ICANCLAVE STE-23-D User manual

STEAM STERILIZERS
Instructions Manual
For Model: STE-23-D


Thank you for choosing our steam sterilizers.
Prior to operating this instrument, please read the operations manual carefully and follow
all installation instructions.
IMPORTANT NOTICE:
If you can't open the door, please unlock the door according to the instruction “How to open the door
in the case of power outage” in the manual.
Need Maintenance
If this picture appears on the screen when power on or E88 appears on the report, please call your
dealer or local service maintenance. Your steam sterilizer needs a regular maintenance.
Instructions Manual
Icanclave STE-23-D
Ningbo Ican Machines Co., Ltd.
No. 77 Yunlin East Road, Gulin Town, Ningbo, China.
www.icanclave.com
European Representative
Icanclave Europe, S.L.
Juan Ramon Jimenez 6,
Quart de Poblet, Valencia, Spain
Document: Version 05D20000 v2.8
Subjects to technical changes

Table of Contents
1. General .................................................................................................................................................... 1
1.1 Scope of manual ........................... ............................................................................................................. 1
1.2 Intended use .............................. ........................................... .................................................................... 1
1.3 General Safety Instructions ........................................... ............................................................................. 1
1.4 Standards and directives ....................... ......................... ......................... ................................................. 1
1.5 Symbols ............................................................ ..... .... ............................................................................... 1
2. Description of the autoclave ........................................ ..... ......... ........................................................... 2
2.1 Autoclave views ........................................ ..... ......... ............................................................. .................. 2
2.2 Control panel ........................................ ..... ......... ................................................. .................................. 2
2.3 Technical Specifications ........................................ .......... ........................................................................... 3
2.4 Packing content ................................... ..... ......... ............................................... ........................................ 3
3. Installation ........................................ ..... ......... ......................................................... .......................... 4
3.1 General conditions ................................... ..... ......... .............................................................................. 4
3.2 Power supply connection .................................. ..... ......... ....................................................................... 4
4. Setup ..................................... . ....................... ....................................................................................... 4
4.1 Basic set ..................... ....................... ................. ..................................................................................... 5
4.2 Fill the distilled Water Tank ....................... ....................... ....................... ................................................ 5
4.3 Preparation of the sterilization materials ................. ....................... ........................................................ 5
5. Operation ....................................... ...................... ...................... .......................................................... 6
5.1 Select the program ....................... ...................... ...................... ............................................................... 6
5.2 Running the sterilization program ......................... ...................... ................. ........................................... 6
5.3 Start the sterilization program ............................ ...................... ............... ................................................ 6
5.4 End of cycle .......................................... ...................... ...................... ........................................................ 6
5.5 Manual abort the program .................. ...................... ................. ................................................................ 7
5.6 Test Programs ...................................... ...................... ...................... ......................................................... 7
5.7 Record of the cycle .................................................... ...................... ............. ............................................ 7
5.8 Printer ...................................... ...................... ...................... .................................................................... 8
5.9 Report ..................................... ...................... ...................... ..................................................................... 8
6. Advance settings ................................................................ .................. ..... ............................................. 10
6.1 Parameter ....................................... ...................... ...................... .............................................................. 10
6.2 Unit ........................................ ...................... ...................... ...................................................................... 10
6.3 Preheat .................................. ...................... ...................... ....................................................................... 10
7. Maintenance ..................................................... ...................... ...................... ........................................ 11
7.1 Clean the distilled water tank ....................... ...................... ...................................................................... 11
7.2 Replacement of the bacteriological filter ......... ...................... ...................... ............................................ 11
7.3 Clean Chamber, door seal, trays and tray Rack ............... ............. ........ ...... ............................................. 11
7.4 Door adjustment ............................. ...................... ...................... ........................................................... 12
7.5 Replacement of the door seal .............. ...................... ...................... ........................................................ 12
7.6 The drain valve .............................. ...................... ...................... ............................................................... 12
8. Error codes ....... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...................................................................... 13
9. Transportation and storage .............. ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ................................... 14
10. Safety devices .................. ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ................................................. 14
Appendix 1 - Water properties / Characteristics ......... ... ... ... ... ... ... ... .. ... ... ... ...... ... ... ... ... .................. 15
Appendix 2 - Diagrams of the sterilization programs ....... .............. ... ... ... ... .................... .......................... 16
Appendix 3 - Wiring diagram ................................ .................... ........... ... ... .............................................. 18
Appendix 4 - Hydraulic diagram .................... .................... ............... ...... ................................................... 19

