ICANCLAVE STE-18-T User manual

Instructions Manual
For Models: STE-18-T
STE-23-T

2
Thank you for choosing our steam sterilizer.
Prior to operating this instrument, please read the operations manual carefully and follow all
installation instructions.
Ningbo Ican Machines Co., Ltd.
No. 77 Yunlin East Road, Gulin Town,
Ningbo, China.
www.icanclave.com
European Representative:
Icanclave Europe S.L
Juan Ramon Jimenez 6
Quart de Poblet, Valencia, Spain.
Document: Version 10T20000V2.10
Subject to technical changes
Save these instructions

3
Content
1. General
4
4 Scope of manual
4 Intended use
4 General safety instructions
4 Standards and directives
5 Symbols
2. Description of the sterilizer
5
5 Control panel
6 Sterilizer views
7 Technical specification
7 Packing content
3. Installation
7
7 General conditions
8 Power supply connection
8 Location requirements
4. Setup
8
8 Fill the distilled water tank
9 Preparation of the sterilization materials
10 Basic set
12 Advance set
5. Operation
14
14 Load the sterilization chamber
14 Select the program
14 Start the sterilization program
15 End of cycle
15 Manually interruption of the cycle
16 Delay start cycle
17 Test Program
18 Descale
19 Data
19 Print report
20 Print Labels
20 Save report
6. Maintenance
20
21 Clean the distilled water tank
21 Replacement of the bacteriological filter
21 Clean Chamber, trays and tray Rack
22 Door adjustment
22 Replacement of the door seal ring
22 The drain valve
7. Troubleshooting
23
23 Troubleshooting
8. Transportation and storage
23
23 Transportation and storage conditions
9. Safety devices
24
24 Safety device description
11. Appendix
25
25 Water properties/characteristics
26 Diagrams of the sterilization programs

4
1 General
Scope of Manual
This manual contains information concerning the installation, operation and maintenance of
the steam sterilizers. To ensure proper performance of the sterilizer, the instructions given in
this manual should be thoroughly understood and followed.
Keep the manual near the sterilizer in an accessible location for future reference.
Intended Use
The steam sterilizer described in this manual is intended for the sterilization in all medical,
dental, beauty, vet and tattoo fields of the following types of instrument loads: solid, porous,
hollow loads type A and hollow loads type B, un-wrapped, single wrapped and double
wrapped, liquid, that are compatible with steam sterilization.
General Safety Instructions
▪Read and understand this manual before attempting to install or operate the sterilizer.
▪Make sure that all the installation conditions are fully complied with.
▪Ensure that the voltage agrees with the supply voltage specified on the supply on the
type plate of the sterilizer.
▪This appliance must be grounded. Connect only to a properly grounded outlet.
▪Do not cover or block any openings on this appliance.
▪Use this appliance only for its intended use as described in this manual.
▪Do not exceed the maximum weight limit of the loads specified in this manual.
▪Do not operate this appliance if it has a damaged cord or plug if it is not working
properly or if it has been damaged or dropped.
▪Never put into the sterilizer inflammables or explosive products.
▪The sterilizer may not be operated in areas in which gas or any other explosive
volatile substance is present.
▪Installation and repair work should only be performed by authorized service
technicians. Work by unqualified persons could be dangerous and may void the
warranty.
Standards and directives
The steam sterilizers were designed and produced in conformity with the following directives
and standards:
Directives:
97/23/CE Pressure equipment.
93/42/EEC Medical devices (class II b).
Standards:
EN 13060 Relative to small steam sterilizers.
EN 61010-1 Safety regulations for laboratory devices - Part 1: General regulations.
EN 61010-2-040 Safety regulations specific to sterilizers used in the processing of
medical material.
EN 61326-1 Electromagnetic compatibility regulations for laboratory devices.

