  9. Katronic KATflow 150 Instruction manual

Katronic KATflow 150 Instruction manual

KATflow 150 Quick Start Manual
1 / 3
Flow in pipe
Use the cursor keys to select
Setup Wizard Single for single
channel measurement (one set
of clamp-on sensors).
Select Setup Wizard Dual, for
dual channel measurement (two
sets of clamp-on sensors).
Confirm selection by pressing
<ENTER>.
If the sensors are recognised,
the serial number will be shown.
If they are not recognised or not
connected, they may be selec-
ted from a list.
Select the units of measurement
using cursor keys and confirm
by pressing <ENTER>.
Choose the pipe material using
cursor keys and confirm
selection by pressing <ENTER>.
Enter outside pipe diameter
using alphanumerical keys and
confirm by pressing <ENTER>.
Use <UP> key as a backspace
to correct for data entry errors.
If “0” is entered, an additional
screen appears that allows entry
of the pipe circumference. Press
<ENTER> to confirm.
Enter pipe wall thickness using
alphanumerical keys and con-
firm by pressing <ENTER>.
Use <UP> key as a backspace
to correct for data entry errors.
Select fluid using cursor keys.
Confirm by pressing <ENTER>.
Enter process temperature using
alphanumerical keys. Confirm by
pressing <ENTER>.
Use <UP> key as a backspace
to correct for data entry errors.
Continue on page 2.
Step 1: Selection of measurement point & pipe preparation
Pic 1: Mounting points Pic 2: Flow direction
Step 2: Keyboard Familiarisation
●Show NEXT (1) available item
●Q
ON
(2) = Start totaliser function
●Show next DISPlay (3)
●Q
OFF
(8) = Stop totaliser function
●DIRECT (9) access to trend plot
●Move menu / selection item UP
●Move menu / selection item DOWN
●ESCape entry without saving
●ENTER selection with saving
After the provisional installation of the sensors, connect the trans-
mitter to a power source. You can then enter the parameters into
the Setup Wizard to prepare the flow meter for measurement.
Step 3: Quick Start Menu and Setup Wizard
Upon first power on and the
boot-up sequence, the main
menu is displayed. Use <UP>
and <DOWN> cursor keys to
select Quick start. Confirm by
pressing <ENTER>.
On subsequent restoration of power the flow meter will be in
measurement mode. Press <ESC> to return to the main menu.
●Avoid installation of sensors in the vicinity of deformations and
pipe defects, near welding seams or where deposits could have
accumulated.
●Select a measuring point with sufficient straight pipe to obtain
accurate measurements, generally a length of ten times the
pipe diameter on the inlet of the measuring point, and a five
times pipe diameter on the outlet.
●For a horizontal pipe, mount the sensors on the side of the
pipe. For a vertical pipe, mount the sensors at a location where
the liquid flows upwards. (Pic 1)
●Mount the sensors in the direction of the flow. (Pic 2)
●Clean the pipe at the measuring point. Remove loose paint and
rust with a wire brush or file.
●Apply coupling paste to the bottom of the clamp-on sensors
before installation. Leave straps slightly loose so sensors can
be moved along the pipe during the positioning process.
Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I [email protected] I www.katronic.co.uk
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2010. I Issue: KF150QuickStartV22E0710
Step 3: Quick Start Function and Setup Wizard (Continued)
Select pipe lining material using
cursor keys.
Confirm by pressing <ENTER>.
Select transducer configuration
(number of passes) using cursor
keys.
Auto: Automatic selection by
meter according to the entered
parameters (number of passes
shown at later stage on Sensor
Positioning Screen).
1: 1 pass, diagonal mode
2: 2 passes, reflection mode
3: 3 passes, diagonal mode
4: 4 passes, reflection mode
5: 5 passes, diagonal mode
6: 6 passes, reflection mode
..etc.
Confirm by pressing <ENTER>.
1. Reflection Mode
The flow meter uses an even number of passes. This is the most
convenient mounting arrangement, as the transducer separation
distance can be measured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3)
2. Diagonal Mode
The signal travels on an odd number of passes through the pipe.
Used for larger pipes where greater signal attenuation might
occur. The sensor distance on this mounting configuration can be
negative (sensors overlapping). (Pic 4)
Pic 3: Reflection mode Pic 4: Diagonal mode
Use cursor keys to select Start
Measurement.
Confirm by pressing <ENTER>.
If Setup Wizard Dual has been chosen, parameter entry for
channel 2 will start after the selection of the number of passes for
channel 1. The respective channel number is indicated on each
screen at the top left corner.
Sensor Positioning Screen
Mount the transducers with the
suggested spacing. This has
been determined by the flow
meter on basis of to the entered
parameters.
The suggested sensor spacing
is the distance from the inside
edge of each sensor.
Use the middle line between the
signal indication bars for the fine
adjustment of the sensor moun-
ting position. A central position
of this line is desired.
If the line between the bars is to
the left-hand side of the lower
middle line, the sensors are too
close to each other. If the line
between the bars is to the right,
the sensors are too far apart.
Observe the upper bar (signal-
to-noise ratio) and lower bar
(signal quality). These should be
of identical length and as far to
the right as possible.
Confirm by pressing <ENTER>
to obtain measurements.
The flow totaliser can be started
or reset by pressing <Q
ON
>. The
totalisers can be stopped with
<Q
OFF
>.
Switch between flow rate and
totaliser without resetting by
pressing <NEXT>.
Press <DISP> in order to switch
between the selected process
value and the diagnostic screen.
The main process value can be
changed in the Quick Start or
Output section (Output-Display-
Units). Press <ESC> to return to
the main menu. Change to the
diagnostic display by pressing
<DISP>. Cycle through available
flow channels using <MUX>.
Process Value Display, Totaliser and Datalogger
The Datalogger function is selected from the main menu. Enable
it by entering a non-zero value (1 s to 999 s) for the interval. The
active process values will be logged when in measurement
mode.
Setting an interval value of zero stops the data logger and retains
measured data. The changing from an interval of zero to a non-
zero value will clear the logger. Ensure that any required data is
downloaded before doing this. Changing between two non-zero
intervals does not clear the logger.
An activated datalogger is indicated by a flashing document icon
at the top left corner of the measurement screen. Measurement
sessions are recorded separately.
2 / 3
KATflow 150 Quick Start Manual
Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I [email protected] I www.katronic.co.uk
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2010. I Issue: KF150QuickStartV22E0710
3 / 3
KATflow 150 Quick Start Manual
RS 485 Serial Interface
The RS 485 interface is used for connecting up to 32 flow meters
to a computer system. The communication protocol used
conforms to the conventions of the Modbus RTU protocol. Please
refer to customer support for further information. The ASCII
printer output can also be transmitted through the RS 485
interface.
Analogue Current Output 0/4 ... 20 mA
The analogue current outputs operate in a 4 ... 20 mA or 0 ... 20
mA span. Current outputs may be assigned to process values in
the “mode” section of the output menu. The current outputs can
be programmed and scaled within the menu structure.
Digital Open-Collector Output
Open-Collector output 1 is assigned to the totalizer function of
channel 1 which is related to the selected process value. Open-
Collector output 2 is assigned in the same way to channel 2 (on a
two-channel system).
Digital Relay Output
Relay outputs may be assigned to process values in the “mode”
section of the output menu. The relay outputs are configured
using the menu structure.
Heat Quantity Measurement (HQM), optional
Where equipped, heat quantity (energy) and heat flow (energy
flow) can be measured on Channel 1 of the flow meter.
If a heat quantity unit is specified for the process value, the
KATflow 150 will prompt the user for the specific heat capacity of
the medium in J/g/K (for example 4.18 J/g/K for water).
The Setup Wizard will then allow the user to select the tem-
perature input source; either PT100 temperature sensors or a
fixed value for measurement against a known inlet or outlet
temperature. Where PT100 sensors are selected, the Wizard will
prompt the user for a temperature offset, which may be useful
where the temperature of the medium differs from the tem-
perature of the pipe wall (for example with unlagged pipes). If a
fixed value is selected, the Wizard will ask the user to specify this
value. The Wizard will then proceed with the installation
parameters as in step 3.
When heat quantity units are selected, these behave as any
other Process Value and may be totalized, logged, or applied to a
Process Output.
Connection of PT100 Sensors and Installation to Pipe
For heat quantity measurements a set of flow sensors connected
to channel no. 1 may be installed on a suitable pipe section along
with two PT100 temperature sensors which are mounted to the
pipe at the inlet and outlet of the chosen location.
Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I [email protected] I www.katronic.co.uk
The PT100 sensors are installed
with the flat side of the sensor
facing the pipe surface. The
sensors can be held in place
using the supplied chains and
springs (Pic 5). Alternatively,
cable straps can be used.
Where no lagging is present, it is
recommended that the PT100
sensors are insulated to obtain
more accurate measurements.
Pic 5: PT100 sensor
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2010. I Issue: KF150QuickStartV22E0710

