HP 412A Service manual

I.
:
HP
Archive
..
.
This vintage Hewlett Packarddocument was
preserved and distributed by
www.
hparchive.com
'
Please visit us on the web
!
..
On-line curator: Tony Gerbic

I
i
i
(r,
,'
a
i
f
OPERATING
AND
SERVICE
MANUAL
(HP
PART
NO.
00412-90003)
MODEL
412A/AR
DC
VACUUM
TUSE
VOLTMETER
SERIALS
PREFIXED: 301-, 3160, 424-,
649-
Copyright Hewlett-Packard
Company
1961
P.
0.
Box
301,
Loveland,
Colorado,
80537
U.
S.A.
1
i
OOT29-5
Printod:
NOV
1968
I

0
Model
4
12A
h
TABLE
OF
CONTENTS
Section Page
I
GENERAL INFORMATION
.................
1-1
1.1
.
Introduction
.........................
1-1
1.7
.
Specifications
........................
I
-I
1.Y
.
instrument and Manual Identification
.....
1-1
.
Section
I
.
?
Page
11
INSTALLATlON
..........................
2-1
2.1
.
Introduction
.........................
2-1
2.3
.
Initial inspection
.....................
2-1
2.5
.
Power Requirements
...................
2-1
2.7
.
Grounding Requirements
...............
2-1
2.10
.
lnstallatiori
..........................
2-1
2.13
.
Repackaging
for
Shipment
..............
2-1
i
i
1
Section Page
I11
OPERATING INSTRUCTIONS
...............
3-1
3.1
.
Introduction
.........................
3-1
3.3
.
Low-Level Electrical Phenomena
.........
3-1
3.7
.
Turn& Procedure
....................
3-1
3-8
.
Voltage Measurement
..................
3-1
3.9
.
Current Measurement
..................
3-1
3-1
0
.
Resistance Measurement
................
3-2
3-12
.
Amplifier Operation
...................
3-3
Y
Section Page
Table
of
Contents
List
of
Tables
Section page
5.1
.
lntroduction
.........................
5-1
5.3
.
Required Test Equipment
...............
5-1
5.5
.
Performance Checks
...................
5-1
5.7
.
Meter Zero
.........................
5-1
5.10
.
Voltmeter Accuracy Check
.............
5-1
5-1
1
.
Milliammeter Accuracy Clieck
..........
5-1
5-1
2
.
Ohmmeter Accuracy Check
............
5-2
5.13
.
Input KcsistanceClieck
...............
5-3
5.14
.
AC Kejection Check
..................
53
5.15
.
Adjustment and Calibration Proccdure
......
53
5.17
.
Cabinet Removal
....................
53
5.18 .
Mechanical Adjustment of Meter Zero
....
5-3
5.20
.
Hum Balance Adjustment
..............
5-5
5.21
.
Cathode
Follower
Bias Adjustrncnt
......
5-5
5.22
.
Meter Electrical Zero Adjustment
........
5-5
5.23
.
5.24
.
V
MAINTENANCE
..........................
5-1
Amplifier Gain Adjustment and
Meter Calibration
...................
5-5
Olimmeter Adjustment
................
5-5
5.25
.
Troubleshooting
......................
54
5.27
.
Front Panel Indications
...............
54
5.29
.
Power Supply andAmplifier Checks
......
56
5.33
.
Modulator Clieck
....................
54
5.34
.
Modulator Replacement
...............
5-7
5.36
.
Demodulator Check
..................
5-7
5.37
.
Demodulator Replacement
.............
54
5.39
.
Switch Checks
......................
5-8
541
.
Adjustment
of
Factory Selcctcd
Components
.......................
54
546
.
Input Cable Keplacemcnt
..............
5-0
Section Pap
VI
REPLACEABLEPARTS
....................
6-1
IV
THEORY
OF
OPERATION
..................
4-1
6.1
.
lntroduction
.........................
6-1
4.1 .
General
.............................
4-1
64
.
Ordering Information
..................
6-1
4.3
.
Block Diagram Analysis
................
4-1
6.6.
Non-Listed Parts
......................
6-1
Section l’ilgc
4.9
.
Modulator. Amplifier. Demodulator.
VI1
CIRCUITDIAGRAMS
......................
7-1
...........
4.5.
Range and Function Selection 4-1
4.7
.
Filters
..............................
4-1
and Chopper
........................
4-1
7.1
.
Introduction
.........................
7-1
4-1
1
.
Cathode Follower and Polarity Selection
...
4-1
LIST
OF
TABLES
Page Number
Pagc
Number
1.1
.
Specifications
.............................
1.0
3.1
.
Voltage and Current in Resistance
Measurements
............................
3-2
5.1
.
Test Equipment Required
....................
54
5.2
.
Millianimeter Accuracy Check
................
5-2
5.3
.
Input Resistance Chccks
.....................
5-3
5.5
.
Tubc Replacement
.........................
541
6-1
.
Rcplaccablc Parts
..........................
6-2
54
.
Power Supply Voltages
.....................
:54
~
..
iii