5
Symbols
For safe operation, please pay close attention to the alert symbols below which can be found
on the sterilizer and throughout this manual.
Important information (Caution) Hot surface Ground connection
2 Description of the sterilizer
Control panel
Door closed / Tap to open the door The door is open
Fill the tank if it is flashing From an external container
Tap until (filling the tanks) appears
Drain the tank if it is flashing The capacity is 1.5 liters
1 Report printer Grey = No connected / no paper
Green =Ready
2 Label printer
Grey = No connected
Red = No paper
Green = Ready
USB memory Grey = No connected
Green =Ready
Door status
Printers
USB memory
Internal temperature
Internal pressure
Used water tank level
Clean water tank level

6
Sterilizer views
1. Chamber 9 . Air filter 17. Power socket
2. Door lock
10. Printer
18. Rating plate
3. Door handle 11. Main swich 19. Distilled water tank (out)
4. Door seal ring
12. USB port
20. Safety valve
5. Door 13. Distilled water tank (Inn) 21. Valves vent
6. Door switch
14. Used water tank
22. Used water tank Vent
7. Distilled water tank 15. Condenser vent
8. Touch screen
16.Circuit breaker
Date of manufacture
SN Manufacture number
Manufacturer Name
Manufacturer Address
EC-Representative Name
EC-Representative Address

7
Technical specifications
Model
STE-18-T
STE-23-T
Chamber (mm)
247 x 350 247 x 450
Overall dimensions (mm)
483(W)*450(H)*635(D) 483(W)*450(H)*635(D)
Net Weight (kg)
50 53
Nominal power (VA)
2300 2300
Rated Voltage
220-240;50 Hz 220-240;50 Hz
Sterilization temperatures
121°C/134°C 121°C/134°C
Capacity of the distilled
water tank
4 L (Water at level Max.) Approx.
0.8L (Water at level Min.) Approx.
Circuit breaker
F16A /400 V
Operation temperature
5°C ~ 40°C
Operation relative humidity
Max. 80%, non-condensing
Max. Noise level
<70 dB
Atmospheric pressure
76 kPa ~ 106 kPa
Packing content
Item
Accessories
Quantity
1 Instruments tray 5
2 Instruments tray rack 1
3 Draining hose 2
4 Door Seal 1
5 Try handle 1
6 Instructions Manual 1
7 Calibration certificate 1
3 Installation
General conditions
▪Position the device on a plane surface with a minimum capacity of 60 kgs.
▪The sterilizer should be placed on a level worktable.
▪Leave at least 10 cm between the device rear part and the wall. The clearance required to open
the door is 40cm.
▪Position the sterilizer at such a height as to make it possible for the operator to check the whole
sterilization chamber and carry out the normal cleaning operations.
▪The room where the device is installed must be sufficiently ventilated.
▪Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
▪Do not lean on the door when it is opened.
▪Do not place trays, papers, fluid containers or other objects on the sterilizer.

8
Power supply connection
Check the label on the back panel of sterilizer to verify the voltage rating for the unit. Failure to connect
the sterilizer to an appropriate power supply could result in damage to the unit and electrical shock to
personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used
for the sterilizer is recommended. Never connect the device pin to reductions of any type.
Location requirements (mm)
4 Setup
Switch On and open the door to remove all of the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. After switching on, the machine turns on the LCD and
shows the door position, water level, working program, date, time, etc.
Fill the distilled water tank
Manual water filling
When the level of distilled water reaches a minimum level, the distilled water tank icon will flash and beep
three times.
Press the button on the tank lid and open it to the maximum position.
Fill it carefully with distilled water.
If exceeds the maximum level, an alarm will sound, and the distilled water tank icon will blink.

