KATflow 200 Quick Start Manual
Flow in pipe
Step 1: Selection of measurement point & pipe preparation
●Avoid installation of sensors in the vicinity of deformations and
pipe defects, near welding seams or where deposits could have
accumulated.
●Select a measuring point with sufficient straight pipe to obtain
accurate measurements. Please consult the manual for the
recommended distances from sources of disturbance.
●For a horizontal pipe, mount the sensors on the side of the
pipe. For a vertical pipe, mount the sensors at a location where
the liquid flows upwards. (Pic 1)
●Mount the sensors in the direction of the flow. (Pic 2)
●Clean the pipe at the measurement point. Remove loose paint
and rust with a wire brush or file.
●Apply coupling paste to the face of the clamp-on sensors before
attaching them to the pipe.
Pic 1: Mounting points Pic 2: Flow direction
Step 2: Keyboard familiarisation
Select Setup Wizard to set up
the flow meter for measurement.
If the sensors are recognised,
the serial number will be shown.
If not, they can be selected.
Select the main measurement
unit using the cursor keys and
confirm with ENTER. This unit
will be displayed in the middle of
the measurement screen. Selec-
ting OFF deactivates the mea-
surement channel.
Select the pipe material using
the cursor keys and confirm with
ENTER.
Enter the outer pipe diameter
using the keypad and confirm
with ENTER. Use UP key as
backspace to correct for entry
errors. If “0” is entered and con-
firmed, an additional screen
appears that allows entry of the
circumference. Press ENTER to
confirm.
1 / 2
●Show NEXT (1) available item
●QON (2) = Start totalizer function
●Show next DISPlay (3)
●QOFF (8) = Stop totalizer function
●DIRECT (9) access to trend plot
●Move menu / selection item UP
●Move menu / selection item DOWN
●ESCape entry without saving
Switch device OFF (press > 2 sec.)
●ENTER selection with saving
Switch device ON (press > 2 sec.)
Enter the outer pipe diameter
using the keypad and confirm
with ENTER. Use UP key as
backspace to correct for entry
errors. If “0” is entered and con-
firmed, an additional screen
appears that allows entry of the
circumference. Press ENTER to
confirm.
Enter pipe wall thickness using
the keypad and confirm with
ENTER. Use UP key as a back-
space to correct for entry errors.
Select fluid using cursor keys.
Confirm by pressing ENTER.
Enter the fluid temperature using
the keypad. Confirm by pressing
ENTER.
Use UP key as a backspace to
correct for entry errors.
The flow meter can be prepared for measurement with the Setup
Wizard as found in the Quick start menu.
Step 3: Quick Start menu and Setup Wizard
Upon first power on and the boot
up sequence, the Main Menu is
displayed.
Use the UP and DOWN cursor
keys to select Quick start and
confirm by pressing ENTER.Select pipe liner material using
cursor keys and confirm by pres-
sing ENTER. If a liner material is
chosen, an additional screen
appears that allows entry of liner
thickness.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2011 I Issue: KF200QuickStartV30E0711
Step 3: Quick Start menu and Setup Wizard (continued)
Select number of sound passes
(sound paths) using cursor keys.
Auto: Selection by flow meter
according to entered parameters
(number of passes shown later
on Sensor Positioning Screen).
1: 1 pass, (diagonal mode)
2: 2 passes (reflection mode)
3: 3 passes (diagonal mode)
4: 4 passes (reflection mode)
5: 5 passes (diagonal mode)
..etc.
1. Reflection mode
The flow meter uses an even number of passes. This is the most
convenient mounting arrangement, as the transducer separation
distance can be measured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3)
2. Diagonal mode
Signal travels on odd number of passes through the pipe. Used
for larger pipes and for dirty/aerated liquids where greater signal
attenuation can occur. The sensor distance on this mounting con-
figuration can be negative (sensors overlapping). (Pic 4)
The main measurement unit is
displayed when first entering the
measurement screen.
Press NEXT to see up to 3 units
displayed with the main one in
the middle.
KATflow 200 Quick Start Manual
With correct pipe parameters entered and the sensors mounted
at the suggested spacing, the mark should be near the central
indication line (see three lines below the bottom signal bar). If the
mark is to the left-hand side of the central indication line, the
sensors are too close to each other. If the mark is to the right, the
sensors are too far apart. Slide one sensor carefully along the
pipe to bring the mark into a more central position. Measure-
ments can be obtained with the mark being between the left and
right indication line. Press ENTER to start measurement.
Two further measurement units can be assigned to that screen
by going to Main Menu - Output - Display .
Even number of passes: Both sensors on same side of pipe. Odd
number of passes: Sensors on opposite sides (see below for
more information). Confirm with ENTER.
Measurement screen
Katronic Technologies Ltd . I Telephone: +44 (0)1926 882 954 I Email: info@katronic.co.uk I Web: www.katronic.co.uk
Select Start Measurement and
confirm with ENTER to start the
sensor positioning procedure.
2 / 2
2. Diagonal mode
Signal travels on odd number of passes through the pipe. Used
for larger pipes and for dirty/aerated liquids where greater signal
attenuation can occur. The sensor distance on this mounting con-
figuration can be negative (sensors overlapping). (Pic 4)
Mount the transducers with the suggested spacing between the
inside of the sensor heads. This distance has been determined
by the flow meter on the basis of the entered parameters. Use
the displayed number of passes to install the sensors on the
correct side of the pipe (see modes above). Observe the upper
bar (signal-to-noise ratio) and lower bar (signal quality). These
should be both filled to about the same level with a filling level of
around 1/3 or more desired. Use the moving mark between the
two signal bars for the fine adjustment of the sensor position.
Internal data logger
The totalizer is shown when in measurement mode after pressing
NEXT three times. It can also be assigned to the three line
display, datalogger or process outputs by selecting a volumetric
unit. The totalizer function is started with QON when in mea-
surement mode (measurement screen displayed). Pressing Q+
resets the total in positive flow direction. Pressing Q- resets the
total in negative flow direction. The totalizer function can be
stopped with QOFF. Pressing QON again will reset the positive,
negative and overall totalizer. Change displays without resetting
the totalizer by pressing DISP or NEXT.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2011 I Issue: KF200QuickStartV30E0711
Sensor Positioning Screen
Pic 3: Reflection mode Pic 4: Diagonal mode
Two further measurement units can be assigned to that screen
by going to Main Menu - Output - Display .
Totalizer
The data logger is reached via Main Menu - Output. It is activated
in Datalogger - Interval by entering and confirming a non-zero
value. Enter “0” and confirm to disable the logger. Up to 10
measurement units can be seleceted for logging under
Datalogger - Selection. Use the cursor keys to highlight a unit
and press ENTER to select it. Press “0” to deselect it. An
activated data logger is indicated by a “document” symbol in the
top left corner of the display. On start of measurement (measure-
ment screen displayed) the logger records the selected measure-
ment units. A blinking “document” symbol indicates a recording
data logger. Separation markers are set by the data logger
whenever a session beginns. Leave the measurement screen by
pressing ESC to stop recording. The recording interval can be
changed in Datalogger - Interval. The data logger is cleared
under Datalogger - Log Erase. Ensure all required data has been
downloaded .

