KATf ow 230 Quick Start Manua www.katronic.co.uk
• Avoid insta ation of sensors in the vicinity of deformations and pipe
defects, near we ding seams or where deposits cou d have accumu-
ated.
• Se ect a measuring point with sufficient straight pipe to obtain ac-
curate measurements. P ease consu t the manua for the recom-
mended distances from sources of disturbance.
• For a horizonta pipe, mount the sensors on the side of the pipe. For
a vertica pipe, mount the sensors at a ocation where the iquid
f ows upwards (Pic. 1).
• Mount the sensors in the direction of the f ow (Pic. 2).
• C ean the pipe at the measurement point. Remove oose paint and
rust with a wire brush or fi e.
• App y coup ing paste to the face of the c amp-on sensors before
attaching them to the pipe.
Se ect
Setup Wizard CH1
to set up the
measurement channe 1, or
CH2
for
channe 2. Confirm se ection by pres-
sing
ENTER
. If the sensors are recog-
nised, the seria number wi be shown.
If not, the type can be se ected.
1. Mounting points 2. F ow direction
Senso Mounting Configu ation
Ref ection Mode
The f owmeter uses an even number
of passes. This is the most conveni-
ent mounting arrangement, as the
transducer separation distance can
be measured very easi y and the
sensors can be accurate y a igned.
Use whenever possib e (Pic. 3).
3. Reflection mode (f om above)
Diagona Mode
Signa trave s on an odd number of
passes through the pipe. A sing e
pass can be used for arger pipes
and for dirty/aerated iquids where
greater signa attenuation can occur.
The sensor distance on this moun-
ting configuration can be negative
(sensors over apping) (Pic. 4).
Step 3: Quick Sta t Menu and Setup Wiza d
At first power on and the boot se-
quence, the
Main Menu
is disp ayed.
Use the
UP
and
DOWN
cursor keys to
se ect
Quick Start
and confirm by
pressing
ENTER
.
Enter the f uid temperature using the
keypad. Confirm by pressing
ENTER
.
Use
UP
key as a backspace to correct
for entry errors.
Se ect f uid using cursor keys. Con-
firm by pressing
ENTER
.
Enter pipe wa thickness using the
keypad and confirm with
ENTER
. Use
UP
key as a backspace to correct for
entry errors.
Enter the outer pipe diameter us-
ing the keypad and confirm with
ENTER
. Use
UP
key as backspace to
correct for entry errors. If "0" is entered
and confirmed, an additiona screen
appears that a ows entry of the cir-
cumference. Press
ENTER
to confirm.
Se ect the pipe materia using the cur-
sor keys and confirm with
ENTER
.
Se ect the main measurement unit
using the cursor keys and confirm
with
ENTER
. This unit wi be dis-
p ayed in the midd e of the measure-
ment screen. Se ecting
OFF
deactivates
the measurement channe .
• Show
NEXT
(1) avai ab e item
•
QON
(2) = Start tota iser function
• Show next
DISP
ay (3)
•
QOFF
(8) = Stop tota iser function
•
DIRECT
(9) access to trend p ot
• Move menu/se ection item
UP
• Move menu/se ection item
DOWN
•
ESC
ape entry without saving
Switch device
OFF
(press > 2 sec.)
•
ENTER
se ection with saving
Switch device
ON
(press > 2 sec.)
Step 1: Measu ement Point and Pipe P epa ation
4. Diagonal mode (f om above)
• The f owmeter can be prepared for measurement with the
Setup
Wizard
as found in the
Quick Start
menu.
Step 2: Keyboa d Familia isation
Se ect pipe iner materia usingcursor
keys and confirm by pressing
ENTER
.
If a iner materia is chosen, an additi-
ona screen appears that a ows entry
of iner thickness.
KATf ow 230
Quick Sta t Manual
Measu ement Sc een
The main measurement unit is dis-
p ayed when first entering the mea-
surement screen. Press
MUX
to switch
between channe s 1 and 2. Press
NEXT
to cyc e through the three ine
disp ay, tota iser and dua measure-
ment screens.
Totalise
The tota iser is shown when in meas-
urement mode after pressing
NEXT
twice. It can a so be assigned to the
three ine disp ay, data ogger or pro-
cess outputs by se ecting a quantity
as the unit.
Inte nal Datalogge
• The data ogger is reached via
