KATflow 200 Quick Start Manual
Flow in pipe
Step 1: Selection of measurement point & pipe preparation
●Avoid installation of sensors in the vicinity of defor ations and
pipe defects, near welding sea s or where deposits could have
accu ulated.
●Select a easuring point with sufficient straight pipe to obtain
accurate easure ents. Please consult the anual for the
reco ended distances fro sources of disturbance.
●For a horizontal pipe, ount the sensors on the side of the pipe.
For a vertical pipe, ount the sensors at a location where the liquid
flows upwards. (Pic 1)
●Mount the sensors in the direction of the flow. (Pic 2)
●Clean the pipe at the easure ent point. Re ove loose paint and
rust with a wire brush or file.
●Apply coupling paste to the face of the cla p-on sensors before
attaching the to the pipe.
Pic 1: Mounting points Pic 2: Flow direction
Step 2: Keyboard familiarisation
Select Setup izard to set up the
flow eter for easure ent. If
the sensors are recognised, the
serial nu ber will be shown. If
not, they can be selected.
Select the ain easure ent unit
using the cursor keys and confir
with ENTER. This unit will be
displayed in the iddle of the
easure ent screen. Selecting
OFF deactivates the ea-
sure ent channel.
Select the pipe aterial using the
cursor keys and confir with
ENTER.
Enter the outer pipe dia eter
using the keypad and confir
with ENTER. Use UP key as
backspace to correct for entry
errors. If “0” is entered and con-
fir ed, an additional screen
appears that allows entry of the
circu ference
.
Press
ENTER
to
1 / 2
Step 2: Keyboard familiarisation
●Show NEXT (1) available ite
●QON (2) = Start totalizer function
●Show next DISPlay (3)
●QOFF (8) = Stop totalizer function
●DIRECT (9) access to trend plot
●Move enu / selection ite UP
●Move enu / selection ite DO N
●ESCape entry without saving Switch
device OFF (press > 2 sec.)
●ENTER selection with saving
Switch device ON (press > 2 sec.)
circu ference
.
Press
ENTER
to
confir .
Enter pipe wall thickness using the
keypad and confir with ENTER.
Use UP key as a back-space to
correct for entry errors.
Select fluid using cursor keys.
Confir by pressing ENTER.
Enter the fluid te perature using
the keypad. Confir by pressing
ENTER.
Use UP key as a backspace to
correct for entry errors.
The flow eter can be prepared for easure ent with the Setup
izard as found in the Quick start enu.
Step 3: Quick Start menu and Setup izard
Upon first power on and the boot
up sequence, the Main Menu is
displayed.
Use the UP and DO N cursor keys
to select Quick start and confir
by pressing ENTER.Select pipe liner aterial using
cursor keys and confir by pres-
sing ENTER. If a liner aterial is
chosen, an additional screen
appears that allows entry of liner
thickness.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2014 I Issue: KF200QuickStartV40E0914
Step 3: Quick Start menu and Setup izard (continued)
Select nu ber of sound passes
(sound paths) using cursor keys.
Auto: Selection by flow eter
according to entered para eters
(nu ber of passes shown later on
Sensor Positioning Screen).
1: 1 pass, (diagonal ode)
2: 2 passes (reflection ode)
3: 3 passes (diagonal ode)
4: 4 passes (reflection ode)
5: 5 passes (diagonal ode)
..etc.
1. Reflection ode
The flow eter uses an even nu ber of passes. This is the ost
convenient ounting arrange ent, as the transducer separation
distance can be easured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3)
2. Diagonal ode
Signal travels on odd nu ber of passes through the pipe. Used for
larger pipes and for dirty/aerated liquids where greater signal
attenuation can occur. The sensor distance on this ounting con-
figuration
can
be
negative
(sensors
overlapping)
.
(
Pic
4
)
The ain easure ent unit is
displayed when first entering the
easure ent screen.
Press NEXT to see up to 3 units
displayed with the ain one in the
iddle.
KATflow 200 Quick Start Manual
With correct pipe para eters entered and the sensors ounted at
the suggested spacing, the ark should be near the central
indication line (see three lines below the botto signal bar). If the
ark is to the left-hand side of the central indication line, the sensors
are too close to each other. If the ark is to the right, the sensors are
too far apart. Slide one sensor carefully along the pipe to bring the
ark into a ore central position. Measure ents can be obtained
with the ark being between the left and right indication line. Press
ENTER to start easure ent.
Two further easure ent units can be assigned to that screen by
going
to
Main
Menu
-
Output
-
Display
.
Even nu ber of passes: Both sensors on sa e side of pipe. Odd
nu ber of passes: Sensors on opposite sides (see below for ore
infor ation). Confir with ENTER.
Measurement screen
Katronic Technologies Ltd . I Telephone: +44 (0)2476 714 111 I Email: info@katronic.co.uk I Web: www.katronic.co.uk
Select Start Measurement and
confir with ENTER to start the
sensor positioning procedure.
2 / 2
figuration
can
be
negative
(sensors
overlapping)
.
(
Pic
4
)
Mount the transducers with the suggested spacing between the
inside of the sensor heads. This distance has been deter ined by the
flow eter on the basis of the entered para eters. Use the displayed
nu ber of passes to install the sensors on the correct side of the pipe
(see odes above). Observe the upper bar (signal-to-noise ratio) and
lower bar (signal quality). These should be both filled to about the
sa e level with a filling level of around 1/3 or ore desired. Use the
oving ark between the two signal bars for the fine adjust ent of
the sensor position.
Internal data logger
The totalizer is shown when in easure ent ode after pressing
NEXT three ti es. It can also be assigned to the three line display,
datalogger or process outputs by selecting a volu etric unit. The
totalizer function is started with QON when in easure ent ode
( easure ent screen displayed). Pressing Q+resets the total in
positive flow direction. Pressing Q- resets the total in negative flow
direction. The totalizer function can be stopped with QOFF. Pressing
QON again will reset the positive, negative and overall totalizer.
Change displays without resetting the totalizer by pressing DISP or
NEXT.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2014 I Issue: KF200QuickStartV40E0914
Sensor Positioning Screen
Pic 3: Reflection mode Pic 4: Diagonal mode
going
to
Main
Menu
-
Output
-
Display
.
Totalizer
The data logger is reached via Main Menu - Output. It is activated in
Datalogger - Interval by entering and confir ing a non-zero value.
Enter “0” and confir to disable the logger. Up to 10 easure ent
units can be selected for logging under Datalogger - Selection. Use
the cursor keys to highlight a unit and press ENTER to select it. Press
“0” to deselect it. An activated data logger is indicated by a
“docu ent” sy bol in the top left corner of the display. On start of
easure ent ( easure ent screen displayed) the logger records
the selected easure ent units. A blinking “docu ent” sy bol
indicates a recording data logger. Separation arkers are set by the
data logger whenever a session begins. Leave the easure ent
screen by pressing ESC to stop recording. The recording interval can
be changed in Datalogger - Interval. The data logger is cleared under
Datalogger - Log Erase. Ensure all required data has been
downloaded .