Other manuals for KATflow 200

4

Other Katronic Measuring Instrument manuals

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 150 User manual

Katronic

Katronic KATflow 150 User manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 230 User manual

Katronic

Katronic KATflow 230 User manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 210 User manual

Katronic

Katronic KATflow 210 User manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Popular Measuring Instrument manuals by other brands

ABB GLA133 Series user manual

ABB

ABB GLA133 Series user manual

Simex SRT-N118-XA user manual

Simex

Simex SRT-N118-XA user manual

EUTECH INSTRUMENTS POTASSIUM EPOXY instruction manual

EUTECH INSTRUMENTS

EUTECH INSTRUMENTS POTASSIUM EPOXY instruction manual

Lignomat Ligno-VersaTec User instructions

Lignomat

Lignomat Ligno-VersaTec User instructions

emitor DIGIAIR user manual

emitor

emitor DIGIAIR user manual

VOLTCRAFT AM-39X14/TEMP Operatoring Instructions

VOLTCRAFT

VOLTCRAFT AM-39X14/TEMP Operatoring Instructions

CONDTROL Qb user manual

CONDTROL

CONDTROL Qb user manual

Fluke ScopeMeter 123B User manaul

Fluke

Fluke ScopeMeter 123B User manaul

AYTU BioSience mioxsys user manual

AYTU BioSience

AYTU BioSience mioxsys user manual

LUDLUM 3030E-2 manual

LUDLUM

LUDLUM 3030E-2 manual

YOKOGAWA ScopeCorder DL750 user manual

YOKOGAWA

YOKOGAWA ScopeCorder DL750 user manual

YOKOGAWA 702915 user manual

YOKOGAWA

YOKOGAWA 702915 user manual

Geokon 4050 instruction manual

Geokon

Geokon 4050 instruction manual

Bushnell Phantom user manual

Bushnell

Bushnell Phantom user manual

Emerson Micro Motion 7835 installation manual

Emerson

Emerson Micro Motion 7835 installation manual

TSI Incorporated Alnor CF920 Operation and service manual

TSI Incorporated

TSI Incorporated Alnor CF920 Operation and service manual

Datalogic VLASE IR 1109-1 42 Series user manual

Datalogic

Datalogic VLASE IR 1109-1 42 Series user manual

Kobold OVZ operating instructions

Kobold

Kobold OVZ operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.