Instructions Manual
1
1. General
1.1 Scope of Manual
This manual contains information concerning the installation, operation and maintenance of the steam sterilizers. To ensure
proper performance of the autoclave, the instructions given in this manual should be thoroughly understood and followed.
Keep the manual near to the sterilization in a readily accessible location for future reference.
1.2 Intended Use
The device designed for total elimination and/or inactivation of microorganisms from medical devices and related products,
placed in sterilization wraps/packaging, using pressurized steam(i.e., moist heat) as the sterilizing agent; it is used for
products non-sensitive to high temperature, water, or steam.
1.3 General Safety Instructions
-Read and understand this manual before attempting to install or operate the sterilizer.
-Make sure that all the installation conditions are fully complied with.
-Ensure that the supply voltage agrees with the supply voltage specified on the type plate of the sterilizer.
-This appliance must be grounded. Connect only to a properly grounded outlet.
-Do not cover or block any openings on this appliance.
-Use this appliance only for its intended use a described in this manual.
-Do not exceed the maximum weight limit of the loads specified in this manual.
-Do not operate this appliance if it has a damaged cord or plug, if it is not working properly, or if it has been damaged or
dropped.
-Never must put into the sterilizer in flammables or explosives products.
-The sterilizer may not be operated in areas in which gas or any other explosive volatile substance is present.
-Installation and repair work should only be performed by authorized service technician. Work by unqualified persons could
be dangerous and void warranty.
1.4 Standards and directives
The steam sterilizers were designed and have been produced in conformity with the following directives and standards:
Directives:
2014/68/EC Pressure equipment.
93/42/EEC Medical devices (ClassⅡb).
Standards:
EN13060 Relative to small steam sterilizers.
EN61010-1 Safety regulations for laboratory devices-Part 1:General regulations.
EN61010-2-040 Safety regulations specific to sterilizers used in the processing of medical material.
EN61326-1 Electromagnetic compatibility regulations for laboratory devices.
1.5 Symbols
For safe operation, please pay close attention to the alert symbols below which can be found in the sterilizer or throughout
this manual.
This symbol represents an electrical caution - ground protection.
Hot Surface
This symbol represents a warning of a potential hot surface.
Important safety information.
This symbol represents a warning for extra caution.

Instructions Manual
2
2. Description of the sterilizer
2.1 Sterilizer views
1. Door
8. LCD
15. Safety valve
2. Seal
9. Control Panel
16. Circuit breaker
3. Door lock
10. Printer (Optional)
17. Power socket
4. Door handle
11. USB port
18. Used water tank vent
5. Chamber
12. Main power switch
19. Condenser vent
6. Door spindle
13. Drain connector (Distilled water tank)
20. Bacteriological filter
7. Distilled water tank
14. Drain connector (Used water tank)
21. Rating plate
2.2 Control panel
.
1
Temperature of the cycle
2
Program
3
Holding time
4
Pressure
5
Fill distilled water tank
6
Drain used water tank
7
Printer is connected
8
Door is closed
9
Time
10
Current temperature
11
Up button
12
Down button
13
Enter button
14
Menu button

