manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Katronic
  6. •
  7. Measuring Instrument
  8. •
  9. Katronic KATflow 150 Instruction manual

Katronic KATflow 150 Instruction manual

Katronic Technologies Ltd . I Telephone: +44 (0)1926 882 954 I Email: info@katronic.co.uk I Web: www.katronic.co.uk
KATflow 150 Quick Start Manual
Flow in pipe
Step 1: Selection of measurement point & pipe preparation
●Avoid installation of sensors in the vicinity of deformations and
pipe defects, near welding seams or where deposits could have
accumulated.
●Select a measuring point with sufficient straight pipe to obtain
accurate measurements. Please consult the manual for the
recommended distances from sources of disturbance.
●For a horizontal pipe, mount the sensors on the side of the
pipe. For a vertical pipe, mount the sensors at a location where
the liquid flows upwards. (Pic 1)
●Mount the sensors in the direction of the flow. (Pic 2)
●Clean the pipe at the measurement point. Remove loose paint
and rust with a wire brush or file.
●Apply coupling paste to the face of the clamp-on sensors before
attaching them to the pipe.
Pic 1: Mounting points Pic 2: Flow direction
Step 2: Keyboard familiarisation
Select Setup Wizard CH1 to set
up measurement channel 1, or
CH2 for channel 2. Confirm
selection by pressing ENTER. If
the sensors are recognised, the
serial number will be shown. If
not, they can be selected.
Select the main measurement
unit using the cursor keys and
confirm with ENTER. This unit
will be displayed in the middle of
the measurement screen. Selec-
ting OFF deactivates the mea-
surement channel.
Select the pipe material using
the cursor keys and confirm with
ENTER.
Enter the outer pipe diameter
using the keypad and confirm
with ENTER. Use UP key as
backspace to correct for entry
errors. If “0” is entered and con-
firmed, an additional screen
appears that allows entry of the
circumference. Press ENTER to
confirm.
1 / 3
●Show NEXT (1) available item
●QON (2) = Start totalizer function
●Show next DISPlay (3)
●QOFF (8) = Stop totalizer function
●DIRECT (9) access to trend plot
●Move menu / selection item UP
●Move menu / selection item DOWN
●ESCape entry without saving
Switch device OFF (press > 2 sec.)
●ENTER selection with saving
Enter the outer pipe diameter
using the keypad and confirm
with ENTER. Use UP key as
backspace to correct for entry
errors. If “0” is entered and con-
firmed, an additional screen
appears that allows entry of the
circumference. Press ENTER to
confirm.
Enter pipe wall thickness using
the keypad and confirm with
ENTER. Use UP key as a back-
space to correct for entry errors.
Select fluid using cursor keys.
Confirm by pressing ENTER.
Enter the fluid temperature using
the keypad. Confirm by pressing
ENTER.
Use UP key as a backspace to
correct for entry errors.
The flow meter can be prepared for measurement with the Setup
Wizard as found in the Quick start menu.
Step 3: Quick Start menu and Setup Wizard
Upon first power on and the boot
up sequence, the Main Menu is
displayed.
Use the UP and DOWN cursor
keys to select Quick start and
confirm by pressing ENTER.Select pipe liner material using
cursor keys and confirm by pres-
sing ENTER. If a liner material is
chosen, an additional screen
appears that allows entry of liner
thickness.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2011 I Issue: KF150QuickStartV30E0711
Step 3: Quick Start menu and Setup Wizard (continued)
Select number of sound passes
(sound paths) using cursor keys.
Auto: Selection by flow meter
according to entered parameters
(number of passes shown later
on Sensor Positioning Screen).
1: 1 pass, (diagonal mode)
2: 2 passes (reflection mode)
3: 3 passes (diagonal mode)
4: 4 passes (reflection mode)
5: 5 passes (diagonal mode)
..etc.
1. Reflection mode
The flow meter uses an even number of passes. This is the most
convenient mounting arrangement, as the transducer separation
distance can be measured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3)
2. Diagonal mode
Signal travels on odd number of passes through the pipe. Used
for larger pipes and for dirty/aerated liquids where greater signal
attenuation can occur. The sensor distance on this mounting con-
figuration can be negative (sensors overlapping). (Pic 4)
The main measurement unit is
displayed when first entering the
measurement screen.
Press MUX to switch between
channels 1 and 2. Press NEXT
to see up to 3 units displayed
with the main one in the middle.
KATflow 150 Quick Start Manual
With correct pipe parameters entered and the sensors mounted
at the suggested spacing, the mark should be near the central
indication line (see three lines below the bottom signal bar). If the
mark is to the left-hand side of the central indication line, the
sensors are too close to each other. If the mark is to the right, the
sensors are too far apart. Slide one sensor carefully along the
pipe to bring the mark into a more central position. Measure-
ments can be obtained with the mark being between the left and
right indication line. Press ENTER to start measurement.
Two further measurement units can be assigned to that screen
by going to Main Menu - Display - Channel 1 / Channel 2.
Even number of passes: Both sensors on same side of pipe. Odd
number of passes: Sensors on opposite sides (see below for
more information). Confirm with ENTER. Your second channel
can be prepared now by selecting Setup Wizard CH1 or CH2.
Measurement screen
Katronic Technologies Ltd . I Telephone: +44 (0)1926 882 954 I Email: info@katronic.co.uk I Web: www.katronic.co.uk
Select Start Measurement and
confirm with ENTER to start the
sensor positioning procedure.
When both channels have been
activated, the procedure for CH2
follows the one of CH1.
2 / 3
2. Diagonal mode
Signal travels on odd number of passes through the pipe. Used
for larger pipes and for dirty/aerated liquids where greater signal
attenuation can occur. The sensor distance on this mounting con-
figuration can be negative (sensors overlapping). (Pic 4)
Mount the transducers with the suggested spacing between the
inside of the sensor heads. This distance has been determined
by the flow meter on the basis of the entered parameters. Use
the displayed number of passes to install the sensors on the
correct side of the pipe (see modes above). Observe the upper
bar (signal-to-noise ratio) and lower bar (signal quality). These
should both be filled to about the same level with a filling level of
around 1/3 or more desired. Use the moving mark between the
two signal bars for fine adjustment of the sensor position.
Internal data logger
The totalizer is shown when in measurement mode after pressing
NEXT three times. It can also be assigned to the three line
