HP 3458A User manual

t
T
.O
.
33K8-4-551-1
TECHNICAL MANUAL
CALIBRATION PROCEDURE
FOR
DIGITAL VOLTMETER
3458A
(HEWLETT-PACKARD)
This publication incorporates Interim Operational Supplement T
.O
. 33K8-4-551-1S-1
dated 27 August 2007.
Distribution Statement C - Distribution authorized to U
. S
. Government agencies and their contractors for
official use or for administrative or operational purposes only, 7 November 1990
. Requests for this
document shall be referred to 562 CBSG/GBHA, 813 Irving-Wick Dr W, Heath, OH 43056-1199.
Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent
disclosure of the contents or reconstruction of the document.
Published under Authority of the Secretary of the Air Force
30 JUNE 2007
CHANGE 1 - 30 SEPTEMBER 2007

it
T
.O
. 33K8-4-551-1
INSERT LATEST CHANGED PAGES
. DESTROY SUPERSEDED PAGES.
LIST OF EFFECTIVE PAGES
NOTE
: The portion of the text and illustrations affected by the changes is indicated by a vertical
line in the outer margins of the page.
Date of issue for original and changed pages are:
Original
0
30 June 2007
Change
l
30 September 2007
TOTAL NUMBER OF PAGES IN THIS PUBLICATION IS 44, CONSISTING OF THE FOLLOWING:
Page
* Change
Page
* Change
Page
* Change
No
.
No
.
No
.
No
.
No
.
No.
Title
1
A
1
1 - 16
0
17
1
18 - 31
0
32 Blank
0
A-I - A-10
0
*Zero in this column indicates an original page
A Change 1
USAF

T
.O
. 33K8-4-551-1
DIGITAL VOLTMETER
3458A
(HEWLETT-PACKARD)
I CALIBRATION DESCRIPTION
:
Table I.
Test Instrument (TI)
Performance
Test
Characteristics
Specifications
Method
NOTE
The Performance Specifications listed below for the TI show
Relative
specifications as found in
the commercial data and
Absolute specifications
. The absolute specifications were calculated by
combining the uncertainty of the relative specifications found in the commercial data with the
uncertainty of the calibration standards listed in Section 2, Equipment Requirements.
DC Voltage
Range
: 0 to 1000 VDC
Compared to a DC
Reference Standard
Accuracy
:
and DC Reference
Relative
: ±(ppm of rdg + ppm of mg)
Divider or DC Voltage
Absolute
: ±[Sgrt(TI ppm of rdg^2 +
Divider
STD 1 ppm of rdg^2 + STD2 ppm of rdg^2)
+ TI ppm of rng]
Standard
Range
Relative
Absolute
100mV
9+10
(9+1
.5+0
.5)+10
lV
8+1
(8+1
.5+0
.2)+1
10V
8+0
.2
(8+1
.5)+0
.2
100V 10+ 0
.3
(10+1
.5+0
.2)+0
.3
1000 V
10+0
.1
(10+ 1
.5+0
.5)+0
.1
Inputs >100 VDC, add
12 ppm X (Vin/1000)
2
Opt 002
Range
Relative
Absolute
100mV
5+10
(5+1
.5+0
.5)+10
1V
4+1
(4+1
.5+0
.2)+1
IOV
4+0
.2
(4+1
.5)+0
.2
1

T
.O
. 33K8-4-551-1
Table 1
. (Cont)
Test Instrument (TI)
Performance
Test
Characteristics
Specifications
Method
DC Voltage
(Cont
.)
Range
: 0 to 1000 VDC
Compared to a DC
Reference Standard
Accuracy
:
and DC Reference
Relative
: ±(ppm of rdg + ppm of mg)
Divider or DC Voltage
Absolute
: ±[Sqrt(TI ppm of rdg
A
2 +
Divider
STD 1 ppm of rdg^2 + STD2 ppm of rdg^2)
+ TI ppm of mg]
Opt 002
(Cont
.)
Range
Relative
Absolute
100 V
6 + 0
.3
(6 + 1
.5 + 0
.2) + 0
.3
1000 V
6 + 0
.1
(6 + 1
.5 + 0
.5) + 0
.1
Inputs >100 VDC, add
12 ppm X (Vinll000)
2
Resistance
Range
Range
: 0 to 1 G52
Compared to a
Resistance Standard
Accuracy:
Relative
: ±(ppm of rdg + ppm of mg),
(Add 250 mf
g
for Two-Wire Ohms)
Absolute: ±[Sgrt(TI ppm of rdg^2 +
STDI ppm of rdg^2) + Ti ppm of mg]
(Add 250 ml) for Two-Wire Ohms)
Relative
Absolute
10 S2
15+5
(15+2
.15)+5
10052
12+5
(12+2
.15)+5
1 kl2
10+0
.5
(10+2
.15)+0
.5
10ki2
10+0
.5
(10+2.1)+0
.5
100kI2
10+0
.5
(10+2.4)+0
.5
1 MSZ
15+2
(15+6)+2
10 MCI
50+10
(50 + 12) + 10
100 MS2
500 + 10
(500 + 50) + 10
1 G1)
5000 + 10
(5000 + 50) + 10
2

T
.O
. 33K8-4-551-1
Table 1
. (Cont
.)
Test Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
DC Current
Range
: 0 to 1 ADC
Compared to a known
DC Current
Accuracy:
Relative
: ±(ppm of rdg + ppm of mg)
Absolute
: ±[Sgrt(TI ppm of rdg^2 +
STD1 ppm of rdg^2 + STD2 ppm of rdg^2)
+ TI ppm of mg]
Range
Relative
100 nA
*
I
30 + 400
1 µA
*
I
20 + 40
10µA *'
20+10
100 µA
20 + 8
1 and 10 mA
20 + 5
100mA
35+5
lA
110+ 10
Absolute
N/A
N/A
N/A
(20 + 5
.277 + 5) + 8
(20 + 5
.277 + 5) + 5
(35+5
.277+5)+5
(110+5
.277+6)+10
AC Voltage
Range
: 0 to 1000 VAC
(Synchronous Sub-sampled Mode)
AC BAND <_2 MHz
Compared to an
AC Measurement
Standard or Calibrator
*2
Accuracy:
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD1 % of rdg^2 +
STD2 % of rdg^2) + TI % of mg +
STD1 µV + STD2 [IV]
Range
Frequency
Relative
Absolute
10 mV
1 to 20 Hz
0
.03 + 0
.03
(0
.03+0+0
.029)+0
.03+0+ 1
.3
20 to 40 Hz
0
.03 + 0
.03
(0
.03+0+0
.019)+0
.03+0+ 1
.3
40 Hz to 1 kHz
0
.02 + 0
.011
(0
.02 + 0 + 0
.011) + 0
.011 + 0 + 1
.3
1 to 20 kHz
0
.03 + 0
.011
(0
.03 +0+0
.011)+0
.011 + 0 + 1
.3
20to50kHz
0
.1+0
.011
(0
.1+0+0
.021)+0
.011+0+2
50 to 100 kHz
0
.5 + 0
.011
(0
.5+0+0
.031)+0
.011+0+2
.5
100 to 300 kHz
4 + 0
.02
(4+0+0
.081)+0
.02+0+4
See footnotes at end of Table
.
3