Instructions Manual
3
2.3 Technical specifications
Item
Parameter
Chamber
φ247mm x 450mm
Rated Voltage
220-240 VAC; 50 Hz
Circuit breaker
F16A/400V
Nominal power
1750VA
Sterilization temperature
121°C /134°C
Capacity of the distilled water tank
Approx 2.5 L (Water at level Max)
Approx 0.5 L (Water at level Min)
Operation temperature
5°C-40°C
Operation relative Humidity
Max. 80%, non condensing
Overall dimensions
490mm(W)*455mm(H)*690mm(D)
Net weight
53 kg
Max. Noise level
<70 dB
Atmospheric pressure
76 kPa - 106 kPa
2.4 Packing content
No.
Accessory
Quantity
1
Steam sterilizer
1
2
Instrument tray
3
3
Instrument tray rack
1
4
Instrument tray handle
1
5
Door adjustment tool
1
6
Draining hose
2
7
Instructions manual
1
8
Door seal
1

Instructions Manual
4
3. Installation
3.1 General conditions
Position the device on a plane surface with minimum capacity 60 kgs.
The sterilizer should be placed on a level worktable.
Improper water level in the chamber could cause a sterilizer malfunction.
Leave at least 10cm between the device rear part and the wall. The clearance required to open the door is 40cm.
Position the autoclave at such a height as to make it possible for the operator to check the whole sterilization chamber and
carry out the normal cleaning operations.
The room where the device is installed must be enough ventilated.
Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
Do not lean on the door when it is opened.
Do not place trays , papers, fluid containers, etc. on the sterilizer.
3. 2 Power supply connection
Check the label on back panel o sterilizer to verify voltage rating for the unit. Failure to connect the autoclave to an
appropriate power supply could result in damage to the unit, and electrical shock to personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used for the sterilizer is
recommended.
Never connect the device pin to reductions of any type.
4 Setup
Open the door and remove all of the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. After switching on, the machine turns on the LCD and shows the door
position, water level, working program, date, time and etc.
Note: The control panel will be locked for the initial 10 seconds after powering up for system initialization.
Notice: Before using the sterilizer or at any time the low water level icon blinks, fill the distilled water tank with
distilled water.
Allow Clearance
On Both Sides

Instructions Manual
5
4.1 Basic Set
The “Basic Set” Menu permits to set the following options:
*Date *Time *Language
Select the “Basic Set” from the main menu and Tap button.
Select the item by taping the button. The unit you selected will be lighted.
Adjust the value by Taping button. Tap button to the next item.
After the language is set, then the data is permitted to be saved by taping button until
the main menu appears.
Note: If you want to cancel the setting and return to the main menu, tap button.
Note: The Counter (cycle No) can not be set by the operator.
Abbreviation of language options
4.2 Fill the distilled water tank
Ensure that the drain valve is closed.
Tap the button and open the water tank cover.
Use only high quality distilled water. (see Appendix 1)
4.3 Preparation of sterilization materials
For the most effective sterilization and to preserve the sample, please follow below:
*Clean instruments immediately after use.
*Treat the instruments by ultrasound cleaner.
*Residual chemicals left over after cleaning and disinfecting process may damage and corrode parts of the autoclave, always
rinse off the instruments using distilled water.
*Follow instrument manufacturer’s guidelines and recommendations for handing and cleaning instruments prior to
sterilization.
*Check the manufacturer’s instructions as to proper procedure for sterilizing of each item.
*Arrange the samples of different materials on different trays or with at least 3cm of space between them.
*Clean and dry instruments thoroughly before placing them into tray.
*Always insert a sterilization paper or cloth between the tray and sample to avoid direct contact.
*Arrange the containers (glasses, cups, test-tubes, etc) on one side or inverted position, avoiding possible water stagnation.
*Don’t stack the trays one above the other or put them in direct contact with the walls of the sterilization chamber.
*Always use the instrument tray handle.
CHN
Chinese
ENG
English
DEU
German
ESP
Spanish
PL
Polish
FR
French
HUN
Hungary
ROM
Romanian
NL
Dutch
LTU
Lithuanian
LAT
Latvian
CZE
Czech
ITA
Italian
RUS
Russian
Program
Basic Set
Report
DATE:23-04-17
TIME:09:10:08
LANGUAGE:ENG
Counter:12