Other manuals for KATflow 150

2

Other Katronic Measuring Instrument manuals

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 230 User manual

Katronic

Katronic KATflow 230 User manual

Katronic KATflow 150 User manual

Katronic

Katronic KATflow 150 User manual

Katronic KATflow 210 User manual

Katronic

Katronic KATflow 210 User manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Popular Measuring Instrument manuals by other brands

Nordson Rhino SD2 manual

Nordson

Nordson Rhino SD2 manual

Rae ppbRAE 3000 user guide

Rae

Rae ppbRAE 3000 user guide

Siemens SIPROCESS GA700 Compact operating instructions

Siemens

Siemens SIPROCESS GA700 Compact operating instructions

Viavi SmartPocket OLS-34V2 operating manual

Viavi

Viavi SmartPocket OLS-34V2 operating manual

Union INCA3051 Translation of the original operating instructions

Union

Union INCA3051 Translation of the original operating instructions

SW Stahl PROFI Tools 72351L instruction manual

SW Stahl PROFI Tools

SW Stahl PROFI Tools 72351L instruction manual

Trotec BF06 operating manual

Trotec

Trotec BF06 operating manual

HELWIG CARBON BPK Probe operating guide

HELWIG CARBON

HELWIG CARBON BPK Probe operating guide

Chauvin Arnoux F 21 user manual

Chauvin Arnoux

Chauvin Arnoux F 21 user manual

BMT Blast Movement Monitor Activator System manual

BMT

BMT Blast Movement Monitor Activator System manual

PKP DP05 instruction manual

PKP

PKP DP05 instruction manual

Brooks Instrument GT1000 Series Installation and operation manual

Brooks Instrument

Brooks Instrument GT1000 Series Installation and operation manual

REED 8706 instruction manual

REED

REED 8706 instruction manual

Aquaread Aquaprobe AP-6000 instruction manual

Aquaread

Aquaread Aquaprobe AP-6000 instruction manual

Arkray AUTION MAX AX-4030 operating manual

Arkray

Arkray AUTION MAX AX-4030 operating manual

JAMAR Technologies DB-100 user manual

JAMAR Technologies

JAMAR Technologies DB-100 user manual

Clas Ohlson UT377A manual

Clas Ohlson

Clas Ohlson UT377A manual

Eyedro EH Series Product guide

Eyedro

Eyedro EH Series Product guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.