Hioki 8730-10 User manual

WAVE COMPARATOR
Instruction Manual
8731-10
October 2008 Revised edition 6 8730A981-06 08-10H
8730-10


Contents
Introduction i
Inspection i
Safety Notes ii
Notes on Use iv
Chapter Summary vi
Chapter 1 Instrument Overview 1
1.1 Major Features 2
1.2 Identification of Controls and Indicators 4
1.3 Dimension Diagrams 7
Chapter 2 Installation and Preparation 11
2.1 Installation of the Unit 11
2.1.1 Installation Orientation 11
2.1.2 Ambient Conditions 12
2.1.3 Installation Procedure 13
2.2 Power Supply and Ground Connections 16
2.2.1 Connecting the AC Power Supply 16
2.2.2 Functional Grounding of the 8730-10, 8731-10 17
2.3 Power ON/OFF 18
2.4 Cable Connection 19
2.5 Notes on Measurement 20
2.6 Maximum Input Voltage 21
2.7 Using a Voltage Transformer 21
2.8 Startup Inspection and Shutdown Inspection 22
2.9 Contrast Adjustment 22
Chapter 3 Basic Operation and Settings 23
3.1 Basic Operation 24
3.1.1 Basic Key Operation 24
3.1.2 Common Operations 26
3.2 Basic Measurement and Setting 30
3.2.1 Operation Sequence 30
3.2.2 Measurement Examples 31
3.3 Guide to Using This Manual 40

Chapter 4 Settings on the Measurement Screen 41
4.1 Overview 45
4.1.1 Measurement Screen Overview 45
4.1.2 Flow of Settings on the Measurement Screen and
Measurement 46
4.2 Details of Measurement Setting 47
4.2.1 Time Axis Range 47
4.2.2 Recording Length 48
4.2.3 Time Axis Magnification Rate 50
4.2.4 Display Format 51
4.2.5 Averaging 52
4.2.6 Overlay Function 54
4.2.7 Information (Left) 55
4.2.8 Information (Upper) 57
4.2.9 Auto Save 58
4.3 Details of Input Setting 62
4.3.1 Range 62
4.3.2 Coupling 63
4.3.3 Filter 63
4.3.4 Voltage Axis Magnification 64
4.3.5 Zero Position 65
4.3.6 Color of Wave Display 67
4.3.7 Graph of Waveform Display 67
4.3.8 Scaling Function 68
4.3.9 Variable Function 73
4.3.10 Entering Channel Comments 75
4.4 Trigger Functions 76
4.4.1 Overview of the Trigger Functions 76
4.4.2 Trigger Mode 77
4.4.3 Pre-trigger 78
4.4.4 Trigger Source AND/OR Linking 79
4.4.5 Channel Trigger 80
4.4.6 External Trigger 87
4.4.7 Timer Trigger 87
4.5 Details of List 90
4.6 Auto save based on comparison results 91

Chapter 5 Comparison and Calculation 93
5.1 Overview 96
5.2 Comparison Details 97
5.2.1 Common Stop Condition 97
5.2.2 Screen Updating 99
5.2.3 Switching Test Modes 99
5.2.4 Registering Test Modes 100
5.2.5 Clearing Test Modes 101
5.2.6 Setting the Beep sound for PASS decisions 101
5.2.7 Setting the Beep sound for FAIL decisions 102
5.2.8 Re-compare 102
5.3 Creating Comparison Area 103
5.3.1 Area Comparison Sequence 103
5.3.2 Creating Area 105
5.3.3 Editor Command Details 107
5.4 Area Comparison Details 122
5.4.1 Area Comparison Setting 122
5.4.2 Area Comparison Conditions 123
5.4.3 Stop Conditions of Area Comparison 125
5.5 Value Comparison Details 126
5.5.1 Flow of Setting Value Comparison and
Measurement 126
5.5.2 Value Comparison 127
5.5.3 Value Calculations 127
5.5.4 Setting Calculation Range 137
5.5.5 Stop Conditions of the Value Comparison 138
5.5.6 Display of value calculation results 139
5.6 Waveform Calculation 140
5.6.1 Setting the Waveform Calculation 140
5.6.2 Setting Type of Calculation 141
5.6.3 Re-calculation 142
5.7 List 143
Chapter 6 Saving and Loading Data 145
6.1 Overview 146
6.2 PC Card 146
6.3 Saveable Data 148

