ICANCLAVE STE-18-D User manual

Instructions Manual
For Models: STE-18-D Pro
STE-23-D Pro

2
Thank you for choosing our steam sterilizer.
Prior to operating this instrument, please read the operations manual carefully and follow
all installation instructions.
Ningbo Ican Machines Co., Ltd.
No. 77 Yunlin East Road, Gulin Town,
Ningbo, China.
www.icanclave.com
European Representative:
Icanclave Europe S.L
Juan Ramon Jimenez 6
Quart de Poblet, Valencia, Spain.
Document: Version 00DP0000V2.9
Subject to technical changes
Save these instructions

3

4
Content
1. General
4
4 Scope of manual
4 Intended use
4 General safety instructions
4 Standards and directives
5 Symbols
2. Description of the sterilizer
5
5 Control panel
6 Sterilizer views
7 Technical specification
7 Packing content
3. Installation
7
7 General conditions
8 Power supply connection
8 Location requirements
4. Setup
8
8 Fill the distilled water tank
9 Preparation of the sterilization materials
10 Basic set
10 Advance set
5. Operation
13
13 Load the sterilization chamber
13 Select the program
13 Start the sterilization program
14 End of cycle
14 Manually interruption of the cycle
15 Delay start cycle
15 Test Program
16 Descale
16 Data
17 Print and save report (Optional)
18 Print Labels (Optional)
6. Maintenance
18
18 Clean the distilled water tank
18 Replacement of the bacteriological filter
19 Clean Chamber, trays and tray Rack
19 Door adjustment
19 Replacement of the door seal ring
19 The drain valve
7. Troubleshooting
20
20 Troubleshooting
8. Transportation and storage
21
21 Transportation and storage conditions
9. Safety devices
21
21 Safety device description
11. Appendix
22
22 Water properties/characteristics
23 Diagrams of the sterilization programs
1 General

5
Scope of Manual
This manual contains information concerning the installation, operation and
maintenance of the steam sterilizers. To ensure proper performance of the sterilizer, the
instructions given in this manual should be thoroughly understood and followed.
Keep the manual near the sterilizer in an accessible location for future reference.
Intended Use
The steam sterilizer described in this manual is intended for the sterilization in all
medical, dental, beauty, vet and tattoo fields of the following types of instrument loads:
solid, porous, hollow loads type A and hollow loads type B, un-wrapped, single wrapped
and double wrapped, liquid, that are compatible with steam sterilization.
General Safety Instructions
Read and understand this manual before attempting to install or operate th e
sterilizer.
Make sure that all the installation conditions are fully complied with.
Ensure that the voltage agrees with the supply voltage specified on the supply
on the type plate of the sterilizer.
This appliance must be grounded. Connect only to a properly grounded outlet.
Do not cover or block any openings on this appliance.
Use this appliance only for its intended use as described in this manual.
Do not exceed the maximum weight limit of the loads specified in this manual.
Do not operate this appliance if it has a damaged cord or plug if it is not working
properly or if it has been damaged or dropped.
Never put into the sterilizer inflammables or explosive products.
The sterilizer may not be operated in areas in which gas or any other explosive
volatile substance is present.
Installation and repair work should only be performed by authorized service
technicians. Work by unqualified persons could be dangerous and may void the
warranty.
Standards and directives
The steam sterilizers were designed and produced in conformity with the following
directives and standards:
Directives:
97/23/CE Pressure equipment.
93/42/EEC Medical devices (class II b).
Standards:
EN 13060 Relative to small steam sterilizers.
EN 61010-1 Safety regulations for laboratory devices - Part 1: General
regulations.
EN 61010-2-040 Safety regulations specific to sterilizers used in the processing
of medical material.
EN 61326-1 Electromagnetic compatibility regulations for laboratory devices.

6
Symbols
For safe operation, please pay close attention to the alert symbols below which can be
found on the sterilizer and throughout this manual.
Important information (Caution) Hot surface Ground
connection
2 Description of the sterilizer
Screen and Control panel
Door closed / Tap to open the
door
The door is open
Fill the tank if it is flashing
Min level of freshwater tank 0.5Liters
Drain the tank if it is flashing
The capacity is 1.5 liters
1 Report printer
Grey = No connected / no paper
Green =Ready
USB memory
Grey = No connected
Green =Ready
1
Time
2
Date
3
Process curve
4
Internal temperature
5
Internal pressure
6
Cycle time
7
Program information
8
Clean water tank
level
9
Used water tank
level
10
Printers
11
Door status
12
USB memory
13
Up button
14
Down button
15
Select button
16
Start button

