hydrafacial Syndeo User manual

P/N 19 0 0104 - B
Rev. 02/22/2022
HydraFacial Syndeo™
User Guide

2
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Copyright © 2022 HydraFacial, LLC. All rights reserved.
Terms and Conditions of Use
HydraFacial™, Vortex-Fusion™, HydroPeel™are U.S. Registered trademarks, and many other referenced marks throughout this User
Guide. For a complete listing of all U.S. Registered trademarks, go to www.hydrafacial.com/patents/trademarks. This product and
its use are covered by one or more U.S. and International patents. Patent info: hydrafacialco.com/patents. These intellectual property
rights (IPR’s) remain the exclusive property of HydraFacial, LLC or its licensors.
Contact Info
HydraFacial, LLC
3600 E. Burnett St.
Long Beach, CA 90815 USA
(800) 603-4996 •(562) 597-0102
contact@hydrafacial.com
hydrafacial.com
E C Rep Ltd.
5 Fitzwilliam Square East,
Dublin 2, D02 R744, Ireland
Tel: +353 89 225 1951
Email: info@ecrep.ie
Responsible Person in the United Kingdom:
Responsible Person Ltd.
Summit House, 4-5 Mitchell Street,
Edinburgh, EH6 7BD, UK
Tel: +44 131 618 1115
Email: info@responsible-person.co.uk
1639

3
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Contents
Before You Start
4 Introduction
5 Technical Specications
5 Environmental Conditions
6 Safety Information
Getting Started
9 Unpacking
10 System Components
12 Set–Up
Operating Instructions
14 On, Off, and Standby/Sleep Modes
15 Touchscreen
17 Gesture Control Functions
18 Perform Treatment
Cleaning and Maintenance
19 Cleaning and Maintenance
20 Handpiece Cleaning
21 Manifold Cleaning
21 Touchscreen Cleaning
22 Troubleshooting
23 Limited Warranty

4
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Introduction
Congratulations on your purchase of the HydraFacial Syndeo™ System! You can have the utmost condence in this leading edge
system designed in the USA by HydraFacial, LLC. The HydraFacial System is a modular system that may include other modes of
operation not available in the EEA. Separate User Guides are provided for all optional modalities.
The System is intended to be used in clinics, hospitals, doctor’s ofces, and aesthetic practices by trained and licensed practitioners,
such as dermatologists, facial plastic surgeons, physicians, and aesthetic professionals – in accordance with their licensing scope
and applicable national/local requirements.
Intended Use
The System is intended to abrade (exfoliate) the upper layers of the skin.
Indication (related to the use of the device in the European Economic Area (EEA):
•Mild to moderate acne, i.e. acne vulgaris, comedonal acne (blackheads and whiteheads).
•Supercial acne scarring.
Contraindications
It is intended for external use on intact skin only, subject to the following contraindications and warnings:
•Do not use over active acne, such as severe, inammatory acne.
•Do not use over active infection or open wounds (e.g., herpes simplex, excoriations, or open acne cysts).
•Do not use on client with autoimmune or communicable disease.
•Do not use on client with cold sores or fever blisters without pre-medication.
•Do not use over abnormal, unidentied facial growth or mark.
Side Effects
Client may experience temporary irritation, tightness, or redness at the treatment area. This usually dissipates within 72 hours
depending on skin sensitivity.
Warnings
•Not recommended for severe acne scarring.
•If the skin becomes irritated beyond normal, discontinue the treatment and seek medical advice.
•Do not use within the orbital area or over any body orice, such as over the eyelids. It is recommended to have the eyes closed
during the procedure. Sterile eye rinse solution should be available in case products accidentally get in the eyes.
•This System is not intended to be used on infants, children, pregnant or lactating women because it has not been evaluated for this
group of users.
•The above have been assessed and supported by clinical evaluation in accordance to Medical Device Directive 93/42/EEC
Annex X. The device complies with the applicable Essential Requirements in Annex I. The HydraFacial Syndeo™is a Class IIa
medical device in the EEA; GMDN Code 11177.
System Specications:
•Unit Size: 117 cm / 46 in (H) x 50.8 cm / 20 in (W) x 45.8 cm /18 in (D)
•Unit Weight: 30 kg (65 lbs)
•Electrical:
220 –240VAC, 50/60Hz, nominal 2A, 8A max 7000079
100 –120VAC, 50/60Hz, nominal 2A, 8A max 7000078