display, datalogger or process outputs by selecting a volumetric
unit. The totalizer function is started with QON when in mea-
surement mode (measurement screen displayed). Pressing Q+
resets the total in positive flow direction. Pressing Q- resets the
total in negative flow direction. The totalizer function can be
stopped with QOFF. Pressing QON again will reset the positive,
negative and overall totalizer. Change displays without resetting
the totalizer by pressing DISP or NEXT.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2011 I Issue: KF150QuickStartV30E0711
Sensor Positioning Screen
Pic 3: Reflection mode Pic 4: Diagonal mode
Two further measurement units can be assigned to that screen
by going to Main Menu - Display - Channel 1 / Channel 2.
Totalizer
The data logger is reached via Main Menu - Output. It is activated
in Datalogger - Interval by entering and confirming a non-zero
value. Enter “0” and confirm to disable the logger. Up to 10
measurement units can be seleceted for logging under
Datalogger - Selection. Use the cursor keys to highlight a unit
and press ENTER to select it. Press “0” to deselect it. An
activated data logger is indicated by a “document” symbol in the
top left corner of the display. On start of measurement (measure-
ment screen displayed) the logger records the selected measure-
ment units. A blinking “document” symbol indicates a recording
data logger. Separation markers are set by the data logger
whenever a session beginns. Leave the measurement screen by
pressing ESC to stop recording. The recording interval can be
changed in Datalogger - Interval. The data logger is cleared
under Datalogger - Log Erase. Ensure all required data has been
downloaded .
Heat quantity measurement (where specified)
After setting up the flow measurement channel go to Main Menu -
In/Output, in order to assign the PT100 inputs to the channel:
1.) Use the cursor keys to select PT100 4 WIRE and confirm
with ENTER.
2.) On the next screen, use the cursor keys to select the
flow channel which the temperature input has to be applied
to. You can choose between Channel 1 or 2. Selecting OFF
will disable the input. Confirm with ENTER.
3.) On the following screen choose PT100 to employ the tem-
perature measured on the pipe. If you wish to enter a fixed
temperature value, select User and enter the value.
4.) As next step define whether the PT100 sensor measures
the inlet or outlet temperature. Use the cursor keys to select
the appropriate and confirm with ENTER.
5.) On the next screen a temperature offset can be applied.
After confirming again with ENTER you reach the Main
Menu.
Once the setup of the first temperature input has been completed
repeat steps 1 to 5 for the configuration of the second input.
Process in/outputs: Setup
KATflow 150 Quick Start Manual
The KATflow 150 is capable of measuring heat flow and heat
quantity. If you wish to measure heat flow, select a heat flow unit
(W, kW, MW) as Middle Unit when going through the Setup
Wizard of the selected measurement channel. If you wish to
measure heat quantity, select a heat quantity unit (J, kJ, MJ).
On selection of one of these units, the flow meter requests the
entry of the specific heat capacity of the medium in J/(g.k). Enter
the value for your fluid and confirm with ENTER. Complete the
remaining steps of the Setup Wizard.
The temperatures at the inlet and outlet of the monitored system
are used to determine heat flow and heat quantity. Connect 4-
wire PT100 temperature sensors to the flow meter as per the
illustration below. (Pic 5)
Temperature and flow measurement
Katronic Technologies Ltd . I Telephone: +44 (0)1926 882 954 I Email: info@katronic.co.uk I Web: www.katronic.co.uk
For separate measurement of flow and temperature select a flow
unit as Middle Unit in the Setup Wizard. Complete the wizard and
then go to Main Menu - In/Output. Assign one PT100 input to the
flow channel as described on the left. Define it as an inlet tem-
perature source.
Now go to Main Menu - Datalogger - Selection and choose Tin
as one of the recorded measurement units. Upon start of each
measurement the flow meter will now record selected flow units
as well as the temperature measured by the PT100 sensor.
In/outputs can be set up in Main Menu - In/Output. Depending on
which in/outputs are available, the displayed list contains one or
more in/outputs. The in/output listed at the very top of the list is
available on Slot 1 inside the connection compartment, the in/
output listed second from top is available on Slot 2 and so on.
After selecting an in/output, it can be assigned to a measuere-
ment channel. The structure of the further steps depends on the
individual type of output.
Process in/outputs: Connection
In/outputs can be connected to devices via screw terminals.
These can be found inside the connection compartment below
the control panel. Remove the screws at the left and right to gain
access to the compartment.
Connection to the outputs can be established by wiring transfer
cables into the respective terminal blocks of the I/O box. The
illustration below gives an overview of the terminal assignment.
(Pic 5)
3 / 3
After setting up the flow measurement channel go to Main Menu -
In/Output, in order to assign the PT100 inputs to the channel:
1.) Use the cursor keys to select PT100 4 WIRE and confirm
with ENTER.
2.) On the next screen, use the cursor keys to select the
flow channel which the temperature input has to be applied
to. You can choose between Channel 1 or 2. Selecting OFF
will disable the input. Confirm with ENTER.
3.) On the following screen choose PT100 to employ the tem-
perature measured on the pipe. If you wish to enter a fixed
temperature value, select User and enter the value.
4.) As next step define whether the PT100 sensor measures
the inlet or outlet temperature. Use the cursor keys to select
the appropriate and confirm with ENTER.
5.) On the next screen a temperature offset can be applied.
After confirming again with ENTER you reach the Main
Menu.
Once the setup of the first temperature input has been completed
repeat steps 1 to 5 for the configuration of the second input.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2011 I Issue: KF150QuickStartV30E0711
In/outputs can be connected to devices via screw terminals.
These can be found inside the connection compartment below
the control panel. Remove the screws at the left and right to gain
access to the compartment.
Connection to the outputs can be established by wiring transfer
cables into the respective terminal blocks of the I/O box. The
illustration below gives an overview of the terminal assignment.
(Pic 5)
Pic 5: Input/output compartment, inside terminals

