manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. HERON
  6. •
  7. Measuring Instrument
  8. •
  9. HERON dipper-T Manual

HERON dipper-T Manual

www.heroninstruments.com
Made in Canada
HERON INSTRUMENTS INC.
447 Moxley Road | Dundas, ON Canada L9H 5E2
1-800-331-2032 | [email protected]
Please visit our website www.heroninstruments.com
for more information on the complete Heron line.
WE ALSO CARRY:
• Data Loggers
• Interface Meters
• Conductivity Meters
• Temperature Meters
• Well Casing Indicators
• Well Depth Indicators
dipper-T Water Level
Meter with drawDown
Operating and Maintenance Instructions
IMPORTANT: ENSURE THAT THE PANEL RETAINING THUMB SCREWS (NUTS)ARE TIGHT BEFORE USE.
Figure 1
Battery Box
Keep Thumb Screws Tight
Keep Thumb Screws Tight
Sensitivity Knob for
cascading water
Please note: this is not an on/off
switch. The water is the switch.
Mode Indicator
Mode Select Button
Buzzer
Signal LED
Figure 2
To test entire system, Hold the center pin on
the probe against the stud on the back of
the axle; touch the probe body against the
screw on the frame. The buzzer will sound if
the system is okay. Make sure sensitivity
dial is turned fully clockwise.
Figure 3
Unique hanger and tape protector
supports the meter on the casing and
protects the tape from sharp edges on
the well casing.
Item Part Numbers
Detachable WLM probe 1403 Detach-WLMP
drawDown Electronic Panel 1408 1408 dD-EP
Thumbscrews Set (2) 3250 TScrewsx2
Operating and Maintenance Instructions inside
Warranty is conditional upon adherence to these guidelines.
General Care of the dipper-T
The dipper-T is a high quality, robust unit that will give many years of reliable
service if these recommendations are followed:
• Avoid sharp edged casing, use the tape guide on the unit.
• Take care to avoid the tape becoming entangled with other equipment in
boreholes or wells, use stilling pipes when possible.
• DO NOT use the dipper-T as a guide to backlling, bentonite sealing or
sand packing in wells. This type of material falls through the water column
at a much slower rate than the dipper-T probe and can result in a trapped
probe and tape.
• Neatly rewind and clean the tape after each use.
Refer to: Cleaning the dipper-T
Warranty is conditional upon adherence to these guidelines
Equipment Check
Before taking the unit into the eld, carry out these simple tests with the
sensitivity knob turned fully clockwise (see Figure 1 panel controls), and the
two panel retaining thumb screws (nuts) are tight.
The sensitivity knob is used to adjust the unit’s response to varying conductivities
to maintain a sharp distinctive signal. Turn the knob clockwise for low
conductivity (pure) water and anti clockwise for high conductivity (dissolved
minerals) water.
NOTE: There is no on/off switch on the instrument. If using in DRAWDOWN
mode, return the unit to STATIC mode to turn the meter off. The dipper-T
consumes no power in STATIC mode when not in water.
To check the condition of the battery and circuit, press the CHANGE MODE
button.
The unit should sound and the INDICATOR light will come on. If the unit does not
respond, replace the battery and try again.
To test the entire unit from the panel to the probe, short the probe across the
stand off screw and the pin on the axle as shown in (Figure 2), the unit should
sound as above.
Use in the Field
The dipper-T will operate in two modes, STATIC level and DRAWDOWN. In
the STATIC mode, the unit will be silent until it touches the ground water. In
DRAWDOWN mode, the unit will be sounding and the light will be on until it
touches water.
Reel the tape down the well carefully avoiding the edge of the casing. Hang the
dipper-T on the casing when possible and run the tape over the guide on the
frame leg to avoid cuts and nicks to the tape. Swivel the probe holder on the
frame to allow the tape free movement down the well. (See Figure 3)
When the unit sounds (in STATIC mode) carefully measure the depth to water
indicated on the tape from your reference point. Raise and lower the tape in
and out of the water to get a consistent result.
NOTE: The inverted triangle on the probe holder serves as a datum point
indicating “top of casing”. In wells that have cascading water that may
give false readings, reduce the sensitivity by turning the sensitivity knob
anti-clockwise.
The dipper-T probe is depth rated to the full length of the tape.
DO NOT use the dipper-T to measure the progress of back lling wells. The tape
and probe may become stuck in the backll material.
DO NOT allow the tape to “freewheel” down the well. It may become caught in
other equipment in the well.
When rewinding the tape remove as much water and debris as possible from
the tape and the probe.
Cleaning the dipper-T
Always clean the meters after use in the eld to maintain optimal performance
and extend the life of the unit.
The dipper-T may be cleaned with any mild household dishwashing detergent
and rinsed with water.
DO NOT use abrasives, partially halogenated hydrocarbons or ketones to
clean the reel. If the electronic panel is removed rst, the reel and tape can be
washed with a power washer.
Remove the panel retaining thumb screws (nuts) to release the panel. Take care
not to lose the thumb screws as the unit will not work without them.
For longer units with aluminium reel discs, the panel is connected to the tapes
with male and female push connectors.
Troubleshooting the dipper-T
Q. No sound or light when the unit is tested.
A. Refer to Equipment Check and follow procedures. Change the battery.
Switch from STATIC to DRAWDOWN and back. If the unit is in
DRAWDOWN mode and it sounds, put the probe in water and the sound
should stop.
Q. Equipment checked, panel okay but probe not working.
A. Make sure both panel retaining thumb screws (nuts) are tight.
Q. Every thing okay, but the probe is still not working.
A. Carry out full continuity test, as shown in (Figure 2).
Q. Still not working in STATIC mode.
A. Adjust the sensitivity setting. Check the probe connection to the tape.
Carefully remove the probe, wipe the male connector and replace the
probe, tighten fully (do not lose the O-ring). If the unit still does not work
check all the connections inside the hub. (Inside the hub polarity is not an
issue as the current is AC.)
Q. The instrument continues to sound when not in water.
A. Press the MODE button. Dry the probe with a clean cloth if the water is
saline. Remove the probe and dry the male connector. If at this stage the
unit stops sounding, replace the probe. If the instrument starts sounding
again there is a short in the probe.
Contact Heron Instruments or your supplier if you cannot isolate the problem.
Warranty (5 years, probe 1 year)
Heron Instruments Inc. warrants to repair or replace any such defective
equipment or part (determined to our satisfaction to have a defect in
workmanship or original material) upon receipt and inspection of such defective
equipment to Heron Instruments Inc. with all shipping pre-paid by the user.
In no event shall Heron be liable for any direct, indirect or consequential
damages, abuse, acts of third parties (rental equipment), environmental
conditions or other expenses which may arise in connection with such
defective equipment.
This warranty shall not apply to damage of equipment caused by improper
installation, usage, storage, alteration or inadequate care.
Heron Warranty coverage does not extend to the following:
• Tape, bag or batteries used with the product
• Products used as rental equipment
• Products contaminated by materials which are known to be hazardous
and; as such, have rendered the unit unserviceable
• Parts failure due to neglect in cleaning or servicing
• Failure of parts caused by misuse
For service information:
• visit www.heroninstruments.com under the CONTACT heading
• email service@heroninstruments.com
• call 1-800-331-2032 or 905-628-4999
dipper-T Water Level Meter Instructions