5
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Technical Specications
Parameter Specication 220–240 VAC Specication 110 –12 0 VAC
7000079 7000078
Voltage 220–240 VAC, 50/60 Hz, nominal 2A, 8A max 100–120 VAC, 50/60 Hz, nominal 2A, 8A max
Power Consumption 142 watts, maximum 142 watts, maximum
Mode of Operation Continuous Continuous
Audio Volume Under 70 dB Sound Pressure Level at 1 meter Under 70 dB Sound Pressure Level at 1 meter
Visual Indicators LCD display with system messages LCD display with system messages
Instrument Maker/Model HydraFacial, LLC / HydraFacial Syndeo™HydraFacial, LLC / HydraFacial Syndeo™
Dimensions 117 cm / 46 in (H) x 50.8 cm / 20 in (W) x 45.8 cm /18 in (D) 117 cm / 46 in (H) x 50.8 cm / 20 in (W) x 45.8 cm /18 in (D)
Weight 30 kg (65 lbs) 30 kg (65 lbs)
Operating Altitude Maximum of 2000 m Maximum of 2000 m
System Controls Touchscreen monitor, buttons Touchscreen monitor, buttons
Interconnects Custom cables Custom cables
Ingress of Fluids Not protected against the ingress of uid. IPX0 Not protected against the ingress of uid. IPX0
Flammability Equipment not suitable for use in the presence of ammable
anesthetic mixture with air or with oxygen or with nitrous oxide.
Equipment not suitable for use in the presence of ammable
anesthetic mixture with air or with oxygen or with nitrous oxide.
Electrical Conformity This medical equipment has passed all required testing for
electric shock, re and mechanical hazards in accordance
with IEC 60601–1, IEC/EN 60601–1.
This medical equipment has passed all required testing for
electric shock, re and mechanical hazards in accordance with
IEC 60601–1, IEC/EN 60601–1.
Electrical Shock Protection – Classication Class I Class I
Electrical Shock Protection – Degree Type BF Equipment Type BF Equipment
Product Life The Console is a non-sterile, durable good. The System
warranty runs for a period of two (2) years from the date of
installation of the System to the original Customer.
The Console is a non-sterile, durable good. The System
warranty runs for a period of two (2) years from the date of
installation of the System to the original Customer.
Applied Part The Handpiece is the client applied part The Handpiece is the client applied part
Separation From Power Mains Power switch on rear panel provides full separation from
power mains.
Power switch on rear panel provides full separation from
power mains.
Fuse Slow-blow fuse. 250V 8A Slow-blow fuse. 250V 8A
Environmental Conditions
Parameter Specication
For Shipping & Storage
Temperature: Normal (–20 to 60 C)
Relative Humidity: 10—90% RH
Atmospheric Pressure: 500—1060 hPa
Operating Conditions Temperature: Normal (10 to 40 C)
Relative Humidity: 30—75% RH

6
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Product Marking Symbols
Symbol Description
Attention: Refer to accompanying documentation.
Separate collection for electronic.
Type BF Applied Part. Debrillation proof.
Direct Current
Alternating Current
ON (Power)
OFF (Power)
STANDBY (Power)
Consult Instructions For Use
Fuse
Pushing Prohibited
Certied by SGS
Connect to supply mains with protective earth (ground)
Keep Dry
Caution and Marking Symbols
Symbol Description
Calls attention to a procedure, practice, or condition that
could possibly cause bodily injury or death.
Calls attention to a procedure, practice, or condition that
could possibly cause damage to equipment or permanent loss
of data.
Do Not Re-use
Manufactured By
Date of Manufacture
1639
Conformité Européenne
Catalogue Number
Serial Number
Lot Number
European Representative
Medical Device
Unique Device Identier
Safety Information