Other manuals for KATflow 150

2

This manual suits for next models

8

Other Katronic Measuring Instrument manuals

Katronic KATflow 150 User manual

Katronic

Katronic KATflow 150 User manual

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 210 User manual

Katronic

Katronic KATflow 210 User manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 230 User manual

Katronic

Katronic KATflow 230 User manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Popular Measuring Instrument manuals by other brands

Avtech TemPageR 3E User's guide & reference manual

Avtech

Avtech TemPageR 3E User's guide & reference manual

Aqua Medic 65910 Operation manual

Aqua Medic

Aqua Medic 65910 Operation manual

Centech 66400 Instructions and precautions

Centech

Centech 66400 Instructions and precautions

LogiCO2 Scout Glow user manual

LogiCO2

LogiCO2 Scout Glow user manual

Dilog DL6799 instruction manual

Dilog

Dilog DL6799 instruction manual

ATO YT-1200H user manual

ATO

ATO YT-1200H user manual

Promax PROMAX-12 user manual

Promax

Promax PROMAX-12 user manual

EMU Check Short manual

EMU

EMU Check Short manual

Prostat PGA-710 How To Measure Guide

Prostat

Prostat PGA-710 How To Measure Guide

Delta RetroSign GR1 quick guide

Delta

Delta RetroSign GR1 quick guide

Challenger Optics CO-OPM-PON Interactive operating manual

Challenger Optics

Challenger Optics CO-OPM-PON Interactive operating manual

Sen Source PC-PIR15 installation guide

Sen Source

Sen Source PC-PIR15 installation guide

Midwest 116 DPI Installation and operating instructions

Midwest

Midwest 116 DPI Installation and operating instructions

J-TECH Solar Auxiliary Power Unit Operator's guide

J-TECH

J-TECH Solar Auxiliary Power Unit Operator's guide

Applied Instruments DP Plus Application note

Applied Instruments

Applied Instruments DP Plus Application note

Long-run LRF-2000S user manual

Long-run

Long-run LRF-2000S user manual

Tek-Trol TEK-DP 1610D quick start guide

Tek-Trol

Tek-Trol TEK-DP 1610D quick start guide

Centech 47257 operating instructions

Centech

Centech 47257 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.