T
.O
. 33K8-4-551-1
Table 1
. (Cont)
Test Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
AC Voltage
Range
: 0 to 1000 VAC
Compared to an
(Synchronous Sub-sampled Mode)
(Cont
.)
AC Measurement
AC BAND 5
.2 MHz
Accuracy
:
Standard or Calibrator
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD 1 % of rdg^2 +
STD2 % of rdg^2) + TI % of mg +
STD1 µV + STD2 µV]
Range
Frequency
Relative
100 mV
1 to 20 Hz
0
.007 + 0
.004
20 to 40 Hz
0
.007 + 0
.004
40 Hz to 1 kHz
0
.007 + 0
.002
1 to 20 kHz
0
.014 + 0
.002
20 to 50 kHz
0
.03 + 0
.002
50 to 100 kHz
0
.08 + 0
.002
100to300kHz
0
.3+0
.01
300 to 500 kHz
1 + 0 .01
500 kHz to 1 MHz
1 + 0
.01
1 to2MHz
1
.5+0
.01
1 V
1 to 20 Hz
0
.007 + 0
.004
20 to 40 Hz
0
.007 + 0
.004
40 Hz to 1 kHz
0
.007 + 0
.002
1 to 20 kHz
0
.014 + 0
.002
20 to 50 kHz
0
.03 + 0
.002
50 to 100 kHz
0
.08 + 0
.002
100 to 300 kHz
0
.3 + 0
.01
300to500kHz
1+0
.01
500 kHz to 1 MHz
1 + 0 .01
1to2MHz
1
.5+0
.01
10 V
1 to 20 Hz
0
.007 + 0
.004
20 to 40 Hz
0
.007 + 0
.004
40 Hz to 1 kHz
0
.007 + 0
.002
1 to 20 kHz
0
.014 + 0
.002
20 to 50 kHz
0
.03 + 0
.002
50 to 100 kHz
0
.08 + 0
.002
100 to 300 kHz
0
.3 + 0
.01
300 to 500 kHz
1 + 0
.01
500 kHz to 1 MHz
1 + 0
.01
1 to 2 MHz
1
.5 + 0
.01
See footnote at end of Table
.
Absolute
(0
.007+0+0
.021)+0
.004+0+1
.5
*2
(0
.007+0+0
.0085)+0
.004+0+1
.5
(0
.007+0+0
.0038)+0
.002+0+1
.5
(0
.014+0+0
.0038)+0
.002+0+1
.5
(0
.03 +0+0
.0069)+0
.002+0+2
(0
.08+0+0
.016)+0
.002+0+2
.5
(0
.3+0+0
.025)+0
.01 +0+4
(1+0+0
.038)+0
.01+0+8
(1+0+0
.06)+0
.01+0+0
(1
.5+0.1+0
.0038)+0
.01+3+1
.5
(0
.007+0+0
.02)+0
.004+0+0
*
2
(0
.007+0+0
.0066)+0
.004 +0+0
(0
.007+0+0
.0024)+0
.002+0+0
(0
.014+0+0
.0024)+0
.002+0+0
(0
.03+0+0
.0046)+0
.002+0+0
(0
.08+0+0
.0071)+0
.002+0+0
(0
.3+0+0
.016)+0
.01+0+0
(1+0+0
.026)+0
.01+0+0
(1+0+0
.06)+0
.01+0+0
(1
.5+0
.1-1-0
.0024)+0
.01+3+0
(0
.007+0+0
.02)+0
.004+0+0
*
2
(0
.007+0+0
.0067)+0
.004+0+0
(0
.007+0+0
.0027)+0
.002+0+0
(0
.014+0+0
.0027)+0
.002+0+0
(0
.03+0+0
.0048)+0
.002+0+0
(0
.08+0+0
.0081)+0
.002+0+0
(0
.3+0+0
.019)+0
.01+0+0
(1+0+0
.04)+0
.01+0+0
(1+0+0
.06)+0
.01+0+0
(1
.5+0.1+0
.0027)+0
.01+3+0
*
2
4

T
.O
. 33K8-4-551-1
Table 1
. (Cont)
Test Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
AC Voltage
Range
: 0 to 1000 VAC
Compared to an
AC Measurement(Synchronous Sub-sampled Mode)
(Cont
.)
AC BAND <_2 MHz
Accuracy
:
Standard or Calibrator
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD1 % of rdg^2 +
STD2 % of rdg
A
2) + TI % of mg +
STD1 µV + STD2 µV]
Range
Frequency
Relative
Absolute
100 V
1 to 20 Hz
0
.02 + 0
.004
(0
.02+0+0
.02)+0
.004+0+0
*
2
20 to 40 Hz
0
.02 + 0
.004
(0
.02+0+0
.0068)+0
.004+0+0
40 Hz to 20 kHz
0
.02 + 0
.002
(0
.02+0+0
.0031)+0
.002+0+0
20 to 50 kHz
0
.035 + 0
.002
(0
.035+0+0
.0069)+0
.002+0+0
50 to 100 kHz
0
.12 + 0
.002
(0
.12+0+0
.0098)+0
.002+0+0
100 to 300 kHz
0
.4 + 0
.01
(0
.4+0+0
.021)+0
.01+0+0
*2
300 kHz to 1 MHz
1
.5 + 0
.01
(1
.5+0+0
.05)+0
.01+0+0
*
2
1000 V
* 3
1 to 20 Hz
0
.04 + 0
.004
(0
.04+0+0
.02)+0
.004+0+0
* 2
20 to 40 Hz
0
.04 + 0
.004
(0
.04+0+0
.0099)+0
.004+0+0
*
2
(<220 V)
40 Hz to 1 kHz 0
.04 + 0
.002
(0
.04+0+0
.0031)+0
.002+0+0
(<_700 V)
40 Hz to 1 kHz 0
.04 + 0
.002
(0
.04+0+0
.0041)+0
.002+0+0
(>700 V) 40 Hz to 1 kHz
0
.04 + 0
.002
(0
.04+0+0
.0038)+0
.002+0+0
(<_700 V)
1 to 20 kHz
0
.06 + 0
.002
(0
.06+0+0
.0041)+0
.002+0+0
(>700 V)
I to 20 kHz
0
.06 + 0
.002
(0
.06+0+0
.0038)+0
.002+0+0
20 to 50 kHz
0
.12 + 0
.002
(0
.12+0+0
.013)+0
.002+0+0
*
2
50 to 100 kHz
0
.3 + 0
.002
(0
.3+0+0
.05)+0
.002+0+0
*
2
AC BAND >2 MHz
Accuracy:
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD1 % of rdg^2 +
STD2 % of rdg^2) + TI % of mg +
STD1
[IV
+ STD2 µV]
Range
Frequency
Relative
Absolute
10 mV
45 Hz to 20 kHz
0
.09 + 0
.06
(0
.09+0+0
.011)+0
.06+0+ 1
.3
20 to 50 kHz
0
.09 + 0
.06
(0
.09+0+0
.021)+0
.06+0+2
50 to 100 kHz
0
.09 + 0
.06
(0
.09 + 0 + 0
.031) + 0
.06 + 0 + 2
.5
100to300kHz
1
.2+0
.05
(1
.2+0+0
.081)+0
.05+0+4
300to500kHz
1
.2+0
.05
(1
.2+0+0
.089)+0
.05+0+8
500 kHz to 1 MHz
1
.2 + 0
.05
(1
.2+0+0
.17)+0
.05+0+8
1 to 2 MHz
7 + 0
.07
(7+0
.1 +0
.011)+0
.07+3+ 1
.3
2to4MHz
7+0
.07
(7+0.2+0
.011)+0
.07+3+ 1
.3
4 to 8 MHz
20 + 0
.08
(20+0
.2 +0
.011)+0
.08 +3 + 1
.3
See footnotes at end of Table
.
5