7
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Safety Guidelines
1. Ensure that all operators of the the HydraFacial Syndeo™System are trained and licensed as required by local/national
regulations. Do not operate the unit before being trained. For any questions regarding training, contact The HydraFacial
Company or your local distributor.
2. Be sure to read the User Guide thoroughly before setting up the System. If you experience mechanical and/or electrical difculties
with your unit, contact your local distributor.
3. Always do a client consultation to determine if the client is a candidate for the procedure. Follow contraindications and warnings
as pre-determination for procedure. Verify any known allergies with the client. Perform a patch test if necessary.
4. Always use clean, unused tips for each procedure. Keep tips in original packaging until ready to use.
Reusing tips could result in skin infection.
5. Use only tips manufactured by HydraFacial, LLC.
6. Use only solutions and formulas authorized by HydraFacial, LLC.
7. Do not use contaminated skin solutions.
Reusing contaminated skin solution could result in skin infection, and will void all warranties. Partially used skin solution
bottles must be capped and stored in accordance with instructions on product label.It is recommended not to use skin solutions
that exceed the "Best Used By" date. If you observe any abnormality with your products, please contact HydraFacial, LLC or your
local distributor.
8. Removing contact lenses prior to procedure is recommended.
9. The client should have a minimum sunscreen SPF 30 applied after the procedure and should use sunscreen on an ongoing basis.
10. Each client’s skin conditions and sensitivity are different. Always begin treatment conservatively. Do a sensitivity test on the neck by
the earlobe rst, and increase or decrease the vacuum level as appropriate. Lower vacuum level is recommended for thin, fragile
skin. Skin conditions requiring more aggressive vacuum levels are at operator’s discretion. Follow recommended protocols and
cautiously consider skin types.
11. Empty the #Gunkie waste canister after each service according to your waste handling protocol and local/national regulations.
Follow the System cleaning instructions in this guide to clean your System and handpiece.
12. Keep the System in accordance to recommended environmental conditions. Liquid should only be suctioned through
the HydraFacial handpiece.
13. Always use solution when performing abrasion. Performing the treatment without skin solutions could result in discomfort to
the client.
14. The System is not spill proof. Avoid spillage of any liquid on the System.
15. The System is not intended to be used in conjunction with any other System, medicine or technology and should only be used
in accordance with these Instructions for Use.
16. Use of non-approved skin solutions will void the warranty and may clog the System.
17. Do not let the #Gunkie waste canister overll. If this happens, a built-in oat device will occlude the vacuum opening and skin
solution ow through the handpiece will cease.
Note: Always remove the #Gunkie waste canister before transporting the unit.

8
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
18. Do not attempt to lift the unit alone or move over uneven or damaged ooring.
19. Wearing gloves is recommended when performing all treatments, system cleaning and preventive maintenance activities.
20.
Device cannot be serviced or maintained while in use.
21.
Unauthorized user modications will void the warranty. HydraFacial, LLC will not be responsible for any injuries
sustained due to unauthorized System modications or use of parts not specied by HydraFacial, LLC. Systems returned to
HydraFacial, LLC with unsafe modications will be returned to their original operating condition at the customer’s expense.
22.
DO NOT suction liquids through the dry port located on the side of the unit. This will damage the System and void
the warranty.
23.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are
operating normally.
24.
If this System does cause interference, which can be determined by turning the System off and on, the interference can
be possibly corrected by one or more of the following measures:
•Relocate the System with respect to the receiver.
•Move the System away from the receiver.
•Plug the System into a different electrical outlet so that the System and receiver are on different branch circuits.
25. Use of accessories, transducers, and cables other than those specied or provided by HydraFacial, LLC could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 Class A).
If it is used in a residential environment (for which CISPR 11 Class B is normally required), this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or
reorienting the equipment.
Emissions Standard Test Method
EN 606061-1-2: 2015
IEC 60601-1-2:2014
CISPR 11:2015 +A1:2016
EN 61000-3-2:2014
EN 61000-3-3:2013
Immunity Standard Test Method
EN 60601-1-2:2015
IEC 60601-1-2:2014
IEC 61000-4-2:2008
IEC 61000-4-3:2010
IEC 61000-4-4:2012
IEC 61000-4-5:2014
IEC 61000-4-6:2013
IEC 61000-4-8:2009
IEC 61000-4-11:2004
ISO 7637-2
26.
Portable RF communications equipment should be used no closer than 15 cm/6 in to any part of the HydraFacial
System, including cables specied by HydraFacial, LLC. Otherwise, degradation of the performance of this equipment
could result.