7
Sterilizer views
1. Door
9. Touch pad
17. Power socket
2. Door seal ring
10. Printer
18. Steam vent
3. Door lock
11. USB port
19. Condenser vent
4. Door handle
12. Main switch
20. Bacterial filter
5. Chamber
13. Distilled water tank
vent
21. Used water tank
vent(optional)
6. Door interlock
14. Used water tank vent
22. Rating plate
7. Distilled water tank
15. Safety valve
8. Screen
16. Circuit breaker

8
Technical specifications
Model
STE-18-D Pro
STE-23-D Pro
Chamber (mm)
250 x 350
250 x 450
Overall dimensions (mm)
490(W)*455(H)*600(D)
490(W)*455(H)*690(D)
Net Weight (kg)
50
53
Nominal power (VA)
1750
1750
Rated Voltage
220-240V / 110-120V ; 50/60 Hz
Sterilization temperatures
121°C/134°C
Capacity of the distilled
water tank
2.5 L (Water at level Max.) Approx.
0.5L (Water at level Min.) Approx.
Circuit breaker
F16A /400 V
Operation temperature
5°C ~ 40°C
Operation relative humidity
Max. 80%, non-condensing
Max. Noise level
<70 dB
Atmospheric pressure
76 kPa ~ 106 kPa
Packing content
Item
Accessories
Quantity
1
Instruments tray
3
2
Instruments tray
rack
1
3
Draining hose
2
4
Door Seal
1
5
Tray handle
1
6
Door adjustment
wrench
1
7
Instructions Manual
1
3 Installation
General conditions
Position the device on a plane surface with a minimum capacity of 60 kgs.
The sterilizer should be placed on a level worktable.
Leave at least 10 cm between the device rear part and the wall. The clearance required
to open the door is 40cm.
Position the sterilizer at such a height as to make it possible for the operator to check the
whole sterilization chamber and carry out the normal cleaning operations.
The room where the device is installed must be sufficiently ventilated.

9
Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
Do not lean on the door when it is opened.
Do not place trays, papers, fluid containers or other objects on the sterilizer.
Power supply connection
Check the label on the back panel of sterilizer to verify the voltage rating for the unit. Failure to
connect the sterilizer to an appropriate power supply could result in damage to the unit and
electrical shock to personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only
used for the sterilizer is recommended. Never connect the device pin to reductions of any type.
Location requirements (mm)
4 Setup
Switch On and open the door to remove all of the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. After switching on, the machine turns on the LCD
and shows the door position, water level, working program, date, time, etc.
Fill the distilled water tank
Manual water filling
When the level of distilled water reaches a minimum level, the distilled water tank icon will flash
and beep three times.
Press the button on the tank lid and open it to the maximum position.
Fill it carefully with distilled water.
If exceeds the maximum level, an alarm will sound, and the distilled water tank icon will blink.

10
Drain the distilled water tank
Attach the drain hose on a fitting connection located on the front of the sterilizer. Pull the
connector to start the draining.
Attention: The capacity of the distilled water tank is approximately 2.5 liters.
Drain the used water tank
Attach the drain hose on a quick fitting connection located inside the service door at the left.
Attention: The capacity of the used water tank is approximately 2 liters.
Preparation of sterilization materials
For the most effective sterilization and to preserve the sample, please follow below:
Clean instruments immediately after use.
Treat the instruments by ultrasound cleaner.
Residual chemicals left over after cleaning and disinfecting process may damage and
corrode parts of the sterilizer, always rinse off the instruments using distilled water.
Follow instrument manufacturer’s guidelines and recommendations for handing and
cleaning instruments prior to sterilization.
Check the manufacturer’s instructions as to proper procedure for sterilizing of each item.
Arrange the samples of different materials on different trays or with at least 3cm of space
between them.
Clean and dry instruments thoroughly before placing them into tray.
Always insert a sterilization paper or cloth between the tray and sample to avoid direct
contact.
Arrange the containers (glasses, cups, test-tubes, etc.) on one side or inverted position,
avoiding possible water stagnation.
Don’t stack the trays one above the other or put them in direct contact with the walls of
the sterilization chamber.
Always use the instrument tray handle.
Wrap the samples one by one or, if more tools have to be set in the same bag, verify that
these are made of the same material.
Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the
sterilizer.
Don’t overload the trays over the stated limit (see appendix 2).

