Honeywell X Series User manual

43-TV-33-76 iss.2GLO Jan 21 UK 1
QX and SX Recorder - Installation Instruction
Backlight Replacement (For X Series Only)
QX Recorder
SX Recorder
This Installation Instruction sheet is intended as a guide for replacing or installing hardware and for setting up
functionality in the recorder. Refer to the User manual for detailed operational requirements.
The QX and SX Recorders are designed for ease of assembly with minimal disturbance to the rest of the unit. If unsure,
please return the unit to your supplier for repair or upgrade.
Removing the case and opening the back of the recorder should only be performed under the following circumstances:
When an item of hardware requires individual replacement.
When an item of hardware is to be retrospectively fitted.
In all other instances it is recommended that the complete unit be returned to an authorised agent or service centre.
For Agency Approved recorders the product needs to be upgraded or repaired by an Authorized Repair facility
WARNING
HAZARDOUS VOLTAGES
Disconnect all power to the recorder before removing the case and attempting any maintenance procedures.
SAFETY TESTS
Upon completion of service procedures detailed in this manual two basic safety tests should be performed in order to
ensure continued safe operation of the instrument.
Earth Resistance; 25 VDC applied between case and protective earth, bonding resistance should be < 0.1 Ohm.
Insulation Resistance; 500Vdc applied between the earth terminal, and the live and neutral terminal shorted together,
insulation resistance to be >2.0 MOhm. (Mega Ohm, NOT milli Ohm)
Failure to comply with these instructions could result in death or serious injury
CAUTION
OBSERVE ANTI-STATIC PRECAUTIONS
Refer to BS EN61340-5-1: 2001. Basic specification. Protection of electrostatic sensitive devices.
Full anti-static precautions MUST be observed when in contact with the electronics of your recorder.
SAVE DATA, SETUPS AND LAYOUTS
Removal of PCBs and battery back-up will result in the loss of all non-volatile data.
Ensure all data and set-ups are saved.
Failure to comply with these instructions may result in product damage.

43-TV-33-76 iss.2GLO Jan 21 UK 2
Before attempting to repair or upgrade a recorder, it is advisable to clear a sufficient work space so components such as
the front panel can be rested on the work surface without getting scratched or damaged.
Panel Mounting
Before a new backlight can be fitted the recorder must be removed from the panel. Loosen the mounting clamp screws,
slide the clamp towards the rear of the recorder and remove the clamps and the recorder.
SX Recorder
To fit a new backlight the case and back panel must be removed. The case and back panel can be removed as a single
assembly. Remove the 2 screws on each side of the case as shown in Fig 1 below
Slide the case and rear panel assembly off the recorder.
Fig 1
Remove 2 screws from
each side of the case
Fig 2

43-TV-33-76 iss.2GLO Jan 21 UK 3
Next disconnect the inverter high voltage cables by pulling upwards. See Fig 5.
Remove the eight self tapping screws that fix the plastic bezel to the mounting plate. A screw driver with a long shaft will
make this operation much easier. The screws are located one in each corner and one in the centre of each side. See
Fig 6.
Fig 5
Inverter cable
Fig 6
8 screws, to
release front
bezel

43-TV-33-76 iss.2GLO Jan 21 UK 4
Once the screws are removed the black plastic bezel and touch screen assembly can be folded down. The touch screen
will still be connected so take care not to damage this lead. Fig 7
To remove the display unscrew the four display fixing screws. As the display comes away feed the inverter cable
through the hole in the metal mounting plate, and disconnect the display cable by gently pulling on the connector
housing.
Once disconnected the display should be free. Put the cable face down ensuring it is on a surface that will not scratch it.
Avoid getting dirt or finger marks on the display.
Fig 8
Touch
screen cable
Four display
mounting
screws
Fig 7

43-TV-33-76 iss.2GLO Jan 21 UK 5
Where the inverter lead comes out of the display there is a clip that will release the backlight. See Fig 9.
Press the clip down and gently pull on the cable to withdraw the backlight assembly. See Fig 10
Once the backlight has been removed the new backlight assembly can be fitted. Avoid touching the tube on the
backlight assembly. Slide the new backlight into the display until it is fully home. Ensure the backlight is fitted the correct
way round with the tubes as shown in Fig 10.
Fig 9
Clip
Fig 10

43-TV-33-76 iss.2GLO Jan 21 UK 6
Re-building the recorder
Rebuilding the recorder is the reverse process.
Offer up the display to the front of the unit, reconnect the display lead and feed the inverter cable through the
rectangular hole in the mounting plate. Fit the 4 screws holding the display in place.
Before refitting the bezel/touch screen assembly clean any dirt or marks form the back of the touch screen or display.
Use a lint free cloth to clean both parts, do not use any solvents.
Refit the 8 screws that hold the front bezel in place.
Plug the inverter lead back into the inverter.
Slide the case back onto the recorder ensuring it is fitted the right way round. The cooling slots should be at the top of
the unit.
Finally fit the four M4 screws that hold the case to the chassis.

