ICANCLAVE STE-8-D User manual

Instructions Manual
For Models:
STE-8-D
STE-12-D
STE-18-D
STE-23-D
STE-29-D

2
Thank you for choosing our steam sterilizer.
Prior to operating this instrument, please read the operations manual carefully and
follow all installation instructions.
IMPORTANT NOTICE
If you can’t open the door, please unlock the door according to the instructions “How
to open the door in case of power outage” in the manual.
Need maintenance
If this picture appears on the screen when power On or E88 appears on the report,
please call your dealer or local maintenance service. Your steam sterilizer needs
general maintenance.
Ningbo Ican Machines Co., Ltd.
No. 77 Yunlin East Road, Gulin Town,
Ningbo, China.
www.icanclave.com
European Representative:
Icanclave Europe S.L
Juan Ramon Jimenez 6
Quart de Poblet, Valencia, Spain.
Document: Version 00D20000V2.10
Subject to technical changes
Save these instructions

3
Content
1. General
4
4 Scope of manual
4 Intended use
4 General safety instructions
4 Standards and directives
5 Symbols
2. Description of the sterilizer
5
5 Sterilizer views
6 Control panel
6 Technical specification
6 Packing content
3. Installation
7
7 General conditions
7 Power supply connection
7 Location requirements
4. Setup
8
8 Fill the distilled water tank
8 Preparation of the sterilization materials
9 Basic set
9. About device
9 Advance set
5. Operation
11
12 Load the sterilization chamber
12 Select the program
12 Start the sterilization program
12 End of cycle
12 Manually interruption of the cycle
13 Test Program
14 Data
14 Save report
14 Printer (optional)
16 Labels (optional)
6. Maintenance
16
16 Clean the distilled water tank
16 Replacement of the bacteriological filter
16 Clean Chamber, trays and tray Rack
17 Door adjustment
17 Replacement of the door seal ring
17 The drain valve
7. Troubleshooting
18
18 Error code description
8. Transportation and storage
19
19 Transportation and storage conditions
9. Safety devices
19
19. Safety device description
11. Appendix
20
20 Water properties/characteristics
21 Diagrams of the sterilization programs

4
1 General
Scope of Manual
This manual contains information concerning the installation, operation and maintenance of
the steam sterilizers. To ensure proper performance of the sterilizer, the instructions given in
this manual should be thoroughly understood and followed.
Keep the manual near the sterilizer in an accessible location for future reference.
Intended Use
The steam sterilizer described in this manual is intended for the sterilization in all medical,
dental, beauty, vet and tattoo fields of the following types of instrument loads: solid, porous,
hollow loads type A and hollow loads type B, un-wrapped, single wrapped and double
wrapped, liquid, that are compatible with steam sterilization.
General Safety Instructions
§ Read and understand this manual before attempting to install or operate the sterilizer.
§ Make sure that all the installation conditions are fully complied with.
§ Ensure that the voltage agrees with the supply voltage specified on the supply on the
type plate of the sterilizer.
§ This appliance must be grounded. Connect only to a properly grounded outlet.
§ Do not cover or block any openings on this appliance.
§ Use this appliance only for its intended use as described in this manual.
§ Do not exceed the maximum weight limit of the loads specified in this manual.
§ Do not operate this appliance if it has a damaged cord or plug if it is not working
properly or if it has been damaged or dropped.
§ Never put into the sterilizer inflammables or explosive products.
§ The sterilizer may not be operated in areas in which gas or any other explosive
volatile substance is present.
§ Installation and repair work should only be performed by authorized service
technicians. Work by unqualified persons could be dangerous and may void the
warranty.
Standards and directives
The steam sterilizers were designed and produced in conformity with the following directives
and standards:
Directives:
97/23/CE Pressure equipment.
93/42/EEC Medical devices (class II b).
Standards:
EN 13060 Relative to small steam sterilizers.
EN 61010-1 Safety regulations for laboratory devices - Part 1: General regulations.
EN 61010-2-040 Safety regulations specific to sterilizers used in the processing of
medical material.
EN 61326-1 Electromagnetic compatibility regulations for laboratory devices.