T
.O
. 33K8-4-551-1
Table 1
. (Cont)
Test Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
AC Voltage
Range
: 0 to 1000 VAC
Compared to an
AC Measurement
(Synchronous Sub-sampled Mode)
(Cont
.)
AC BAND >2 MHz
Accuracy
:
Standard or Calibrator
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg
A
2 + STD1 % of rdg^2 +
STD2 % of rdg
A
2) + TI % of mg +
STD1 p
.V + STD2
.tV]
Range
Frequency
Relative
Absolute
100 mV
45 Hz to 20 kHz
0
.09 + 0
.06
(0
.09+0+0
.0038)+0
.06+0+ 1
.5
20 to 50 kHz
0
.09 + 0
.06
(0
.09 + 0 + 0
.0069) + 0
.06 + 0 + 2
50 to 100 kHz
0
.09 + 0
.06
(0
.09 + 0 + 0
.016) + 0
.06 + 0 + 2
.5
100 to 300 kHz
2 + 0
.05
(2+0+0
.025)+0
.05+0+4
300 to 500 kHz
2 + 0
.05
(2+0+0
.038)+0
.05+0+8
500 kHz to 1 MHz 2 + 0
.05
(2+0+0
.06)+0
.05+0+0
1 to 2 MHz
4 + 0
.07
(4+0
.1 +0
.0038)+0
.07 +3 + 1
.5
2to4MHz
4+0
.07
(4 + 0
.2 + 0
.0038) + 0
.07 + 3 + 1
.5
4to8MHz
4+0
.08
(4 + 0.2 + 0
.0038) + 0
.08 + 3 + 1
.5
8to10MHz
15+0
.1
(15+0
.2+0
.0038)+0
.1+3+1
.5
1 V
45 Hzto20kHz
0
.09+0
.06
(0
.09+0+0
.0024)+0
.06+0+0
20 to 50 kHz
0
.09 + 0
.06
(0
.09 +0+0
.0046)+0
.06+0+0
50 to 100 kHz
0
.09 + 0
.06
(0
.09+0+0
.0071)+0
.06+0+0
100to300kHz
2+0
.05
(2+0+0
.016)+0
.05+0+0
300to500kHz
2+0
.05
(2+0+0
.026)+0
.05+0+0
500 kHz to 1 MHz
2 + 0
.05
(2+0+0
.06)+0
.05+0+0
1 to2MHz
4+0
.07
(4+0.1+0
.0024)+0
.07+3+0
2to4MHz
4+0
.07
(4+0.2+0
.0024)+0
.07+3 +0
4to8MHz
4+0
.08
(4+ 0
.2 + 0
.0024) + 0
.08 + 3 +
8to10MHz
15+0
.1
(15+0
.2+0
.0024)+0
.1 +3 +0
10V
45 Hz to 20 kHz
0
.09 + 0
.06 (0
.09+0+0
.0027)+0
.06+0+0
20 to 50 kHz
0
.09 + 0
.06
(0
.09+0+0
.0048)+0
.06+0+0
50 to 100 kHz
0
.09 + 0
.06 (0
.09+0+0
.0081)+0
.06+0+0
100to300kHz
2+0
.05
(2+0+0
.019)+0
.05+0+0
300 to 500 kHz
2 + 0
.05
(2+0+0
.04)+0
.05+0+0
500 kHz to 1 MHz
2 + 0
.05
(2+0+0
.06)+0
.05+0+0
1 to2MHz
4+0
.07
(4+0.1+0
.0027)+0
.07+3+0
*
2
2 to 4 MHz
4 + 0
.07
(4+0.2+0
.0027)+0
.07+3+0
*
2
4to8MHz
4+0
.08
(4+0.2+0
.0027)+0
.08+3+0
*
2
8to 10 MHz
15+0
.1
(15+0.2+0
.0027)+0
.1+3+0
*
2
100 V
45 Hz to 20 kHz
0
.12 + 0
.002
(0
.12+0+0
.0031)+0
.002+0+0
20to50kHz
0
.12+0
.002
(0
.12+0+0
.0069)+0
.002+0+0
50 to 100 kHz
0
.12 + 0
.002
(0
.12+0+0
.0098)+0
.002+0+0
1000V
*
3
(<_220 V)
45 Hz to 20 kHz
0
.3 + 0
.01
(0
.3+0+0
.0031)+0
.01+0+0
(<_700 V)
45 Hz to 20 kHz
0
.3 + 0
.01
(0
.3+0+0
.0041)+0
.01+0+0
(>700 V)
45 Hz to 20 kHz
0
.3 + 0
.01
(0
.3+0+0
.0038)+0
.01+0+0
20 to 50 kHz
0
.3 + 0
.01
(0
.3+0+0
.013)+0
.01+0+0
*
2
50 to 100 kHz
0
.3+0
.01
(0
.3+0+0
.05)+0
.01+0+0
See footnotes at end of Table.
6