6.4 File Screen 149
6.4.1 Loading the Data 150
6.4.2 Saving the Setting conditions (SET) 151
6.4.3 Saving the Waveform Data (MEM) 152
6.4.4 Saving the Text Data (CSV) 153
6.4.5 Saving the Setting & Area (S_A) 154
6.4.6 Saving the Area & Modes (ROM) 155
6.4.7 Automatic Loading of Setting Data
(START UP) 156
6.4.8 Creating the Directory (DIR) 157
6.4.9 Delete files 158
6.4.10 Deleting all files 158
6.4.11 File Details 159
Chapter 7 Adjustment and Setting of the Unit 163
7.1 Overview 165
7.2 Setting / Check 166
7.2.1 Zero Adjustment 166
7.2.2 Clock setting 167
7.2.3 Language 168
7.2.4 Beep Sound 168
7.2.5 Start Backup 169
7.2.6 Contrast 170
7.2.7 Back Light Saver 171
7.2.8 System Reset 172
7.2.9 ROM/RAM Check 172
7.2.10 Key Check 173
7.2.11 Display Check 173
7.3 LAN 174
7.3.1 Local host 174
7.3.2 Gateway 176
7.3.3 DHCP 177
7.3.4 DNS 178
7.3.5 Server (9333) 179
7.3.6 TCP/IP Port 180
7.3.7 Delimiter 180
7.3.8 LAN Restart 181

7.4 RS-232C 182
7.4.1 Transmission Speed 182
7.4.2 Data Bit 182
7.4.3 Parity 183
7.4.4 Stop Bit 183
7.4.5 Flow Control 184
7.4.6 Delimiter 184
7.5 External Control Terminals 185
7.5.1 EXT_OUT1 to EXT_OUT5 186
7.5.2 EXT_IN 188
7.5.3 Range-over Detection 189
Chapter 8 External Control Terminals 191
8.1 Overview 191
8.2 Terminals 192
8.3 Power Supply Terminal 193
8.3.1 External Power Supply 193
8.3.2 Internal Power Supply 194
8.4 Input Terminals 195
8.4.1 Start/Stop Measurement 196
8.4.2 External Trigger 197
8.4.3 External Sampling 198
8.4.4 Selecting a Test Mode 199
8.4.5 Multi-function terminal (Input) 201
8.4.6 Holding Comparison Results 202
8.4.7 Canceling Output of Comparison Result 203
8.5 Output Terminals 204
8.5.1 Output of Value Comparison Results 205
8.5.2 Output of Area Comparison Results 208
8.5.3 Comparison Result Being Output 212
8.5.4 Trigger Output 213
8.5.5 Error Signals Output 213
8.5.6 Processing 214
8.5.7 Measuring 215
8.5.8 Multi-Function Terminals (Output) 216
8.6 Example of Control via the External Control
Terminals 217
8.7 External Control Connection Example 218
Chapter 9 Interface 219
9.1 LAN Interface 220
9.2 RS-232C Interface 225