5
Symbols
For safe operation, please pay close attention to the alert symbols below which can be found
on the sterilizer and throughout this manual.
Important information (Caution) Hot surface Ground connection
2 Description of the sterilizer
Sterilizer views
1. Door
9. Control panel
16. Circuit breaker
2. Door seal ring
10. Printer
17. Power socket
3. Door lock
11. USB port
18. Used water tank vent
4. Door handle
12. Main switch
19. Condenser vent
5. Chamber
13. Distilled water outlet /
20. Bacteriological filter
6. Door switch
Distilled water inlet*
21. Distilled water drain*
7. Distilled water tank
14. Used water tank outlet
22. Rating plate
8. LCD
15. Safety valve
*Models equipped with external water filling function.
Date of manufacture
SN
Manufacture number
Manufacturer Name
Manufacturer Address
EC-Representative Name
EC-Representative Address

6
Control panel
Technical specifications
Model
STE-8-D
STE-12-D
STE-18-D
STE-23-D
STE-29-D
Chamber (mm)
Φ170 x 320
φ200 x 360
φ247 x 350
φ247 x 450
φ247 x 625
Overall dimensions (W*H*D)
420*370*525
420*370*595
490*455*600
490*455*690
490*455*890
Net Weight (kg)
37
41
47
53
65
Nominal power (VA)
1750
1750
1750
1750
2300
Rated Voltage
230-240V;50 Hz
Sterilization temperatures
121°C/134°C
Capacity of the distilled
water tank
2.5 L (Water at level Max.) Approx.
0.5 L (Water at level Min.) Approx.
Circuit breaker
F16A /400 V
Operation temperature
5°C ~ 40°C
Operation relative humidity
Max. 80%, non-condensing
Max. Noise level
<70 dB
Atmospheric pressure
76 kPa ~ 106 kPa
Packing content
Item
Accessories
Quantity
1
Instruments tray
STE-8-D: 2 units
STE-(12/18/23/29)-D: 3 units
2
Instruments tray rack
1
3
Draining hose
2
4
Door Seal
1
5
Try handle
1
6
Instructions Manual
1
1
Program temperature
2
Program
3
Holding time
4
Pressure
5
Distilled water / Fill it flashing.
Filling the tank
Bad water quality
6
Used water tank full / Drain if flashing
7
Printer status
8
Door open
Door closed
Door blocked
9
Time
10
Current internal temperature
11
Up button
12
Down button
13
Enter button
14
Menu button

7
3 Installation
General conditions
§ Position the device on a plane surface with a minimum capacity of 60 kgs.
§ The sterilizer should be placed on a level worktable.
§ Leave at least 10 cm between the device rear part and the wall. The clearance required to open
the door is 40cm.
§ Position the sterilizer at such a height as to make it possible for the operator to check the whole
sterilization chamber and carry out the normal cleaning operations.
§ The room where the device is installed must be sufficiently ventilated.
§ Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
§ Do not lean on the door when it is opened.
§ Do not place trays, papers, fluid containers or other objects on the sterilizer.
Power supply connection
Check the label on the back panel of sterilizer to verify the voltage rating for the unit. Failure to connect
the sterilizer to an appropriate power supply could result in damage to the unit and electrical shock to
personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used
for the sterilizer is recommended. Never connect the device pin to reductions of any type.
Location requirements and dimensions (mm)
Model
STE-8-D
STE-12-D
STE-18-D
STE-23-D
STE-29-D
A
100
100
100
100
100
B
542
582
595
685
875
C
552
592
605
695
885
D
350
350
400
400
400
E
100
100
100
100
100
F
367
367
450
450
450
G
50
50
50
50
50
H
410
410
483
483
483
I
50
50
50
50
50
J
130
130
135
130
130
K
360
400
405
500
690
L
52
52
55
55
55
M
330
330
403
403
403
N
330
330
403
403
403

