OPERATING INSTRUCTIONS
EWL 5837 5
A 6 Operating the unit
All recommendations for the applications of the unit, hether oral, ritten or in the form of practical instructions, are based on our
o n experience and tests and can, therefore, only be considered as guidelines.
A 6.1 Operation: Vacuum-mixing
- Set a mixing time of up to 240 seconds at the time control - stirrer s itches automatically to standby – and the time is stored until
adjusted.
- For manual s itch-on of the stirrer, turn the time control to “0“ and set the mixing time after the vacuum has built up.
- Fill the mixing bo l ith the material.
- Add the liquid and mix by hand until the ingredients are ell integrated.
- Close the bo l lid.
- Fix the bo l to the unit from belo , thus automatically releasing the vacuum generation. Hold the bo l firmly until the vacuum has built
up.
- Stirrer automatically operates for the pre-set time, and at the end an acoustic signal sounds.
- Hold the bo l firmly before turning off the main s itch (Vacuum).
- After the vacuum has diminished, the bo l is no longer secured to the unit.
• To ensure trouble-free operation, the sealing surfaces of the bo l and lid - including the top of the lid and the tang of the rotor- must be
free from contamination.
• Fill the bo l only to the maximum marking, other ise the vacuum ill dra excessive material into the system.
• After the vacuum has diminished, the bo l is automatically released. In order to prevent damage to the seal, never pull at the bo l.
A 6.2 Operation: Pressure-investing
- Set a time of up to 60 minutes at the time control.
- Open the black door – turn the door against the direction of the arro – do not pull!
- Insert the parts – maximum 3 large casting rings.
- Close the pressure chamber and actuate the green s itch (Investing).
- Pressure ill build up according to the setting of the manometer display.
- At the end of the pre-set time the pressure chamber is automatically ventilated, and the parts can be removed. By 7.7.7. – turning off the
Investing s itch – the pressure-investing process can be interrupted at any time, i.e. the pressure chamber ill be ventilated.
• Should the pressure in the compressed air supply line reduce during the processing time, a non-return valve secures the pressure already
built-up.
• Please note that the pressure chamber can only be opened first hen the pressure has completely diminished. Other ise, a premature
opening can cause the rubber ring to be ripped out of its groove, and become damaged.
A 6.3 Operation: Vibrator
- Press main s itch (Vibrator)
- Vibrator arm and rubber ball ill vibrate.
- Hold impression tray (or small impression) against the ball, and fill up ith mixed material (plaster, etc.) in small portions. Thereby,
even small jobs can be accomplished using the unit.