituntilthestabilitysymbolontheLCDisturnedoff.
• TheLCDshowsthepHvalue(automaticallycompensatedfor
temperature) on the primary LCD, while the secondary LCD
showsthetemperatureofthesolution.
• Ifmeasurementsaretakenindifferentsamplessuccessively,rinse
theprobetipthoroughlytoavoidcontaminations.Aftercleaning,
rinsetheprobetipwithsomeofthesampletobemeasured.
pH calibration
For better accuracy, a frequent electrode cleaning (see
also pag. 15) and meter calibration is recommended.
In addition, the instrument must be recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where high accuracy is required.
d) At least once a month.
• Fromnormalmode,press&holdON/OFF/MODEuntilOFFon
thesecondarydisplayisreplacedbyCAL.Releasethebutton.
• TheLCDentersthecalibrationmode,displaying“pH7.01USE”
(or“pH6.86USE”iftheNISTbuffersetwasselected).After1
secondthemeteractivatestheautomaticbufferrecognitionfea-
ture.Ifavalidbufferisdetected,thenitsvalueisshownonthe
primary display, and REC appears on the secondary LCD. If
novalidbufferisdetected,themeterkeepstheUSEindication
activefor12seconds,andthenreplacesitwithWRNG,indicat-
ingthatthesamplebeingmeasuredisnotavalidbuffer.
• For a single-point calibration with buffers pH 4.01, 9.18 or
10.01,themeterautomaticallyacceptsthecalibrationwhenthe
readingisstable;themeterwillshowontheprimarydisplaythe
accepted buffer, with the message “OK 1” on the secondary
display,andanaudiblesignalisproduced.After1secondthe
meterautomaticallyreturnstothenormalmeasuringmode.
Ifasingle-pointcalibrationwithbufferspH7.01or6.86isde-
sired,thenafterthecalibrationpointhasbeenacceptedpress
the ON/OFF/MODE button in order to return to the normal
measuringmode.Afterthebuttonispressed,themetershows
OPERATING THE pH METER pH MEASUREMENT & CALIBRATION