9
From an external container
Connect the supplied tube to the front fitting of the sterilizer.
Insert the other end of the tube with the filter into the container with distilled water.
Tap the distilled water tank “fresh water” icon to operate the water pump and maintain it pressed until the
message (filling the tank) appears.
The pump charges the clean water tank located at the top of the sterilizer, the capacity of the tank is 4.0
liters approx.
If after 240 seconds the tank has not reached the maximum level, the pump will stop, and it will be
necessary to press again the distilled water tank icon to finish filling the tank.
Once the maximum level has been reached, the pump stops automatically.
Fixed water feeding
When the Icanclave water system is connected, the device will fill the distilled water tank every time it
reaches its minimum level.
To install the water system, follow the instructions given together with the Icanclave water device.
Drain the distilled water tank
Attach the drain hose on a fitting connection located on the back of the sterilizer. Pull the connector to
star the draining.
Attention: The capacity of the distilled water tank is approximately 4 liters.
Drain the used water tank
Attach the drain hose on a quick fitting connection located inside the service door at the left.
Attention: The capacity of the used water tank is approximately 2 liters.
Preparation of sterilization materials
For the most effective sterilization and to preserve the sample, please follow below:
▪Clean instruments immediately after use.
▪Treat the instruments by ultrasound cleaner.
▪Residual chemicals left over after cleaning and disinfecting process may damage and corrode
parts of the sterilizer, always rinse off the instruments using distilled water.
▪Follow instrument manufacturer’s guidelines and recommendations for handing and cleaning
instruments prior to sterilization.
▪Check the manufacturer’s instructions as to proper procedure for sterilizing of each item.
▪Arrange the samples of different materials on different trays or with at least 3cm of space
between them.
▪Clean and dry instruments thoroughly before placing them into tray.
▪Always insert a sterilization paper or cloth between the tray and sample to avoid direct contact.
▪Arrange the containers (glasses, cups, test-tubes, etc.) on one side or inverted position,
avoiding possible water stagnation.
▪Don’t stack the trays one above the other or put them in direct contact with the walls of the
sterilization chamber.
▪Always use the instrument tray handle.
▪Wrap the samples one by one or, if more tools have to be set in the same bag, verify that these
are made of the same material.
▪Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the sterilizer.
▪Don’t overload the trays over the stated limit (see appendix 2).

10
Basic Set
After you turns the sterilizer On, the main menu
will appear, select “Settings” then "Basic Set”.
Time & Date
Select “Time & Date”
The screen will show the current configuration.
Tap “ ”to enter the menu.
Touch the parameter that needs to be change, it
will light orange, by pressing (+/−) or by using
the numbers adjust it. To save press “↵”
Language
Select “Language”
The screen will show the current configuration
Tap “ ”to enter into the menu.
Select the language, tap “<” to save and back.

11
Unit
Select “Unit”.
The screen will show the current configuration.
Tap “ ”to enter the menu.
Select the parameter, the selected parameter will
light orange.
Tap “<” to save and back.
.
Pre-Heating
Tap “Pre-heating” button.
The screen will show the current configuration.
Tap “ ”to enter the menu.
Tap ON or OFF to select.
tap “<” to save and back.
Fill
Select “Fill”.
Tap “ ”to enter the menu.
Tap the number 0~9 to set the automatic water
filling time. (We recommend to set at “05:00”).
If set at “00:00”, the automatic water filling
function will be OFF.

12
Advance Set
Tap “Settings”
then "Advance Set”.
Enter passwords
“ 111111”
and tap “↵”enter the
menu.
Parameter
Select “Parameter”.
Tap “Click here to set
cycle parameters” to
enter the menu.
The “Parameter” menu
permits to set the
following options:
“Sterilization time” and
“Dry time”.
Select the cycle name,
for example “Quick”.
The screen will show the
current configuration.
Tap “ ”to enter the menu.
Touch the parameter or by pressing “ ” “ ” select the number that needs to be changed, it
will light red, or by pressing the numbers adjust it. To save press “↵”
Expiry date for label report
Select “Expiry date”
Tap “Click here to set expiry date” to enter the
menu.
By pressing “ ” “ ” select the number that to
be changed, it will light red, or by pressing the
numbers adjust it. To save press “↵”.