Instructions Manual
6
*Wrap the samples one by one or, if more tools have to be set in the same bag, verify that these are made of the same
material.
*Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the autoclave.
*Don’t overload the trays over the stated limit (see appendix 2).
5. Operation
5.1 Select the program
Tap button to the main menu, select “Program”. You will see the available sterilization programs. See Appendix 2.
Select the program by buttons and confirm it by Taping button. If you don’t want to select a program you may
Tap button to exit.
5.2 Running the sterilization program.
After selecting program, the materials to be sterilized can now be placed on the tray, placed
inside the chamber by the tray handle.
After the instruments are loaded, you may close the door.
5.3 Start the sterilization program.
After the button is Taped, the stage and the status of the current cycle will appear on
the display. The sterilizers will begin the cycle automatically. It will take 30-75 minutes (see appendix 2).
5.4 End of cycle
After cycle is completed, the printer will be activated and print out a report of the cycle (if the optional printer has been
connected) or save the report in the USB drive (optional).
Caution: Always use the tray handle to load or unload the tray into the autoclave.
Failure to do so can result in burning.
Program
Basic Set
Report
SOLID(121℃)
SOLID(134℃)
WRAPPED(121℃)
WRAPPED(134℃)
Total time or count
down until completion

Instructions Manual
7
5.5 Manual abort of the program
It is possible to interrupt a started cycle prematurely. If you need to interrupt a cycle and remove the items urgently, you
may hold the button for 3 seconds during the drying time to skip the dry cycle.
If you interrupt a cycle before it reaches the “Drying” step, then the items inside the autoclave must be considered
not sterile.
If you need to interrupt a cycle after the holding time of the sterilization cycle and during the drying step then the
items inside the autoclave can be considered sterile.
Caution: Depending on the status of the Cycle, steam can escape from the sterilization chamber when you open the
door.
5.6 Test programs
Bowie & Dick test
Select ‘programs’ from the main menu, Tap button and select the “B&D” Test and Tap
button to confirm.
Put the Bowie-Dick pack in to the chamber and follow the instructions of the test pack
manufacturer. Then close the door and Tap button.
After the cycle is finished you may check the indicator and evaluate the result.
Helix Test
Select the “Helix” test from the "Programs" menu.
Following the instructions of the Helix test manufacturer. Put the indicator paper in the capsule.
Put the Helix test into the chamber, then close the door and Tap button.
After finish the cycle you may check the indicator and evaluate the result according with the
instructions of the test manufacture.
Vacuum Test
Select the Vacuum Test cycle from the "Programs" menu.
Close the door and tap button.
In compliance with EN 13060, the test requires that the air leakage rate less than or equal to 0.13
kPa/min. During 10 minutes.
If leakage rate is not greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will
show void. That means the result of the test is fail. You need run the vacuum test again after the
chamber has cooled down.
Caution: The “Vacuum” test must be carried out with the chamber of the sterilizer dry and cold.
5.7 Record of the cycle
USB Flash memory (Optional)
A USB drive can be used as a method of storing a report of the cycle. To do so, insert the USB drive into the slot located on
the service door of the sterilizer.
The information will automatically output directly to the USB drive after the cycle has completed. The name of the file is
determined by the serial number of the machine and the cycle number.
For example:
The serial number is E00001. The cycle number is 0012.
The file name in the USB stick is 01001200.txt.
The first two numbers represent machine number.
PRION
B&D TEST
HELIX TEST
VACUUM TEST

Instructions Manual
8
The middle four numbers represent cycle number.
The last two numbers represent error code.
E.g. 00:no error;01: error E01
5.8 Printer (Optional)
If installed you can see the Icon in the screen stop flashing (see Installation), at the end of each cycle
The printer (Optional) will produce a report of the cycle that just ended. t the end of each cycle the
printer will print out a report of the cycle.
5.9 Report
Internal Memory
In this menu you can read the latest information of the last 20 cycles stored in the internal
memory of the sterilizer.
5.9.1 Select “Report” from the main menu and Tap button, you will see the list of records.
5.9.2 Select the record by Taping button.
5.9.3 Tap button.
Then the record is printed(if the optional printer has been connected) the report or/and is
saved in the USB drive (optional).
Note: It can save in the internal memory only the reports of the last 20 cycles.
Note: The storage system is based on the principle of “first In-first-\out”.
Tap button to exit.
When vising printed data records, refer to the diagram below:
Program
Basic Set
Report
00012
00013
00014
00015