List
of
Illustrations
n
n
Number
1-1.
Model
412A
DC
Vacuum Tube
Voltmeter
.
.
.
.
.
..
..
.
..
.
.
..
.
4-1.
Block Diagram
..
.
.
.
..
..
...
.
.
5-1.
Input Resistance Check
.
..
.
...
.
5-2.
AC
RejectionCheck
.
....
.
.
.
.
.
53.
Location
of
Adjustments
.
..
..
.
54.
Input Circuit Board
.
.
......
.
.
.
5-5.
ModulatorWaveform
...
.
...
.
.
54.
DemodulatorAssembly
....
....
5-7.
input Cable Connections.
.
.
...
.
5-8.
Simplified Schematic,Voltmeter
Switching
....
.
.
.
.
..
.
.
.
.
.
..
5-9.
Simplified Schematic,Ammeter
Switching
..
*
.
.
.
..
.
.
.
.
,
..
.
.
.1
t,p’
8
LIST
...........
...........
...........
...........
...........
...........
...........
...........
...........
...........
.....e.....
’
OF
ILLUSTRATIONS
Page Number
,
.
,
1-1
,
.
,
4-1
.
.
.
5-2
..
.
53
...
54
...
56
.
..
5-7
...
s-8
...
59
.
.5-10
.
.
.5-11
5-1O.Simplified Schematic,Ohmmeter
Switching
.
...
.
...
.
..
...
.
....
6-1.
Location
of
Mechanical Parts,
Cabinet Model
.
.
...
.
.
.
..
.
.
.
.
.
.
6-2.
Location
of
Mechanical Parts,
Rack Model
.
.
.
.
...
.
.
.
.
....
.
.
.
6-3.
Voltage
Probe
Assembly
.
.
.
..
.
.
.
.
7-1.
Location of Chassis Mounted.Parts,
Cabinet Model
....
.
...
.
.
.
.
.
.
..
7-2.
Location
of
Chassis Mounted
Parts,
Rack Model
.
.
.
.
.
.
..
.
.
.
.
.
.
.
...
7-3.
Location
of
Switch
Components
.
..
74.
SchematicDiagram
.
.
.
.
.
.
.
.
.
.
..
.
7-5
Power
Supply Backdating
.
....
.
..
Model
4
12A
..........
..........
..........
..........
..........
e.........
..........
..........
..........
Page
.
.5-I
2
..
6-5
.
.
6.6
..6-7
.
.
7-2
..
7-3
..
74
.
.
7-5
.
.
7-7

b
'5
Section
I
Model
4
12A
Table
1-1.
Specifications
VOLTMETER
Voltage Range: Positive and negative voltages from
1
millivolt
fuull
scale to
1000
volts full scale in
13
9
ranges
Accuracy
:
+
1
%
of full scale onany range
Input Resistance:
10
megohms
1%
on
1
mV,
3
mV,
b
and
10
mV ranges
30
megohms
21%
on
30
mVrange
100
megohmsf.l%
on
100
mV range
200
megohms
t
1
%
on
300
mV range and above
AC Rejection: A voltage atpower line
or
twice power
line frequency
40
dB
greater
than full scale affects
reading
less
than
1%.
Peak voltage must not exceed
1500
volts.
AMMETER
Current Range: Positive and negative currentsfrom
1
microampere full scale to 1 ampere full scale in
thirteen ranges
Accuracy
:
2
2%offull scale on any range
v
tnput Resistance
:
tnternal Shunt Full
Scale
Range Resistance* Voltage
Drop
.001
mA
.003
mA
.01
mA
.03
mA
.I
mA
.3
mA
1
mA
3
mA
10
mA
30
mA
100
mA
300
mA
1000
mA
d
I000
ohms
3
I6
ohms
100
ohms
3
I
.6
ohms
IO
ohms
3.16
ohms
1
ohm
.3
16
ohm
.1
ohm
.I
olim
.I
olim
.I
ohin
.I
ohm
1
mV
0.9486
mV
1
mV
0.9486
mV
1mV
0.9486
mV
1
mV
0.9486
mV
1
mV
3
mV
10
mV
30
mV
100
mV
*For
total insertion resistance add
0.07
ohms copper lead
rcsistance at
25
C
OHMMETER
Hesistaticc Ibiige: Rcsistancc from
I
olim
center-scale to
IO0
niegolims center-scale in nine
decade
raiiges
a
Accuracy:
f-
5%
of
reading froni
0.3
to
3
on Ohms
scale
'I
14
Voltages and Currents:
Open Circuit ShortCircuit
Current
Range
Volts
XI
10
mV
10
mA
XI0
100
mV
10
mA
XI00
IV
10
mA
XlOOo
1v 1
mA
XlOK
1v
100~
A
XlOOK
1v
lop
A
XIM
XlOM
1v
.lp
A
Xl00M
1v
.Olp
A
1v
'P
A
AMPLIFIER
Voltage Cain:
1000
maximum
AC
Rejection:
3
dB at
1
Hz,
80
dBat
50
and
60
Hz
Output: Proportional
to
meter indication;
1
volt at
full scale; maximum current,
1
mA.
(Full
scale
corresponds to
1
.O
on upper scale.)
Output Impedance:
Less
than
2
ohmsatdc
Noise:
Less
than
5
mV
RMS
on
any range
Drift: negligible
GENERAL
Isolation Resistance: At least
100
megohms shunted
by 0.1~
F
between common terminal and case
(power line) ground
Common Mode Rejection: May be operated up to
500
Vdc
or
130
Vac from ground
Power: 115/230V
f10%,
5040
Hz,
35
W
Dimensions: Cabinet Mount:
11-1/2
in. high,
7-1/2
in. wide,
10
in. deep
Rack Mount:
001
29-5

Model
412A
1-1.
INTRODUCTION.
Section
I
SECTION
I
GENERAL
INFORMATION
1-2.
The Hewlett-Packard Model
412A/AR
DC
Vacuum
Tube Voltmeter is
a
multifunction instrument which
measures the entire range
of
dc voltages, current, and
resistance normally encountered in electronic equipment.
1-3.
The Model
412AIAR
measures dc voltage from
1.0
mV full scale to
IO00
V full scale in thirteen ranges
arranged in a
1,3,
10
sequence. Voltmeter accuracy is
1%
of
full scale. The input circuit
is
isolated from the case and
from
power line ground.
14.
The Model
412AlAR
measures positive
or
negativedc
current from
1.OpA
full scale
to
1.0
A
full scale with an
accuracy
of
2%
of
full scale. Current ranges are also in a
1,
3,
10
sequence.
1-5.
Resistance measurement capability
of
the Model
412AIAR
is
from
1.0
ohm center scale
to
I00
megohms
center scale in nine decade rangcs. Resistaiiccnieasureiiieiit
accuracy
is
5%
between
.3
and
3
on
the resistance scale.
16.
The Model
412AIAR
has
a
dc aiiiylifier output
voltage proportional to the input. Ainplificr giiiil is froiii
-1000
to
t1000,
depending on range selected.
1-7.
SPEC1
FlCATlONS.
1-8.
Complete specifications
for
the dip- Modcl41
2AIAR
are listed in Table
1-1.
1-9.
INSTRUMENT AND MANUAL IDENTIFICATION.
1-10.
Hewlett-Packard uses
a
twoscctioii cigltbdigit scrial
number
(OOO-00000).
If
the first tlircc digits
of
tlic
scrial
number
of
your instrumcnt do not agrcc witti thosc
on
tlic
title page of this manual, changc sliects supplicd
with
tlic
manual
will
explain the diffcrenccs bctwccii
your
instrument and the Model
412AIAll
dcscribcd
in
lliis
manual.
'
412A-A-30188
Figure
1.1.
Model
412A
DC
Vacuum Tube Voltmctcr
00
1
295