11
Basic Set
The “Basic Set” menu permits to set the following
options:
*Date *Time *Language
Select the “Basic Set” from the main menu by
pressing button.
Select the item by pressing button. The unit
you selected will be lighted.
Adjust the value by pressing buttons. Press
button to select the next item.
Press button to save and back.
Note: The Counter (cycle No) cannot be set by the
operator.
Advance Set
The “Setup” menu permits to set the following
options:
*Parameter *Unit *Preheat *Expire date (labels)
*Water quality (sensor) *Preheat *Factory reset
and see the information of: *Last error
Select “Setup” from the main menu by pressing
button.
Input the password digit to digit by pressing
and button to go next.
Password: 1111
Parameter
The “Parameter” menu permits to set the following
options:
*Holding time *Dry time
Select “Parameter” from the menu by pressing
button.
Select the program by pressing then press
button.
Select the parameter by pressing button. Adjust
it by pressing .
Press button to save and back.
Unit

12
Select “Unit” from the menu by pressing button.
Select the parameter by pressing button. Adjust
it by pressing .
Press button to save and back.
Preheat
When this mode is activated, the chamber and
steam generator start to warm until it reaches the
minimum temperature to begin a sterilization cycle,
this helps to reduce the total cycle time and the
drying efficiency. The “Preheat” mode will be
deactivated after one hour of inactivity.
Select “Preheat” from the menu by pressing
button.
Adjust it by pressing .
Press button to save and back.
Expiry date (Optional)
To modify the expiration date of the labels, select
“Expire date” from the menu by pressing button.
Adjust it by pressing .
Press button to save and back.

13
Water quality (optional)
If your sterilizer is equipped with a water
quality sensor and you want to deactivate it,
select “Water quality” from the menu by pressing
button.
Adjust it by pressing .
Press button to save and back.
Last Error
In order to help the technical assistance
process, the most relevant information
corresponding to the last error can be
displayed on the screen.
Select “Last error” from the menu by pressing
button.
Press button to back.
Factory reset
Select “Factory reset”
Press “Yes” to reset all settings
Press button to back.
5 Operation

14
Load
After selecting program, the materials to be sterilized can now be placed
on the tray and placed inside the chamber by the tray handle.
After the instruments are loaded, you may close the door.
Select the program
Enter to the main menu by pressing button,
Select “Program”. Select the program by
pressing then press to confirm program,
in the screen will appear the information of
selected program as the temperature and
sterilization time (holding time), also th e date,
time, current pressure and current
temperature.
Start the sterilization program
Press to start the cycle. The stage, conditions
and the status of the cycle will appear on the
display. The sterilizers will perform the program
automatically. (see appendix 2).

15
End of the cycle
Once the cycle is completed, “End” will appear at the end of the graphic,
the printer will print out and the digital report saved in the USB memory is
these are connected.
Caution: Always use the tray handle to load or unload the tray
into the sterilizer. Failure to do so can result in burning.
Manually interruption of the cycle
To interrupt a started cycle prematurely, hold for 3 seconds.
If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer
may be considered sterile and considering that the cycle has been interrupted during the drying
phase the materials and instruments inside the chamber may be wet.
Note: If the cycle is manually interrupted before it reaches the drying phase, the items inside the
sterilizer must be considered not sterile. N20 will appear on the screen. (see Error code
description).
Caution: Depending on the phase of the cycle, steam and water can escape from the
sterilization chamber when you open the door.
If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer
may be considered sterile.
Note: Considering that the cycle has been interrupted during the drying phase the materials and
instruments inside the chamber may be wet.
If the cycle is manually interrupted before it reaches the drying phase, the items inside the
sterilizer must be considered not sterile.
Caution: Depending on the phase of the cycle, steam and water can escape from the
sterilization chamber when you open the door.
Delay start cycle

16
Select “Delay start”. Select the program by pressing then press to confirm.
Select the parameter by pressing button. Adjust it by pressing .
Press button to save and back.
Test Programs
Helix Test
Put the Helix test device into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu,
select “Helix test”; in the screen will appear the information as temperature and
sterilization time (holding time), also the date, time, current pressure and current
temperature.
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.
B&D test
Put the B&D test package into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu,
select B&D test, in the screen will appear the information as the temperature and
sterilization time (holding time), also the date, time, current pressure and current
temperature.
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.
Vacuum Test