T
.O
. 33K8-4-551-1
Table 1
. (Cont)
Test
Instrument (TI)
Characteristics
Performance
Specifications
Test
Method
AC Voltage Range
: 0 to 1000 VAC
Compared to an
(Analog Mode)
AC Measurement
AC BAND <_2 MHz
Accuracy
:
Standard or Calibrator
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg
A
2 + STD 1 % of rdg^2 +
STD2 % of rdg
A
2) + TI % of rng +
STD1 µV + STD2 µV]
Range
Frequency
Relative
Absolute
10mV
10 to 20 Hz
0
.4+0
.32
(0
.4+0+0
.029)+0
.32+0+ 1
.3
20 to 40 Hz
0
.15 + 0
.25
(0
.15+0+0
.019)+0
.25+0+ 1
.3
40 to 100 Hz
0
.06 + 0
.25
(0
.06+0+0
.011)+0
.25+0+ 1
.3
100 Hz to 20 kHz
0
.02 + 0
.25
(0
.02+0+0
.011)+0
.25+0+ 1
.3
20to50kHz
0
.15+0
.25
(0
.15+0+0
.021)+0
.25+0+2
50 to 100 kHz
0
.7+0
.35
(0
.7+0+0
.031)+0
.35+0+ 2
.5
100 to 250 kHz
4+0
.7
(4+0+0
.081)+0
.7+0+4
100 mV
10 to 20 Hz
0
.4 + 0
.02 (0
.4+0+0
.021)+0
.02+0+ 1
.5
20 to 40 Hz
0
.15+0
.02 (0
.15 +0+0
.0085)+0
.02+0+ 1
.5
40 to 100 Hz
0
.06 + 0
.01
(0
.06+0+0
.0038)+0
.01 +0+ 1
.5
100 Hz to 20 kHz
0
.02 + 0
.01
(0
.02+0+0
.0038)+0
.01 +0+ 1
.5
20 to 50 kHz
0
.15 + 0
.04
(0
.15 +0+0
.0069)+0
.04+0+2
50 to 100 kHz
0
.6 + 0.08 (0
.6+0+0
.016)+0
.08+0+ 2
.5
100 to 2501(Hz
2+0
.5
(2+0+0
.025)+0
.5+0+4
250 to 300 kHz
3 + 0
.6
(3+0+0
.025)+0
.6+0+4
300 to 500 kHz
3 + 0
.6
(3+0+0
.038)+0
.6+0+8
500 kHz to 1 MHz
5 + 2
(5+0+0
.06)+2+0+0
1to2MHz
10+5
(10+0
.1 +0
.0038)+5 +3 + 1
.5
1 V
10 to 20 Hz
0
.4+0
.02
(0
.4+0+0
.02)+0
.02+0+0
20 to 40 Hz
0
.15 + 0
.02
(0
.15+0+0
.0066)+0
.02+0+0
40 to 100 Hz
0
.06 + 0
.01
(0
.06+0+0
.0024)+0
.01 +0+0
100 Hz to 20 kHz
0
.02+0
.01
(0
.02+0+0
.0024)+0
.01 +0+0
20 to 50 kHz
0
.15 + 0
.04
(0
.15+0+0
.0046)+0
.04+0+0
50 to 100 kHz
0
.6 + 0
.08
(0
.6+0+0
.0071)+0
.08+0+0
100to250kHz
2+0
.5
(2+0+0
.016)+0
.5+0+0
250 to 300 kHz
3+0
.6
(3+0+0
.016)+0
.6+0+0
300 to 500 kHz
3 + 0
.6
(3+0+0
.026)+0
.6+0+0
500 kHz to l MHz
5 + 2
(5+0+0
.06)+2+0+0
1to2MHz
10+5
(10+0
.1+0
.0024)+5+3+0
7

T
.O
. 33K8-4-551-1
Table 1
. (Cont
.)
Test
Instrument (TI) Performance
Test
Characteristics
Specifications
Method
AC
Voltage Range
: 0
to 1000
VAC
Compared
to an
(Analog Mode) (Cont
.)
AC
Measurement
AC
BAND
<_2
MHz
Accuracy
:
Standard
or
Calibrator
Relative
:
±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI
% of rdg^2 + STD 1 % of rdg^2 +
STD2
% of
rdg^2)
+
TI
% of mg +
STD
1 µV + STD2 µV]
Range
Frequency
Relative
Absolute
10
V
10 to
20
Hz
0
.4 + 0
.02
(0
.4+0+0
.02)+0
.02+0+0
20
to
40
Hz
0
.15 + 0
.02
(0
.15 +0+0
.0067)+0
.02+0+0
40 to
100 Hz
0
.06+0
.01
(0
.06 + 0 + 0
.0027) + 0 .01 + 0 +
0
100
Hz to 20
kHz
0
.02 + 0
.01
(0
.02 + 0 + 0 .0027) + 0
.01 + 0 + 0
20to50kHz
0
.15+0
.04
(0
.15 +0+0
.0048)+0
.04+0+0
50 to 100 kHz
0
.6 + 0
.08
(0
.6+0+0
.0081)+0
.08+0+0
100 to 250
kHz
2 + 0
.5
(2+0+0
.019)+0
.5+0+0
250 to 300 kHz
3 + 0
.6
(3+0+0
.019)+0
.6+0+0
300to500kHz
3+0
.6
(3+0+0
.04)+0
.6+0+0
500
kHz to 1 MHz 5 + 2
(5+0+0
.06)+2+0+0
1
to2MHz
10+5
(10+0
.1+0
.0027)+5+3+0
*
2
100 V
10 to 20 Hz
0
.4 + 0
.02 (0
.4+0+0
.02)+0
.02+0+0
20
to 40
Hz
0
.15 + 0
.02 (0
.15+0+0
.0068)+0
.02+0+0
40
to 100 Hz
0
.06 + 0
.01
(0
.06+0+0
.0031)+0
.01
+0+0
100 Hz to
20
kHz
0
.03 +
0
.01
(0
.03+0+0
.0031)+0
.01+0+0
20to50kHz
0
.15+0
.04 (0
.15+0+0
.0069)+0
.04+0+0
50
to 100 kHz
0
.6 + 0
.08
(0
.6+0+0
.0098)+0
.08+0+0
100
to 250
kHz
2+0
.5
(2+0+0
.021)+0
.5+0+0
250
to
300 kHz
3 + 0
.6
(3+0+0
.021)+0
.6+0+0
300
to
500
kHz
3 + 0
.6
(3+0+0
.05)+0
.6+0+0
500
kHz
to
1 MHz 5 + 2
(5+0+0
.06)+2+0+0
*
2
1000
V
*3
lO
to 20 Hz
0
.42+0
.03
(0
.42+0+0
.02)+0
.03+0+0
*2
20 to 40
Hz
0
.17+0
.03
(0
.17+0+0
.0099)+0
.03+0+0
*
2
(<_700 V)
40 to
100 Hz
0
.08 + 0
.02
(0
.08+0+0
.0041)+0
.02+0+0
(>700 V)
40 to
100 Hz
0
.08 + 0
.02
(0
.08+0+0
.0038)+0
.02+0+0
(0
.06+0+0
.0041)+0
.02+0+0
(5700 V)
100 Hz
to 20
kHz
0
.06 + 0
.02
(>700 V)
100
Hz
to 20
kHz
0
.06 + 0
.02
(0
.06+0+0
.0038)+0
.02+0+0
20 to 50
kHz
0
.15+0
.04 (0
.15+0+0
.013)+0
.04+0+0
*
2
50 to 100 kHz
0
.6+0
.2
(0
.6+0+0
.05)+0
.2+0+0
See footnotes at end of
Table.
8