0
h
cy
P
Y
4b
4
J
Model
4
12A
2-1.
INTRODUCTION.
SECTION
II
IN
STACLATIO
N
2-2. Tliis section contains instructions necessary for
installation
of
the Model 412AIAR
DC
Vacuum Tube
Voltmeter. Included are initial inspection procedures,
power and grounding requirements, installation
information, and instructions for repackaging for shipment.
2-3.
INITIAL INSPECTION.
2-4.
This
instrument was carefully inspected both
mechanically
and
electrically before shipment. It should be
free
of
mars or scratches, and in perfect electrical order
upon receipt.
To
confirm this, the instrument should be
inspected for physical damage
in
transit.
Also
test the
clcctrical
performance
of
tlic instrument, using the
procedure outlined in Paragraph
5-5.
If
there is damage or
deficiency, see tlic warranty on the inside front cover of
this manual.
2.5.
POWER
REQUIREMENTS.
2-6.
The Model 412AIAR can be operated from any
source
of
115
or
230
volts at
50
or
60
Hz.
The
115/230
V
slide switch
on
the rear panel selects the desired voltage.
Power dissipationis approximately
35
watts.
2-7.
G
R
0
U
N
D
I
NG
REQUl
R
EMENTS.
2-8.
To protect operating personnel, the National
Electrical Manufacturer’s Association
(NEMA)
recommends
that tltc instrument panel and cabinet
be
grounded. This
instrument is equipped with a threeconductor power cable
wluch, when plugged into an appropriate receptacle.
grounds the instrument. The offset pin
on
the connector is
tlie ground wire.
2-9.
To
preserve the protection feature when operating the
instrument from a twocontact outlet, use a threeprong
to
two-prong adapter and connect the green pigtail
on
the
adapter toground.
NOTE
Operating the Model 412A/AR without
grounding the third (power line ground)
wire
of
the power cable will result
in
a
zero offset
of
the meter and oscillation
of
the pointer.
2-1
0.
INSTA
1
LATlON.
Section
I1
2-1
I.
The Model 412A is intended
for
use
as
a
bcncli
instrument, having a top-mounted handle and rubbcr fcct.
2-12. The Model 412AR is intended for rack mounting.
Front panel height
is
5-7/32 inches.
2-13.
REPACKAGING
FOR
SHIPMENT.
2-14.
The following paragraphs contain
a
general guidc for
repackaging
of
tlie instrument. Rcfcr to Paragraph
2-15
if
the original container is
to
be
uscd;
2-16
if
it
is
not.
If
you
have any questions, contact your
local
-lip-
Salcs
and
Service Office. (SCCAppendix
B
for office locatiwis.)
NOTE
If
the instrumcnt is to be sliippcd
to
I
Icwlctt-i’ackard for scrvicc
or
repair.
attach a tag to tlic instrumcnt idcntifying
the owncr and indicating tlic scrvicc
or
repair
to
be accomplislicd. lncludc tlic
model number
of
tltc instrumcnt.
In
any
corrcspondcnce, identify tlic instrumcnt
by modcl numbcr, scrial numbcr prcfix,
and scrial numbcr.
2-15. Place the instrumcnt
in
original conhincr with
suitable packing material and
scat
wcll
with strong tapc
or
metal bands.
If
original containcr
is
not aviilablc, onc
CBR
be
purchased from your ncarcst
hp
Salcs
and
Scrvicc
Office.
2-16.lf original container
is
not to hc uscd, procccd
as
follows:
a. Wrap instrument
in
hcavy
papcr or plastic hcfurc
placing in an inncr containcr.
b.
Place packing material around a11
sidcs
of
instrumcnt and protcct panel facc
with
cardhturd
strips
.
c. Place instrument and inncr containcr
in
a hcavy
carton
or
wooden box and scal wcll with strong
tapc
or
metal hands.
d. Mark shipping containcr
“DI~LICATE
INSTIWMENT,”
“FRAGILE,” ctc.
00
1
29-5
2-1