13
Reset all settings
Select “Reset All Settings”
Tap “Click here to reset all settings”
The screen will show ”confirm to reset all
settings?
To confirm press “ ”
To deny press “ ”
Maintenance
Tap “Settings”
then "Maintenance”.
You can see the
information of how many
days or cycles left for
maintenance.

14
5 Operation
Load
After selecting program, the materials to be sterilized can now be placed on the
tray and placed inside the chamber by the tray handle.
After the instruments are loaded, you may close the door.
Select the program
After you turns the sterilizer On, the main menu
will appear, select “Programs”.
Tap the program to select, in the screen will
appear the information of selected program as
the pressure, temperature, time (this may vary
depending on the environmental conditions and
the load), vacuum times, Sterilization time
(holding time) and drying time.
Start the sterilization program
After closing the door, tap “ ” to start the cycle.
The stage, conditions and the status of the cycle will
appear on the display. The sterilizers will perform the
program automatically. (see appendix 2).

15
End of the cycle
Once the cycle is completed, “End” will appear at the end of the graphic, the
printer will print out and the digital report saved in the USB memory is these are
connected.
To open the door, tap to the door status icon.
If the label printer is connected (Optional) (See label printer), the icon will appear
next to the door status icon.
Caution: Always use the tray handle to load or unload the tray into the
sterilizer. Failure to do so can result in burning.
If the label printer is connected
At the end of a
successful cycle, if the
label printer is connected,
the printer icon will
appear. Tap the icon to
print the labels. Tap
labels icon and select the
number of copies, tap
printer icon to print the
labels.
Manually interruption of the cycle
To interrupt a started cycle prematurely, tap “Stop”
and “✓” to confirm.

16
If the cycle is manually interrupted after it reaches the
drying phase, the items inside the sterilizer may be
considered sterile.
Note: Considering that the cycle has been interrupted
during the drying phase the materials and instruments
inside the chamber may be wet.
If the cycle is manually interrupted before it reaches
the drying phase, the items inside the sterilizer must
be considered not sterile.
Caution: Depending on the phase of the
cycle, steam and water can escape from
the sterilization chamber when you open
the door.
Delay start cycle
Tap “Delay Start” to schedule or modify a delay start cycle. The screen will show the current
configuration.
Tap “ ”to enter in the configuration menu.
Select the program then tap “Tap to set a delay cycle”.
Select the program then tap on the time and date and adjust it using the numeric keypad.
Tap “Disable” to set the delay start cycle once or daily. Tap “↵”to save and back.

17
Once the cycle has been programmed a
message icon will appear on the main screen.
Tap Message to review the parameters of the
programed cycle or/and to cancel it.
Test Programs
Helix Test
Tap Test from the main menu, select Helix test, in the
screen will appear the information of selected Test
program as the pressure, temperature, time (this may
vary depending on the environmental conditions and
the load), vacuum times, Sterilization time (holding
time) and dry time.
Following the instructions of the test manufacturer.
Put the Helix test into the chamber, then close the
door and Tap “ ”
After finishing the cycle, you may check the indicator
and evaluate the result according with the instructions
of the test manufacturer.
B&D test
Tap “Test” from the main menu, select “B&D test”, in
the screen will appear the information of selected Test
program as the pressure, temperature, time (this may
vary depending on the environmental conditions and
the load), vacuum times, Sterilization time (holding
time) and drying time. Following the instructions of the
test manufacturer.
Put the B&D test into the chamber, then close the
door and Tap “ ”
After finishing the cycle, you may check the indicator
and evaluate the result according with the instructions
of the test manufacturer.

