Indes Esense Rise Series Manual

Leaflet esense Rise series EN I R 2.00 | 28-01-2021
Product leaflet esense Rise patient lift series
Product description
The Rise is an
active
patient lift. With this type of lift the patient has an active role in the transfer. Rise patient lifts are
standard equipped with esense power assist drive system but are also available as manual version.
Use instructions
•Always check the patients’ medical file to learn if the patient has any limitations regarding patient lift transfers.
•Esense lifts are guaranteed to be safe to use in combination with ‘esense’ slings, identified by the label. When using slings
of other brands, make sure the clips are compatible with the geometry of the lugs on the lifting frame; if in doubt about
compatibility, check with the sling manufacturer. Specifications of the lugs can be found in the full user guide.
•Use the hand control to control the electrical adjustments: the leg adjustment and height adjustment of the lifting
frame. When the button is released the corresponding movement stops immediately.
•Pressing the red emergency switch immediately switches off
all
motors.
•In case a system fault occurs during a transfer, the patient can manually be lowered using the emergency lowering
function (pull the red/grey knob at the bottom of the actuator upwards).
•For patient lifts equipped with esense drive system –The power assist functionality is activated by sensitive
sensors in the pushbar. Ensure at all times that the pushbar can move freely and that no external forces are
exerted on the pushbar, other than the push and pull forces required for manoeuvring the lift.
Important safety instructions
•Ensure that you never lift patients weighing over the maximum weight capacity. Never lift more than one patient at
a time and never leave a patient in the lift unsupervised.
•When using the patient lift ensure that no lateral forces are being exerted on the patient, the lifting frame or the
mast: this may lead to decreased stability of the lift.
•The patient lift is not suited to be exposed to large quantities of water; e.g. it is not suited for use under a shower.
•Do not expose the lift to temperatures below -5°C or over 50°C.
•Before every use always check the slings –in particular the straps and clips –for any signs of wear or damage.
•Regularly check the lift for loosened parts (bolts, axles) and wear of e.g. rubbers, wiring, straps or plastic covers. In
case of any irregularities please stop using the lift and contact your supplier/service department immediately.
•For patient lifts equipped with esense drive system –The power assist functionality contains moving parts required
for driving and steering of the drive wheel. These components are protected by the blue cover at the back of the lift:
prevent hands or fingers from getting stuck underneath the cover and always make sure to switch off the system
before carrying out any maintenance.
Certification
All esense products are marked with a CE certification. The products are classified as class I of the Medical Devices
directive MDD 93/43/EEC. All esense products are tested and meet the requirements of NEN EN ISO 10535 ‘Hoist for the
transfer of disabled persons –Requirements and test methods’ and NEN EN ISO 60601-1 ‘Medical electrical equipment’.
Product lifespan
If used and maintained correctly the average lifespan of the patient lift shall be 8 years minimum.
Cleaning
For cleaning it is best to use regular, non-aggressive and non-abrasive cleaning products (PH 5,5-8). Wipe the lift with a
well-wrung damp cloth using clean water to remove all disinfectant residue.
Always read the full user guide before use.
The user guide is available on www.esense-moves.com on the product page of the esense Rise (“downloads” tab.
Leaflet esense Rise series EN I R 2.00 | 28-01-2021
Service and (periodic) maintenance
In accordance with NEN-EN ISO 10535 the patient lift needs to be inspected at least once a year by someone qualified by
the manufacturer. In accordance with NEN-EN ISO 10535 the slings need to be inspected by an authorised person at
least twice a year. It is required to keep a maintenance log in which all relevant observations regarding defects, damage
and wear are reported.
Any defects, damage or wear need to be reported to the manufacturer or supplier immediately and the concerned lift
needs to be taken out of order until further inspection and until after reparation.
In case an error occurs, first check the quick debugging guide in the full user guide. In general, please contact your service
department or supplier in case a system fault/error occurs, for maintenance and service (e.g. availability of spare parts).
Technical specifications
#
Lift dimensions
Rise
A
Height of the base
11,5 cm
B
Min. clearance to floor
3,7 cm
C
Min. internal width
46 cm
D
Max. internal width frame –closed legs
52,7 cm
E
Max. internal width frame –open legs
84,6 cm
F
Distance front mast to 700mm internal width reference point
(open legs)
44,1 cm
G
Max. distance front mast to sling attachment point
41,3 cm
H
Min. height sling attachment point
88,5 cm
I
Max. height sling attachment point
163,5 cm
J
Vertical range
75 cm
-
Weight capacity
180 kg
-
Weight patient lift (incl. battery)
64 kg
-
Weight esense drive system
14 kg
-
Protection grade –patient lift and hand control
IPx2
-
Protection grade –esense drive system
IPx3
Manufacturer and supplier
esense is a product by
Your supplier
Indes Production Management B.V.
Name:
Pantheon 28
Address:
7521 PR
Postal code:
Enschede
City:
The Netherlands
Country:
+31 (0)53 4803920
Phone:
info@esense-moves.com / www.esense-moves.com
Email:

Other Indes Medical Equipment manuals

Indes esense Rise User manual

Indes

Indes esense Rise User manual

Indes esense Line User manual

Indes

Indes esense Line User manual

Popular Medical Equipment manuals by other brands

Medite TFB 35 instruction manual

Medite

Medite TFB 35 instruction manual

Dräger X-plore 8000 Instructions for use

Dräger

Dräger X-plore 8000 Instructions for use

MEDsan ULTRA SBMS 24 Quick operation guide

MEDsan

MEDsan ULTRA SBMS 24 Quick operation guide

Innova 3160e Quick reference guide

Innova

Innova 3160e Quick reference guide

Akron STREAMLINE 2221 user manual

Akron

Akron STREAMLINE 2221 user manual

Mediana HeartOn A16 Series Operator's manual

Mediana

Mediana HeartOn A16 Series Operator's manual

Welch Allyn VSM 300 Series Training handbook

Welch Allyn

Welch Allyn VSM 300 Series Training handbook

NeuroNexus SmartBox Pro quick start guide

NeuroNexus

NeuroNexus SmartBox Pro quick start guide

AGFA DX-D 100 user manual

AGFA

AGFA DX-D 100 user manual

Itamar Medical OM2196390 Instruction and Illustrated manual

Itamar Medical

Itamar Medical OM2196390 Instruction and Illustrated manual

MedaCure AirStream AS36 user manual

MedaCure

MedaCure AirStream AS36 user manual

Direct Healthcare Group Linido LI2613.000 Series manual

Direct Healthcare Group

Direct Healthcare Group Linido LI2613.000 Series manual

Comeg SOPRO 218 user manual

Comeg

Comeg SOPRO 218 user manual

CMC Patient Tie-In System Important information

CMC

CMC Patient Tie-In System Important information

Gima SP-10 user manual

Gima

Gima SP-10 user manual

Trumpf TruSystem 7000 Service manual

Trumpf

Trumpf TruSystem 7000 Service manual

ADC Diagnostix 700 Series Use, care & maintenance

ADC

ADC Diagnostix 700 Series Use, care & maintenance

ResMed 33963 User instructions

ResMed

ResMed 33963 User instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.