17
Select “Program” from the main menu by
pressing then to enter in the menu,
select “Vacuum test”.
After closing the door, press to start the cycle.
The stage, conditions and the status of the cycle
will appear on the display. The sterilizers will
perform the program automatically. (see appendix
2).
In compliance with EN 13060, the test requires that
the air leakage rate less than or equal to 0.13
kPa/min during 10 minutes.
If leakage rate is not greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will
show void. That means the result of the test is fail. You need run the vacuum test again after the
chamber has cooled down.
Descale
Mix the descaling powder with distilled water to make descaler according with the
instruction of the descaling powder manufacturer.
From the Programs menu select descale.
After closing the door, press to start the cycle. The stage, conditions and the status of the
cycle will appear on the display. The sterilizers will perform the program automatically. (see
appendix 2).
Follow the instructions in the screen.
Remove all instruments inside the chamber a nd close the door.
Empty the freshwater tank and fill with descaler which prepared as above, to 2/3 of the water
tank.
Press , start descaling, it will take around 25 minutes.
After descaling, empty the freshwater tank and fill with distilled water.
Press , start flushing, it will take around 8 minutes.
After flushing, the cycle “Descale” is end.
Data
The internal memory will store the information of the last 9999 cycles.

18
Print report (Optional)
Internal Memory
In this menu you can get the information of all the cycles stored in the internal memory of the
sterilizer.
Select “Report” from the main menu and press button, you will see the list of records.
Select the record by pressing button.
Press button to print and save the report.
Press button to exit.
E.g.
If the serial number of the steam sterilizer is
E00001 and the cycle number is 0012.
The file name of the file will be 01001200.log
The first two numbers represent machine
number.
The middle four numbers represent cycle
number.
The last two numbers rep resent error code.
00: no error; 01: error E01
Sample of a printer report
==============================
==
Program: WRAPPED
Temperature: 134C
Pressure: 206.0 kPa
Drying Time: 08Min
Holding Time: 4.0Min
--------------------------------
Time Temp. Pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -075.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -075.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature:135.2C
Min. Temperature:134.3C
Max. Pressure:214.0kPa
Min. Pressure:204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
End 13:04:07 102.4C
--------------------------------
Cycle No.: 00017
Ster. Value: Success
Date: 2017-06-07
SN:E54723
Operator:
v 2B00V2.5
==============================
==

19
Print labels (Optional)
Select “labels”, it will show cycle numbers.
Select the cycle number, press button to
print the labels.
6 Maintenance
To ensure proper operation and maximum steam sterilizer life, carefully follow all
recommendations for periodic maintenance.
One of the most important steps you can take to prevent problems with your sterilizer is to use
only distilled water.
Frequency
Number of cycles
Maintenance operation
Monthly
50
Clean the door seal
Clean the filter inside the chamber and in the
clean water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months
200
Clean the distilled water tank
Every year
800
Replace the door seal
Select Settings from the main menu and tap “Maintenance”, to enter in the menu. Select
the item to check the information.
Clean the distilled water tank
Disconnect the main cable.
Drain the tank completely using the drain connector at the back of the sterilizer and leave it
connected into the connector in an open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas difficult to reach
using and a distilled water.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled water,
and put it back in to the position.
Replacement of the bacteriological filter.
The bacteriological filter is in the back of the sterilizer. Unscrew
the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.

20
Clean chamber, door seal ring, trays, and tray rack.
Remove the trays and tray rack from the chamber.
Clean trays, rack and inside of the chamber with mild soap.
Rinse the trays, rack and inside of the chamber with a smooth cloth and distilled water.
Examine door seal for possible damage.
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials or substances in the chamber.
Failure to comply may cause damage to the chamber and/or other components.
Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the
surface.
Door adjustment
Under normal circumstances, the chamber door does not require
adjustments. However, if the seal fails (resulting in steam leaking from the
front of the chamber), you may adjust it.
Open the door, insert your finger into the bottom of the door, and pull the
ring to unlock the mechanism. Turn the lid counterclockwise to tighten it. If it
is too tight, you may also turn the lid clockwise to loosen it.
Replacement of the door seal ring
Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and tap in sequence as follows:
Press in the top and bottom
of the door seal.
Press in the left and right
sides of the door seal.
Press the remaining sections
of the seal.
Caution: Please ensure the chamber and the door are cold prior to replacing the seal ring.
The drain valves
To drain the water from the used water tank or to fill the tank from an external container, insert the
connector attached to the silicone tube until a click is heard. To disconnect the connector, press
the metal button.
This manual suits for next models
1
Table of contents
Other ICANCLAVE Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Thermo Scientific
Thermo Scientific Cimarec HP194515 Operation manual and parts list

Hanna Instruments
Hanna Instruments HI 731312 instruction manual

Bruker BioSpin
Bruker BioSpin NMR BVT3200 Technical manual

ThermoFisher Scientific
ThermoFisher Scientific QuantStudio 12K Flex OpenArray AccuFill... user guide

Lumen Dynamics Group
Lumen Dynamics Group OmniCure R2000 user guide

D-Link
D-Link DKT-200 Quick installation guide