Instructions Manual
9
================================
Program: WRAPPED
Temperature: 134C
Pressure: 206.0 kPa
Drying Time: 08Min
Holding Time: 4.0Min
--------------------------------
Time Temp. Pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -075.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -075.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature:135.2C
Min. Temperature:134.3C
Max. Pressure:214.0kPa
Min. Pressure:204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
End 13:04:07 102.4C
--------------------------------
Cycle No.: 00017
Ster. Value: Success
Date: 2017-06-07
SN:E54723
Operator:
v 2B00V2.5
================================
=======================
Program:Vacuum test
Tp:1℃
P1:-75.0kPa
P2:-74.0kPa
rate of pressure rise:0.10
Start Time:08:22
End Time:09:01
Date:2017-07-19
Test Value:Success
SN:E00001
Operator:
========================

Instructions Manual
10
6. Advance setting
The advance setting interface permits to set the following
Options:
*Parameter: Permits to change the time of Holding time and Drying time.
*Unit: Permits to change the unit of measure temperature and pressure.
*Preheat: This option allows you to maintain the required temperature in the sterilization chamber and the steam generator
to start a new cycle immediately for the next 60 minutes so to run a new cycles should expect preheating time from 3 to 5
minutes upon start.
If the option is disabled (OFF) once the sterilization cycle end immediately the Autoclave is no longer heated, so when you
start a new cycle should expect a preheating time between 3 to 5minutes upon start.
Note: To maintain the temperature for longer time it is recommended that after each cycle hold the door closed.
Enter the setting
Power on the machine while holding the button and hold for 5 seconds. This will
enter into the advanced setting mode.
6.1 Parameter
Tap button to select Parameter. Tap button to enter the menu.
Select the program that you need to change by Taping button. Tap button to
enter the setting.
Adjust the parameter by taping the button.
Drying time:0-20 minutes
121℃holding time: 20-60 minutes
134℃holding Time: 4-20 minutes
After you finish adjusting the parameter Tap button to save and return to the
above menu. Tap button to cancel and exit.
6.2 Unit
Select the “Unit” to adjust the unit of temperature and pressure.
Tap button to enter the menu.
Select the item by Taping button.
Change the unit by Taping button.
Pressure: kPa/bar/psi
Temperature: ℃/F
Tap button to save and exit. Cancel and Exit by Taping button.
6.3 Preheat
Select the “Preheat” to adjust the Preheat.
Tap button to enter the menu.
If you don’t want to preheat after switch on, set the value to OFF by Taping button.
Tap button to save. Tap button to cancel and exit.
Parameter
Unit
Preheat
SOLID(121℃)
SOLID(134℃)
WRAPPED(121℃)
WRAPPED(134℃)
Holding Time:20.0
Dry Time:02
Pressure: kPa
Temperature: C
Preheat: ON

Instructions Manual
11
7.Maintenance
To assure proper operation and maximum autoclave life, carefully follow all recommendations for periodic maintenance.
One of the MOST important steps you can take to prevent problems with your sterilizer is to use ONLY distilled water.
Frequency
Number of cycles
Maintenance operation
Monthly
50
Clean the door seal
Clean the filter inside the chamber and in the clean
water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months
200
Clean the distilled water tank
Every year
800
Replace the door seal
7.1 Clean the distilled water tank
Disconnect the main cable.
Drain the tank completely using the drain tube and leave it connected into the connector in a
open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas that are
difficult to reach using and a mild soap.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled
water and put it back in to the position.
7.2 Replacement of the bacteriological filter.
The bacteriological filter is in the back of the sterilizer. Unscrew
the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.
7.3 Clean Chamber, door seal, trays and tray Rack.
Remove the trays and tray rack from the chamber.
Clean trays, rack and inside of chamber with mild soap.
Rinse the trays, rack and inside of chamber with a smooth cloth and
distilled water.
Examine door seal for possible damage.
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials / substances
in chamber. Failure to comply may result in damage to the chamber and/or other components.
Caution: To prevent burns, let unit to cool before cleaning gaskets and touch the surface.
7.4 Door adjustment
Under normal circumstances the chamber door does not require adjustments. However, if
the seal fails (resulting in steam leaking from the front of the chamber), you may use the
spanner tool to tighten the door seal.