Other HERON Measuring Instrument manuals

HERON LongLiner 4000 Series User manual

HERON

HERON LongLiner 4000 Series User manual

HERON dipper-Temp Manual

HERON

HERON dipper-Temp Manual

HERON H.0IL Manual

HERON

HERON H.0IL Manual

HERON dipper-T Installation guide

HERON

HERON dipper-T Installation guide

HERON dipper-T Installation and operation manual

HERON

HERON dipper-T Installation and operation manual

HERON dipperLog SMART User manual

HERON

HERON dipperLog SMART User manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments HI715 instruction manual

Hanna Instruments

Hanna Instruments HI715 instruction manual

IKA C 1 operating instructions

IKA

IKA C 1 operating instructions

IMC CAEMAX DX manual

IMC

IMC CAEMAX DX manual

Solinst 615 ML instructions

Solinst

Solinst 615 ML instructions

Extech Instruments RH300 user manual

Extech Instruments

Extech Instruments RH300 user manual

Volth S-32 Operation manual

Volth

Volth S-32 Operation manual

Master Meter SONATA ALLEGRO installation guide

Master Meter

Master Meter SONATA ALLEGRO installation guide

BRONKHORST ES-FLOW 1 C Series instruction manual

BRONKHORST

BRONKHORST ES-FLOW 1 C Series instruction manual

enphase RGM Hardware installation and operation manual

enphase

enphase RGM Hardware installation and operation manual

ELEKTRO PARTNER TwinProbe user manual

ELEKTRO PARTNER

ELEKTRO PARTNER TwinProbe user manual

Nedo COMMANDER2 HVA2N operating instructions

Nedo

Nedo COMMANDER2 HVA2N operating instructions

KERN DAB-SH-e-1610 Service manual

KERN

KERN DAB-SH-e-1610 Service manual

Atlas Copco ACTA 4000 user guide

Atlas Copco

Atlas Copco ACTA 4000 user guide

DH Instruments PG7307 Operation and maintenance manual

DH Instruments

DH Instruments PG7307 Operation and maintenance manual

Endress+Hauser Memosens CPS16D operating instructions

Endress+Hauser

Endress+Hauser Memosens CPS16D operating instructions

WIKA CPH6300 operating instructions

WIKA

WIKA CPH6300 operating instructions

PCB Piezotronics LARSON DAVIS 730 manual

PCB Piezotronics

PCB Piezotronics LARSON DAVIS 730 manual

AST EL50-H+ user manual

AST

AST EL50-H+ user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.