hlodcl4
12A
SECTION
111
0
PER
AT1
N
G
3-1.
INTRODUCTION.
3-2.
The
-lip-
Model 412A/AR
DC
Vacuum Tube
Voltmeter nieasures dc voltages from
1.0
mV full scale to
lo00 V
full
scale with
an
accuracy of
1%
of
full
scale.
Current measurement ranges are from
0.001
mA
full
scale
to
1000
mA
full
scale, with accuracy
of
2%
of
full scale.
The Model
41
2A/AR makcs resistaricc measurements from
1.0
ohm center
scale
to
100
megohms center scale.
Resistance measurenicnt accuracy is
5%
of
reading between
.3
and
3
on
the ohms scale.
A
dc amplifier output is also
provided.
Voltage
gain of tlic amplificr is from
-1000
to
+1000,
depending
on
range selcctcd. Amplifier output is
1
.O
V
at full scale. and maxinium current load is
1
.O
mA.
3-3.
LOW-LEVEL ELECTRICAL PHENOMENA.
34.
Stray low-level electrical phenomena may cause error
in
low voltage measurements. When using the lower voltages
of the Model 412A, consider the possibility
of
low-level
voltages produced by thermoelectric effects, residual
charges on capacitors, battery action,
or
flexing
of
coaxial
cables,
3-5.
The voltagc probe, currentlresistance lead, and
common lead
of
the Model 412A/AR are all designed to
have a very low thermoelectric effect with copper, the most
commonly used conductor. However, some component
leads such as transistors and reed relays may have leads
made
of
such alloys as nickel-iron, which produce a
thermoelectric voltage when joined with copper. Whenever
possiblc, connect tlic
412A
leads to copper. If this
is
not
possible, the voltage probe and common lead should both
In:
conncctcd
to
the
smic
kind
of
inctal compncnt lead,
and both connection points maintained at the same
tc
111
pcra
t
urc.
34.
To minimize tlic possibility
of
battery action when
making
low
voltage mcasurcments, make sure the 412A
probe tip and common lcad alligatorclip are clean.
Also
be
surc
tlic component leads or other points of connection are
clean. The prcsence
of
sonic chemicals such
as
solder flux
niay cause noticeable nieasiirement error due to battery
action.
NOTE
Operating the Model 412A/AR without
grounding the third (power line ground)
wire
of
the power cable will result in a
zero offset
of
the meter and oscillation
of
the pointer.
IN
STRU
CTiO
NS
3-7.
TURN-ON PROCEDURE.
Section
111
BEFORE CONNECTING PRIMARY
POWER, MAKE SURE THE REAR
PANEL 115/230 V SLIDE SWITCH
IS
SET
TO
THE POWER LlNE
VOLTAGE
TO
BE
USED.
INCORRECT SETTING MAY
RESULT IN DAMAGE TO THE
INSTRUMENT.
a. Make sure the proper fuse is installed for operation
with the power line voltage to be used. A
0.6
A
slow-blow fuse should be used for
I
IS
V
operation, and a
0.4
A slow-blow fuse for
230
V
operation.
b. Connect
to
ac power and turn instrument ON.
Allow
a
S-minute warm-up period.
3-8.
VOLTAGE
MEASUREMENT.
a.
Set
FUNCTION switch to VOLTS, and select
proper polarity.
b. Set RANGE switch to desired range.
If
in doubt,
set
switch to high range and downrange as
necessary.
c.
Connect VOLTS and
COM
leads across circuit
or
component, and read voltage.
DO
NOT OVEKLOAD THE
INSTRUMENT. MOMENTARY
OVERLOADS UP TO TEN TIMES
FULL SCALE WILL NOT DAMAGE
THE INSTRUMENT; HOWEVER,
CURRENT SHUNTS, AND
INTERNAL RESISTANCE
STANDARDS ARE NOT
PROTECTED FROM EXTREME
OVERLOAD.
33.
CURRENT MEASUREMENT.
a. Set FUNCTION switch to
MA.
b.
Set RANGE switch to desired range.
If
in doubt,
select high range and downrangeas necessary.
00
1
29-5
3-1

section
111
Resistance Range
.
c. Deenergizecircuit
to
be
measured.
OpenCircuit Volts
(at
1
on top scale)
d. Connect
412A
into circuit,
using
MA/OHMS
and
COM leads.
e.
Energize circuit,
set
POLARITY
switch
for
up-scale indication, and read current.
NOTE
When making current measurements,
make
sure
there
is
noconnection between
the circuit common
6)
and power line
ground (&)terminals on rear panel.
3-10.
RESISTANCE
MEASUREMENT.
a.
Set
FUNCTIONswitch
to
OHMS.
b.
If resistance
to
be
measured
is
part of a circuit,
make sure
power
is
turned
off.
c.
Connect
412A
across circuit
or
component
to
be
measured,using
MA/OHMS
and COM leads.
XI
XI0
XI00
X1
K
XlOK
XlOOK
XIM
XlOM
XlOOM
Model
41
2A
d.
Set
RANGE switch
so
that meter indication is as
near center scale as possible, and read resistance on
OHMS
scale.
3-1
1.
When measuring the resistance
of
non-linear devices
such as transistors and diodes, it is often desirable
to
know
what voltage and current were present in obtaining the
resistance measurement. Table
3-1
lists the open-circuit
voltage and shortcircuit current availableat
the
412A
leads
during resistance measurements. Voltage and current are
read on the top scale
of
the meter
for
all
resistance ranges.
The meter reading is directly proportional
to
the voltage
across the device being measured, and inversely
proportional
to
the current.
As
an example, assume that
a
device being measured on the
XI0
range gives a reading
of
.25
on
the resistance scale. The meter
also
indicates
.2
on
the
top
scale. Since full-scale voltage for the
XI0
range is
100
mV (Table
34),
this indicates a voltage
of
20
mV
across the device. To determine the current, subtract the
meter reading
of
.2
from the full scale reading
of
1,
and
multiply by the shortcircuit current listed in Table
3-1
.
Since the current shown
for
the
XI0
range
is
10
mA,
this
indicates
a
current
of
8
mA through the device being
measured. The resistance measurement, then, indicates an
equivalent resistance
of
2.5
ohms, with an applied voltage
of
20
mV, and a current
of
8
mA.
Table
3-1.
Voltage and Current in Resistance Measurements
1
10
mV
100
mV
IV
1v
1v
IV
IV
1v
1v
ShortCircuitCurrent
(at
0
on tot, scale)
10
mA
10
mA
10
mA
1
mA
l00pA
IO
pA
lPA
0.1
pA
0.01
pA
3-2
00
1
29-5

Model
4
12A
Section
111
3-12.
AMPLIFIER OPERATION.
at full scale; maximum rated load current is
1
.O
mA.
Load
currents
in
excess
of
1.0
mA (less than
1000
ohms load
3-13.
The Model
412A/AR
DC
Amplifier Output may be resistance) may cause errors in meter indication and
used todrive devices
such
as an analog recorder toprovide a amplifier
gain.
Amplifier gain is from
+lo00
on the
,001
permanent record
of
measurements. Output level is
1
.O
volt range
to
-1000
on the
loo0
range.
4
Y
00
1
29-5
I‘
3-3