18
Vacuum Test
Remove all instruments
inside chamber before
doing Vacuum Test.
Tap “Test” from the main
menu, select vacuum test.
Tap “Vacuum test” to
select, in the screen will
appear the information of
selected test program as
the pressure, temperature,
time (this may vary
depending on the
environmental conditions
and the load), vacuum
times, Sterilization time (holding time) and drying time. In compliance with EN 13060, the test requires
that the air leakage rate less than or equal to 0.13 kPa/min. During 10 minutes: If leakage rate is not
greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will show
void. That means the result of the test is fail. You need run the vacuum test again after the chamber has
cooled down.
Descale
Mix the descaling powder
with distilled water to make
descaler according with the
instruction of the descaling
powder manufacturer.
Tap “Test” from the main menu,
select descale.
Tap “Descale” to select, and tap
“ ”. Remove all instruments
inside the chamber and
close the door.
Empty the fresh water tank and
fill with descaler which prepared
as above, to 2/3 of the water
tank.
Tap , start descaling, it will
take around 25 minutes.
After descaling, empty the
fresh water tank and fill with
distilled water.
Tap , start flushing, it will take
around 8 minutes.
After flushing, the cycle
“Descale” is end.

19
Data
The internal memory will store the information of the last 9999 cycles.
Print report
Select Data from the
main menu and tap Print
report, it will show the
current status.
Tap “ ”to enter in the
menu.
Select the cycle and Tap
Print.
Sample of a printer report
==============================
==
Program: WRAPPED
Temperature: 134C
Pressure: 206.0kPa
Drying Time: 08Min
Holding Time:4.0Min
--------------------------------
Time Temp. Pressure
Start 12:28:17 089.0C
T1: 12:31:32 087.1C -075.0kPa
T2: 12:33:43 110.2C 052.0kPa
T3: 12:36:37 088.9C -075.0kPa
T4: 12:39:20 114.7C 053.7kPa
T5: 12:43:37 087.9C -075.0kPa
T6: 12:50:40 134.8C 206.0kPa
TS: 134.7C 209.5kPa
Max. Temperature:135.2C
Min. Temperature:134.3C
Max. Pressure:214.0kPa
Min. Pressure:204.9kPa
T7: 12:54:39 134.4C 211.4kPa
T8: 12:57:36 102.1C -060.0kPa
T9: 12:59:54 098.2C -060.0kPa
End 13:04:07 102.4C
--------------------------------
CycleNo.: 00017
Ster. Value: Success
Date: 2017-06-07
SN:E54723
Operator:
v2B00V2.5
==============================
==

20
Print labels
Select Data from the main menu and tap “Print labels”, it will show the current status.
Tap “ ”to enter in the menu. Select the cycle number and select the number of copies, tap
printer icon to print the labels.
Save Report
Select Data from the main menu and tap “Save Report”, it will show the current status.
Tap “ ”to enter in the menu.
E.g.
If the serial number of the steam sterilizer is E00001 and the cycle number is 0012.
The file name of the file will be 01001200.log
The first two numbers represent machine number.
The middle four numbers represent cycle number.
The last two numbers represent error code.
00: no error; 01: error E01
6 Maintenance
To ensure proper operation and maximum steam sterilizer life, carefully follow all recommendations for
periodic maintenance.
One of the most important steps you can take to prevent problems with your sterilizer is to use only
distilled water.
This manual suits for next models
1
Table of contents
Other ICANCLAVE Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Honeywell
Honeywell Solaris 7980g quick start guide

Nippon Genetics
Nippon Genetics FastGene NG040 instruction manual

Orbis
Orbis AirSTAR Operation manual

IKA
IKA H 44 Mounting instructions

Monmouth Scientific
Monmouth Scientific Ductaire Pro DP700 operating & maintenance manual

Opticon
Opticon NLV-5201 Specification manual

TFT
TFT Max Series INSTRUCTIONS FOR SAFE OPERATION AND MAINTENANCE

DEMA
DEMA 162HCGHT installation instructions

Gilson
Gilson PIPETMAN M CONNECTED quick guide

Northwest Marine Technology
Northwest Marine Technology Multishot instruction manual

Masterclock
Masterclock CS5-LAN user manual

IKA
IKA RET control-visc operating instructions