Instructions Manual
12
Open the door.
Insert the spanner tool in the gap beneath the plastic cover; use the spanner to grip the
adjusting nut. Turn the nut counter clockwise as the figure below. This will tighten the sealing
plate.
Turn the nut until the sealing plate is tight. If the door knob is too tight, you may also turn
the nut clockwise to loosen it.
Caution: Never adjust the chamber door while the door is closed.
7.5 Replacement of the door seal
Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and Tap in sequence as follows:
1) Press in the top and bottom
of the door seal.
2) Press in the left and right
sides of the door seal.
3) Press the remaining
sections of the seal .
Caution: Please ensure the chamber and the door have cooled prior to replacing the seal ring.
7.6 The drain valve
7.7 How to open the door in the case of power outage
1.Press the included hose on
to the drain valve firmly.
2.Pull the drain valve
outward to drain the tank.
3. Push the drain valve back
after draining the tank.
1.Put the spanner into the door.
2.Press the button by the spanner and pull
the door handle at the same time.

Instructions Manual
13
8.Error codes
Code
Description
Proposed solution
E1
Steam generator temperature sensor
error
Power off & run a new cycle
Contact your supplier if error persists.
E2
Inner temperature sensor error
Power off & run a new cycle
Contact your supplier if error persists.
E3
Temperature sensor of the chamber
wall error
Carefully ensure that the chamber wall is heated and
contact your supplier
E5
Fail to release the pressure
Power off & run a new cycle
Contact your supplier if error persists.
E6
Door lock problem during the cycle
Make sure you had closed the door properly.
check the door switch
E9
Failure to hold temperature
Ensure the distilled tank isn't empty. Check the inner
temperature sensor. Check somewhere for leaking.
E10
The door locking system doesn’t work.
The electromagnet of locking system doesn’t work.
The switch of locking system doesn’t work.
E11
Failure to preheat the steam
generator
Power off & run a new cycle
Contact your supplier if error persists.
E12
Failure to preheat the chamber
Power off & run a new cycle
Contact your supplier if error persists.
E13
Vacuum failed
Power off & run a new cycle
Contact your supplier if error persists.
E20
Program manually interrupted
Holding the button for 3 seconds
E21
Failure to reach the holding time.
(sterilization time)
Check somewhere leaking inside the autoclave.
Contact your supplier if error persist
E22
Vacuum test failure
Somewhere is leaking. Check the door seal.
Or contact your supplier if error persists.
E23
Result of vacuum test is void
The temperature of the chamber is high.
Try again after the chamber has cooled down.
E24
It takes too long time to enter the next
status.
Check somewhere leaking.
Or contact your supplier if error persists.
E27
The vacuum test is forbade
Switch off. Then switch on after the chamber cool
down and try again.
E28
The pressure is overload.
Power off and contact your supplier if error persists.
Caution: You may cancel the voice of alarm by taping any button. And cancel the alarm by holding the button for 3
seconds after you repair it. Then swith off and switch on again.