Model
412A
SECTION
IV
THEORY
OF
OPERATION
4-1.
GENERAL.
4-2.
The kp- Model
412AIAR
uses a chopper-stabilized
amplifier circuit to produce a meter indication proportional
to the voltage,current, or resistance being measured.
4-3.
BLOCK
DIAGRAMANALYSIS.
44.
Figure
4-1
is
a block diagram
of
the Model
412AIAR.
The following paragraphs describe operation of the various
circuits.
4.5.
RANGE AND FUNCTION SELECTION.
4-6.
The Voltage Divider. Current Shunts, and Resistance
Standards circuits provide a
0.9
mV output for full-scale
input in all ranges and functions. Full scale
is
defined as a
reading
of
1
on
the top scale of the meter for all ranges.
Simplified schematics
of
the switchingcircuits
for
all ranges
and
functionsare shown in Figures
5-8
through
5-10,
47.
FILTERS.
4-8.
input filtering is provided to attenuate any ac
components present in the input signal. The second fdter
section filters the Demodulator output.
RANOC
SELtCllOW
r
Section
1V
4-9.
MODULATOR, AMPLIFIER, DEMODULATOR,
AN0
CHOPPER.
4-10. A
light-beam chopper driven by a synchronous motor
sets the Modulator/Demodulator frequency at
516
power
line frequency. The Modulator output is an ac signal
proportional
to
the Model
412A
input. The amplifier
gain
is
stabilized at
1
,111
by the feedback network, resultingin an
output
of
1
.O
volt for a full-scaleinput. The Demodulator
is
synchronized with the Modulator
so
that the amplified ac
signal
is
converted to
a
steady dc, which
is
applied
to
the
Cathode Follower. This voltage is the same polarity as the
input signal.
4-11.
CATHODE FOLLOWER AND POLARITY
SELECTION.
4-12.The Cathode Follower drives the Meter and the
DC
Amplifier Output. Polarity selection provides upscale
meter reading for either polarity input. The Polarity switch
is disabled in the
OHMS
function.
I
Figure
4-1.
Block Diagram
00
129-5 4-1

6
Section V Model 412A
INSTRUMENT
TYPE
DC
Voltage Standard
Variable Line Transformer
DC
Voltmeter
AC
Voltmeter
~ ~ ~~
Oscilloscope
Precision Resistors
Oscillator
Table
5-1.
Test Equipment Required
REQUIRED CHARACTERISTICS
Range:
0-300V
Accuracy:
0.1%
Current Rating:
10
mA
Output Voltage:
to
127
Vac
(or
253 Vac)
Range:
1
V
to
300
V
Resolution
:
10
mV
Input Resistance:
100
Ma
Range: to
300
V
Range: to
10
Ma
Sensitivity:
0.01
Vlcm
Resistance:
1
10
100
1K
10
K
30K
5W
100K
1W
300
K
1/2
W
1M
10
M
100
M
10
M
Accuracy
:
i
1.0%
2
1.0%
f
1.0%
f
1.0%
+,
1.0%
?
0.1%
2
0.1%
2
0.1%
2
0.1%
2
1.0%
2
1.0%
k
0.1%
-~
~
Frequency
:
60Hz
Output:
IO
V peak
hp-738BR Voltmeter
Calibrator
1
Superior Electric
Powerstat 3PF116
(for
1
15
V line) 3PF216
(for 230
V
line)
hp-427A Voltmeter
hp-
427A Voltmeter
hp-427A Voltmeter
hp
120B
Oscilloscope
Part Number:
-hP-
08
1
I-OoFiO
-hP-
hP-
07574346
0757.040
1
-hp-
07574280
-hP- 07574442
Daven
DAS-5
Dale
MF-I
Dale MFS-112
-hP
0698416
1
0733401
7
-hP
0698-5049
-hP-
-hP-
08
1
1-2475
-hp-
200CD Wide Range
Oscillator

Model
4
12A
SECTION
V
MAINTENANCE
5-1.
INTRODUCTION.
5-2.
This section contains information necessary to
maintain the kp- Model 412A/AR. The following
paragraphs describe the Performance Checks, Calibration
Procedures, Troubleshooting, and ServicingProcedures.
5-3.
REQUIRED TEST EQUIPMENT.
54.
Recommended test equipment for maintaining and
checking performance
of
the Model 412A is listed in Table
5-1.
Test
instruments other than those recommended may
be
used if their specifications equal
or
exceed the required
characteristics.
55.
PERFORMANCE CHECKS.
5-6.
Use
the following procedures to verify proper
operation of the Model 412A. The 412A and test
equipment should
be
operated at 115/230 Vac unless
otherwise specified. A Performance Test Card is provided at
the end of this section for recording the performance of the
Model 412A. The card may
be
removed from the manual
and used as a permanent record
of
the incominginspection
or
of
a routine performance check. If the Model 412A does
not meet specificationsat
any
point in this procedure, refer
to Paragraph
5-1
5,
Adjustment and Calibration procedure.
5-7.
METER
ZERO.
5-8.
The meter pointer should rest over the zero
calibration mark when the instrument is at normal
operating temperature, is turned off, and is in the normal
operating position.
If
the pointer is off
zero,
perform the
mechanical adjustment procedure outlined in Paragraph
5-1
8
before proceeding with the performance checks.
5-9.
Turn the Model 412A POWER switch ON. After a
5-minute warm-up period, the meter pointer should rest on
zero
when the input leads are shorted. If not, perform the
electrical zero adjustment outlined in Paragraph 5-22.
Section
V
5-10.
VOLTMETER ACCURACY CHECK.
a.
A
dc voltage standard and a variable line
transformer are required
for
this chcck.
If
an -tip-
Model 738BR Voltmeter Calibrator is used, it
should be alloweda 30-minute warm-up period.
b. Connect 412A
to
variable line transformer, and set
line voltage switch (1
15/230)
tocorrespond toline
voltage being used. Turn 412A ON and allow
5-minute warm-up period.
c.
Set 412A FUNCTION to VOLTS, POLARITY to
+.
d. Connect VOLTS and COM leads to dc standard
output.
e.
Check 412A indication againstdcstandard output
at full scale
on
each 412A range. Maximum error
should be
no
greater than
f
1%
of full scale (one
small division
on
topscale).
NOTE
When checking the Modcl 412A
on
the .001 and
.003
volt ranges, turn
off the dc standard output and note
any 412A
zero
offset due
to
thermoelectric voltage. Add
or
subtract
as
necessary any offset from
the 412A reading.
f. Check 412A meter tracking
on
the
1
V rangat
all
cardinal points on the meter scale. Maximum crror
should be no greater than
0.01
V
with cithcr
polarity input.
.
g. Repeat step e on
1
V
range at line voltagesof
103
and I27volts
(or
207and 253 V).
NOTE
5-11.
MILLIAMMETER ACCURACY CHECK.
Operating the Model 412A/AR without
grounding the third (power line ground)
wire of the power cable will result in a
zero offset of the meter and oscillation
of
the printer.
a. A dc voltage standard and the prccision rcsistors
listed in Table
5-2
rcquircd
for
this chcck.
b. Set 412A FUNCTION toMA. I’OLARITY to
+.
001
29-5
5-1