8
4 Setup
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. Open the door to remove all of the inner contents for
unpacking. After switching on, the machine turns on the LCD and shows the door position, water level,
working program, date, time, etc.
Fill the distilled water tank
Manual water filling
When the level of distilled water reaches a minimum level, the distilled water tank icon will flash and beep
three times.
Press the button on the tank lid and open it to the maximum position.
Fill it carefully with distilled water.
If exceeds the maximum level, an alarm will sound, and the distilled water tank icon will blink.
Drain the used water tank
Attach the drain hose on the drain connector located inside the service
door at the left.
Attention: The capacity of the used water tank is approximately 1.5 liters
Preparation of sterilization materials
For the most effective sterilization and to preserve the sample, please follow below:
§ Clean instruments immediately after used.
§ Treat the instruments by ultrasound cleaner.
§ Residual chemicals left over after cleaning and disinfecting process may damage and corrode
parts of the sterilizer, always rinse off the instruments using distilled water.
§ Follow instrument manufacturer’s guidelines and recommendations for handing and cleaning
instruments prior to sterilization.
§ Check the manufacturer’s instructions as to proper procedure for sterilizing of each item.
§ Arrange the samples of different materials on different trays or with at least 3cm of space
between them.
§ Clean and dry instruments thoroughly before placing them into tray.
§ Always insert a sterilization paper or cloth between the tray and sample to avoid direct contact.
§ Arrange the containers (glasses, cups, test-tubes, etc.) on one side or inverted position,
avoiding possible water stagnation.
§ Don’t stack the trays one above the other or put them in direct contact with the walls of the
sterilization chamber.
§ Always use the instrument tray handle.
§ Wrap the samples one by one or, if more tools have to be set in the same bag, verify that these
are made of the same material.
§ Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the sterilizer.
§ Don’t overload the trays over the stated limit (see appendix 2).

9
Basic Set
From the main menu,, select "Basic Set”.
The “Basic Set” menu permits to set the following options:
*Date *Time *Language
Select the “Basic Set” from the main menu by pressing button.
Select the item by pressing button. The unit you selected will be lighted.
Adjust the value by pressing buttons. Press button to select the next item.
Press button to save and back.
Abbreviation of language options
Note: The Counter (cycle No) cannot be set by the operator.
About device
Select “About device” from the main menu then press button.
Press button to back.
Advance Set
The “Setup” menu permits to set the following options:
*Parameter *Unit *Preheat *Expire date (labels) *Water quality (sensor)
and see the information of: *Last error
Select “Setup” from the main menu by pressing button.
Input the password digit to digit by pressing and button to go next.
Password: 1111
Parameter
CHN
Chinese
ENG
English
DEU
German
ESP
Spanish
PL
Polish
FR
French
HUN
Hungary
ROM
Romanian
NL
Dutch
LTU
Lithuanian
LAT
Latvian
CZE
Czech
ITA
Italian
RUS
Russian
PT
Portuguese
HR
Croatian
Program
Basic Set
Report
Label
Date: 11-07-2019
Time: 12:05:35
Language: ENG
Counter: 00000
Parameter
Unit
Preheat
Expiry date
Water quality
Last error
About device
Password
1111
Device Info.
3BB12B 11111110
V2.9.0.1—00
SN: A09999B12
Basic Set
Report
Label
My device

10
The “Parameter” menu permits to set the following options:
*Holding time *Dry time
Select “Parameter” from the menu by pressing button.
Select the program by pressing then press button.
Select the parameter by pressing button. Adjust it by pressing .
Press button to save and back.
Unit
Select “Unit” from the menu by pressing button.
Select the parameter by pressing button. Adjust it by pressing .
Press button to save and back.
Preheat
When this mode is activated, the chamber and steam generator start to warm until it reaches the
minimum temperature to begin a sterilization cycle, this helps to reduce the total cycle time and the
drying efficiency. The “Preheat” mode will be deactivated after one hour of inactivity.
Select “Preheat” from the menu by pressing button.
Adjust it by pressing .
Press button to save and back.
Parameter
Unit
Preheat
Expiry date
Solid (121°C)
Solid (132°C)
Wrapped (121°C)
Wrapped (134°C)
Parameter
Unit
Preheat
Expiry date
Pressure: kPa
Temperature: °C
Parameter
Unit
Preheat
Expiry date
Preheat: off
Holding time: 20.0
Dry time: 03.5