43-TV-33-76 iss.2GLO Jan 21 UK 7
QX Recorder
To fit a new backlight the case and back panel must be removed. The case and back panel can be removed as a single
assembly. Remove the 2 screws on each side of the case as shown in Fig 11 below
Slide the case and rear
panel assembly off the
recorder.
Fig 11
Remove 2 screws from
each side of the case
Fig 12

43-TV-33-76 iss.2GLO Jan 21 UK 8
To remove the Front surround assembly four self tapping screws must be removed. See Fig 13
Once these have been removed the front surround assembly can be carefully drawn forward. Disconnect the speaker
cable from the mother board. Disconnect the display cable, inverter cable and touch screen cable from the processor
board. Take care not to pull on the cable, only on the connector body.
The display cable is unplugged by gently pulling up the top part of the connector until the cable is unclamped and can
easily be pulled free.
Fig 13
4 screws to
release Front
surround. 2 on
each side
Speaker
connector

43-TV-33-76 iss.2GLO Jan 21 UK 9
To remove the front bezel and touch screen assembly there are 4 self tapping screws that need to be removed, see Fig
14. Two screws are at the top of the assembly, and two screws are fitted one from each side. When these are removed
the display assembly can be separated.
Take care not to touch the display face or the back of the touch screen, any dirt or finger marks will show when the unit
is re-assembled.
Lift the display assembly out of the front bezel and place it face down. Ensure it is on something that will prevent it
getting scratched or dirty.
Fig 14
4 screws to
release display
assembly
Fig 15
Unplug backlight cables from
inverter and feed through the
holes as shown

43-TV-33-76 iss.2GLO Jan 21 UK 10
There are 2 backlights on the QX recorder display. These should be changed as a pair. To get the backlights out use a
small screwdriver to push back the clip as shown in Fig 16 and gently pull on the leads to withdraw the backlight from
the display. Remove both tubes, noting their orientation. Fig 17.
Insert the new backlights and push into the display until the clips latch them in.
Fig 16
Push clip up and
pull on cables
Fig 17
Withdraw
backlight tube

43-TV-33-76 iss.2GLO Jan 21 UK 11
Re-building the recorder
Rebuilding the recorder is the reverse process.
Feed the backlight leads through the holes in the moulding, and re-connect to the inverter. Pull any loose cable through
the holes.
Before refitting the bezel/touch screen assembly clean any dirt or marks form the back of the touch screen or display.
Use a lint free cloth to clean both parts, do not use any solvents.
Take the display assembly and place it back into the bezel/touch screen assembly. Fit the 4 self tap screws that attach
the two parts.
Offer up the front surround to the unit, reconnect the display lead, pulling up the top of the connector to allow the lead to
be seated, and then push the top down to clamp the lead in place. Re-connect the inverter lead, touch screen lead and
speaker lead.
Slide the front surround assembly onto the chassis plates and fit the 4 self tap screws that hold the unit together.
Slide the case back onto the recorder ensuring it is fitted the right way round. The cooling slots should be at the top of
the unit.
Finally fit the four M3 screws that hold the case to the chassis.
Copyrighted Materials. For more information please contact your supplier.
Other manuals for X Series
3
This manual suits for next models
2
Table of contents
Other Honeywell Voice Recorder manuals

Honeywell
Honeywell DR Graphic User manual

Honeywell
Honeywell DR4300 Series Guide

Honeywell
Honeywell DPR180 User manual

Honeywell
Honeywell DR4500A Manual

Honeywell
Honeywell DR4300 Series User manual

Honeywell
Honeywell DR4300 Series Guide

Honeywell
Honeywell X Series User manual

Honeywell
Honeywell DR4500A User manual

Honeywell
Honeywell DR4300 Series User manual

Honeywell
Honeywell DR4500A Manual