Instructions Manual
14
9. Transportation and storage
9.1 Switch off the sterilizer before transportation or storage.
9.2 Pull out the plug. Let the machine cool down.
9.3 Drain the distilled water tank and the used water tank.
Condition for transport and storage
Temperature: -20°C ~ +50°C
Relative humidity: ≤ 85%
Atmospheric pressure: 50kPa~ 106kPa.
10. Safety devices
1.Main fuses: Protection the instrument against possible failures of the heating resistor.
Action: Interruption of the electric power supply.
2.Thermal cutouts on the main transformer winding: protection against possible short circuit and main transformer primary
winding overheating
Action: Temporary interruption of winding.
3.Safety valve: Protection against possible sterilization chamber over-pressure.
Action: Release of the steam and restoration of the safety pressure.
4.Safety micro-switch for the door status: Comparison for the correct closing position of the door.
Action: Signal of the wrong position of the door
5.Thermostat on chamber heating resistors: Protection for possible over heating of the chamber heating resistors.
Action: Interruption of the power supply of the chamber resistors.
6.Thermostat on steam generator heating resistors: Protection for possible overheating of the steam generator heating
resistors.
Action: Interruption of the power supply of the steam generator resistors.
7.Door safety lock: Protection against accidental opening of the door.
Action: Impediment of the accidental opening if the door during the program.
8.Self-leveling hydraulic system: Hydraulic system for the natural pressure leveling in case of manual cycle interruption,
alarm or black-out.
Action: Automatic restoration of the atmospheric pressure inside chamber.

Instructions Manual
15
Appendix 1
Water properties / Characteristics
Description
Feed water
Condensate
Evaporate residue
≤ 10mg/ I
≤ 1.0mg/kg
Silicium oxide sio2
≤ 1mg/ I
≤ 1.0mg/kg
Iron
≤ 0.2mg/ I
≤ 0.1mg/kg
Cadmiun
≤ 0.005mg/ I
≤ 0.05mg/kg
Lead
≤ 0.05mg/ I
≤ 0.1mg/kg
Rest of heavy metals
≤ 0.1mg/ I
≤ 0.1mg/kg
Chloride
≤ 2mg/ I
≤ 0.1mg/kg
Phosphates
≤ 0.5 mg/ I
≤ 0.1mg/kg
Conductivity
≤ 15μs /cm
≤ 3 μs /cm
PH Value
5 – 7.5
5-7
Appearance
Colorless, clean
Colorless, clean
Hardness
0.02 mmol/ I
0.02 mmol/ I

Instructions Manual
16
Appendix 2
Diagrams of the sterilization programs
The time required for sterilizer to be ready for routine use after the power is switched is less than 5
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
Temperature (°C)
Pressure (kPa)
Holding time
(min)
Total time (min)
Type
Max load (kg)
Max load per tray
(kg)
SOLID
134
210
4
25-45
Unwrapped solid material
5.00
1.50
121
110
20
30-60
WRAPPED
134
210
4
35-60
Unwrapped solid material
5.00
1.50
121
110
20
35-65
Single-wrapped solid or hollow
material
4.00
1.20
TEXTILE
134
210
8
45-65
Unwrapped porous material
1.25
0.40
Single-wrapped porous material
1.10
0.30
121
110
30
50-75
Dual-wrapped porous material
0.75
0.25
Single-wrapped hollow material
4.00
1.25
Dual-wrapped solid and hollow
material
2.00
0.60
PRION
134
210
18
50-75
Unwrapped porous material
1.25
0.40
Single-wrapped porous material
1.10
0.30
Dual-wrapped porous material
0.75
0.25
Single-wrapped hollow material
4.00
1.25
Dual-wrapped solid and hollow
material
2.00
0.60
LIQUID
(optional)
134
210
10
35-60
Liquid
1.20
0.40
121
110
30
35-65
Drying
(optional)
―
―
―
1-20
―
―
―
B&D test
134
210
3.5
28-35
―
―
―
Helix test
134
210
3.5
28-35
―
―
―
Vacuum
test
―
―
―
20-25
―
―
―
Table of contents
Other ICANCLAVE Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

dosatron
dosatron MKD14MZ3000AF manual

Barska
Barska MICROi AY12586 manual

OHAUS
OHAUS Frontier FC5513 instruction manual

Biobase
Biobase BSC-4FA2-GL user manual

Thermo Scientific
Thermo Scientific Barnstead EASYpure RODI 1332 Series Operation manual and parts list

Tuttnauer
Tuttnauer 5050 Operation and maintenance manual