manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. isweek
  6. •
  7. Thermometer
  8. •
  9. isweek AquaTuff FW2000 User manual

isweek AquaTuff FW2000 User manual

This manual suits for next models

1

Popular Thermometer manuals by other brands

Benetech GM900 instruction manual

Benetech

Benetech GM900 instruction manual

Parton PTD-400 manual

Parton

Parton PTD-400 manual

GAME 15900 owner's manual

GAME

GAME 15900 owner's manual

scigiene FP-180 instruction manual

scigiene

scigiene FP-180 instruction manual

Ametek Land UNO U1 600/1600C-L user guide

Ametek Land

Ametek Land UNO U1 600/1600C-L user guide

Delton 981A instruction manual

Delton

Delton 981A instruction manual

Taylor 5286498 instruction manual

Taylor

Taylor 5286498 instruction manual

Baker MM2P instruction manual

Baker

Baker MM2P instruction manual

Chino IR-FA Series instructions

Chino

Chino IR-FA Series instructions

Vive DMD1010 user manual

Vive

Vive DMD1010 user manual

Tarmo 320281 manual

Tarmo

Tarmo 320281 manual

Citizen CTD 504 instruction manual

Citizen

Citizen CTD 504 instruction manual

TFA 14.1500 instruction manual

TFA

TFA 14.1500 instruction manual

Gotemp THB4002 user guide

Gotemp

Gotemp THB4002 user guide

GREISINGER G 1202 Quick reference guide

GREISINGER

GREISINGER G 1202 Quick reference guide

Amprobe TMD-53W user manual

Amprobe

Amprobe TMD-53W user manual

Trotec BP10 operating manual

Trotec

Trotec BP10 operating manual

Radiance instruments RT616 Series product manual

Radiance instruments

Radiance instruments RT616 Series product manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

AquaTuff™
FW2000MK
Digital Thermocouple
Thermometer
Warranty
®
AquaTuff™
FW2000
5 YEAR
WARRANTY
Any instrument which proves
to be defective in material or
workmanship within five years
of original purchase will be
repaired or replaced without
charge upon receipt of the
unit prepaid with proof of
purchase. This Limited
Warranty does not cover
damage in shipment or failure
caused by tampering, obvious
carelessness, or abuse, and
is the purchaser’s sole
remedy.
MMaannuuffaaccttuurreeddiinntthheeUUSSAAbbyyCCOOOOPPEERR--AATTKKIINNSSCCOORRPPOORRAATTIIOONN
AAnnIISSOO99000011RReeggiisstteerreeddFFaacciilliittyy
27-958489E
Press to activate / deactivate the unit.
Note: Unit will automatically shut off after
1-1/2 minutes of inactivity.
Press to select Fahrenheit scale.
During Calibration, press to increase the
temperature displayed.
Press to select Celsius scale.
During Calibration, press to decrease the
temperature displayed.
Atkins Key. Used as a “Hold” button
during calibration only.
1. Press the ON/OFF Key to turn on the unit.
(No other button will activate the FW2000 / FW2000MK.)
2. Select the temperature scale you wish to have the unit
display.
-- Press the FKey to display Fahrenheit readings or
-- Press the CKey to display Celsius readings.
3.Insert probe into item to be tested.
Special notes:
• If the measured temperature is stable (doesn’t change by
more than 3°F (2°C)) for 1-1/2 minutes, the unit will shut off
automatically.
• If the battery needs to be replaced, three blinking decimal points
(. . . ) appear on the LCD display. This means that 8 hours of
battery life remain. As battery voltage becomes less, “b.A.t.”
will appear along with 3 blinking decimal points.
• If the unit gets a reading that is outside of its range
(-100° to 999°F )(-73° to 537°C), it will display either “HI” or “LO”,
as appropriate.
• The unit has a “continuous on” feature. To activate, the unit
must be on. Hold down the Atkins Key and press and release
the ON/OFF Key. To deactivate, press the ON/OFF Key and
the unit will turn off.
To calibrate the unit, the temperature of the test solution you are
using must be 32°F (0°C) ± 5°F (±2°C)
or
must be greater than
60°F (16°C). For best results, use an ice bath which consists
almost entirely of crushed ice, with only a little water. Proceed as
follows:
1. Turn unit on.
2. Place the probe into the test solution.
3. Hold down both the F and C keys until “CAL” is displayed.
This takes approximately 3 seconds. You are now in
calibration mode.
OPERATING
INSTRUCTIONS
4. Release both keys. “CAL” will be replaced by “Hold” on
the display.
5. Press and hold down the Atkins Key. “Hold” will be replaced
by three decimal points ( . . . ) with the current temperature
reading.
6. While holding the Atkins Key down, press the up or down
arrow keys (F or C) as appropriate. For example, if you are
using the ice bath method, and your reading is 35°F, press and
release the down arrow key as many times as needed until
“32F” displays on the LCD.
7. When you are done making adjustments, or if no
calibration is necessary, simply keep the Atkins Key
pressed until “End” displays on the LCD (approximately 7
seconds) and the unit returns to NORMAL mode. Release
the Atkins Key.
Special notes:
• Failure to keep the Atkins Key pressed down at all times during
calibration will take you out of the mode and read “Abrt”. To
correct this simply restart the calibration process from
Calibration step 1.
• If the unit cannot be recalibrated by using the ice bath
method you can reset the unit to factory settings. To do so,
while in the CAL mode, still holding the Atkins Key, press both
the F and C keys for 3 seconds to select the factory defaults.
“FAC” will be displayed on the LCD for 2 seconds.
HOLD MODE
If “Hold” is displayed on the LCD when the unit is turned on, it
means an error has occurred. To correct this error, press the Atkins
Key. This will reset the unit to its previous calibration.
THE KEYS
NORMAL OPERATION
CALIBRATION
CALIBRATION - Continued
ON
OFF
F
C
SPÉCIFICATIONS / FONCTIONS
FW2000 Temperature Range:
FW2000MK Temperature Range:
Probe Temperature Range:
Ambient Operating Range:
Accuracy from -35° to 350°F (-37° to 177°C)
in Ambient 60° to 80°F (16° to 27°C):
Accuracy from All Other Ranges:
Temperature Display Resolution:
Battery:
Battery Life:
Low Battery:
Automatic Shut Off:
LCD:
-100° to 500°F (-73° to 260°C)
-100° to 999°F (-73° to 537°C)
-100° to 500°F (-73° to 260°C)
0° to 115°F (-18° to 46°C)
±1F° (±0.5C°)
±2F° (±1C°)
1°F or 1°C
9 volt alkaline NEDA 1604
Minimum of 500 hours ON time
LCD displays 3 blinking decimal
points when 8 hours of battery
life remain
After 1-1/2 minutes of non-use
Liquid crystal display, with
0.5” (1.27 cm) high
digits; viewing area
1.5” x 0.7” (3.81 x 1.77 cm)