9
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Unpacking
Inspect the Shipping Carton
Your System was shipped in a custom foam insulated box that has been tested to UPS standards. If the shipping carton is damaged,
inspect the contents for visible damage such as dents, scratches or any other obvious signs of damage. If the System is damaged,
notify HydraFacial, LLC or your local distributor and they will assist you in the repair or replacement of your System.
Do not return your unit without rst contacting your distributor and receiving an RMA (Return Material Authorization) number.
Unpack the unit, accessories and all documents. Retain the original packaging in case you need to ship the System for any reason.
Verify that all appropriate items have been received.
The unit is large and heavy and should not be lifted alone.
Package Contents
•Syndeo System
•Accessory Box
◦ Power cord
◦ #Gunkie waste canister
◦ Puncture tool
◦ Handpiece caps
•Handpiece Kit
◦ Black plastic handpiece
◦ Handpiece tubing
◦ Cleaning tools
•Lymphatic kit
◦ Facial lymphatic handpiece
◦ Body lymphatic handpieces
◦ Tubing
◦ Split Cup Tubing
27.
To prevent electric shock, do not remove the instrument cover. There are no user-serviceable parts inside. Routine
maintenance or cleaning of internal parts is not necessary. Avoid the use of cleaning agents or chemicals on the instrument.
Some chemicals may damage plastic parts or lettering. Any external cleaning should be done with a clean, dry, or slightly
damp cloth.
28. Do not rest cups or any liquid containing vessels that could spill on the monitor or in the backpack.
29. The use of LightStim Elipsa™in conjunction with HydraFacial products and protocols is at the discretion of the esthetician. Products
that can cause irritation are recommended to be used after LightStim Elipsa™such as: Retin-A, benzoyl peroxide, harsh acids,
potential reactive forms of vitamin C and A, and products containing metals.
30. Do not locate machine in a position that blocks access to the rear power switch.
31. End of life disposal instructions: Do not dispose of this product with ordinary waste. Please refer to your local, state, or national
authorities for directions on how to comply with local electronic waste disposal requirements.

10
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
System Components
Tray
Handpiece cradle
Manifold buttons
For solution bottles
Qr code camera
Gesture control
sensors
Facial lymphatic
cradle
Wheels
(front wheels
with brakes)
Pick-up tubes
For skin solution
and system
cleaning
solution bottles
Touchscreen

11
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
#Gunkie canister
Back panel:
USB port
Ethernet port
Power
switch
Power button for
#Gunkie light magnier

