InTemp CX600 Dry Ice/CX700 Cryogenic Logger Manual
1-508-743-3309 (U.S. and International) 4 www.onsetcomp.com
6. Tap Disconnect.
Only the phone or tablet used to set the passkey can then
connect to the logger without entering a passkey; all other
mobile devices will be required to enter the passkey. For
example, if you set the passkey for the logger with your tablet
and then try to connect to the device later with your phone,
you will be required to enter the passkey on the phone but not
with your tablet. Similarly, if others attempt to connect to the
logger with different devices, then they would also be required
to enter the passkey. To reset a passkey, connect to the logger,
tap Set Logger Passkey, and select Reset Passkey to Factory
Default.
Downloading Data from the Logger
You can download data from the logger to a phone or tablet
and generate reports that include logged data and alarm
information. Reports can be shared immediately upon
download or accessed later in the InTemp app.
InTempConnect users: Privileges are required to download,
preview, and share reports in the InTemp app. Report data is
automatically uploaded to InTempConnect when you download
the logger. Log in to InTempConnect to build custom reports
(requires privileges). In addition, InTempConnect users can also
download CX loggers automatically on a regular basis using the
CX Gateway. Or, if the logger was configured with a logger
profile in which InTempVerify was enabled, then anyone can
download the logger with the InTempVerify app. For details on
the gateway and InTempVerify, see www.intempconnect/help.
To download the logger with the InTemp app:
1. Press the button on the logger to wake it up.
2. Tap the Devices icon and connect to the logger.
3. Tap Download.
4. Choose a download option:
Important: You cannot restart CX602 and CX702 loggers.
If you want a CX602 or CX702 logger to continue logging
after the download is complete, select Download &
Continue.
•Download & Continue. The logger continues logging once
the download is complete.
•Download & Restart (CX603 models only). The logger
starts a new data set using the same profile once the
download is complete. Note that if the logger was
originally configured with a push button start, you must
push the start button for logging to restart.
•Download & Stop. The logger stops logging once the
download is complete.
To download multiple loggers with the InTemp app:
1. Tap Devices, then Bulk Download.
2. The screen changes to Bulk Download mode. This
changes the way the screen behaves when you tap on
a logger tile. Tap on a tile to select it for bulk
download. You can select up to 20 loggers. Text at the
bottom of the page updates to indicate how many
loggers are selected.
3. Tap Download X Loggers to begin the download.
•Download & Continue. The loggers continue logging
once the download is complete.
•Download & Restart (CX400, CX450, CX503, CX603,
and CX703 models only). The logger starts a new
configuration using the same profile once the
download is complete. You will be prompted to
enter trip information again (if applicable). Note that
if the logger profile is set up to start with a button
push, you must push the button on the logger for
logging to restart.
•Download & Stop. The logger stops logging once the
download is complete. The downloads begin and run
one after another. The screen displays the download
queue.
4. Click Cancel to cancel the downloads and return to
the Devices screen, not in Bulk Download mode.
5. The screen displays Done when all the loggers are
downloaded.
A report of the download is generated and is also uploaded to
InTempConnect if you are logged in to the InTemp app with
your InTempConnect user credentials. If you are using the bulk
download feature, one report per logger is generated.
In the app, tap Settings to change the default report type
(Secure PDF or XLSX) and report sharing options. The report is
also available in in both formats for sharing later. Tap the
Reports icon to access previously downloaded reports. See
www.intempconnect.com/help for details on working with
reports in both the InTemp app and InTempConnect.
Logger Events
The logger records the following events to track logger
operation and status. These events are listed in reports
downloaded from the logger.
Event Name Definition
Configured The logger was configured by a user.
Connected The logger was connected to the InTemp
app.
Started The logger started logging.
Stopped The logger stopped logging.
Downloaded The logger was downloaded.
Alarm Out of
Range/In Range
An alarm has occurred because the reading
was outside the alarm limits or back within
range. Note: Although the reading may
return to a normal range, the alarm
indicator will not clear in the InTemp app
and the alarm LED will continue to blink.
Safe Shutdown The battery level dropped below a safe
operating voltage and performed a safe
shutdown.
Deploying the Logger
Use the mounting loop on the logger to secure it to a shipment
or other application you are monitoring. You can also remove
the backing on the tape that is adhered to the top and bottom
of the logger to mount it on a flat surface.