Hobo Pro v2 User manual

HOBO®Pro v2 User’s Manual
(Part # U23-00x)
© 2010 Onset Computer Corporation. All rights reserved. Patent #: 6,826,664.
Onset, HOBO, and HOBOware are registered trademarks of Onset Computer Corporation. Other products and brand names may be trademarks or registered trademarks of their
respective owners.
Inside this package:
•HOBO®Pro v2 logger
•Clamp and mounting screws
Doc #10694-H, MAN-U23
Onset Computer Corporation
Thank you for purchasing a HOBO data logger. With proper
care, it will give you years of accurate, reliable measurements.
The HOBO Pro v2 logger’s environmentally rugged case is
designed for years of reliable use in outdoor applications. It has
enough memory to record over 42,000 12-bit measurements.
The U23-001 and U23-002 models also feature user-replaceable
RH sensors.
The logger uses an optical USB communications interface (via a
compatible shuttle or base station) for launching and reading out
the logger. The optical interface allows the logger to be
offloaded without compromising the electronics. The USB
compatibility allows for easy setup and fast downloads.
HOBOware®software version 2.2.1 or higher is required for
logger operation. Visit www.onsetcomp.com for compatible
software.
Specifications
Temperature Sensor
Operation range Internal sensors: -40° to 70°C (-40° to 158°F)
U23-002 external temperature sensor: -40° to 70°C (-40° to 158°F)
U23-003 and U23-004 external sensors: -40° to 100°C
(-40° to 212°F), with tip and cable immersion in fresh water
up to 50°C (122°F) for one year
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
-40 -20 0 20 40 60 80 100
Temperature (˚C)
Accuracy/Resolution (˚C)
Accuracy
Resolution
Plot A
Accuracy 0.2°C over 0° to 50°C (0.36°F over 32° to 122°F); see Plot A
Resolution 0.02°C at 25°C (0.04°F at 77°F); see Plot A
Response time
(typical to 90%)
U23-001 Internal sensor: 40 minutes in air moving 1 m/sec
U23-002 external temperature sensor: 5 minutes in air moving
1 m/sec
U23-003 and U23-004 external sensors: 3 minutes in air moving
1 m/sec; 30 seconds in stirred water
Stability (drift) < 0.1°C (0.18°F) per year
Relative Humidity Sensor (U23-001, U23-002 only)
Operation range 0-100% RH, -40° to 70°C (-40° to 158°F)
Exposure to conditions below -20°C (-4°F) or above 95% RH may
temporarily increase the maximum RH sensor error by an
additional 1%
Plot B
Accuracy ±2.5% from 10% to 90% RH (typical), to a maximum of ±3.5%.
See Plot B for full range.
Resolution 0.03%
Response time
(typical to 90%)
U23-001: 40 minute in air moving 1 m/sec with protective cap
U23-002: 5 minutes in air moving 1 m/sec with protective cap
Stability (drift) < 1% per year typical; hysteresis 1%
Logger
Operation range -40° to 70°C (-40° to 158°F)
Real-time clock ± 1 minute per month 0° to 50°C (32° to 122°F)
Battery 1/2 AA, 3.6 Volt lithium, user-replaceable (part # HP-B)
Battery life (typical use) 3 years with 1 minute or greater logging interval
Memory (non-volatile) 64K bytes memory (approx. 21,000 temperature and RH measurements)
HOBO Pro v2 User’s Manual
2
Materials All models: ASA styrene polymer housing and mounting clamp; polypropylene protective cap; Buna-N o-ring(s);
U23-001, U23-002 only: ASA styrene polymer RH sensor cap; modified hydrophobic polyethersulfone membrane
Cables U23-001: No cables
U23-002: One 184 cm (6 ft.) PVC cable; sensor diameter 1 cm (0.38 in.)
U23-003: Two 184 cm (6 ft.) PVC cables; sensor diameter 0.5 cm (0.20 in.)
U23-004: One 184 cm (6 ft.) PVC cable; sensor diameter 0.5 cm (0.20 in.)
Environmental rating Electronics housing is NEMA 6P equivalent (tolerant of brief submergence);
Units with RH sensors are NEMA 4 equivalent (splash-resistant)
Launch modes Immediate start; delayed start
Logging interval Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from
1 second to 18 hours (refer to HOBOware software manual)
Offload modes Offload while logging; stop and offload
Battery indication Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
Weight U23-001: 57 g (1.5 oz); U23-002: 118 g (3.1 oz); U23-003: 138 g (3.7 oz); U23-004: 102 g (2.7 oz)
Dimensions Housing measures 10.2 × 3.8 cm (4.0 × 1.5 in.)
NIST certificate Temperature certificate available for additional charge
The CE Marking identifies this product as complying with the relevant directives in the European Union (EU).
Accessories available
•RH sensor replacement kit for U23-001
(Part # HUM-RHPCB-1)
•RH sensor replacement kit for U23-002
(Part # HUM-RHPCB-2)
Connecting the logger
The HOBO Pro v2 requires a coupler (Part # COUPLER2-E)
