1.4.4 Temperature difference / (L in2)
To display the temperature difference push (ENTER/MENUE). Use the up and
down keys ( ) to select L in2. Push again (ENTER/MENUE). On the right
corner of the display appears a T1-T2. Use the up and down keys ( ) to
adjust the requested selection. Push (ENTER/MENUE) to confirm. Push (ESC)
to revert back to the measuring mode.
Note: Both channels have to be activated for displaying temperature difference.
Push (ENTER/MENUE) to confirm. Push (ESC) to revert back to the measuring
mode.
Note: Please check the probe selection to be sure that the correct probe is
entered. If a measurement value is changed with the ENTER-key at the menu
Prob, the standard calibration will be automatically used.
Indications for the combination probes (temperature and humidity):
Please adjust relatively humidity as measurement value for the channel on which
the combination probe is plugged. If you want that the measured temperature
of the plugged combination probe is shown as well, you have to deactivate the
channel on which no probe is plugged on (see chapter 1.4.6./Chnl)
Unit Prob L in 2 CAL ChnI Lo6 bLE
MENUE
1.4.5 Calibration function / (CAL)
Despite high quality manufacturing techniques, each probe is slightly different
from specified standards. To eliminate inaccuracies caused by exchanging or
ageing of probes, the instrument offer easy calibration functions which guarantee
that the system accura cy is always as good as if the instrument was specifically
calibrated to the individual probes in our laboratory.
The instruments offer ve different calibration options:
1) (CoFF): Standard characteristic curve
(e.g. Pt100-resistance according EN60751)
2) (oP 1): Calibration by code (2 x four digit code) is equivalent to a 2-point
calibration. The code is marked clearly by a label on each standard
probe.
3) (oP 2): Calibration by physical standard references
(1-point, 2-point or 3-point calibration)
4) (oP 3): Calibration according to Coefficients of EN60751 (RO, ABC)
5) (oP 4): Smart EEprom probes with internal calibration
CAL = calibration
Push (ENTER/MENUE) to calibrate the instrument with sensor. Use the up and
down keys ( ) to select CAL. Push again (ENTER/ MENUE). On the left
corner of the display appears a small 1, which indicates the selected channel. To
change the channel (1 or 2) use the up and down keys ( ). Push (ENTER/
MENUE) to confirm.
Use the up and down keys ( ) to select the requested calibration option.
Push (ENTER/MENUE) to confirm.
Unit Prob L in 2 CAL ChnI Lo6 bLE
MENUE
CoFF oP 1 oP 2 oP 3 oP 4
CAL
1) Standard calibration according to DIN EN60751 / (CoFF)
Use the up and down keys ( ) to select (CoFF). Push (ENTER/MENUE) to
confirm. Push (ESC) to revert back to the measuring mode.
2) CaIibrationby code / oP1
Use the up and down keys ( ) to select oP 1. Push (ENTER/ MENUE) to
confirm. On the bottom of the display appears a small 1, after this number a
four-digit number (Hex-Code/0..F) is dis played. For changing the number use the
up key ( ). To step to the next number use the down key ( ). lf the requested
number is complete then push (ENTER/MENUE) to confirm. At the bottom of the
display a very small 2 appears, after which a second four-digit number is displayed.
For changing the number please follow the manual as before. Push (ESC) to
revert back to the measuring mode.
Note:
After confirmin oP 1 by pushing (ENTER/MENUE) the function oP 1 (calibration
by code) is activated, even though you leave the menu by pressing (ESC).
Display-indication with active calibration code (oP 1):
The CAL-segment and the small 1 indicates to the user
that oP 1 is activated.
Measurement Probe selection LC-Display Suitable for these types
(Prob) of measuring instruments:
...............................................................................................................................
Temperature Pt100 (RTD) P700/P705/P750/P755/
P770/P755-LOG/P770-LOG
...............................................................................................................................
Temperature Fe-CuNi Typ J for all P700-instruments
...............................................................................................................................
Temperature NiCr-Ni Typ K for all P700-instruments
...............................................................................................................................
Temperature Fe-CuNi Typ L for all P700-instruments
...............................................................................................................................
Temperature NiCrSi-NiSi Typ N for all P700-instruments
...............................................................................................................................
Temperature Pt13Rh-Pt Typ R P700/P705/P750/P755/
P770/P755-LOG/P770-LOG
...............................................................................................................................
Temperature Pt10Rh-Pt Typ S P700/P705/P750/P755/
P770/P755-LOG/P770-LOG
...............................................................................................................................
Temperature Cu-CuNi Typ T for all P700-instruments
...............................................................................................................................
Humidity %rF P700/P705/P750/P755/
P770/P755-LOG/P770-LOG
...............................................................................................................................