Step 3: Quick Start Function and Setup Wizard (Continued)
Select pipe lining material using
cursor keys.
Confirm by pressing <ENTER>.
Select transducer configuration
(number of passes) using cursor
keys.
Auto: Automatic selection by
meter according to the entered
parameters (number of passes
shown at later stage on Sensor
Positioning Screen).
1: 1 pass, diagonal mode
2: 2 passes, reflection mode
3: 3 passes, diagonal mode
4: 4 passes, reflection mode
5: 5 passes, diagonal mode
6: 6 passes, reflection mode
..etc.
Confirm by pressing <ENTER>.
1. Reflection Mode
The flow meter uses an even number of passes. This is the most
convenient mounting arrangement, as the transducer separation
distance can be measured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3)
2. Diagonal Mode
The signal travels on an odd number of passes through the pipe.
Used for larger pipes where greater signal attenuation might
occur. The sensor distance on this mounting configuration can be
negative (sensors overlapping). (Pic 4)
Pic 3: Reflection mode Pic 4: Diagonal mode
Use cursor keys to select Start
Measurement.
Confirm by pressing <ENTER>.
If Setup Wizard Dual has been chosen, parameter entry for
channel 2 will start after the selection of the number of passes for
channel 1. The respective channel number is indicated on each
screen at the top left corner.
Sensor Positioning Screen
Mount the transducers with the
suggested spacing. This has
been determined by the flow
meter on basis of to the entered
parameters.
The suggested sensor spacing
is the distance from the inside
edge of each sensor.
Use the middle line between the
signal indication bars for the fine
adjustment of the sensor moun-
ting position. A central position
of this line is desired.
If the line between the bars is to
the left-hand side of the lower
middle line, the sensors are too
close to each other. If the line
between the bars is to the right,
the sensors are too far apart.
Observe the upper bar (signal-
to-noise ratio) and lower bar
(signal quality). These should be
of identical length and as far to
the right as possible.
Confirm by pressing <ENTER>
to obtain measurements.
The flow totaliser can be started
or reset by pressing <Q
ON
>. The
totalisers can be stopped with
<Q
OFF
>.
Switch between flow rate and
totaliser without resetting by
pressing <NEXT>.
Press <DISP> in order to switch
between the selected process
value and the diagnostic screen.
The main process value can be
changed in the Quick Start or
Output section (Output-Display-
Units). Press <ESC> to return to
the main menu. Change to the
diagnostic display by pressing
<DISP>. Cycle through available
flow channels using <MUX>.
Process Value Display, Totaliser and Datalogger
The Datalogger function is selected from the main menu. Enable
it by entering a non-zero value (1 s to 999 s) for the interval. The
active process values will be logged when in measurement
mode.
Setting an interval value of zero stops the data logger and retains
measured data. The changing from an interval of zero to a non-
zero value will clear the logger. Ensure that any required data is
downloaded before doing this. Changing between two non-zero
intervals does not clear the logger.
An activated datalogger is indicated by a flashing document icon
at the top left corner of the measurement screen. Measurement
sessions are recorded separately.