Main Menu
-
Output
. It is activated
in
Data ogger
-
Interva
by entering and confirming a non-zero
va ue and se ecting units to be recorded.
• Enter "0" and confirm to disab e the ogger. Up to ten measure-
ment units can be se ected for ogging under
Data ogger
-
Se ec-
tion
.
• Use the cursor keys to high ight a unit and press
ENTER
to se ect
it. Press "0" to dese ect it.
• An activated data ogger is indicated by a “document” symbo in
the top eft corner of the disp ay. On start of measurement (mea-
surement screen disp ayed) the ogger records the se ected mea-
surement units.
• A b inking “document” symbo indicates a recording data ogger.
Separation markers are set by the data ogger whenever a session
begins.
• Leave the measurement screen by pressing
ESC
to stop recording.
• The recording interva can be changed in
Data ogger
-
Interva
.
The data ogger is c eared using
Data ogger
-
Log Erase
. Ensure
a required data has been down oaded.
Wall thickness gauge (WTG) (optional)
Optiona sensor probes to measure
pipe wa thickness are avai ab e. En-
sure the correct pipe materia is se ec-
ted. Connect to the sensor input
(Channe 1) and se ect
Start Measure-
ment
.
• The tota iser function is started with
QON
when in measurement
mode (measurement screen disp ayed). Pressing
Q+
resets the
tota in positive f ow direction. Pressing
Q-
resets the tota in neg-
ative f ow direction. The tota iser function can be stopped
with
QOFF
.
• Pressing
QON
again wi reset the positive, negative and overa
tota iser. Change disp ays without resetting the tota iser by
pressing
DISP
or
NEXT
.
1: 1 pass (diagona mode)
2: 2 passes (ref ection mode)
3: 3 passes (diagona mode)
4: 4 passes (ref ection mode) etc.
Se ect
Start Measurement
and con-
firm with
ENTER
to start the sensor
positioning procedure. When both
channe s have been activated, the
procedure for
CH2
fo ows the one of
CH1
.
Senso Positioning Sc een
• Mount the transducers with the suggested spacing between the
sensor heads. This distance has been determined by the f ow-
meter on the basis of the entered parameters.
• Use the disp ayed number of passes to insta the sensors on the cor-
rect side of the pipe (see pictures 3 and 4).
• Observe the upper bar (signa -to-noise ratio) and ower bar (signa
qua ity). These shou d be both fi ed to about the same eve with a
fi ing eve of around 1/3 or more desired.
• Even number of passes: Both sensors on same side of pipe (see pic. 3).
• Odd number of passes: Sensors on opposite sides (see pic. 4). Con-
firm with
ENTER
. The second channe can now be prepared by
se ecting
Setup Wizard CH2
.
Step 3: Quick Sta t Menu and Setup Wiza d (continued)
• Use the cursor between the two signa bars for the fine adjustment
of the sensor position. With correct pipe parameters entered and
the sensors mounted at the suggested spacing, the mark shou d be
near the centra indication ine (see three ines be ow the bottom
signa bar).
• If the mark is to the eft-hand side of the centra indication ine, the
sensors are too c ose to each other. If the mark is to the right, the sen-
sors are too far apart. S ide one sensor carefu y a ong the pipe to
bring the mark into a more centra position. Measurements can be
obtained when the mark is between the eft and right indication
ine. Press
ENTER
to start measurement.
5. Sensor positioning screen
2/3
• Se ect number of sound passes (sound paths) using cursor keys.
•
Auto
: Se ection by f owmeter according to entered parameters (num-
ber of passes shown ater on
Sensor Positioning Screen
).
• The f owmeter wi recognise the probe and disp ay the
measurement screen. Wa thickness wi be shown when the
sensor is in good acoustic contact with the pipe.
• Two further measurement units can be assigned to the three ine
screen by going to
Main Menu
-
Output
-
Disp ay
-
Channe 1/
Channe 2
.
www.katronic.co.uk
KATf ow 230 Quick Start Manua
KATf ow 230
Quick Sta t Manual
• For separate measurement of f ow and temperature se ect a