12
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Set-Up
#Gunkie Waste Canister
1. Attach the #Gunkie waste canister by aligning the protruding tab on the #Gunkie with the release button on the system and
press the #Gunkie upwards until the latch connects.
2. Make sure the waste canister is held securely. Pull down on the #Gunkie. If it does not pull out, it is engaged.
Note: Always remove the #Gunkie waste canister before transporting the unit.
3. To remove the #Gunkie, press the release button and pull down on the #Gunkie.
HydraFacial Handpiece
1. Connect the blue twist lock (Return Line) to the blue connector on the bottom of the handpiece and secure by twisting until hand
tight. Do not overtighten.
2. Connect the clear twist lock (Supply Line) to the cartridge and secure it by twisting the lock until hand tight. To prevent the lock
from breaking, do not overtighten the twist lock.
3. Snap tubing connector onto the Syndeo Device. Press metal tab on the device to allow for connection. This connects both tubes
to the Syndeo device.
4. Insert the cartridge into the handpiece. Align the arrow on the cartridge with the unlock symbol on the handpiece. Push rmly
until it snaps into place, then twist the cartridge up until the arrow aligns with the lock symbol.
Note: If the cartridge is difcult to insert, drizzle a small amount of water into the cartridge to lubricate it, and try again.
5. To remove the cartridge and insert skin solution vials (sold separately), twist the cartridge until the arrow aligns with the unlock
symbol, then pull the cartridge straight out. Next, insert the vial into the handpiece. Push rmly until it snaps into place.
6. Turn the ow control valve toward the “+” sign to increase the serum ow or towards the “–” sign to decrease the ow.

13
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
HydroPeel Tip
1. Align the HydroPeel tip and attach it to the handpiece.
Caution: Avoid dropping or damaging the disposable tip prior to treatment. Disposable tips suspected of being
damaged should be discarded and replaced. Damaged tips may cause scratches or discomfort to the client.
Single use only.
Solution Bottles (Sold separately)
1. Remove the cap of the skin solution bottle and use the puncture tool to pierce the seal inside the bottle.
2. Press and hold each manifold button as you insert each bottle into the manifold, and then release the button.
3. The Syndeo system solution bottle controls are located on the bottom of the screen within the treatment steps. The Syndeo system
automatically selects and deselects the recommended solution bottle based on the respective treatment step. To override the
system selection, simply select and deselect bottles by pressing the buttons on the touchscreen. Long press a button to select
multiple bottles at one time.

14
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
On, Off, and Sleep Modes
Turn ON the System
1. Connect the power cord to the Syndeo system and connect the plug into a grounded wall outlet.
2. Switch the power entry module switch to the ON ( I ) position.
3. The Syndeo system will boot up and the Start Up Screen will appear.
Sleep Mode
1. The Syndeo system enters sleep mode after one hour of idle time (time since last touching of the touchscreen). Idle time can be
adjusted in Settings.
2. To wake from sleep mode simply touch the monitor screen.
Turn OFF the System
To turn OFF the unit simply switch the power entry module to the OFF (O) position.
Note: The Syndeo system can typically be left on and connected to the internet.
Connectivity
•To ensure you receive regular software updates, connect your Syndeo unit to the internet (WiFi).
•If Wi-Fi is unavailable, internet connection can be achieved via ethernet or cellular service.
•Internet connection can be established during the initial setup process by following the onscreen prompts. Alternatively, the system
connectivity can be adjusted in Settings.
Prep Tray
The tray is enabled with smart technology that scans consumables placed in the tray and displays them in Pre-Treatment Screens under
the Personalization section. To easily personalize a treatment:
1. Place consumable vials (e.g. Boosters, Peels, etc.) into the tray prior to starting a treatment.

15
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Touchscreen
This section shows how to operate the touchscreen and navigate the system software. The user interface is designed to be intuitve
and consistent on each screen, therefore, only some screens will be shown below as an example. The touchscreen has been preset
for optimal performance with the HydraFacial program. Thus, you cannot and will not need to adjust the monitor.
1. Access each page by lightly tapping the buttons displayed with your ngertip, not ngernail. If the touchscreen does not respond,
verify that you have actually pressed on the button and did not miss it.
Note: There may be a slight delay when you press the touchscreen; do not press repeatedly if it does not respond immediately.
Initial Setup
Start-Up Page
Click the Get Started button to begin the onboarding process.
You will be asked to:
•Select language
•Connect to the internet
•Login using your HydraFacial Nation App user prole or continue as a guest
Follow the prompts on each screen to complete the setup process.
Navigating the System
Login
Navigate to the login screen and select your name from the available user proles.
Select New User and follow the prompts to create a new user prole or continue
as a guest. Click Continue to be brought to the Home screen.
Note: Select Edit Proles to delete proles from the login screen
From the Home
Screen, you can
initiate a treatment,
scan in a client,
explore the
HydraFacial treatment,
and consumable
options, or pull up
the menu.
Wi signal
Clock
Logged in user
Scan client
Access HydraFacial catalog
Select a treatment
Home
Menu
Home Screen