T
.O
. 33K8-4-551-1
Table 1
. (Cont
.)
Test Instrument (TI)
Characteristics Performance
Specifications
Test
Method
Range
: 0 to 1000 VAC
Accuracy:
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD1 % of rdg^2 +
STD2 % of rdg
A
2) + TI % of mg +
STD1
.tV + STD2 µV]
Range
Frequency
Relative
Absolute
10 mV
20 to 40 Hz
0
.5 + 0
.02
(0
.5 +0+0
.019)+0
.02+0+ 1
.3
40 Hz to 20 kHz
0
.5 + 0
.02
(0
.5 +0+0
.011)+0
.02+0+ 1
.3
20 to 50 kHz
0
.5 + 0
.02
(0
.5 +0+0
.021)+0
.02+0+2
50 to 100 kHz
0
.5 + 0.02 (0
.5 +0+0
.031)+0
.02+0+ 2
.5
100 to 300 kHz
4 + 0
.02
(4+0+0
.081)+0
.02+0+4
100 mV
20 to 40 Hz
0
.08 + 0
.002
(0
.08 + 0 + 0
.0085) + 0
.002 + 0 + 1
.5
40 Hz to 20 kHz
0
.08 + 0
.002
(0
.08 + 0 + 0
.0038) + 0
.002 + 0 + 1
.5
20 to 50 kHz
0
.08 + 0
.002
(0
.08 + 0 + 0
.0069) + 0
.002 + 0 + 2
50 to 100 kHz
0
.08 + 0
.002
(0
.08+0+0
.016)+0
.002+0+ 2
.5
100to300kHz
0
.3+0
.01
(0
.3 +0+0
.025)+0
.01 +0+4
300to500kHz
1+0
.01
(1+0+0
.038)+0
.01+0+8
500 kHz to 1 MHz
1 + 0
.01
(1+0+0
.06)+0
.01+0+0
1 to2MHz
1
.5+0
.01
(1
.5 +0.1 +0
.0038)+0
.01 +3 + 1
.5
1 V
20 to 40 Hz
0
.08 + 0
.002
(0
.08+0+0
.0066)+0
.002+0+0
40 Hz to 20 kHz
0
.08 + 0
.002
(0
.08+0+0
.0024)+0
.002+0+0
20 to 50 kHz
0
.08 + 0
.002
(0
.08+0+0
.0046)+0
.002+0+0
50 to 100 kHz
0
.08 + 0
.002
(0
.08+0+0
.071)+0
.002+0+0
100 to 300 kHz
0
.3 + 0
.01
(0
.3+0+0
.016)+0
.01 +0+0
300to500kHz
1+0
.01
(1+0+0
.026)+0
.01+0+0
500 kHz to 1 MHz
1 + 0
.01
(1+0+0
.06)+0
.01+0+0
1to2MHz
1
.5+0
.01
(1
.5 +0.1 +0
.0024)+0
.01 +3 +0
10V
20 to 40 Hz
0
.08 + 0
.002
(0
.08+0+0
.0067)+0
.002+0+0
40 Hz to 20 kHz
0
.08 + 0
.002
(0
.08+0+0
.0027)+0
.002+0+0
20 to 50 kHz
0
.08 + 0
.002
(0
.08+0+0
.0048)+0
.002+0+0
50 to 100 kHz
0
.08 + 0
.002
(0
.08+0+0
.081)+0
.002+0+0
100 to 300 kHz
0
.3+0
.01
(0
.3 +0+0
.019)+0
.01 +0+0
300 to 500 kHz
1 + 0
.01
(1+0+0
.04)+0
.01+0+0
500 kHz to 1 MHz
1 + 0
.01
(1+0+0
.06)+0
.01+0+0
*2
1 to2MHz
1
.5+0
.01
(1
.5+0
.1 +0.0027)+0
.01+3+0
See footnote at end of Table.
AC Voltage
(Random Sampled Mode)
AC BAND _<2 MHz
Implicitly calibrated
during Synchronous
Sub-sampled Mode
Calibration
9

T
.O
. 33K8-4-551-1
Table 1
. (Cons)
Test Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
AC Voltage
Range
: 0 to 1000 VAC
Implicitly calibrated
(Random Sampled Mode)
(Cont
.)
during Synchronous
AC BAND 52 MHz
Accuracy
:
Sub-sampled Mode
Relative : ±(% of rdg + % of mg)
Calibration
Absolute:
±[Sqrt(TI % of rdg^2 + STD 1 % of rdg^2 +
STD2 % of rdg
A
2) + TI % of mg +
STD1 µV + STD2 µV]
100 V
20 to 40 Hz
0
.12 + 0
.002
40 Hz to 20 kHz
0
.12 + 0
.002
20 to 50 kHz
0
.12 + 0
.002
50 to 100 kHz
0
.12 + 0
.002
100 to 300 kHz
0
.4 + 0
.01
300 to 500 kHz
1
.5 + 0
.01
500 kHz to 1 MHz
1
.5 + 0
.01
1000 V *
3
20 to 40 Hz
0
.3 + 0
.01
(5700 V)
40 Hz to 20 kHz
0
.3 + 0
.01
(>700 V)
40 Hz to 20 kHz
0
.3 + 0
.01
20 to 50 kHz
0
.3 + 0
.01
50 to 100 kHz
0
.3+0
.01
(0
.12+0+0
.0068)+0
.002+0+0
(0
.12+0+0
.0031)+0
.002+0+0
(0
.12+0+0
.0069)+0
.002+0+0
(0
.12+0+0
.0098)+0
.002+0+0
(0
.4+0+0
.021)+0
.01 +0+0
(1
.5+0+0
.05)+0
.01+0+0
(1
.5+0+0
.06)+0
.01+0+0
*
2
(0
.3+0+0
.0099)+0
.01+0+0
*
2
(0
.3+0+0
.0041)+0
.01 +0+0
(0
.3+0+0
.0038)+0
.01+0+0
(0
.3+0+0
.013)+0
.01+0+0
*
2
(0
.3+0+0
.05)+0
.01+0+0
*
2
AC BAND >2 MHz
Accuracy:
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD 1 % of rdg^2 +
STD2 % of rdg^2) + TI % of mg +
STD1
.tV + STD2
.tV]
Range
Frequency
Relative
Absolute
10 mV
20 Hz to 20 kHz
0
.1 + 0
.05
20 to 50 kHz
0
.1 + 0
.05
50 to 100 kHz
0
.1 +0
.05
100to300kHz
1
.2+0
.05
300 to 500 kHz
1
.2 + 0
.05
500 kHz to 1 MHz
1
.2 + 0
.05
l to2MHz
7+0
.07
2to4MHz
7+0
.07
4 to 8 MHz
20 + 0
.08
100 mV
20 Hz to 20 kHz
0
.1 + 0
.05
20 to 50 kHz
0
.1 + 0
.05
50 to 100 kHz
0
.1 + 0
.05
100to300kHz
2+0
.05
300 to 500 kHz
2 + 0
.05
500 kHz to 1 MHz 2+0
.05
1 to 2 MHz
4 + 0
.07
2 to 4 MHz
4 + 0
.07
4to8MHz
4+0
.08
8to 10 MHz
15+0
.1
See footnotes at end of Table
.
(0
.1 +0+0
.011)+0
.05 +0+ 1
.3
(0
.1 + 0+0
.021)+ 0
.05 +0+ 2
(0
.1 +0+0
.031)+0
.05 +0+2
.5
(1
.2+0+0
.081)+0
.05+0+4
(1
.2+0+0
.089)+0
.05+0+8
(1
.2+0+0
.17)+0
.05+0+8
(7+0
.1 +0
.011)+0
.07+3+ 1
.3
(7+0
.2+0
.011)+0
.07+3+ 1
.3
(20+0
.2 +0
.011)+0
.08+3 + 1
.3
(0
.1 +0+0
.0038)+0
.05 +0+ 1
.5
(0
.1 +0+0
.0069)+0
.05+0+2
(0
.1 +0+0
.016)+0
.05+0+2
.5
(2+0+0
.025)+0
.05+0+4
(2+0+0
.038)+0
.05+0+8
(2+0+0
.06)+0
.05+0+0
(4+0
.1 +0
.0038)+0
.07+3 + 1
.5
(4 + 0 .2 + 0
.0038) + 0 .07 + 3 + 1
.5
(4+0.2+0
.0038)+0
.08 +3 + 1
.5
(15+0
.2+0
.0038)+0
.1 +3+ 1
.5
10