Chapter 10 A/B Cursors and Scrolling the
Waveform 227
10.1 A/B Cursors 228
10.1.1 Vertical Cursor 228
10.1.2 Horizontal Cursor 229
10.1.3 Trace Cursor 229
10.1.4 Using the A/B Cursors 230
10.2 Scrolling the Waveform 232
Chapter 11 Specifications 233
11.1 General Specifications 233
11.1.1 Basic Specifications 233
11.1.2 Display 234
11.1.3 External Data Storage 234
11.1.4 Operating Unit 234
11.1.5 External Control Terminals 235
11.1.6 External Interface 235
11.2 Input Unit Specifications 236
11.3 Trigger Unit 237
11.4 Memory Recorder Function 238
11.5 Comparison Function 239
11.5.1 Waveform Area Comparison 239
11.5.2 Value Comparison 239
11.6 Auxiliary Functions 240
11.7 Others 240
Chapter 12 Maintenance and Service 241
12.1 Maintenance and Inspection 241
12.2 Removing the Battery Before Discarding
the Unit 242
12.3 Troubleshooting 244
Chapter 13 Appendix 245
13.1 Error Messages 245
13.2 Reference 247
13.2.1 Sampling 247
13.2.2 Aliasing 247
13.2.3 Measurement Limit Frequency 248
13.2.4 Averaging Equations 248
13.2.5 SI Prefix 249

13.3 Size of a File 250
13.3.1 Screen Data (BMP Format) 250
13.3.2 Waveform Data (CSV) 250
13.3.3 Waveform Data (MEM) 250
13.3.4 Other Data 251
13.4 Glossary 252
13.5 Waveform Viewer (Wv) 253
13.5.1 Starting the Waveform Viewer 254
13.5.2 Waveform Viewer Menus 255
13.5.3 Using the Waveform Viewer 257
13.5.4 Conversion to CSV Format 260
13.5.5 Batch Conversion 261
Index INDEX 1


i
_____________________________________________________________________________________________
Inspection
______________________________________________________________________________________________
External Control Terminals label
Jumper connector
Introduction
Inspection
Thank you for purchasing the HIOKI "8730-10, 8731-10 WAVE
COMPARATOR". To obtain maximum performance from the instrument,
please read this manual first, and keep it handy for future reference.
When you receive the instrument, inspect it carefully to ensure that no
damage occurred during shipping. In particular, check the accessories, panel
switches, and connectors. If damage is evident, or if it fails to operate
according to the specifications, contact your dealer or Hioki representative.
Accessories
Power cord 1
Jumper connector 1
External control terminals label 1
Instruction manual 1
Communications command manual (FD) 1
Application disk (CD-R) 1
Options
9626 PC CARD 32M
9627 PC CARD 64M
9726 PC CARD 128M
9727 PC CARD 256M
9728 PC CARD 512M
9729 PC CARD 1G
9637 RS-232C CABLE (9pin-9pin/1.8m)
9638 RS-232C CABLE (9pin-25pin/1.8m)
9642 LAN CABLE
9333 LAN COMMUNICATOR

ii
_____________________________________________________________________________________________
Safety Notes
______________________________________________________________________________________________
WARNING
This instrument is designed to conform to IEC 61010 Safety Standards,
and has been thoroughly tested for safety prior to shipment. However,
mishandling during use could result in injury or death, as well as
damage to the instrument. Be certain that you understand the
instructions and precautions in the manual before use. We disclaim any
responsibility for accidents or injuries not resulting directly from
instrument defects.
DANGER
Indicates that incorrect operation presents an extreme hazard that
could result in serious injury or death to the user.
WARNING
Indicates that incorrect operation presents a significant hazard that
could result in serious injury or death to the user.
CAUTION Indicates that incorrect operation presents a possibility of injury to the
user or damage to the instrument.
NOTE Advisory items related to performance or correct operation of the
instrument.
Safety Notes
The symbol printed on the instrument indicates that the user should
refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
In the manual, the symbol indicates particularly important information
that the user should read before using the instrument.
Indicates a grounding terminal.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Safety symbols (Used on the 8730-10, 8731-10 and other units)
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using the instrument, be sure to carefully read the following safety notes.
Symbols Used in This Manual
The following symbols in this manual indicate the relative importance of
cautions and warnings.

