DIGITAL CALIBRATION
~ha diiitat calibrationfeature eliminatescalibrationpotentiometers. All
calibrationfactorsfor each rangeare storedinnon-volatileRAM;The rn-
strument can be calibratedfrom the front panel or with the optional
4353 IEEEinterface. With the Model 4653 IEEE-488 interface installed,
the Model 465 can be calibrated much faster using an automated
system. Front panel digitalcalibrationisoutlinedas follows:
NOTE
The following is an actual calibrationprocedure for
the 2Ofi range. If recalibrationisnot necessary,then
do not movethe calibrationjumper. Leavethe jumper
inthe disableposition.Referto the MaintenanceSec-
tion of the 465 InstructionManual for jumper place-
ment.
1. Move the calibrationjumper to the enable position.
2. Turn on the Model 465.
3. Press the REL and LOG buttons simultaneouslyuntil the message
“CAL” is displayed. Releasethe buttons. The Model 465 is now in
the calibrationmode as indicatedby the “CAL” annunciator.
4. Apply the proper signal or level to the Model 465 input. Example:
place the Model 465 to the 2Ofi range and apply 19.066~.
NOTE
Consultthe Model465 InstructionManualfor detailed
methods of creatingthe proper calibrationlevels.
5. Adjustthe displayaccordinglywith the usaof the STOKLR and RCL
buttons. The STOKLR button increments the displayed reading.
The RCL button decrements the displayedreading.
6. Press the LOG and REL button simultaneouslyuntil the massage
“STOR” isdisplayed.The “new” calibrationpointforthe 2&A range
is now permanently stored. This new constant remains as is until
changad.
7. To return to normaloperation, turn off the instrumentmomentarily,
then back on.
8. Turn the instrumentoff and placethe calibrationjumper inthe disable
positionto prevent inadvertent calibrationof the instrument.
Consultthe Model465 InstructionManual for completedetailsconcern-
ing Model465 calibration.The precedingprocedureispresentedhereto
remind the user that front panel digital calibrationis availableon the
Model 465.
IEEE-468 option: Allows the user to calibratethe instrumentover the
IEEE-468 bus. With the IEEE48 interface installed(option4653). the
Model 465 can be calibratedon most automated calibrationsystems.