Section
V
DC
Standard
output
IV
3v
.O
1
10
v
.03
30
V
.I
100
v
.3
300
v
1
300
V
3
300
V
IO
300
V
30
300
V
3
Range Series
Resistance
I.00M
t
0.1%
1
.OOM
t
0.1%
1.00M~O.I%
1
.OOM
2
0.1%
I
.OOM
2
0.1%
1.00M?
0.1%
300K
iO.l%
1/2
W
lOOK
20.1%
1
W
30K
2
0.1%
5
W
30K
to.]%
5
W
Model
4
12A
Table
5-2.
Milliammeter Accuracy Check
Model
4
1
2A Reading
(
2%
of full scale)
1
.o
3
.O
1
.o
3
.O
1
.o
3
.O
1
-0
3
.O
1
.o
1
.O
(on
3
scale)
c.
Connect Precision Resistor in series with MA lead
and positive side
of
dc standard output. Connect
412A
COM lead to dc standard negative output
terminal.
Paragraph
5-10.
changes. Figure
5-9
shows simplified ammeter switching
circuits.
5-12.
OHMMETER
ACCURACY
CHECK.
d. Perform checkslisted in Table
5-2.
a.
Set
FUNCTION
selector
to
OHMS.
NOTE
It is not necessary
to
check higher current
ranges.
On
current ranges
10
through
1000, the current shunt does not change.
Only the voltage divider, checked in
b. Connect, in turn, resistor
of
1,10,100,1
K,
10
K,
100
K,
1
M,
10
M,
and
100
M
ohms+I.O%
between MA/OHMS and COM leads. With
appropriate RANGE setting, meter should indicate
between
0.95
and
1.05
in each case.
-hp-
MODEL
412A
DC
STANDARD
-hp-
MODEL
738BR
I
I
~
Figure
5-1.
Input Resistance Check
5-2
001
29-5

J
i
h
9
c
D
J
b
1
412A
Input
'
Resistance
Model 4
12A
5-13.
INPUT RESISTANCE
CHECK.
a.
Connect the Model 412A/AR, dc standard and
10
rnegolim resistor
as
shown
in
.Figure
5-1.
b.
Set 412A
FUNCTION
to
VOLTS,
POLARITY to
+.
c. Adjust
412A
RANGE and dc standard output to
voltages listed in Table
5-3.
A
correct meter
reading in each case verifies the input resistance
where:
Ein
=
Eapplied
[
&&]
Table
5-3.
Input Resistance Checks
100
mV
412A
Meter
Reading
(?
1
%)
5.0
mV
22.5
mV
90.9
mV
0.95
V
IOMQ
30
Ma
100
Ma
200
Ma
Section
V
c. Set ac voltmeter FUNCTION
to
AC
VOLTS,
RANGE to
IO
V.
d. Adjust oscillator frequency to
60
Hz,
and adjust
output for an indication
of
7.07
V
on
ac
voltmeter. This corresponds
to
a
IO
V
pcak
;IC
noise signal at the
412A
input.
e. The 412A meter rcadingshould
be
02
I%(l
minor
division
on
upper scale)..
5-15.
ADJUSTMENT AND CALIBRATION
PROCEDURE.
5-16.
The followingproccdures should bc pcrfornicd
only
if
it has been determined by the Performance Chccksoutlincd
in
Paragraph
5-5
that the Model
412A/AR
docs
not
iiicct
specifications.
5-17.
CABINET REMOVAL.
a. Disconnect power cord from powcr source.
b.
On cabinet model, rcmovc two rctaining
scrcws
near center of rear panel. On rack modcl, rciiiovc
three retaining screws
in
rcar
of
covcr.
c.
Slideinstrument chassis forward
out
of
cabinct.
5-18.
MECHANICAL ADJUSTMENT
OF
METER
ZERO.
5-14.
AC
REJECTION
CHECK.
5-19.
If
mechanical zero
of
mctcr is corrcct, pointcr rcsts
over the zero calibration mark
on
thc nwtcr scalc wlicii
voltmeter as shown in Figure
5-2.
instrument is at normal opcrating tcmpcraturc,
in
its
normal operating position, and turncd
off.
If
nicclianical
zero is not corrcct, perform thc following adjuslment
procedure.
a.
Connect the Model
412A/AK,
oscillator, and ac
b.
Set 412A FUNCTION to
VOLTS,
POLARITY to
+,
and RANGE to
0.1
V.
-hp-
MODEL
412A
OSCILLATOR
hp
200CD
Figure
5-2.
AC
Rejection Check
001
29-5
5
-3

J
Section
V
19
Model
4
I
2A
*7
3.
C
R46
-
METER
CAL
-
4
10,
CAL
-
R126
HUM
BAL
n
GAIN
CAL
R126
HUM
BAL
Figure
5-3.
Location
of
Adjustments
00
I
29-5