iii
_____________________________________________________________________________________________
Safety Notes
______________________________________________________________________________________________
CAT I Secondary electrical circuits connected to an
AC electrical outlet through a transformer or
similar device.
CAT II Primary electrical circuits in equipment
connected to an AC electrical outlet by a power
cord (portable tools, household appliances, etc.)
CAT III Primary electrical circuits of heavy equipment
(fixed installations) connected directly to the
distribution panel, and feeders from the
distribution panel to outlets.
CAT IV The circuit from the service drop to the service
entrance, and to the power meter and primary
overcurrent protection device (distribution
panel).
Measurement categories (Overvoltage categories)
This instrument complies with CAT I safety requirements.
To ensure safe operation of measurement instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as CAT I to
CAT IV, and called measurement categories. These are defined as follows.
Higher-numbered categories correspond to electrical environments with
greater momentary energy. So a measurement device designed for CAT III
environments can endure greater momentary energy than a device designed
for CAT II.
Using a measurement instrument in an environment designated with a
higher-numbered category than that for which the instrument is rated could
result in a severe accident, and must be carefully avoided.
Never use a CAT I measuring instrument in CAT II, III, or IV environments.
The measurement categories comply with the Overvoltage Categories of the
IEC60664 Standards.

iv
_____________________________________________________________________________________________
Notes on Use
______________________________________________________________________________________________
DANGER
Probe Connection, Measurement Voltage input
Maximum input voltage ratings for the CH1, CH2 and the external control
terminals are shown below. To avoid the risk of electric shock and
damage to the unit, take care not to exceed these ratings.
The maximum rated voltage to earth of the CH1, CH2 and external
control terminals (voltage between each channel and the 8730-10,
8731-10, between channels, and between an external control terminal
and the 8730-10, 8731-10) is shown below. To avoid the risk of electric
shock and damage to the unit, take care that voltage between
channels and between a channel and ground does not exceed these
ratings.
The maximum rated voltage to earth rating applies also if an input
attenuator or similar is used. Ensure that voltage does not exceed
these ratings.
Power cord Connection
Before using the instrument, make sure that the insulation on the
power cord is undamaged and that no bare conductors are
improperly exposed. Using the instrument under such conditions
could result in electrocution. Replace the power cord specified by
Hioki .
Input/Output Terminals Maximum Input Voltage Maximum raged voltage
to earth
CH1, CH2 Input Terminals 30 V rms or 60 V DC 30 V rms or 60 V DC
External Control Terminals 30 V DC 30 V rms or 60 V DC
WARNING
Power Supply Connections
Before turning the instrument on, make sure the source voltage
matches that indicated on the instrument's power connector.
Connection to an improper supply voltage may damage the
instrument and present an electrical hazard.
Grounding the Unit
To avoid electric shock and ensure safe operation, connect the power
cable to a grounded (3-contact) outlet. (See 2.2)
Before Powering on
Check that the power supply is correct for the rating of the unit.
(The AC fuse is integrated in the unit.)
Notes on Use
Follow these precautions to ensure safe operation and to obtain the full
benefits of the various functions.
Preliminary Checks
Before using the instrument the first time, verify that it operates normally to
ensure that the no damage occurred during storage or shipping. If you find
any damage, contact your dealer or Hioki representative.

v
_____________________________________________________________________________________________
Notes on Use
______________________________________________________________________________________________
CAUTION Installation Environment
This instrument should be installed and operated indoors only, between 5
and 40 and 30 to 85% RH.
Do not store or use the instrument where it could be exposed to direct
sunlight, high temperature or humidity, or condensation. Under such
conditions, the instrument may be damaged and insulation may deteriorate
so that it no longer meets specifications.
Carrying
Grasp the handle when carrying the 8730-10, 8731-10 to avoid dropping it
accidentally. (See 1.2 )
Shipping
To avoid damage to the instrument, Remove the PC card from the unit.
Use the original packing materials when reshipping the instrument, if
possible.
Handling the CD-R
Always hold the disc by the edges, so as not to make fingerprints on
the disc or scratch the printing.
Never touch the recorded side of the disc. Do not place the disc
directly on anything hard.
Do not wet the disc with volatile alcohol or water, as there is a
possibility of the label printing disappearing.
To write on the disc label surface, use a spirit-based felt pen. Do not
use a ball-point pen or hard-tipped pen, because there is a danger of
scratching the surface and corrupting the data. Do not use adhesive
labels.
Do not expose the disc directly to the sun's rays, or keep it in
conditions of high temperature or humidity, as there is a danger of
warping, with consequent loss of data.
To remove dirt, dust, or fingerprints from the disc, wipe with a dry
cloth, or use a CD cleaner. Always wipe radially from the inside to
the outside, and do no wipe with circular movements. Never use
abrasives or solvent cleaners.
Hioki shall not be held liable for any problems with a computer
system that arises from the use of this CD-R, or for any problem
related to the purchase of a Hioki product.
Others
In the event of problems with operation, first refer to Section 12.3.
Carefully read and observe all precautions in this manual.