and USB-Optic Base Station (Part # BASE-U-4) or HOBO
Waterproof Shuttle (Part # U-DTW-1) to connect to the
computer.
1. Install the logger software on your computer before
proceeding.
2. Follow the instructions that came with your base station or
shuttle to attach the base station or shuttle to a USB port on
the computer.
3. Make sure the logger’s communications window is clean and
dry. (Use a clean, nonabrasive cloth, if necessary.) If the
logger is damp, wipe off excess moisture.
4. Attach the coupler to the base station or shuttle, then insert
the logger into the coupler with the ridge on the logger
aligned with the ridge on the coupler.
5. If you are using the HOBO Waterproof Shuttle, briefly press
the coupler lever to put the shuttle into base station mode.
6. If the logger has never been connected to the computer
before, it may take a few seconds for the new hardware to be
detected by the computer.
7. Use the logger software to launch the logger, check the
logger’s status, read it out, stop it manually with the software,
or let it continue to record data until the memory is full. Or,
use the HOBO Waterproof Shuttle to read out and relaunch
the logger in the field.
Refer to the software user’s guide for complete details on
launching, reading out, and viewing data from the logger.
Important: USB communications may not function properly
at temperatures below 0°C (32°F) or above 50°C (122°F).
Note: The first time you launch the logger, the deployment
number will be greater than zero. Onset launches the loggers to
test them prior to shipping.
HOBO Pro v2 User’s Manual
3
Operation
A light (LED) in the communications window of the logger
confirms logger operation. (In brightly lit areas, it may be
necessary to shade the logger to see the LED blink.) The
following table explains when the light blinks during logger
operation:
When: The “OK” light:
The logger is logging Blinks once every one to four
seconds (the shorter the logging
interval, the faster the light blinks);
blinks when logging a sample
The logger is awaiting a
start because it was
launched in Start At
Interval or Delayed Start
mode
Blinks once every eight seconds
until logging begins
Sample and event logging
The logger can record two types of data: samples and events.
Samples are the sensor measurements recorded at each logging
interval (for example, temperature every minute). Events are
independent occurrences triggered by a logger activity, such as
Bad Battery or Host Connected. Events help you determine what
was happening while the logger was logging.
The logger stores 64K of data, and can record over 42,000 12-bit
measurements.
Deploying and protecting the logger
•To clean the logger’s case, use a sponge with warm, soapy
water.
•Use the included clamp to mount the
logger to a surface. The clamp has two
holes for the screws, 44 mm (1.7 inches)
apart.
The clamp is slightly tapered to
accommodate the logger. Install the
clamp so the logger fits better with its
communication window facing up. This
will prevent condensation from pooling
around the sensor and/or grommet.
•A solar shield is recommended if the
logger will be exposed to sunshine.
•Periodically inspect the three desiccant packs located in the
logger cap. If they are not bright blue, dry them following the
instructions below.
To dry a desiccant pack, remove it from the logger cap and
leave in a warm (≤70oC (158oF), dry location until the bright
blue color is restored. (Refer to the “Battery” section for
instructions on removing and replacing the logger cap.).
If a desiccant pack remains pink and will not turn blue, replace
it with a new desiccant pack (Onset Part #: DESICCANT1)
Replacing the RH sensor
The RH sensor (on models U23-001 and U23-002) is protected
by an ASA styrene polymer cap and a modified hydrophobic
polyethersulfone fluid barrier membrane that allows vapor to
penetrate while protecting the sensor from condensation.
RH sensor performance may degrade over time. To replace the
RH sensor in your logger, refer to the diagram and instructions
for your logger:
U23-001
1. Turn the RH sensor cap counter-clockwise slightly and pull
to remove it. Discard the sensor cap, membrane, and o-ring.
Clean the sensor end of the logger.
2. There may be a spacer installed on the RH Sensor Board.
Remove and discard the spacer.
3. Note the orientation of the small circuit board containing the
RH sensor. With a pair of needle-nose pliers, grip the sensor
board pins. Pull out and discard the board.
4. Use needle-nose pliers to hold the pins on the new sensor
