Step 3: Quick Start menu and Setup izard (continued)
Select nu ber of sound passes
(sound paths) using cursor keys.
Auto: Selection by flow eter
according to entered para eters
(nu ber of passes shown later on
Sensor Positioning Screen).
1: 1 pass, (diagonal ode)
2: 2 passes (reflection ode)
3: 3 passes (diagonal ode)
4: 4 passes (reflection ode)
5: 5 passes (diagonal ode)
..etc.
1. Reflection ode
The flow eter uses an even nu ber of passes. This is the ost
convenient ounting arrange ent, as the transducer separation
distance can be easured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3)
2. Diagonal ode
Signal travels on odd nu ber of passes through the pipe. Used for
larger pipes and for dirty/aerated liquids where greater signal
attenuation can occur. The sensor distance on this ounting con-
The ain easure ent unit is
displayed when first entering the
easure ent screen.
Press NEXT to see up to 3 units
displayed with the ain one in the
iddle.
KATflow 200 Quick Start Manual
With correct pipe para eters entered and the sensors ounted at
the suggested spacing, the ark should be near the central
indication line (see three lines below the botto signal bar). If the
ark is to the left-hand side of the central indication line, the sensors
are too close to each other. If the ark is to the right, the sensors are
too far apart. Slide one sensor carefully along the pipe to bring the
ark into a ore central position. Measure ents can be obtained
with the ark being between the left and right indication line. Press
ENTER to start easure ent.
Two further easure ent units can be assigned to that screen by
Even nu ber of passes: Both sensors on sa e side of pipe. Odd
nu ber of passes: Sensors on opposite sides (see below for ore
infor ation). Confir with ENTER.
Measurement screen
Katronic Technologies Ltd . I Telephone: +44 (0)2476 714 111 I Email: info@katronic.co.uk I Web: www.katronic.co.uk
Select Start Measurement and
confir with ENTER to start the
sensor positioning procedure.
2 / 2
Mount the transducers with the suggested spacing between the
inside of the sensor heads. This distance has been deter ined by the
flow eter on the basis of the entered para eters. Use the displayed
nu ber of passes to install the sensors on the correct side of the pipe
(see odes above). Observe the upper bar (signal-to-noise ratio) and
lower bar (signal quality). These should be both filled to about the
sa e level with a filling level of around 1/3 or ore desired. Use the
oving ark between the two signal bars for the fine adjust ent of
the sensor position.
Internal data logger
The totalizer is shown when in easure ent ode after pressing
NEXT three ti es. It can also be assigned to the three line display,
datalogger or process outputs by selecting a volu etric unit. The
totalizer function is started with QON when in easure ent ode
( easure ent screen displayed). Pressing Q+resets the total in
positive flow direction. Pressing Q- resets the total in negative flow
direction. The totalizer function can be stopped with QOFF. Pressing
QON again will reset the positive, negative and overall totalizer.
Change displays without resetting the totalizer by pressing DISP or
NEXT.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2014 I Issue: KF200QuickStartV40E0914
Sensor Positioning Screen
Pic 3: Reflection mode Pic 4: Diagonal mode
Totalizer
The data logger is reached via Main Menu - Output. It is activated in
Datalogger - Interval by entering and confir ing a non-zero value.
Enter “0” and confir to disable the logger. Up to 10 easure ent
units can be selected for logging under Datalogger - Selection. Use
the cursor keys to highlight a unit and press ENTER to select it. Press
“0” to deselect it. An activated data logger is indicated by a
“docu ent” sy bol in the top left corner of the display. On start of
easure ent ( easure ent screen displayed) the logger records
the selected easure ent units. A blinking “docu ent” sy bol
indicates a recording data logger. Separation arkers are set by the
data logger whenever a session begins. Leave the easure ent
screen by pressing ESC to stop recording. The recording interval can
be changed in Datalogger - Interval. The data logger is cleared under
Datalogger - Log Erase. Ensure all required data has been
downloaded .