f ow unit as
Midd e Unit
in the
Setup Wizard
. Comp ete the
wizard and then go to
Main Menu
-
In/Output
.
• In the
In
/
Output
menu assign the temperature input to channe 1,
and se ect
In et
or
Out et
from the menu.
• In the
Disp ay
menu set the
Top Line
or
Bottom Line
to either
Tin
or
Tout
depending on the previous se ection.
• In the
Data ogger
menu se ect either
Tin
or
Tout
from the ist of
variab es depending on the previous se ection.
• When measuring, the f owmeter wi be ogging the PT100 input
and can be viewed by pressing
1
/
NEXT
on the keypad.
• This data can a so be assigned to a process output in the
In/
Output
menu fo owing the same process.
P ocess Outputs: Setup
• Configured process outputs can be set up in
Main Menu - In/Output
:
-
RELAY OUT
(re ay output)
-
I OUT ACTIVE
(0/4 … 20 mA)
-
PULSE OUT
(open-co ector)
can be se ected.
• After se ecting an output, it can be assigned to a measurement
channe . Se ecting
OFF
disab es the output.
• On the next screen, the unit that wi appear on the output can
be se ected.
• The type of output wi determine the remaining screens.
P ocess Outputs: Connection
• Process outputs can be connected to third party devices using
the dedicated input/output box. This connects to the KATf ow
230 via a circu ar p ug which is found on the bottom of the f ow-
meter (Pic. 6).
• The i ustration be ow gives an overview of the output termina
a ocation.
PT100 1
RS 232
Expansion
box
Input/output
box
PT100 2
Heat Quantity Measu ement (whe e specified)
• The KATf ow 230 is capab e of measuring heat f ow and heat quantity.
• To measure heat f ow, se ect a heat f ow unit (W, kW, MW) as
Midd e
Unit
when going through the
Setup Wizard
of the se ected measure-
ment channe .
• To measure heat quantity, se ect a heat quantity unit (J, kJ, MJ).
• On se ection of one of these units, the f owmeter requests the entry
of the specific heat capacity of the medium in [J/(g.K)]. Enter the
va ue of the f uid and confirm with
ENTER
. Comp ete the remaining
steps ofthe
Setup Wizard
.
• The temperatures at the in et and out et of the monitored system are
used to determine heat f ow and heat quantity.
• Connect the supp ied PT100 temperature sensors to the bottom of the
f owmeter by p ugging them into the eft and right sockets (Pic. 6).
6. Bottom connectors
• After setting up the f ow measurement channe go to
Main Menu
-
In/Output
, in order to assign the PT100 inputs to the channe .
• Use the cursor keys to se ect
PT100 4 WIRE
and confirm with
ENTER
.
• On the next screen, use the cursor keys to se ect the f ow channe to
which the temperature input is app ied. Choose
Channe 1
or
2
.
Se ecting
OFF
wi disab e the input. Confirm with
ENTER
.
• On the fo owing screen choose
PT100
to use the temperature meas-
ured on the pipe. To enter a fixed temperature va ue, se ect
User
and
enter the va ue.
• Define whether the PT100 sensor measures the in et or out et tempera-
ture. Use the cursor keys to se ect as appropriate and confirm with
ENTER
.
• On the next screen a temperature offset can be app ied. After confirm-
ing again with
ENTER
the meter returns to the
Main Menu
.
• Once the setup of the first temperature input has been comp eted, re-
peat these steps for the configuration of the second input.
Tempe atu e and Flow Measu ement
3/3
© Copyright Katronic Techno ogies Ltd. 2015 | Issue: QS_KF230_V52EN_1512 | A rights reserved.
7. Input/output box, connection termina s
Medium at 25 °C
(un ess stated different y)
Specific Heat Capacity
[J/(g.K]
Ethano (0 °C) 2.30
Ethy ene g yco (100 %) 2.36
25 % Ethy ene g yco /75 % water (26.7 °C) 3.85
30 % Ethy ene g yco /70 % water (26.7 °C)
3.77
50 % Ethy ene g yco /50 % water (26.7 °C)
3.41
65 % Ethy ene g yco /35 % water (26.7 °C) 3.11
R22 refrigerant* (30 °C) 6.60
R134a refrigerant* 8.87
Minera oi 1.67
Water 4.18
KATf ow 230
Quick Sta t Manual
OUTPUT
TYPE TERMINAL
BLOCKS CONNECTION
* Liquid under pressure