16
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
QR Reader
The QR reader allows you to scan a QR code to:
•Login as Provider or Esthetician
•Load a Client’s prole information
•Redeem a Gift Card
Note: Logging in as a provider or esthetician allows that user to receive credit for treatments performed. Scanning in a client
before a treatment allows the user to:
•View the client’s treatment history and sele assessment results.
•Add delivered treatment to the client’s treatment history.
Pre-Treatment Screens
Get to know your client’s history and skin status. Personalize and prep for their
treatment in Syndeo’s Pre-treatment Screens.
•Select Add a Treatment to add on another treatment for this visit.
•Select Edit Treatment to adjust the treatment plan for this visit.
•Select Edit Personalization to add or remove boosters and peels from the
treatment plan.
•View current Solution Levels in the attached solution bottles.
Client prole information
(accessible during client treatment only).
System Rinse Menu
Catalog
Settings Menu
Current Notications
Manual Mode (without Treatment Guidance).
Access QR Code Reader to Redeem Gift Card.
Sign Out of User Account
Access treatment overview screen to edit treatment
Treatment Type
Treatment Step
Adjust vacuum
Turn On/Off vacuum
Next/Previous Step
Access more information about displayed element
Treatment Screens
Menu

17
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Gesture Control Functionality
System Rinses
Access all System Rinse options by selecting System Rinses in the Menu.
Select the appropriate system rinse type and follow the on-screen prompts.
To minimize screen touching, gesture control functionality allows for intuitive and
quick in-treatment navigation and vacuum control.
Treatment Step Navigation
•Navigate to the next step in a treatment: Wave your hand from right to left
above one of the sensors.
•Navigate to the previous step in a treatment: Wave your hand from left to right
above one of the sensors.
Vacuum Control
•Turn on the vacuum: Wave your hand from bottom to top over the sensor on
the right or left side of the monitor.
•Subsequent waves from bottom to top over the sensor increases the vacuum
level one point.
•Subsequent waves from top to bottom over the sensor decreases the vacuum
level one point.
Pop-Up Windows
If you see a pop-up window, follow the on-screen instructions and select the
appropriate button to continue.
In-treatment Rinses
Run a quick system rinse when prompted during a treatment.
Select Begin Rinse to start the rinse.
Vacuum Setting
Vacuum Setting
Treatment Step
Note: Any movement within 15 inches of the sensors may be read. Optimal reading range is within 2 inches of the sensors.
Note: The sensors do not recognize black gloves due to 'stealth behavior' of the black color. Avoid wearing black gloves. Clear, pink,
blue or purple gloves may be worn.

18
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Perform Treatment
Wearing gloves is recommended when performing all treatments.
Pretreatment Screens
To begin a New Treatment select one from the Home Screen. Follow the prompts
to scan a client’s QR code. The QR code can be found in the prole section of the
client’s HydraFacial Nation app. The client’s information will be displayed. Select
Continue to proceed to the Your Treatment Overview screen.
Repeat a Treatment
To repeat a treatment from a client's history, select Scan QR Code on the Home Screen.
Follow the prompts
to scan a client’s QR code. The QR code can be found in the
prole section of the client’s HydraFacial Nation app. The client’s information will be
displayed. Select Treatments to view treatment history. Select a treatment from the
list. The Treatment
History details will be displayed. Select Do it Again to perform this
treatment again
and proceed to the Your Treatment Overview screen.
Adjust User Session Timeout
Change Language
Change WiFi
Add Credit Code
Access Software Version
View User Manual
View System Status
Software Updates
Note: To receive the latest software update, internet connection is required.
A notication will appear along the top of the screen when a new update is
available. Press the notication or select Menu then Notications and follow the
prompts to learn about, and update to, the latest software version.
Turn On/Off Gesture Control
Adjust Machine Color
Adjust Screen Brightness
Adjust Sound
Adjust Business Hours
Set Date & Time
Adjust Screen Idle Time
Settings