board, push the board gently but firmly and install it in the
same orientation as the old board. Make sure it engages with
the pins inside the logger housing. Do not touch the sensor
itself; only touch the sides of the board.
5. Install the new spacer on the sensor board by placing the
spacer onto the third pin from the left (use the topmost hole
on the spacer instead of the center hole).
6. Make sure the o-ring is clean and seated properly, and set the
protective membrane on top (either side can face up).
7. Put the sensor cap back on. Push down and turn it slightly
clockwise to close it securely. Do not force it. If the cap does
not go on easily, the sensor may be installed incorrectly.
Check the sensor orientation and try again.
8. Check logger status in HOBOware to verify the RH reading.
Logger
RH Sensor Board
Spacer
O-ring
Protective Membrane
RH Sensor Cap
WARNING: Do not touch
the actual sensor (black
chip). Doing so may
compromise accuracy.
Face-on view with spacer installed
RH Sensor
HOBO Pro v2 User’s Manual
4
U23-002
1. Grasp the cap and membrane and pull firmly. Discard them.
2. Note the orientation of the small circuit board containing the
RH sensor. Pull it out and discard it.
Cable from logger
Receptacle RH sensor
RH sensor cap
& protective membrane
3. Holding the sides of the board only, push gently but firmly to
install the new sensor (Onset part # HUM-RHPCB-2) in the
same orientation. WARNING: Do not touch the actual
sensor (black chip) itself; doing so may compromise
accuracy.
4. Put the new sensor cap and membrane on. Do not force the
cap. If it does not go on easily, the sensor may be installed
backwards. Reverse the sensor and try again.
Battery
Typical battery life is about three years. Actual battery life is a
function of the number of deployments, logging interval, and
operation/storage temperature of the logger. To obtain a three-
year battery life, use a logging interval of one minute or greater,
and operate and store the logger at temperatures between 0° and
40°C (32° and 104°F).
Frequent deployments with logging intervals of less than one
minute, and continuous storage/operation at temperatures above
40°C, will result in significantly lower battery life. For example,
continuous logging at a one-second logging interval will result in
a battery life of approximately one month.
The logger can report and log its own battery voltage. If the
battery falls below 3.1 V, the logger will record a “bad battery”
event in the datafile. If the datafile contains “bad battery” events,
or if logged battery voltage repeatedly falls below 3.3 V, the
battery is failing and should be replaced before the next
deployment. To change the battery:
1. Turn slightly counter-clockwise and pull to remove the
protective cap. Loosen the cable grommet on the opposite
end, if the logger has an external sensor.
2. Carefully pull out the circuit board containing the battery. (If
the logger has an external sensor, you will probably find it
easier to push the cable(s) into the case to push the circuit
board out.)
3. Examine the desiccant packs that were packed into the case.
If the desiccant is not bright blue, put the desiccant packs in a
warm, dry place until the blue color is restored.
Circuit board
Protective cap
Battery Desiccant packs
Cable from sensor
(if applicable)
Grommet for cable
(if applicable)
4. Install a new 1/2 AA, 3.6 Volt lithium battery (part # HP-B).
The positive end of the battery should face towards the
communication LEDs.
5. Use a clean, dry cloth to wipe away any moisture inside the
case.
6. Push the board and the desiccant packs back into the case,
taking care not to bend the communication LEDs. Align the
board with the grooves inside the case. (If you try to put the
board in upside-down, the battery will get in the way.)
7. Make sure o-ring on the protective cap is still in place. It
should not be pinched, twisted, or trapping dirt or lint, which
could interfere with the protective cap.
8. Line up the bumps on the protective cap with the notches in
the logger’s case. Push and turn the cap slightly clockwise.
Pull the slack in the sensor cable(s) and tighten the grommet,
if applicable.
WARNING: Do not cut open, incinerate, heat above
100°C (212°F), or recharge the lithium battery. The battery
may explode if the logger is exposed to extreme heat or
conditions that could damage or destroy the battery case. Do
not dispose of the logger or battery in fire. Do not expose the
contents of the battery to water. Dispose of the battery
according to local regulations for lithium batteries.