Model
4
12A
a. Turn 412A ON and allow a 20-minute warm-up
period. This
allows
meter movement
to
reach
normal operating tcniperaturc.
b. Turn instrumcnt OFF and allow
30
seconds for all
capacitors
to
discharge.
c.
Rotate mechanical zero-adjustment screw
clockwise until meter pointer is toleft of zero and
moving upscale toward zero.
d. Continue to rotate adjustment screw clockwise
and stop when pointer
is
exactly over zero mark.
If pointer overshoots zero, repeat steps
c
and d.
e. With pointer exactly on zero, rotate adjustment
screw counterclockwise approximately
10
degrees.
This frees adjustment screw from meter
suspension. If pointer moves during this step,
repeat steps
c
through e.
5-20.
HUM BALANCE ADJUSTMENT.
a. Turn
412A
ON and allow 5-minute warm-up
period.
b.
SetFUNCTION
to
VOLTS.
c.
Connect oscilloscope to
DC
AMPLIFIER
OUTPUT.
d. Adjust HUM BAL (R126, Figure
5-3)
for
minimum
10
Hz
signal
on
oscilloscope. (If power
line frequency is
50
Hz, beat frequency will be
8-1/3
Hz instead of
10
Hz.)
NOTE
The Hum
Bal.
control docs not affect
the 120
Hz
ripple. Adjust only for
minimum 10
Iiz
(or
8-1/3
Hz) signal.
With the VOLTS lead shorted
to
COM
lead and POLARITY set
to
t,
the 412A meter will show a
minimum deflection when the Hum
Bal.
is
correctly adjusted.
5-21.
CATHODE FOLLOWER BIAS ADJUSTMENT.
a.
Set
412A
RANGE
switch fully clockwise (one step
beyond 1000).
b. Adjust rear panel BIAS
ADJ
(R116)
to
set
meter
pointer approximately on zero. This adjustment is
not critical, since
any
deviation from zero is
reduced more
than
100
times when the RANGE
switch is in any operating position.
00
129-5
Section
V
5-22.
METER ELECTRICAL ZERO ADJUSTMENT.
NOTE
Meter electrical zero adjust circuit
may not
be
installed in 412A cabinet
model instruments with serial
number 424-14482 and below, or
in
41
2AR rack mount instruments with
serial number 424-15082 and below.
a. Turn 412A ON and allow
a
30-minute warm-up
period.
b. Set FUNCTION
to
VOLTS, RANGE
to
.001.
c.
Short VOLTS probe
to
COM lead.
d. Adjust ZERO ADJ (R129) for zero dcflcction of
the meter pointer. Check the zcro adjustmcnt by
switching POLARITY from
t
to
-
When zero
is
adjusted properly, this action should produce a
minimum amount of deflection.
5-23.
AMPLIFIER
GAIN
ADJUSTMENT AND METER
CALI
B
R
AT1
0
N
.
a.
A
dc standard and
a
de voltmeter arc rcquircd for
these adjustments.
b.
Connect
4
12A VOLTS probe and COM lcad
to
dc
standard output.
Set
FUNCTION
to
VOLTS,
RANGE
to
1
V,
POLARITY to
t.
c.
Connect dcvoltmeter betwccn dc standard output
(t)
and412A
DC
AMPLIFIER OUTPUT liigli.
d.
Set
dcstandard output to
+I
V.
e.
Adjust 412A GAIN
CAL
(RI
19)
for zcro
indication on test voltmctcr.
A
10
mV rcading
on
test voltmcter indicates
1%
crror in
4
12A gain
calibration.
f. Adjust METER CAL (R46) to sct 412A meter
pointer on
1.0
on top scale. Disconnect test
voltmctcr.
g.
Chcck 4l2A calibration at full scalc on all ranges
as outlined
in
Paragraph
5-10.
If
any rcading
is
in
error morc than
I%,
rcadjusl
GAIN
CAL
(It1
19)
to
bring
all
readings within
f
1
X.
5-24.
OHMMETER ADJUSTMENT.
a. Set 412A FUNCTION toOIIMS, ItANGIi to
XI
K.
Make ccrtain
MA/OlIMS
and (:OM
lcads
are not
connected through wmccxtcrnal rcsistancc.
5-5

Section
V
Check Point
b. Adjust rear panel OHMS ADS (R36) for meter
indication
of
aD
(corresponding to
1.0
on upper
scale).
Voltage
525.
TROUBLESHOOTING.
5-26. When the Model 412A/AR operates improperly, first
adjust and calibrate it according
to
the procedures outlined
in Paragraph
5-15.
If
calibration
is
not possible, procecd
with the troubleshooting steps.
CONTAMINATION OR
FINGERPRINTS ON THE RANGE
SWITCH
OR
THE COMPONENTS
MOUNTED
ON
THE RANGE SWITCH
MAY REDUCE ACCURACY OF
INSTRUMENT.
5-27.
FRONT
PANEL INDICATIONS.
5-28.
If the malfunction is common
to
all
ranges and
functions, the trouble may be in the amplifier or power
supply areas. If the instrument operates improperly on only
one range
or
function, the trouble
is
probably in the
switchingsection
or
input cables.
5-29.
POWER SUPPLY
AND
AMPLIFIER CHECKS.
5-30.
Check power supply voltages at pointslisted in Table
5-4.
Voltages are measured with respect
to
circuit common
(A).
Figure
7-2
shows the location
of
components.
53
1.
Tubes are
a
coinmon source
of
trouble and are most
easily checked by substitution. Following replacement
of
a
particular tube, iiiake the adjustrneiits indicated
in
Table
5-5.
5-32.
Additional
checks
of
amplifier operation niay be
made
by
cliecking
voltages shown
on
the amplifier section
of
tlic
sclrwutic diagram.
Be
sure
to
observe conditions
in
Note
I1
in
Paragraph
7-2.
Allow
3
tolerance
of
210%.
5-6
Model
4
12A
Table
5-5.
Tube Iteplacement
Tube Iteplaced
VI01 (12AX7) Amplifier
V102 (6AU8) Amplifier/
Cathode Follower
V103 (6x4) Rectifier
V104 (OA2) Regulator
VI05 (OB2) Regulator
5-33.
MODULATOR
CHECK.
Adjustment
Huni
Bal.
Par.
5-20
Gain Cal. Par.
5-23
Bias
Adj.
Par.
5-21
Cain
Cal.
Par.
5-23
None
Ohms Adj.
Par.
5-24
-71
a.
An oscilloscope and
a
10
Maresistor are required
for this test.
b.
Disconnect 412A from power source and remove
cabinet.
c.
Remove VlOl and reconnect 412A to power
source.
d. Connect
a
clip lead
from
the center arm
of
R116
(rear panel
BIAS
ADJ) to the termination
of
the
VOLTS input cable (terminal
4
in Figure
5-7).
e. Set 412A FUNCTION to VOLTS, RANGE to
.001.
f.
Set oscilloscope input to
DC.
g.
Connect a
10
Mfi
resistor in series with
oscilloscope vertical input. Using a clip lead,
connect resistor
to
junction
of
RlOl and ClOl
(input circuit board, Figure 54). Connect
oscilloscope common to 412A COM lead. Set
Vertical Sensitivity
to
0.2
V/cm.
I
Figure
54.
Input
Circuit
Board
00
1
29-5