Other manuals for KATflow 200

4

Other Katronic Measuring Instrument manuals

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 230 User manual

Katronic

Katronic KATflow 230 User manual

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 210 User manual

Katronic

Katronic KATflow 210 User manual

Katronic KATflow 150 User manual

Katronic

Katronic KATflow 150 User manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments HI 718 quick start guide

Hanna Instruments

Hanna Instruments HI 718 quick start guide

Sealey TA203.V2 instructions

Sealey

Sealey TA203.V2 instructions

Dwyer Instruments VP2 Specifications-installation and operating instructions

Dwyer Instruments

Dwyer Instruments VP2 Specifications-installation and operating instructions

MRM L Series user manual

MRM

MRM L Series user manual

GATRIK R12-800SC user manual

GATRIK

GATRIK R12-800SC user manual

SIRETTA SNYPER-LTEM quick start guide

SIRETTA

SIRETTA SNYPER-LTEM quick start guide

Lutron Electronics PM-9100HA Operation manual

Lutron Electronics

Lutron Electronics PM-9100HA Operation manual

ABB PositionMaster EDP300 Operating	 instruction

ABB

ABB PositionMaster EDP300 Operating instruction

TAC H2S/H2O/O2 instruction manual

TAC

TAC H2S/H2O/O2 instruction manual

Hoyt 2940 CL instruction manual

Hoyt

Hoyt 2940 CL instruction manual

Hauber Elektronik HE200 Series manual

Hauber Elektronik

Hauber Elektronik HE200 Series manual

EDM SP011AU quick guide

EDM

EDM SP011AU quick guide

Microsensor MPM380 Operation manual

Microsensor

Microsensor MPM380 Operation manual

Newport 1825-C Operator's manual

Newport

Newport 1825-C Operator's manual

VOLTCRAFT AN-10 operating instructions

VOLTCRAFT

VOLTCRAFT AN-10 operating instructions

Kett PM-650 operating manual

Kett

Kett PM-650 operating manual

Honeywell AD1200 product manual

Honeywell

Honeywell AD1200 product manual

Test Products International Vib Meter 9085 user guide

Test Products International

Test Products International Vib Meter 9085 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.