T
.O
. 33K8-4-551-1
Table 1
. (Cons)
Test
Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
AC
Voltage
Range
: 0 to
1000
VAC
Implicitly calibrated
(Random
Sampled Mode)
(Cont
.)
during
Synchronous
AC
BAND
>2
MHz
Accuracy
:
Sub-sampled
Mode
Relative
:
±(% of
rdg
+ % of mg)
Calibration
Absolute:
±[Sgrt(TI
% of rdg^2 +
STD 1
% of rdg^2 +
STD2
% of rdg^2) +
TI
% of mg +
STD1
+
STD2
µV]
1
V
20
Hz
to 20
kHz
0
.1
+ 0
.05
20 to 50 kHz
0
.1 + 0
.05
50 to
100 kHz
0
.1 + 0
.05
100to300kHz
2+0
.05
300 to 500
kHz
2 + 0
.05
500
kHz
to 1
MHz
2 + 0
.05
1
to 2
MHz
4
+ 0
.07
2to4MHz
4+0
.07
4to8MHz
4+0
.08
8tol0MHz
15+0
.1
10
V
20
Hz to
20
kHz
0
.1 + 0
.05
20 to 50
kHz
0
.1
+ 0
.05
50 to
100 kHz
0
.1
+ 0
.05
100to300kHz
2+0
.05
300 to 500
kHz
2 + 0
.05
500
kHz
to
1 MHz
2 + 0
.05
1 to 2
MHz
4
+ 0
.07
2to4MHz
4+0
.07
4 to
8
MHz
4
+ 0
.08
8to 10 MHz
15+0
.1
100
V
20 to 40
Hz
0
.12
+ 0.002
40
Hz to
20
kHz
0
.12
+ 0.002
20to50kHz
0
.12+0
.002
50 to 100 kHz
0
.12 + 0
.002
1000 V
*
3
20 to 40
Hz
0
.3 +
0 .01
(<_220 V)
40
Hz to
20
kHz
0
.3 + 0
.01
(<
_700 V)
40
Hz
to 20
kHz
0
.3 +
0
.01
(>700 V)
40
Hz to
20
kHz
0
.3 +
0 .01
20
to50kHz
0
.3 +
0 .01
50 to
100 kHz
0
.3+0
.01
Frequency
Range
: 0 to
10 MHz,
VAC;
0
to 100 kHz,
AAC
(0
.1
+0+0
.0024)+0
.05+0+0
(0
.1
+0+0
.0046)+0
.05+0+0
(0
.1 +0+0
.0071)+0
.05+0+0
(2+0+0
.016)+0
.05+0+0
(2+0+0
.026)+0
.05+0+0
(2+0+0
.06)+0
.05+0+0
(4+
0
.1 +0.0024)+0
.07+3+0
(4+0
.2+0
.0024)+0
.07+3+0
(4+0.2+0
.0024)+0
.08+3+0
(15 +0
.2+0
.0024)+0
.1
+3 +0
(0
.1
+0+0
.0027)+0
.05 +0+0
(0
.1 +0+0
.0048)+0
.05 +0+0
(0
.1 +0+0
.0081)+0
.05
+0+0
(2+0+0
.019)+0
.05+0+0
(2+0+0
.04)+0
.05+0+0
(2+0+0
.06)+0
.05+0+0
(4+0
.1
+0
.0027)+0
.07 +3 +0
(4+0
.2+0
.0027)+0
.07+3+0
(4+0
.2+0
.0027)+0
.08+3+0
(15
+0
.2
+0
.0027)+0
.1
+3 +0
(0
.12+0+0
.0068)+0
.002+0+0
(0
.12+0+0
.0031)+0
.002+0+0
(0
.12
+ 0 + 0
.0069) + 0
.002 + 0 + 0
(0
.12+0+0
.0098)+0
.002+0+0
(0
.3+0+0
.0099)+0
.01+0+0
(0
.3+0+0
.0031)+0
.01
+0+0
(0
.3+0+0
.0041)+0
.01
+0+0
(0
.3+0+0
.0038)+0
.01+0+0
(0
.3+0+0
.013)+0
.01+0+0
*
2
(0
.3+0+0
.05)+0
.01+0+0
*
2
Compared
to a
Synthesized Signal
Generator
*2
*2
*2
*2
Accuracy
:
1
to 40
Hz,
±0
.05% of rdg;
40
Hz
to 10
MHz, ±0
.01%
of
rdg
See footnotes at end of Table
.
11

T
.O
. 33 K8-4-551-1
Table 1
. (Cont
.)
Test Instrument (TI)
Characteristics
Performance
Test
Specifications
Method
AC Current
Range
: 0 to 1 AAC
Accuracy:
Relative: ±(% of rdg + % of mg)
Absolute
: ±[Sgrt(TI % of rdg^2 +
STD 1 % of rdg^2 + STD2 % of rdg^2 +
STD3 % of rdg
A
2) + TI % of rng]
Implicitly calibrated
during DC Current and
AC Voltage Calibration
Range
Frequency
100 µA 10 to 20 Hz
20 to 40 Hz
40 to 45 Hz
45 to 100 Hz
100 Hz to 1 kHz
1 mAto
10to20Hz
100 mA
20 to 40 Hz
40 to 45 Hz
45 to 100 Hz
100 Hz to 5 kHz
5 to 20 kHz *
20 to 50 kHz
*
1
50 to 100 kHz *
1 A
10 to 20 Hz
20 to 40 Hz
40 to 45 Hz
45 to 100 Hz
100 Hz to 5 kHz
5to20kHz
*
I
20 to 50 kHz *
Relative
Absolute
0.4 + 0
.03
(0
.4 + 0
.02 + 0 .0005277 + 0 .0005) + 0
.03
0
.15 + 0
.03
(0
.15 + 0
.0066 + 0
.0005277 + 0
.0005) + 0
.03
0
.15 + 0
.03
(0
.15 + 0
.0024 + 0.0005277 + 0
.0005) + 0
.03
0
.06 + 0
.03
(0
.06 + 0
.0024 + 0.0005277 + 0
.0005) + 0
.03
0
.06 + 0
.03
(0
.06 + 0
.0024 + 0
.0005277 + 0 .0005) + 0
.03
0
.4 + 0
.03
(0
.4 + 0
.02 + 0
.0005277 + 0
.0005) + 0
.03
0
.15 + 0
.03
(0
.15 + 0
.0066 + 0
.0005277 + 0 .0005) + 0
.03
0
.15 + 0
.03
(0
.15 + 0
.0024 + 0
.0005277 + 0 .0005) + 0
.03
0
.06 + 0
.03
(0
.06 + 0
.0024 + 0
.0005277 + 0 .0005) + 0
.03
0
.06 + 0
.03
(0
.06 + 0
.0024 + 0
.0005277 + 0
.0005) + 0
.03
0
.06 + 0
.02
(0
.06 + 0
.0024 + 0
.0005277 + 0 .0005) + 0
.02
0
.4 + 0
.04
(0.4 + 0
.0046 + 0
.0005277 + 0
.0005) + 0
.04
0
.55 + 0
.15
(0
.55 + 0
.0071 + 0
.0005277 + 0
.0005) + 0
.15
0
.4 + 0
.02
(0.4 + 0
.02 + 0
.0005277 + 0
.0006) + 0
.02
0
.16 + 0
.02
(0
.16 + 0
.0066 + 0
.0005277 + 0
.0006) + 0
.02
0
.16 + 0
.02
(0
.16 + 0
.0024 + 0
.0005277 + 0
.0006) + 0
.02
0
.08 + 0
.02
(0
.08 + 0
.0024 + 0
.0005277 + 0
.0006) + 0
.02
0
.1 + 0
.02
(0.1 + 0
.0024 + 0.0005277 + 0
.0006) + 0
.02
0
.3 + 0
.02
(0
.3 + 0
.0024 + 0
.0005277 + 0 .0006) + 0
.02
1 + 0
.04
(1 + 0
.0046 + 0
.0005277 + 0
.0006) + 0
.04
*
1
*2
*
3
Typical accuracies, will not be calibrated.
Calibration of the complete Voltage and/or Frequency ranges at these points are not possible due to the
capability limitations of the Calibration Standards . The TI does not require a limited certification for these
Standards limitations.
Option HO1 1000 V rms max (All other units 700 V rms max).
12

