I “PERKrINGINSTRUCTIONS
---.-____I
HODELS 164, 164~
2-5. MODESOF OPERATION. The Model 164 permits meas-
urements in either of two modes of operation, 1, Multf-
meter ?lode (DMM) or 2) Ohmmeter-Milliobmmeter Mode (n,
mn).
a. Elulrimeter Mode (DMM .1 I" the DMMmode, rbe
Model lb4 permifs measurement of uolrage, current,
and resiseance (LO 0.U per digit) for conventional
cAeas*relne*ts. To select the DMNmode, set the Mode
Switch to the DMMposition. men set rhe Range
Switch to the appropriate function and range.
b. Ohmmeter-Milliohmmeter Mode (0, mi2). I" either
thz1 or mil modes, the Model 164 permits mea~"=ement
Of resistance using a four-terminal voltage-current
method. To select the n or mii modes, set the Mode
Smirch to the desired "ohms" position. Separate
rerminals ace provided for cu==ent and voltage se"-
sing. rhe Model 1641 Kelvin Test Lead Set ifurnish-
ed with the Model 164) simplifies connections to the
resiSta"ce to be measured. The terminals identified
as "m;2 SOURCE" provide a c"==ent in five decades
from 10 microamperes to 100 milliamperes, co==eg-
pending to positions on the Made Switch identi-
fied as IOK?, IOn, I!,, 100m0, o= 10mrt. The ter-
minals normally used for "MB measurements are for
voltage sensing when making four terminal resis-
tance meusurements. The Range Switch Of the Model
164 nl"Sf he set to rile lrn" function when using the
oirmmeter-milliohmmeter modes.
2-6. VOLTMETEROPERATION @MMNode). This mode of b. Co""ecfio*s.
operation enables voltage meaS"=eme"eS from il micro- Connect the input twminals so as
to place the Multimeter in series with the current to
volt per digit (1 millivolt full range) to tlOO0 volts. be measured.
Range Selection. To select "oltmeLe= operation
se:'the !Iode Switch t" DMM, then set the Range Switch
to the appropriate full range voltage. as a "olr-
meter, the Model 164 provides seven ranges as given
in Table 2-l.
C. Measurement Accuracy. The current accuracy is (?
specified as '0.2% of reading, fl digit on all ranges:
Since the accuracy is specified ac the input terminals,
the loading effects should also be considered. See
Table 2-2 for values of shu"t resisfors.
TABLE z-1.
"olmeter operation d. Off-Ground Measurements. The Model 164 low
terminal can be floated t500 volts above CASE ground
for current measu=eme"ts. Refer to paragraph 2-11
for complete instructions.
I El
Range setcin Full P.a"Se Display Overrange Display
I
1 In” 1.000 In” 1.999 In"
10 In" 10.00 rn" 19.99 rn"
100 II" 100.0 mv 199.9 in"
1 " 1.000 " 1.999 "
10 v 10.00 v 19.99 "
100 " 100.0 v 199.9 "
1000 " 1000. " 1000. v*
*Maximum allowable input is 1000 volts continuous in-
p"t on 1-"OlC to 1000~volt ranges. 1000 volts momen-
tary, 300 volts co"rin"ous on lower ranges.
b. Measurement Accuracy. The voltage a~curac,. is
t(O.l% of reading + 1 digit) on all ranges.
C. Overload Recovery. The instrument will recover
from 300-volt overloads within five seconds on the
lm" range. "p CO 1000 volts peak may be applied
momentarily on any range without damaging the i"st=u-
ment .
d. Innut resistance. The input haa a reeist="Ce
"f 10 m=whms 0" the 10 millivolt and higher =a"~*,
1 megohm on the l-millivolt range. FO= lomv and
higher =a"Ses, a 1000 ohm source eestatance will
introduce only 0.01% e==o=. 'ro maintain rated acw-i
racy o" the 1mV =a"~. the source reaiseance should
not exceed 100 ohms.
e. Off-Ground Measurements. The instrument low
terminal can be floated *500 volts above CASE ground
for voltage ,~~~%S"re?me"t~. Refer to paragraph 2-11
for complete instructions.
f. AC Voltage Adapter Probe. The Model 1601 AC/DC
Probe permits convenient a-c voltage meaguremenfs from
0.1 volt to 250 volts rms o"e= a span of 4s HZ to 45
kkh. A probe-mounted switch provides convenient se-
lection of,a-c o= d-c operation without disconnecring
the probe. Therefore, the probe may be pe=manently
attached.
2-7. aMMETEROPERATION (DMMMode). This mode of
operation enables current meaS"=eme*ts from 0.1 "ano-
ampere per digit (100 nanoamperes full range) to 2
amperes with 100% overranging on all ranges.
a. Range Selection. To select ammeter operarion,
set the Mode Switch to DMM, the" set the RanSe Switch
to the appropriate full range cu==e"t. As an ammeter,
the Model 164 provides eight ranges as given in Table
2-2.
2-S. OWMMETEROPERATION @MMMode). TXhis mode of
operaeion enables resistance meaS"=eme"eS from 0.1
ohm per digit (100 ohms full range) to 2000 megohms.
2,. Range Selection. TO select ollmmeter operation
@MMmode), set the Mode Switch to DMM, then set the
Range Switch to the appropriate full range resistance.
As an ohmmeter, the Model 164 provides eighr ranges
as given in Table 2-3.
b. Voltage Across the Unknown. "olfage drop is
100 millivolts at full range with 1.5 V max. into
a" open circuit. Thus the current through the "n-
known is determined by dividing the 100 millivolts
by the full range resistance.
c. Maximum Overload. Maximum voltage ac=oss the
input tennina1s is 20 voles momentary and 1 Volt
co"ti""o"s.
d. Accuracy. As shown in Table 2-4.
6 0474