11
Expiry date (Optional)
To modify the expiration date of the labels, select “Expire date” from the menu by pressing button.
Adjust it by pressing .
Press button to save and back.
Water quality (optional)
If your sterilizer is equipped with a water quality sensor and you want to deactivate it,
select “Water quality” from the menu by pressing button.
Adjust it by pressing .
Press button to save and back.
Last Error
In order to help the technical assistance process, the most relevant information
corresponding to the last error can be displayed on the screen.
Select “Last error” from the menu by pressing button.
Press button to back.
5 Operation
Prerequisites
Switch On.
Check the status of the icons in the screen . (Refer to point 2. Description of
the sterilizer)
Parameter
Unit
Preheat
Expiry date
Month: 03
Unit
Preheat
Expiry date
Water quality
W. quality: OFF
Preheat
Expiry date
Water quality
Last Error
Last error: E20
13-07-2019 17:00
PC:01 ST:00 CN:00042
Pressure: 100kPa
T1:070.8°C T3:040.0°C
T2:033.7°C T4:244.0°C
Initialize…
P: 07kPa T: 28.1°
Wrapped
134°C/3.5 Min
16-07-2019
P: 0.0kPa T:28.2°C
10:45:05

12
Load
Open the door then placed the trays inside the chamber by the tray handle.
After the instruments are loaded, you may close the door.
Select the program
Enter to the main menu by pressing button, Select “Program”. Select the program by
pressing then press to confirm program, in the screen will appear the information of
selected program as the temperature and sterilization time (holding time), also the date, time,
current pressure and current temperature.
Start the sterilization program
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
End of the cycle
Once the cycle is completed, “End” will appear at the end of the graphic, the printer will print out and the
digital report saved in the USB memory is these are connected.
Caution: Always use the tray handle to load or unload the tray into the sterilizer. Failure to do
so can result in burning.
Manually interruption of the cycle
To interrupt a started cycle prematurely, hold for 3 seconds.
If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer may be
considered sterile and considering that the cycle has been interrupted during the drying phase the
materials and instruments inside the chamber may be wet.
Program
Basic Set
Report
Label
Solid (121°C)
Solid (134°C)
Wrapped (121°C)
Wrapped (134°C)
Wrapped
134°C/3.5 Min
16-07-2019
P: 0.0kPa T:28.2°C
10:45:05
134°C Wrapped
P: 0.0kPa T:28.2°C
00:00:05
134°C Wrapped
P: -0.2kPa T:87.2°C
00:00:17
134°C Wrapped END
P: -0.2kPa T:87.2°C
00:00:17

13
Note: If the cycle is manually interrupted before it reaches the drying phase, the items inside the sterilizer
must be considered not sterile. N20 will appear on the screen. (see Error code description).
Caution: Depending on the phase of the cycle, steam and water can escape from the
sterilization chamber when you open the door.
Test Programs
Helix test
Put the Helix test device into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu, select
“Helix test”; in the screen will appear the information as temperature and sterilization time
(holding time), also the date, time, current pressure and current temperature.
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.
B&D test
Put the B&D test package into the chamber, then close the door.
Select “Program” from the main menu by pressing then to enter in the menu, select B&D
test, in the screen will appear the information as the temperature and sterilization time
(holding time), also the date, time, current pressure and current temperature.
Press to start the cycle. The stage, conditions and the status of the cycle will appear on the
display. The sterilizers will perform the program automatically. (see appendix 2).
After finishing the cycle, you may check the indicator and evaluate the result according with the
instructions of the test manufacturer.
Vacuum Test
Select “Program” from the main menu by pressing then to enter in the menu, select
“Vacuum test”.
After closing the door, press to start the cycle. The stage, conditions and the status of the cycle
will appear on the display. The sterilizers will perform the program automatically. (see appendix 2).
In compliance with EN 13060, the test requires that the air leakage rate less than or equal to 0.13
kPa/min. during 10 minutes.
If leakage rate is not greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will show
void. That means the result of the test is fail. You need run the vacuum test again after the chamber has
cooled down.

14
Data
The internal memory will store the information of the last 9999 cycles.
USB Flash memory (Optional)
A USB drive can be used as a method of storing a report of the cycle. To do so, insert the USB drive
into the slot located on the service door of the sterilizer.
The information will automatically output directly to the USB drive after the cycle has completed. The
name of the file is determined by the serial number of the machine and the cycle number.
For example:
The serial number is E00001. The cycle number is 0012.
The file name in the USB stick is 01001200.txt.
The first two numbers represent machine number.
The middle four numbers represent cycle number.
The last two numbers represent error code.
E.g. 00:no error; 01: error E01
Printer (Optional)
If installed, you can see the icon in the screen stop flashing.
At the end of each cycle the printer will print out a report of the cycle.
Note: is there is no paper inside the printer; the icon will flash.
Report
Internal Memory
In this menu you can get the information of all the cycles stored in the internal memory of the sterilizer.
Select “Report” from the main menu and press button, you will see the list of records.
Select the record by pressing button.
Press button to print and save the report.
Press button to exit.
Program
Basic Set
Report
Label
00012
00011
00010
00009