T
.O
. 33K8-4-551-1
2 EQUIPMENT REQUIREMENTS:
Noun
Minimum Use
Specifications Calibration
Equipment
Sub
Item
2
.1
DC REFERENCE
Output
: 10 VDC
Fluke
STANDARD 734A/AF
2 .2
DC REFERENCE
Accuracy
: ±1
.5 ppm of output
*
I
.
*
2
FlukeRatio
: 10
:1, ±0 .2 ppm
;
*I
DIVIDER
*
I
100
:1, ±0
.5 ppm
752A
2 .3
SHORTING BARS Range
: N/A
ESI
(2 EA)
SB 103
Accuracy
: N/A
Range
: 0 to I VDC
Hewlett-Packard
H-P
2
.4
DIGITAL
VOLTMETER
3458A OPT 002
3458A
Accuracy:
±(4
.277 ppm of rdg + 1 ppm of mg) *
3
for 3458A OPT 002;
±(8
.142 ppm of rdg + 1 ppm of mg) *
4
for 3458A
2.5
2 .6
CALIBRATOR
AC MEASUREMENT
STANDARD
Range
: 0 to 1000 VDC and VAC,
20 Hz to 10 MHz
; 1 VAC at 3, 7
and 10 MHz
; 0 to 1 ADC
Accuracy : VDC and ADC, N/A;
1 VAC at 3, 7 and 10 MHz,
±(0
.2% of output + 3 µV)
Range
: 0 to 1000 VAC
Accuracy
: ±(ppm of rdg + µV)
Fluke
5700A
W/5725A
Fluke
5790A
Range
Frequency
±(ppm of rdg + ]AV)
22 mV
20 to 40 Hz
±(190
+ 1
.3) (TAR 2
.5
:1)
40 Hz to 20 kHz
±(110
+ 1
.3) (TAR 2
:1)
20 to 50 kHz
±(210
+ 2) (TAR 3
.3
:1)
50 to 100 kHz
±(310
+ 2
.5)
100 to 300 kHz
±(810
+ 4)
See footnotes at end of Equipment Requirements
.
13

T
.O
. 33K8-4-551-1
Noun
Minimum Use
S ' ecifications
Calibration
Eiui'ment
Sub
Item
2
.6 AC MEASUREMENT
Range
: 0 to 1000 VAC
STANDARD
(Cont
.)
Accuracy
: ±(ppm of rdg +µV)
Fluke
5790A
Range
Frequency
220mV 20to40Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
100 to 300 kHz
300 to 500 kHz
500 kHz to 1 MHz
2
.2 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
100 to 300 kHz
300 to 500 kHz
500 kHz to 1 MHz
22 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100kHz
100 to 300 kHz
300 to 500 kHz
500 kHz to 1 MHz
220 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
±(ppm of rdg + µV)
±(85 + 1 .5) (TAR 1
.7
:1)
±
(38 + 1
.5) (TAR 2
.2
:1)
±(69 + 2) (TAR 3
.9
:1)
±
(160+2
.5)
±(250 + 4)
±(380 + 8)
±
600
±66 (TAR 2
.1
:1)
±
24 (TAR 3
.9
:1)
±
46
±71
±160
±
260
±600
±
67 (TAR 2
:1)
±27 (TAR 3
.5
:1)
±
48
±81
±190
±
69
±
98
14

T
.O
. 33K8-4-551-1
Minimum Use
Noun
Specifications
Calibration
Equipment
Sub
Item
2
.6 AC MEASUREMENT
Range
: 0 to 1000 VAC
Fluke
STANDARD
(Cont
.)
5790A
Accuracy
: ±(ppm of rdg + µV)
Range
Frequency
700 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
1000V 20to50kHz
Range
: N/A
Accuracy
: N/A
Range
: 1 S2 to 10 MS
-
2
RESISTORS
Accuracy
: AFPSL Certified
±2
.15 ppm, 1 S2 through 1 kS2;
±2
.1 ppm, 10 kS2;
±2
.4 ppm, 100 kS2;
±
6 ppm, 1 MS2, (TAR 3 .0
:1);
±
12 ppm, 10 M52
2
.9 DMM SWITCH
Range
: N/A
Accuracy
: N/A
2
.10 RESISTANCE
Range
: 10 S2 to 1 S2/step
ESI
STANDARD
SR 1060
Accuracy
: Charted value *5
10 S2/step,
±6
ppm;
100 S2/step,
±5
ppm,
(TAR 3
.6
:1 @ 1 kS2
w/3458A OPT 002),
*3
(TAR 2
.7
:1 @ 1 k)2 w/3458A)
;
*
4
1 1d2/step,
±5
ppm,
(TAR 3
.6
:1 @ 100 S2
w/3458A OPT 002),
*
3
(TAR 2
.7:1 @ 100 S2 w/3458A),
*4
(TAR 3
.0
:1 @ 10 Id-2 w/3458A)
*
4
2
.11 RESISTANCE
Range
: 10 MCI/step
STANDARD
Accuracy
: 100 MS
-
2, -1-50 ppm
See footnotes at end of Equipment Requirements.
2
.7 RESISTANCE
MEASUREMENT
SYSTEM
2
.8 STANDARD
±(ppm of rdg + uV)
±
99
±
41
±
130
±500
±
130
Guildline
6625AF
ESI
SR1050
1
Measurements
International
4310AF
Guildline
66252
15

