9VENTURE® HOMEFILL II COMPRESSOR
EXTERNAL EXAMINATION
1. Examine the outside of the cylinder for the following conditions, and replace the cylinder if
they exist:
A. Dents or dings
B. Arc Burns
C. Oil or Grease
D. Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for
use.
2. Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or
blistering of the paint, or other protective coating or heat sensitive indicator. If fire or
thermal damage is found, replace the cylinder.
. Inspect the Invacare valve for the following, and replace if found:
A. Debris, oil or grease
B. Noticeable signs of damage
C. Signs of corrosion inside the valve
D. Signs of excessive heat or fire damage
FIGURE 1 - CYLINDER PREFILL INSPECTION
COMPRESSOR OPERATION PROCEDURE 3
COMPRESSOR OPERATION
OXYGEN CYLINDERS FILLED BY THE VENTURE HOMEFILL
SHALL BE USED FOR PERSONAL USE ONLY. NOT TO BE FILLED
FOR RESALE OR USE BY PROFESSIONAL USERS
WARNING:
HIGH PRESSURE OXIDIZING GAS VIGOROUSLY ACCEL-
ERATES COMBUSTION.
CAUTION: For medi al appli ations. Use only as dire ted by a li ensed
pra titioner. Uninterrupted use of high on entrations of oxygen over a long
duration, without monitoring its effe t on oxygen ontent of arterial blood,
may be harmful.
CAUTION: DO NOT HANDLE CYLINDER OR USE CONTENTS UNTIL YOU
ARE TRAINED TO USE CYLINDER AND CONTENTS, INCLUDING EMER-
GENCY PROCEDURES.
CAUTION: IMPROPER FILLING OR ABUSE OF THIS CYLINDER, OR
FAILURE TO HEED THIS CAUTION MAY CAUSE SERIOUS INJURY OR
LOSS OF LIFE. DO NOT ALTER OR MODIFY CYLINDER OR RELATED
COMPONENTS. KEEP OUT OF THE REACH OF CHILDREN. Se ure
ylinder during storage and use. No smoking in ylinder area. Open valve
slowly. Close valve after ea h use and when empty. Use in a ordan e with
Venture HomeFill Operators Manual. Cylinder temperature should not ex-
eed 125°F (52°C). Keep away from heat, flame and spark. Do not drop.
Use No Oi or Grease. KEEP CYLINDER, CYLINDER VALVE, ALL EQUIP-
MENT AND CONNECTIONS FREE OF OIL AND GREASE TO AVOID
VIOLENT IGNITION.
TRANSFILLING OF THIS GAS IS PERFORMED BY VENTURE
HOMEFILL ONLY. Cylinders that have been refinished and/or exposed to
elevated temperatures must be hydrostati ally tested before filling. Do not
use austi paint strippers. Aluminum ylinders subje t to the a tion of fire
or heated to temperatures in ex ess of 350°F must be withdrawn from ser-
vi e and ondemned by trained personnel. Valve and safety relief devi es
must be removed and repla ed by trained and authorized personnel. Do
not a ter or change va ve assemb y.
THIS CYLINDER IS TO BE USED WITH THE INVACARE
VENTURE HOMEFILL SYSTEM. THIS CYLINDER
CONTAINS 93+3% OXYGEN PRODUCED BY AN
OXYGEN CONCENTRATOR USING THE PRESSURE
SWING ADSORPTION PROCESS.
DO NOT REMOVE THIS PRODUCT LABEL
Not for Commercia Resa e
Invacare Corporation
699 Aero Lane
Sanford, FL 32771
1-800-832-4707
OXYGEN
2
P/N: 1082951