15
Sample of a printer report
When reading printed data records, refer to the diagram below:
================================
Program: WRAPPED
Temperature: 134C
Pressure: 206.0 kPa
Drying Time: 08Min
Holding Time: 4.0Min
--------------------------------
Time Temp. Pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -075.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -075.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature:135.2C
Min. Temperature:134.3C
Max. Pressure:214.0kPa
Min. Pressure:204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
End 13:04:07 102.4C
--------------------------------
Cycle No.: 00017
Ster. Value: Success
Date: 2017-06-07
SN:E54723
Operator:
v 2B00V2.5
================================
=======================
Program:Vacuum test
Tp:1℃
P1:-75.0kPa
P2:-74.0kPa
rate of pressure rise:0.10
Start Time:08:22
End Time:09:01
Date:2017-07-19
Test Value:Success
SN:E00001
Operator:
========================

16
Print labels
Select “Labels” from the main menu and press button to enter in the menu.
Select the cycle number by pressing button. Choose the labels quantity by pressing
then press button to print.
Press button to exit.
6 Maintenance
To ensure proper operation and maximum steam sterilizer life, carefully follow all recommendations for
periodic maintenance.
One of the most important steps you can take to prevent problems with your sterilizer is to use only
distilled water.
Frequency
Number of cycles
Maintenance operation
Monthly
50
Clean the door seal
Clean the filter inside the chamber and in the clean
water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months
200
Clean the distilled water tank
Replace the bacteriological filter
Every year
800
Replace the door seal
Clean the distilled water tank
Disconnect the main cable.
Drain the tank completely using the drain connector at the back of the sterilizer and leave it connected
into the connector in an open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas difficult to reach using
and a distilled water.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled water, and put it
back in to the position.
Replacement of the bacteriological filter.
The bacteriological filter is in the back of the sterilizer. Unscrew
the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.
Clean chamber, door seal ring, trays, and tray rack.
Remove the trays and tray rack from the chamber.
Clean trays, rack and inside of the chamber with mild soap.
Rinse the trays, rack and inside of the chamber with a smooth cloth and distilled water.
Examine door seal for possible damage.
Program
Basic Set
Report
Label
00012
00011
00010
00009
QT: 01

17
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials or substances in the chamber. Failure to
comply may cause damage to the chamber and/or other components.
Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the surface.
Door adjustment
Under normal circumstances, the chamber door does not require
adjustments. However, if the seal fails (resulting in steam leaking from the
front of the chamber), you may adjust it.
Open the door.
Insert the spanner tool in the gap beneath the plastic cover; use the spanner
to grip the adjusting nut. Turn the nut counter clockwise as the figure below.
This will tighten the sealing plate.
Turn the nut until the sealing plate is tight. If the door knob is too tight, you may also turn
the nut clockwise to loosen it.
Caution: Never adjust the chamber door while the door is closed.
Replacement of the door seal ring
Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and tap in sequence as follows:
Press in the top and bottom of
the door seal.
Press in the left and right sides
of the door seal.
Press the remaining sections of
the seal.
Caution: Please ensure the chamber and the door are cold prior to replacing the seal ring.
The drain valves
1. Press the included hose
on to the drain valve firmly.
2. Pull the drain valve
outward to drain the tank.
3. Push the drain valve back
after draining the tank.

