
NOTE: For increasing the dosing precision, please refer to the
actual dosing liquid volume for calibration. The calibration liquid
volumes can be set to 10,20,30,40,50,60,70,80,90,100 (ml), and
the calibration duration will be stored in the memory. The closer
the calibration volume is to the actual dosing liquid volume, the
more accurate the actual dosing liquid volume will be.
For example: If user want the pump automatically dose about
10~30ml of the liquid, the calibration liquid volume can be
entered 20 (ml), and the channel is selected to 1. This calibration
duration will be automatically stored in the memory.
Calibration Method: Make sure the pump tubing is primed with
the liquid first by manual mode. Using a measuring cylinder (20ml
or larger), put the channel 1 outlet hose into the cylinder, and put
the channel 1 inlet hose into the liquid. Tap the “Start calibration”
switch to be on, then the channel 1 calibration starts.