Other Hobo Other manuals

Hobo H08-006-04 User manual

Hobo

Hobo H08-006-04 User manual

Hobo MX series User manual

Hobo

Hobo MX series User manual

Hobo UX90-004 User manual

Hobo

Hobo UX90-004 User manual

Hobo Pendant UA-001-08 User manual

Hobo

Hobo Pendant UA-001-08 User manual

Hobo RX3000 Series Operator's manual

Hobo

Hobo RX3000 Series Operator's manual

Hobo U30/GSM User manual

Hobo

Hobo U30/GSM User manual

Hobo MX2001-0 Series User manual

Hobo

Hobo MX2001-0 Series User manual

Hobo U10-003 User manual

Hobo

Hobo U10-003 User manual

Hobo U12 User manual

Hobo

Hobo U12 User manual

Hobo MX2301 User manual

Hobo

Hobo MX2301 User manual

Hobo U12 User manual

Hobo

Hobo U12 User manual

Hobo Pendant MX Temp User manual

Hobo

Hobo Pendant MX Temp User manual

Hobo U20L-01 User manual

Hobo

Hobo U20L-01 User manual

Hobo U24 User manual

Hobo

Hobo U24 User manual

Hobo U12-015 User manual

Hobo

Hobo U12-015 User manual

Hobo U12-008 User manual

Hobo

Hobo U12-008 User manual

Popular Other manuals by other brands

Versiton BHR40-V851 Manual guide

Versiton

Versiton BHR40-V851 Manual guide

Dometic CT 4050 operating manual

Dometic

Dometic CT 4050 operating manual

Siemens SIMATIC RF1000 Series operating instructions

Siemens

Siemens SIMATIC RF1000 Series operating instructions

Clarke CAR8MC user guide

Clarke

Clarke CAR8MC user guide

Anviz C2 user manual

Anviz

Anviz C2 user manual

JBJ 28g Nano Cube CF-QUAD owner's manual

JBJ

JBJ 28g Nano Cube CF-QUAD owner's manual

Cardin Elettronica EL Series instruction manual

Cardin Elettronica

Cardin Elettronica EL Series instruction manual

GRASS VALLEY Kaleido-X16 Hardware description and installation manual

GRASS VALLEY

GRASS VALLEY Kaleido-X16 Hardware description and installation manual

PianoDisc HD troubleshooting guide

PianoDisc

PianoDisc HD troubleshooting guide

GUTMANN mega macs 66 user manual

GUTMANN

GUTMANN mega macs 66 user manual

Siemens SIRIUS 3R Assembly instructions

Siemens

Siemens SIRIUS 3R Assembly instructions

IDTECK Star SR10 user manual

IDTECK

IDTECK Star SR10 user manual

Gantner GT7.2500 manual

Gantner

Gantner GT7.2500 manual

Rheem P-M206 instructions

Rheem

Rheem P-M206 instructions

IBM HS21 XM BLADECENTER - FOR SPECJBB2005 specification

IBM

IBM HS21 XM BLADECENTER - FOR SPECJBB2005 specification

Honeywell LonWorks quick start guide

Honeywell

Honeywell LonWorks quick start guide

Microelettrica Scientifica MG30 Operation manual

Microelettrica Scientifica

Microelettrica Scientifica MG30 Operation manual

Robinson R-66 Series Lithium-Ion installation instructions

Robinson

Robinson R-66 Series Lithium-Ion installation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.