IRadimed MRIDIUM 3860+ User manual

MRidiumMRidium®®38603860++MRI Infusion PumpMRI Infusion Pump
Reference GuideReference Guide
www.IRadimed.comwww.IRadimed.com

12
Initial Setup
Ready to Prime Push In Pull Through
1.2 | Proper Flow Clamp Positioning for Priming the IV Set
Initial Setup
When priming, DO NOT STRETCH the Silicon Segment of
the IV Set as this can cause freeow issues during infusions.
112233
1.1| Selecting the Proper IV Set
Post PrimePost Prime
a. Select the 1056 IV Set when infusing from a bag or bottle uid source (18mL).
b. Select the 1057 IV Set when infusing from a syringe uid source (15mL).
c. Select the 1058 IV Set as an extension or adapter for current IV Infusions (4mL).
Flow Open Flow Closed
44Ensure all roller and/or slide clamps are in the closed position.

34
1.3 | Power ON the Pump
Press the ( I ) button as shown to power on the pump.
Initial Setup
1.4 | Open the Door
“Push Down” on the Purple Button,“Pull Out”on the Handle.
Purple
Button
Handle
1.5 | Loading the IV Set
a. Insert the round Disk on the IV set into the “Disk Slot” on top of the pump.
b.Tilt up and insert the top of the Black Flow Clamp into the pump, then push
to insert as shown. The clamp will click when secured.
- DO NOT Push the Black Flow Clamp which triggers a Check Door Alarm
and cause possible Freeow.
Initial Setup
c. Secure tubing inside Bubble Detector
Door Pin
d. Firmly CLOSE THE DOOR
Ensure Door Handle is FULLY Latched, Open All Clamps, Vent Caps.
No uid ow should be in the drip chamber of the IV set or bubbles
in a syringe. If uid ow is present, clamp the IV tubing and repeat
steps 1.2 on page 2.
Note: Contact your biomed department if the door pin does not look as shown.

5
Before Programming Channels A or B
2.1 | Selecting Infusion Type: Drug Libraries A or B
a. Press the UP or DOWN arrow keys to scroll through listed
medications. Press ENTER when the proper medication
is highlighted.
d. To start infusion, press the Start/Stop A Button.
b. Primary is highlighted, press ENTER again.
Program entries with Numeric Keypad such as:
- Dose; ENTER
- Weight (kg); ENTER
- VTBI; ENTER
Programming Channel A
a. Press the softkey next to New Patient (Drug Library A)
or
Press the softkey next to New Patient (Drug Library B)
b. Press the softkey next to Dose Rate Calc. A
or
Press the softkey next to Dose Rate Calc. B
6
2.2 | Selecting Medication: Drug Library A
Use Library A for Channel A, and Library B for Channel B
c. Push Start A will appear when programmed
correctly.
Startup Screen Menu Screen

2.3 | Selecting Medication: Drug Library B
a. Press the MENU button.
c. Press the UP or DOWN arrow keys to scroll
through listed medication. Press ENTER when
the proper medication is highlighted.
f. To start infusion, press the Start/Stop B Button.
d. Primary is highlighted, press ENTER again.
Program entries with Numeric Keypad such as:
- Dose; ENTER
- Weight (kg); ENTER
- VTBI; ENTER
Programming Channel B
7
Programming Basic Infusions
3.1 | Rate Mode: Channel A
d. Press the Start/Stop A Button to start infusion.
(as shown in 2.1)
a. Press the softkey by “New Patient- Rate Mode”
on Main Menu Screen.
b. Press the softkey by RATE on the screen
-Program RATE number using the Numeric Keypad
-Press the ENTER Key
c. Press VTBI Softkey
- Program the VTBI using the Numeric Keypad
- Press the ENTER key
Ready to Start will appear at bottom of the screen.
8
e. Push Start B icon will appear when programmed correctly.
b. Press Dose Rate Cal. B (Reference 2.1)