Model
4
12A
h. Turn 412A ON and allow 30-second warm-up
period. Observe de voltage amplitude on
oscilloscope.
i. Move oscilloscopc input (with series resistor) to
Modulator output at junction of green lead from
Modulator and C104
(0.01
Microfarads).
j.
Waveform should be similar
to
that shown in
Figure
5-5.
Peak-to-peak amplitude should be
approximately equal
to
dcvoltage observed in step
h. Signal frequency sliould be
5/6
power line
frequency.
k. If waveform is not satisfactory, check chopper
motor and assembly or Modulator and input
circuit as indicated by deficiency in waveform.
p-20
MSEC-
--)1
Figure
55.
Modulator Waveform
5-34.
MODULATOR
REPLACEMENT.
5-35.
The photoconductor cells in the Modulator
(VllOA/B) are not separately replaceable.
If
a
cell
is
defective, the Modulator block assembly should be
replaced, using the following procedure.
a. Disconnect 412A from power source and remove
cabinet.
CONTAMlNATION OR
FINGERPRINTS
ON
THE RANGE
SWITCH OR COMPONENTS
MOUNTED
ON
THE RANGE
SWlTCII MAY KEDUCE
ACCURACY
OF
THE
INSTRUMENT.
b.
In 412A cabinet model, remove side shield.
c.
Note and record color code of wires before
removing from Modulator assembly printed circuit
board.
001
29-5
Section
V
d. Remove four screws holding light-beam chopper
assembly. It is not necessary to disconnect wires.
e. Spread grip rings on Modulator light rods and slide
rods out
of
Modulator block.
f. Remove four screws holding modulator assembly
to
chassis.
NOTE
Be
sure
to
seat the four insulating
shoulder washers in mounting holcs
when replacing the assembly.
g.
Lift out Modulator assembly. If it is necessary to
replace Modulator block asscmbly only, rcmovc
four screws holding printed circuit board
to
block.
Note color and location
of
wircs between block
and printed circuit board before disconnccting.
5-36.
DEMODULATOR
CHECK.
a.
Disconnect 412A from powcr sourcc and removc
cabinet.
b. Remove V102 (on Amplifier Assembly).
c.
On
iight-beam chopper assembly, rcmove uppcr
lamp nearest front of instrument. This
lamp
illuminates
VI
1
I
A.
d. Connect a
1.0
microfarad capacitor across input
terminals of test ohmmeter.
e. Connect test ohmmeter common lead
to
Demodulator terminal with pinklorange wirc.
Sec
Figure
56.
f. Connect other
test
ohmmeter lcad to tcrminal
with whitelorange wire.
g.
Connect 412A
to
power source and turn
ON.
.
h. Resistance indicated on
test
ohmmeter
should
bc
I
to 2 megohms.
i. Turn
41
2A
OFF.
j.
Replace lamp in light-beam chopper and remove
upper lamp nearest rear of instrument. This lamp
illuminates
VI
I
1
B.
k. Connect test ohmmeter common lcad to
Demodulator terminal with brownlorange wirc.
5
-7

.
.
,
.
. -.
.
,
.
.
._...
.
..
-
-
..
.,
.
. --*w
”-
Section
V
b
I.
Connect other test ohmmeter lead to terminal
with whitelorange wire.
m.Turn412AON.
n. Resistance indicated on test ohmmetershould be
I
b
to2 megohms.
0.
Turn
41
2A OFFand replace lamp and
V102.
Iy
5-37. DEMODULATOR REPLACEMENT.
5-38.
The photoconductors in the Demodulator are not
separately replaceable. If a
cell
is defective, the
Demodulator assembly should
be
replaced, using the
following procedure:
a. Disconnect instrument from power source and
remove cabinet.
b.
Disconnect the three wires
to
Demodulator
assembly. Figure
5-6
shows wire
color
codes for
both cabinet (4 12A)and rack (4 12AR)models.
c. Remove four screws holding light-beam chopper.
It
is
not necessary todisconnect wires.
d.
Remove nuts from four screws holding
.I
Demodulator assembly and remove assembly from
instrument.
5-39.SWITCH CHECKS.
CLEAN RUBBER GLOVES SHOULD
BE WORN WHEN WORKING ON
THE
RANGE SWITCH. CONTAMINATION
6
5-8
412A
(CABINET
MODEL)
412A-8-50160
Model 412A
OR FINGERPRINTS MAY KEDUCE
ACCURACY OF INSTRUMENT. IN
ADDITION, LOW ROSIN CONTENT
SOLDER SHOULD BE USED WHEN
REPLACING COMPONENTS. USE A
MINIMUM OF HEAT WHEN
SOLDERING, AND REMOVE TRACES
OF
FLUX
FROM CONNECTIONS. IF
NECESSARY, RANGE SWITCH MAY
BE CLEANED USING A SOLUTIONOF
MILD DETERGENT AND WARM
WATER. RINSE THOROUGHLY IN
DISTILLED WATER AND
DRY
IMMEDIATELY.
540.If
the malfunction is peculiar to one range
or
function, trouble may be in the switching circuits. Either a
component
or
switch contact may be defective. Figures
58
through
5-10
show simplified schematics of range and
function switching, and may be used toisolate the trouble.
For example, if the instrument operates properly on all
voltage rangesexcept
.003;
R1, R9,
or
the associated switch
contacts may be defective. Figure
7-3
identifies the
componentsmountcd on the
rmgc
switch.
5-41.
ADJUSTMENT
OF
FACTORY
SELECTED
COMPONENTS.
542. R34*.
5-43.
The value of
R34*
is selected tobring the OHMS ADJ
potentiometer within the proper range, and is dependent
upon the power supply voltage and the value
of
other
resistors in the ohmmeter circuit. Replacement
of
R34*
should not be necessary unless V104 has been replaced
or
the range switch components have been changed.
If
the
VI
412AR
(RACK
MODEL)
Figure
56
Demodulator Assembly
TO
Cl13
TO
C114
AND
Rll7
TO
RI
16
00
1
29-5
Other manuals for 412A
1
This manual suits for next models
1
Table of contents
Other HP Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Stoneridge
Stoneridge SE5000 quick guide

Elster
Elster BK-G Series operating instructions

PCB Piezotronics
PCB Piezotronics IMI SENSORS 625B00 Installation and operating manual

APS
APS LSE-D16-BS manual

Continental Refrigerator
Continental Refrigerator ContiPressureCheck Series Brief instructions

Seametrics
Seametrics iMAG 4700r quick start guide