18
7 Troubleshooting
Code
Description
Proposed solution
E1
Steam generator temperature sensor
error.
Power off & run a new cycle
Contact your supplier if error persists.
E2
Inner temperature sensor error.
Power off & run a new cycle
Contact your supplier if error persists.
E3
Temperature sensor of the chamber
wall error.
Carefully ensure that the chamber wall is heated and
contact your supplier.
E5
Fail to release the pressure.
Power off & run a new cycle
Contact your supplier if error persists.
E6
Door lock problem during the cycle.
Make sure you had closed the door properly.
check the door switch.
E7
Error between temperature and
pressure correlation.
Power off & run a new cycle
Contact your supplier if error persists.
E8
Error between temperature and
pressure correlation.
Power off & run a new cycle
Contact your supplier if error persists.
E9
Failure to hold temperature.
Ensure the distilled tank isn't empty. Check the inner
temperature sensor. Check somewhere for leaking.
E10
The door locking system doesn’t
work.
The electromagnet of locking system doesn’t work.
The switch of locking system doesn’t work.
E11
Failure to preheat the steam
generator.
Power off & run a new cycle
Contact your supplier if error persists.
E12
Failure to preheat the chamber.
Power off & run a new cycle
Contact your supplier if error persists.
E13
Vacuum failed.
Power off & run a new cycle
Contact your supplier if error persists.
E15
Inner temperature sensor error #2*
Power off & run a new cycle
Contact your supplier if error persists.
E16
Pressure error
Replace the air filter
Power off & run a new cycle
Contact your supplier if error persists.
N20
Program manually interrupted
Reset the error from main screen.
E22
Vacuum test failure
Somewhere is leaking. Check the door seal.
Or contact your supplier if error persists.
N23
Result of vacuum test is void
The temperature of the chamber is high.
Try again after the chamber has cooled down.
E24
It takes too long time to enter the next
status.
Check somewhere leaking.
Or contact your supplier if error persists.
N27
The vacuum test fails.
Switch off. Then switch on after the chamber cool
down and try again.
E28
The pressure is overload.
Power off and contact your supplier if error persists.
E30
Vacuum failed.
Power off & run a new cycle
Contact your supplier if error persists.
E31
Inner temperature sensor error #2*
Power off & run a new cycle
Contact your supplier if error persists.

19
8 Transportation and storage
Switch off the sterilizer before transportation or storage.
Pull out the plug. Let the machine cool down.
Drain the distilled water tank and the used water tank.
Conditions for transport and storage
Temperature: -20°C ~ +50°C
Relative humidity: ≤85%
Atmospheric pressure: 50kPa~ 106kPa.
9 Safety devices
1. Main breaker: Protection of the instrument against possible failures of the heating resistor.
Action: Interruption of the electric power supply.
2.Thermal cutouts on the main transformer winding: protection against possible short circuit and main
transformer primary winding overheating
Action: Temporary interruption of winding.
3.Safety valve: Protection against possible sterilization chamber over-pressure.
Action: Release of the steam and restoration of the safety pressure.
4.Safety micro-switch for the door status: Comparison for the correct closing position of the door.
Action: Signal of the wrong position of the door
5.Thermostat on chamber heating resistors: Protection for possible overheating of the chamber heating
resistors.
Action: Interruption of the power supply of the chamber resistors.
6.Thermostat on steam generator heating resistors: Protection for possible overheating of the steam
generator heating resistors.
Action: Interruption of the power supply of the steam generator resistors.
7.Door safety lock: Protection against accidental opening of the door.
Action: Impediment of the accidental opening if the door during the program.
8.Self-leveling hydraulic system: Hydraulic system for the natural pressure leveling in case of manual
cycle interruption, alarm or blackout.
Action: Automatic restoration of the atmospheric pressure inside the chamber.

20
Appendix 1
Water properties / Characteristics
Description
Feed water
Condensate
Evaporate residue
≤10mg/ I
≤1.0mg/kg
Silicium oxide sio2
≤1mg/ I
≤1.0mg/kg
Iron
≤0.2mg/ I
≤0.1mg/kg
Cadmiun
≤0.005mg/ I
≤0.05mg/kg
Lead
≤0.05mg/ I
≤0.1mg/kg
Rest of heavy metals
≤0.1mg/ I
≤0.1mg/kg
Chloride
≤2mg/ I
≤0.1mg/kg
Phosphates
≤0.5 mg/ I
≤0.1mg/kg
Conductivity
≤15μs /cm
≤3 μs /cm
PH Value
5 – 7.5
5-7
Appearance
Colorless, clean
Colorless, clean
Hardness
0.02 mmol/ I
0.02 mmol/ I
This manual suits for next models
4
Table of contents
Other ICANCLAVE Laboratory Equipment manuals