Programming Basic Infusions
3.2 | Rate Mode: Channel B
d. Press the Start/Stop B Button to start infusion.
a. Press the softkey by “New Patient- Rate Mode”
on Main Menu Screen.
b. Press the Channel B softkey
-Program RATE number using the Numeric Keypad
-Press the ENTER Key
c. Press VTBI Softkey
- Program the VTBI using the Numeric Keypad
- Press the ENTER key
Ready to Start will appear at bottom of the screen.
9
Titrating Basic Infusions
3.3 | Rate Mode: Channel A or B During Infusion In Progress
a. Press the softkey by Channel A or B at the bottom of the screen.
c. Press VTBI Softkey
- Program the VTBI using the Numeric Keypad
- Press the ENTER key
10
(as shown in step 2.1)
b. Press Rate softkey
-Program RATE number using the Numeric Keypad
-Press the ENTER Key
OR

4 | Proper DERS Bolus Programming*
a. To program a bolus, press Bolus Softkey A/B or
press Menu Button.
e. PRESS the TIME button, program time
using Numeric Keypad.
f. Press ENTER to Start Bolus.
11
Programming a Bolus
d. Program dose using Numeric Keypad
c. Press Dose Softkey.
b. From Menu screen, press Softkey by
Bolus A or Bolus B.
- Enter the infusion time of 6 seconds (0.1 minutes) to 99
minutes.
12
KVO (Keep Vein Open)
e. Press Start/Stop A or B Button to resume infusion.
5 | Resuming Infusion During KVO
a. Press Start/Stop A or B(depending on channel in use) to
return to running screen.
b. Press the VTBI Softkey.
c. Program new VTBI using Numeric Keypad
d. Press ENTER
Ready to Start will appear at bottom of the screen.
Note: Primary infusion will resume following completion of bolus.
*Bolus option ONLY available if enabled by facility pharmacy.

13
First, Review ALARM Messages, then Press the ALARM SILENCEALARM SILENCE
button before proceeding.
6.1| Check Door Alarm
14
Alarm Messages
a. If Door is OPEN
- Pull BLACK FLOW CLAMP, Press ALARM SILENCE.
- Set BLACK FLOW CLAMP to Pull Through Position and Reload.
b. If Door is CLOSED, verify Door Handle is FULLY Latched.
- Press in on Door Handle, DO NOT push down on Purple Button.
- Press ALARM SILENCE.
c. If Door is CLOSED and FULLY Latched, Check Door Pin Condition.
- If Door Pin damaged or missing, remove pump from use.
- Contact facility Biomedical Department for repair.
Alarm Messages
6.2| PT Occlusion Alarm
Alarm will sound if occlusion between the pump and patient occurs.
Check between the pump and patient to assure that All Clamps have
been Opened and uid ow is permitted.
6.3| Inlet Occlusion Alarm
6.4| Bubble Detected Alarm
Alarm will sound if an air bubble is in the IV Set. Open the pump door, re-
move the set, follow the facility’s protocol on removing air. Re-install IV set
by repeating Steps 1.5 and secure tubing through Bubble Detector.
Alarm will sound if an occlusion has been detected between the uid
source and the pump. Check between the IV uids and the pump to assure
that all Clamps/Vents are Opened and no tubing is Kinked or Restricted.
Vent Clamp

15
Alert Messages
7.1| Rate Too High
Visual Message
Resolution
- During Bolus Setup: Check “Time”settings and adjust as necessary. Typically the “Time
value needs to be reviewed and increased if appropriate.
7.2| Rate Too Low
Visual Message
Resolution
- During Bolus Setup: Check “Time”settings and adjust as necessary. Typically the “Time
value needs to be reviewed and decreased if appropriate.
16
If ALARMS persist or other technical issues not mentioned in this guide, Remove Pump from use and
contact your facility’s Biomedical Department and/or IRadimed Technical Support at: 1-866-677-8022
Notes
_____________________________________________________
_____________________________________________________
_____________________________________________________
_____________________________________________________
_____________________________________________________
_____________________________________________________
_____________________________________________________
_____________________________________________________
_____________________________________________________
___________________________________________________

Full product technical specications are available upon request.
Refer to MRidium1138 Operation Manual for further instruction.
Information subject to change.
©IRadimed Corporation 2020 - All rights reserved
LB026-A 1025 Willa Springs Drive, Winter Springs, FL 32708 | (407) 677-8022
Always be aware of the Flow Clamp Position
Ready to Prime Push In Pull Through
112233
Scan For
Training Videos
Open Tubing Partially Clamped Clamped Tubing
Other manuals for MRIDIUM 3860+
1
Table of contents
Other IRadimed Medical Equipment manuals