vi
_____________________________________________________________________________________________
Chapter Summary
______________________________________________________________________________________________
Chapter Summary
Chapter 1 Instrument Overview
Contains an overview of the unit and its features.
Chapter 2 Installation and Preparation
Explains how to set the unit up for measurement.
Chapter 3 Basic Operation and Settings
Explains how to operate the keys and the Jog/Shuttle control for carrying out
basic measurement functions.
Chapter 4 Settings on the Measurement Screen
Explains measurement settings.
Chapter 5 Comparison and Calculation
Explains the settings for comparison and calculation. Also explains how to
create a comparison area.
Chapter 6 Saving and Loading Data
Explains how to store, recall, and delete measurement data and measurement
settings.
Chapter 7 Adjustment and Setting of the Unit
Explains how to adjust and diagnose the 8730-10, 8731-10 and set up the
interface and external control terminals.
Chapter 8 External Control Terminals
Gives specifications and usage details of the external control terminals.
Chapter 9 Interface
Explains about LAN and RS-232C interface.
Chapter 10 A/B Cursors and Scrolling the Waveform
Explains how to perform waveform scrolling, how to use the A/B cursors
and so.
Chapter 11 Specifications
Contains general specifications and detailed function specifications.
Chapter 12 Maintenance and Service
Describes maintenance procedures.
Chapter 13 Appendix

1
_____________________________________________________________________________________________
______________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
Chapter 1
Instrument Overview

2
_____________________________________________________________________________________________
1.1 Major Features
______________________________________________________________________________________________
1.1 Major Features
(1) Waveform Comparison Function
The 8730-10, 8731-10 Waveform Comparator are designed specifically to
compare waveforms, which is a function of our MEMORY HiCORDERs.
The 8730-10 is the one-channel model; the 8731-10 is the two-channel
model.
Two methods of comparison are available: Area Comparison (which uses a
comparison area created for the waveform), and Value Comparison (which
uses upper and lower limits). Both methods can be used simultaneously.
In Area Comparison, separate areas can be used for CH1 and CH2, with
independent comparison possible. (This is only available with the 8731-
10.)
Value Comparison is based on the result of value calculation (e.g.,
maximum, minimum, or average value).
(2) Forms of Installation
Both 8730-10, 8731-10 models have handles and rubber feet. The units
are portable and can be installed on a workbench or shelf.
The handles and rubber feet are easily secured with screws and can be
easily removed. The 8730-10, 8731-10 can be assembled on a rack
(compliant with JIS or EIA standard) or a panel. A custom-made
mounting bracket is optionally available. Please contact your vendor or
our agent for details.
(3) High-Speed Sampling
The 8730-10 records the waveform of one channel and the 8731-10
records the waveforms of two channels simultaneously at a maximum
speed of 1MS/s.
Given the high-speed sampling of 1,000,000 times per second, even an
instantaneous change in input signal is not missed.
(4) Saving Data
Not only waveform comparison, but also waveform data can be saved on a
PC card.
Data is saved in binary format for reading with the 8730-10, 8731-10, in
CSV format for reading with a PC, or in BMP format for screen images.
Comparison area and waveform data (up to 500DIV when using the 8936)
saved on the 8835-01 can be read with the 8730-10, 8731-10.
(5) Color LCD
A 7.2-inch (640 x480 dots) STN color LCD is used. The waveforms of
CH1 and CH2, or a waveform and its comparison area are clearly
distinguished.
The screen background is blue when comparison is not being conducted.
The background changes to green when the comparison result is PASS or
red when the result is FAIL.