T
.O
. 33K8-4-551-1
Noun
Minimum Use
Specifications Calibration
Sub
Equipment
Item
2
.12 NULL DETECTOR
2
.13 DC VOLTAGE
DIVIDER
2
.14 SYNTHESIZED
SIGNAL GENERATOR
2
.15 PARALLEL
COMPENSATION
NETWORK
2
.16 TERMINATION
I
2
.17 NANO VOLT/
MICRO OHM METER
Range
: 0 to 1000 VDC
Accuracy
: N/A
Range
: 0 to
.8000000
Terminal Linearity
: ±0
.2 ppm
*2
Range
: 2 Hz to 10 MHz
Accuracy
: ±0
.0025% of setting
Range
: N/A
Accuracy
: N/A
Range
: 50 Q
Accuracy
: NIA
Range
: 0 VDC
; 0 Cl
Accuracy
: N/A VDC
;
±2
ppm S2
Fluke
845AB
Fluke
720A
Hewlett-Packard
3325B
ESI
PC 101
As Available
Agilent
34420A
*' The following TAR is calculated using the RSS accuracy of the DC Reference Standard and DC Reference
Divider
. For high stability TIs, 1 (TAR 3
.5
:1) and 10 VDC range (TAR 3
:1).
*2
The following TAR is calculated using the RSS accuracy of the DC Reference Standard and DC Voltage
Divider
. For high stability TIs, 10 VDC range (TAR 3
:1).
*5
The following TAR is calculated using the RSS accuracy of the Digital Voltmeter and Resistance
Standard
. For TI 1 mA and 10 mA DC ranges (TAR 3
.6
:1).
*
4
The following TAR is calculated using the RSS accuracy of the Digital Voltmeter and Resistance
Standard
. For Tl 100 µA DC range (TAR 3
:1), 1 mA and 10 mA DC ranges (TAR 2
.7
:1).
*
5
See step 3
.9.
3 PRELIMINARY OPERATIONS:
3
.1 Review and become familiar with the entire procedure before beginning the Calibration Process.
WARNING
Unless otherwise designated, and prior to beginning the Calibration Process,
ensure that all test equipment voltage and/or current outputs are set to zero (0) or
turned off, where applicable
. Ensure that all equipment switches are set to the
proper position before making connections or applying power.
3
.2 Use low-thermal, shielded leads with spade-lug connectors whenever possible during the Calibration Process.
16

T
.O
. 33K8-4-551-1
3
.3 If the 66252 DMM Switch is not available, disconnect the Automatic Resistance Bridge (P/O 2
.7) rear panel Rs
and Rx connections (terminals P1, P2, Cl and C2) . Connect the Rs connections (P1, P2, Cl and C2) to the Rx
connections (P 1, P2, Cl and C2), respectively. Ensure connections are restored to original configuration upon
completion of the Calibration Process.
3
.4 Connect all test equipment to appropriate power sources
. Set all POWER switches to ON and allow test equipment
warm-up period per manufacturer requirements.
3
.5 Ensure that AC power is applied to the temperature stabilized Standard Resistors
. Manufacturers suggested
minimum warm-up time is 24 hours
.
NOTE
If at any time power is removed from the Standard Resistors for over 1 minute,
there is a minimum 1 hour warm-up time for the oven to stabilize.
3
.6 Connect the Nano Volt/Micro Ohm Meter to the temperature stabilized Standard Resistors temperature monitor
terminals and turn TEMP DEV switch to ON
. Using TEMP DEV dial to obtain a VDC null reading on the Nano
Volt/Micro Ohm Meter, record temperature offset of the TEMP DEV dial before and after calibration
. TEMP DEV
dial before and after values should be between -20 m°C to +20 m°C (i
.e
., on scale).
3
.7 Connect TI to appropriate power source . Set ON/OFF switch to ON and allow a 4 hour warm-up period.
3
.8 Use only that portion of the Calibration Process that pertains to the TI being calibrated.
3
.9 Calibrate Resistance Standard (2
.10), required to perform the calibration, IAW T
.O
. 33K8-4-1-1 before use
. See
T
.O
. 33K8-4-1-1 for clarification of (at time of calibration)
. Verify accuracies at time of calibration in
T
.O
. 33K8-4-1-1 still meet the requirements.
3
.10 Perform the DC Reference Divider Self-Calibration IAW the Commercial Data prior to use.
3
.11 Perform the DC Voltage Divider Self-Calibration IAW the MTO prior to use.
3
.12 Verify all TI inputs (including rear input) are disconnected and execute the TEST function
. The TI display must
read SELF TEST PASSED.
3
.13 Ensure the TI is in the FULL MENU mode by pressing:
3
.14 Check the TI Internal Temperature by pressing the TI keys in the following sequence:
Menu
Enter
p
in
.
)
431 e)
v
4iHF11
4310
T
ESNer
(record this value).
3
.15 Press the TI keys in the following sequence to obtain CAL? 58:
. The TI display must be within ±5 °C of the value recorded in
step 3
.14.
3
.16 Repeat step 3
.15 for CAL? 59 and CAL? 60
.
NOTE
If TI displays greater than 5 °C difference from the internal temperature
recorded in step 3
.14, CAL 0, CAL 10 and CAL 10 k should be performed
IAW the Manufacturers Calibration Manual.
a
5
C
6HF1,
431me)
V
Change 1 17

T
.O
. 33K8-4-551-1
3
.17 Perform ACAL on the TI before beginning calibration by pressing the TI keys in the following sequence:
until ACAL ALL is displayed, then
Ganma)
3
9
5
8
EntEr
NOTE
(,),
3458 is the Security Code for the TI when shipped from the manufacturer
. If
the Security Code has been changed, that Security Code will be entered in its
place
. If the TI has been unsecured, the Security Code will not be entered.
Refer to the TI Calibration Manual, Chapter 1, for more information.
3
.18 Set the TI for DCV, 8 digit display and NPLC at 100 by pressing the TI keys in the following sequence:
, use the menu scroll keys
or
ocv
6H£1)
Elie)
N
8
Enter
NPLC
1
0
0
Enter
3
.19 Connect a low thermal short (12 to 14 Gauge solid copper wire) across the Tl Input and SI Sense HI and LO
terminals, allow 5 minutes for stabilization.
3
.20 Set the TI to each range listed in the Range column of Table 2.
3
.21 The TI should indicate within the corresponding values listed in the Limits column of Table 2.
Table
2.
Range (VDC)
Limits (VDC)
100 m
0 ±000 .00106 m
0 ±0
.00000106
10
0 ±00
.0000023
100
0 ±000 .000036
1000
0 ±0000
.00010
3
.22 If the TI DC Voltage zero is >90% of the value listed in the Limits column of Table 2 on any range, enter a
K action in the Maintenance Data Collection System
. If the TI DC Voltage zero is <_90% of the value listed in the
Limits column of Table 2 on all ranges, performing the following TI CAL 0 steps is not a K action.
NOTE
The TI zero offset adjustment (CAL 0) should be performed individually for
both front and rear terminals
. Once CAL 0 has started, move away from
the TI very slowly until the TI is done (about 5 minutes)
. If drafts exist, it may
be necessary to cover the TI INPUT terminals.
3
.23 Set the TI to execute CAL 0 by pressing the TI keys in the following sequence:
(s-l
q
(Eke)
C
0
(fie)
f
345 8
18
Other manuals for 3458A
2
Table of contents
Other HP Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PDK
PDK PDR100 user manual

Brooks Instrument
Brooks Instrument MT3809G Installation and operation manual

Digital Watchdog
Digital Watchdog Blackjack Ai DW-BJAiARA2 quick start guide

Laser 2000
Laser 2000 ADM-1000 operating instructions

Komshine
Komshine KLS-25m instruction manual

Velleman
Velleman INDC003 user manual