Other manuals for KATflow 230

2

Other Katronic Measuring Instrument manuals

Katronic KATflow 210 User manual

Katronic

Katronic KATflow 210 User manual

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 150 User manual

Katronic

Katronic KATflow 150 User manual

Katronic KATflow 170 Quick guide

Katronic

Katronic KATflow 170 Quick guide

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 230 Instruction manual

Katronic

Katronic KATflow 230 Instruction manual

Katronic KATflow 200 User manual

Katronic

Katronic KATflow 200 User manual

Katronic KATflow 230 User manual

Katronic

Katronic KATflow 230 User manual

Katronic KATflow 150 Instruction manual

Katronic

Katronic KATflow 150 Instruction manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Katronic KATflow 200 Instruction manual

Katronic

Katronic KATflow 200 Instruction manual

Popular Measuring Instrument manuals by other brands

OHAUS STARTER 5000 instruction manual

OHAUS

OHAUS STARTER 5000 instruction manual

Zeiss MCS 600 user manual

Zeiss

Zeiss MCS 600 user manual

LOVATO ELECTRIC DME D111 installation manual

LOVATO ELECTRIC

LOVATO ELECTRIC DME D111 installation manual

DynAmp LKAT2 Installation, operation and service manual

DynAmp

DynAmp LKAT2 Installation, operation and service manual

Agilent Technologies N9039A Service guide

Agilent Technologies

Agilent Technologies N9039A Service guide

CARLO GAVAZZI EM21 72D manual

CARLO GAVAZZI

CARLO GAVAZZI EM21 72D manual

CST/BERGER Lasermark LD500 quick guide

CST/BERGER

CST/BERGER Lasermark LD500 quick guide

Powerfix Profi HG00384C Operation and safety notes

Powerfix Profi

Powerfix Profi HG00384C Operation and safety notes

Badger Meter Hedland T303 user manual

Badger Meter

Badger Meter Hedland T303 user manual

RS PRO RS-92 instruction manual

RS PRO

RS PRO RS-92 instruction manual

Thyracont VD8 Series operating instructions

Thyracont

Thyracont VD8 Series operating instructions

Keysight Technologies 34420A Service guide

Keysight Technologies

Keysight Technologies 34420A Service guide

RKI Instruments OX-500 Manual supplement

RKI Instruments

RKI Instruments OX-500 Manual supplement

Emerson Rosemount 8800D Series quick start guide

Emerson

Emerson Rosemount 8800D Series quick start guide

HEDÜ TALMETER TM 2 m manual

HEDÜ

HEDÜ TALMETER TM 2 m manual

Vernier HD-BTA owner's manual

Vernier

Vernier HD-BTA owner's manual

Mastech MS8332C user manual

Mastech

Mastech MS8332C user manual

Kobold DAG-A44 Series operating instructions

Kobold

Kobold DAG-A44 Series operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.