19
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Cleaning & Maintenance
Wearing gloves is recommended when performing all cleaning and maintenance activities.
Select System Rinses in the Menu to view.
#Gunkie Waste Canister Rinse
•After each treatment empty the waste from the #Gunkie waste canister. Rinse with water.
•DO NOT let the waste canister ll above the “Please Empty Me” line printed on the canister.
•At the end of each day, rinse the #Gunkie waste canister with soap and water. Let it air dry.
In-Treatment 20-Second Rinse
•Perform a 20-Second System Rinse after the Extraction step of each treatment as prompted in the Treatment ow. Ensure the
cleaning cap is attached to the handpiece.
After Each Treatment 20-Second Rinse
•Perform a 20-Second System Rinse after each treatment as prompted in the Treatment ow. Ensure the cleaning cap is attached
to the handpiece.
•Don’t forget to empty the #Gunkie waste canister and rinse it with water.
After Each Lymphatic Treatment
•Clean the lymphatic handpiece(s) with a germicidal solution on any area that makes contact with the skin.
End of Day 60-Second Rinse
•Perform a 60-Second System Rinse at the end of the day
•Don’t forget to rinse the #Gunkie waste canister with soap and water. Let it air dry.
Prepare for the treatment on the Your Treatment Overview Screen. You can edit the
personalization options and treatment directly by selecting the respective options.
Select Conrm Treatment when you are ready to begin the treatment.
Treatment
Note: Each treatment guides you through the suggested protocol steps, stroke patterns,
formulas, and vacuum settings.
1. Cleanse client's face.
2. Place a tip on the handpiece.
3. Turn on the vacuum.
4. Place tip against skin to occlude and dab skin solution onto skin.

20
User GuideHYDRAFACIAL Syndeo™
1900104-B
en
Handpiece Cleaning
The HydraFacial™handpiece tubing chambers must be cleaned and disinfected using the cleaning brush.
1. On the screen, select Menu. Then, select Handpiece Rinse from the System Rinses.
2. Open the control valve on the handpiece.
3. Detach the handpiece from all tubing by gently twisting the knobs counter-clockwise.
4. Insert the cleaning brush into the vacuum (smaller) opening on the face of the handpiece. Push the brush in and out fully while
rotating three times. Repeat on the blue twist lock connector opening on the back of the handpiece.
5. Wash the brush bristles in soapy water between each handpiece opening insertion to clean off debris.
Optional: Insert the compressed air dispenser into the blue twist lock connector and administer two sprays.
6. Clean the entire outer surface of the handpiece with a germicidal wipe, removing all visible soil. Use a second germicidal wipe
to thoroughly wet the surface of the handpiece and ensure the handpiece remains visibly wet for 2 minutes.
Note: Barbicide is a recommended disinfectant (follow Barbicide’s instructions for use). Isopropyl alcohol is not suitable
as a disinfectant.
7. Re-attach the tubing and cleaning cap to the handpiece.
8. Select Continue To Rinse on the screen to run a system rinse and follow the on-screen prompts.
This manual suits for next models
2
Table of contents
Other hydrafacial Personal Care Product manuals
Popular Personal Care Product manuals by other brands

Scarlett
Scarlett TOP Style SC-HAS73I07 instruction manual

ORTHOSERVICE RO+TEN
ORTHOSERVICE RO+TEN epi320 quick start guide

Beurer
Beurer US FW Instructions for use

HairMax
HairMax LaserComb Advanced 7 user manual

StarTech.com
StarTech.com WRSTRST quick start guide

Do Beauty
Do Beauty Starter Lizzy ND602 product manual