3
_____________________________________________________________________________________________
1.1 Major Features
______________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
(6) Data Display
A gauge (for graduations) or an individual comparison result is displayed
at the left of the screen.
Comments are shown at the top of the screen.
(7) Interface
The 8730-10, 8731-10 is equipped with LAN (10BASE-T) and RS-232C
interfaces.
The 9333 LAN COMMUNICATOR enables data transfers and remote
control.
(8) External Control Terminals
The external control terminals are used to output comparison results
(PASS or FAIL) and to control the 8730-10, 8731-10.
The external control terminals are insulated from the power supply and
GND of the 8730-10, 8731-10.
Value comparison and area comparison results are output from different
terminals. (The 8731-10 compares data for CH1 and CH2 separately, then
outputs the results separately.)
(9) Dust-proof Structure
Flat sheet keys eliminate gaps where dust can gather.
The front panel has a simplified dust-proof structure. The PC card slot has
acover.

4
_____________________________________________________________________________________________
1.2 Identification of Controls and Indicators
______________________________________________________________________________________________
14 3 4
12
8
9
65 7
12 10 1113
Front Panel
1DISPLAY Key Displays the measurement screen and shows a waveform.
This screen lets you set parameters for waveform display and
measurement.
The tag at the top of the screen will show the settings for
measurement.
2COMP SETTING Key The tag at the top of the screen will show the settings for
comparison.
This screen lets you set parameters for waveform comparison.
3AREA EDIT Key Displays the area creation screen.
This screen lets you create, edit, or save a comparison area used
for area comparison.
4FILE Key Displays the file screen.
This screen lets you save data on or read data from the PC card.
It is also used for file-related operations.
5WAVE SCROLL Key Used to scroll back and forth across the waveform.
6CURSOR SCROLL Key Used to move cursors A and B.
7ADJUST/SYSTEM Key Sets up the zero-adjustment and clock and checks the 8730-10,
8731-10.
Sets up the external control terminals, LAN, and RS-232C.
8CURSOR Key Moves the selector cursor up and down and right and left on the
screen.
Used to select a tag or setting and to move a waveform or cursor.
9ENTER Key Opens the selected setting parameter.
Confirms the selected value.
10 STOP Key Interrupts ongoing measurement.
Terminates measurement when pressed twice.
11 START Key Starts measurement.
12 LCD Cover Protective cover for the LCD
13 PC Card Cover Cover for the PC card slot. Open the cover to expose the PC card
slot.
14 Model Name Model No. (For the 8731-10, this number is "8731-10.")
1.2 Identification of Controls and Indicators
This manual suits for next models
1
Table of contents
Other Hioki Measuring Instrument manuals

Hioki
Hioki PQ3198-92 User manual

Hioki
Hioki 9651 User manual

Hioki
Hioki IM3536-01 User manual

Hioki
Hioki SM7420 User manual

Hioki
Hioki CHT3548 User manual

Hioki
Hioki PW3360-21 User manual

Hioki
Hioki RM3545 User manual

Hioki
Hioki PW3365-20 User manual

Hioki
Hioki HiTester 3443-01 User manual

Hioki
Hioki CT9691 User manual

Hioki
Hioki 3196 Instruction manual

Hioki
Hioki CM3286 User manual

Hioki
Hioki 9196 User manual

Hioki
Hioki CM4141-90 User manual

Hioki
Hioki CT6845 User manual

Hioki
Hioki IM7580A-1 Use and care manual

Hioki
Hioki CM3286 User manual

Hioki
Hioki CT6700 User manual

Hioki
Hioki HITESTER 3184 User manual

Hioki
Hioki IM3533 User manual