iHealth PT3SBT User manual

1
iHealth® PT3SBT
Wireless No-Touch Forehead Thermometer
FCC ID:SLRPT3SBT IC:10913A-PT3SBT

2
iHealth®
Wireless No-Touch Forehead Thermometer (PT3SBT)
OWNER’S MANUAL
Table of content
INTRODUCTION.................................................................................................................................3
Indication for use ..............................................................................................................................3
Safety Precautions............................................................................................................................. 3
Overall Description............................................................................................................................ 6
Product performance........................................................................................................................7
Expected use and scope of applications ...........................................................................................8
Contraindications ..............................................................................................................................8
Instruction For Use............................................................................................................................8
SYNC TIME AND RESULTS VIA BLUETOOTH .....................................................................................11
Product errors and troubleshooting ...............................................................................................12
Body temperature........................................................................................................................... 12
Cleaning and Care ...........................................................................................................................13
Care the probe ................................................................................................................................13
Maintenance ...................................................................................................................................14
Disposal ........................................................................................................................................... 14
Operating, shipping, and storage conditions ..................................................................................14
Accessories......................................................................................................................................14
Standard icons.................................................................................................................................15
Calibration....................................................................................................................................... 15
Storage ............................................................................................................................................15
OTHER STANDARDS AND COMPLIANCES .........................................................................15
ELECTROMAGNETIC COMPATIBILITY INFORMATION ......................................................................16
WARRANTY..................................................................................................................................17

3
INTRODUCTION
Thank you for choosing our product.
This product is a high-tech infra-red (IR) thermometer designed to take human
body temperature by measuring the energy of IR emitted from the forehead
and it can transmit the temperature to a smart device with Bluetooth. The
product will help you to assess you and your family members’ health
conditions easily and quickly.
Product Name: Wireless No-Touch Forehead Thermometer
Product Model: PT3SBT
Indication for use
The Wireless No-Touch Forehead Thermometer is intended for the intermittent
measurement of body temperature from the forehead on people of all ages. It
can transmit the temperature to a smart device with Bluetooth and can be used
by consumers in the household environment and by healthcare providers.
Safety Precautions
Warning
Use of this thermometer is not intended as a substitute for consultation
with your physician. Please consult your doctor if you have any doubt
about the temperature reading.
Keep the thermometer out of reach of children. For accidental swallowing
of the battery or other components, please contact emergency services
immediately.
Batteries must not be thrown into an open fire or short circuited.
Caution
(1) Measurements
1. Thermometer readings should be regarded as a reference. Do not attempt
self-diagnostics or self-treatment using the temperature readings. Please seek
professional medical advice when necessary.
2. There is no absolute standard for human body temperature. Knowing your
own normal body temperature range is important to accurately determine if
you have a fever.
3. Make sure that the forehead of the subject is free from sweat, cosmetics, dirt,
or grease before measuring.
4. Patients should not drink, eat, or be physically active before/while taking a
reading. Wait 30 minutes before taking a measurement. Temperature readings
taken when a body is in a state of stable equilibrium is more accurate and
useful as a reference.
5. Do not take temperature measurement over scar tissue, open sores or
abrasions.
6. Do not measure body temperature from scar tissues on the forehead as

4
such tissues will affect temperature conduction of the body.
7. If there is a temperature difference between the thermometer storage area
and the new ambient environment around the subject, please let the
thermometer sit within the new environment for 30 minutes before taking the
measurement.
8. Do not measure body temperature immediately after consuming a drug that
raises body temperature. Temperature readings taken at this time will not be
accurate.
9. It is normal for readings taken from continuous measurements to fluctuate
within a small range. During continuous measurements, the subject’s body
temperature may be transmitted to the thermometer, affecting measurement
accuracy. We recommend taking only up to 3 continuous readings within a
short period.
10. Do not directly face the sun or an air outlet of an air conditioning or radiator
device during the measurement as this will cause changes to the forehead
temperature. Measurements should be taken in a stable environment where
possible.
11. Do not measure body temperature in an environment with strong EM
interference (examples include places close to a working microwave, induction
cooker, or cellphone in-use) as EM interference may cause errors in the
reading or even device failure.
12. Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the product. Otherwise, degradation of the
performance of this equipment could result.
13. This product should be considered a personal device. Clean and sanitize
the product properly to prevent cross contamination.
14. To clean a dirty thermometer probe, gently swipe the probe using a cotton
swab dipped in 70% isopropyl. Let the cleaned thermometer sit for at least 15
minutes before taking more measurements.
15. This infrared thermometer meets requirements established in ISO
80601-2-56:2017/AC:2018. It displays subject’s temperature over a range of
89.6℉~109.4℉(32℃~42.9℃). Full responsibility for the conformance of this
product to the standard is assumed by (manufacturer: ANDON HEALTH CO.,
LTD. Add: No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190,
China).
(2) About the product
1. This product is a precision device. Return the product to its original
packaging for proper storage after use. To ensure accurate measurements,
avoid the device or probe contacting any liquid or droplets. Avoid tiny particles
(such as dust or powder) falling into the probe.
2. Avoid dropping or subjecting the product to external forces. Do not
disassemble or re-assemble the product on your own.

5
3. Do not directly touch the probe with your fingers or blow on it.
Measurements taken using a damaged or dirty IR probe may be inaccurate.
4. Keep the product at a place inaccessible to children to prevent children from
swallowing the batteries or small parts.
5. Do not throw the thermometer or batteries into fire to prevent explosions.
6. Remove the batteries from the thermometer if the device will not be used for
more than one month.
7. If you are allergic to plastic/rubber, please don’t use this device.
8. The materials (ABS, TPU, metal) of expect contact with patient had passed
the ISO 10993-5 and ISO 10993-10 standards test, no toxicity, allergy and
irritation reaction. They are in compliance with the MDD requirements. Based
on the current science and technology, other potential allergic reactions are
unknown.
9. The patient can be an intended operator. All functions of the unit can be
used safely by patient.
10. This product should be considered a personal device. Clean and sanitize
the product properly to prevent cross contamination.
If performance of the instrument may be adversely affected should one or
more of the following occurred:
1) Operation outside of the manufacturer-specified subject temperature
range.
2) Operation outside of the manufacturer-specified operating temperature
and humidity ranges.
3) Storage outside of the manufacturer-specified ambient temperature
and humidity ranges.
4) Mechanical shock.
5) Manufacturer-defined soiled or damaged infrared optical components.
11. The patient is an intended operator.
12. This product contains batteries and recyclable electronic waste. To
protect the environment, do not dispose of it in the household waste, but take it
to appropriate local collection points.
13. This equipment complies with FCC RF radiation exposure limits set forth
for an uncontrolled environment.
14. This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference
that may cause undesired operation.
15. The manufacturer is not responsible for any radio or TV interference
caused by unauthorized modifications to this equipment. Such modifications
could void the user’s authority to operate the equipment.
16. This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
Table of contents