kENDALL SCD EXPRESS Operation manual

kENDALL
SCD EXPRESSTM
Compression System
tyco
SYSTÈME DE COMPRESSION
Manuel D’utilisation Et D’entretien
KOMPRESSIONSSYSTEM
Betriebs- Und Serviceanleitung
SISTEMA DI COMPRESSIONE
Manuale Operativo E Di Manutenzione
SISTEMA DE COMPRESIÓN
Manual De Operación Y Servicio
KOMPRESSIONSSYSTEM
Användar- Och Servicemanual
COMPRESSIESYSTEEM
Handleiding Voor Bediening En Onderhoud
SISTEMA DE COMPRESSÃO
Manual De Funcionamento E Assistência
KOMPRESSIOJÄRJESTELMÄ
Käyttö- Ja Huolto-opas
KOMPRESSIONSSYSTEM
Betjenings- Og Servicehåndbog
Operation and Service Manual
ΣΎΣΤΗΜΑ ΣΥΜΠΊΕΣΗΣ
Εγχειρίδιο Λειτουργίας Και Σέρβις
TLAKOVÝ SYSTÉM
Provozní A Servisní Příručka
KOMPRESSZIÓS RENDSZER
Kezelési És Karbantartási Kézikönyv
СИСТЕМА ПНЕВМАТИЧЕСКОЙ
КОМПРЕССИИ
Руководство По Эксплуатации И
Обслуживанию
SYSTEM UCISKOWY
Instrukcja Obsługi I Serwisowania
KOMPRESYON SISTEMI
Operasyon Ve Servis Kılavuzu

TABLE OF CONTENTS
Indications
..........................................................................................................................
1
Contraindications
................................................................................................................
1
Cautions
..............................................................................................................................
1
Explanation
of Symbols Used
................................................................................................
2
Front Panel Display
..............................................................................................................
2
Section I -
General Op
erating
Instructions
.............................................................................
3
Se
t up
.............................................................................................................................................................................
3
Start-up
.........................................................................................................................................................................
3
Garment Configuration & Detection
.............................................................................................................................
3
Normal Operation
..........................................................................................................................................................
4
Pressure Settings
...........................................................................................................................................................
4
Vascular Refill Detection
...............................................................................................................................................
4
Garment
Compatibility
..................................................................................................................................................
4
Tubing Set Com
patibility
...............................................................................................................................................
4
Portable Controller Configuration
.................................................................................................................................
4
Section II -
Battery Operation
...............................................................................................
4
Battery Indicators Status
..............................................................................................................................................
5
Charging the Battery
.....................................................................................................................................................
5
Battery Warnings
...........................................................................................................................................................
6
Section III -
Fault Conditions and Troubleshooting
.................................................................
6
Watchdog Circuit
...........................................................................................................................................................
6
Section IV - Service and Maintenance
....................................................................................
8
Introduction
...................................................................................................................................................................
8
Warranty and Factory Service
.......................................................................................................................................
8
Fan Screen and Ventilation
............................................................................................................................................
9
Fuses
..............................................................................................................................................................................
9
Electrical Safety
.............................................................................................................................................................
9
Suggested Preventative Maintenance Schedule
..........................................................................................................
9
Cleaning
......................................................................................................................................................................
1
0
Electrical/Electronics Description
...............................................................................................................................
1
0
Pneumatic Operation Description
..............................................................................................................................
1
0
Section V -
Test Methods and Cal
ibration
.............................................................................
1
0
Test Mode 01 - Burn-In
..............................................................................................................................................
1
0
Test Mode Look up Char
t
...........................................................................................................................................
1
1
Test Mode 02 - General Function Test
........................................................................................................................
1
1
Test Mode 03 - Pressure Transducer Calibration
........................................................................................................
1
1
Test Mode 04 - Pressure Transducer Calibration Verification
.....................................................................................
1
1

TABLE OF CONTENTS
Test Mode 05 -
Self Test
.............................................................................................................................................
1
2
Test Mode 06 - Leak Test
............................................................................................................................................
1
2
Test Mode 07 - Performance Test
...............................................................................................................................
1
2
Test Mode 08
-
Manufacturing Test
...........................................................................................................................
1
2
Section VI - General Disassembly / Reassembly
....................................................................
1
3
Battery Pack (Removal / Installation)
........................................................................................................................
1
3
Compressor (Removal / Installation)
.........................................................................................................................
1
3
Valve Manifold and Bed Hook (Removal / Installation)
............................................................................................
1
3
Muffler (Removal / Installation)
................................................................................................................................
1
3
Power Supply Board (Removal / Installation)
...........................................................................................................
1
4
Fan and Screen (Removal / Installation
)
...................................................................................................................
1
4
Main Controller Board (Removal/Installation)
.........................................................................................................
1
4
Section VII -
Specifications
..................................................................................................
1
5
Section VIII -
Schematics
.....................................................................................................
1
6
Figure 1 -
P
arts Assembly Diagram Exp
loded View
..................................................................................................
1
6
C
ontroller Parts List
....................................................................................................................................................
1
6
Figure 2 -
Pneumatic & Electric
al Schematic
............................................................................................................
1
8
Section IX -
Pump Diagrams
................................................................................................
2
0
D
iagram 0
1
.................................................................................................................................................................
2
0
D
iagram 0
3
.................................................................................................................................................................
2
0
D
iagram 02
.................................................................................................................................................................
2
1
D
iagram 0
4
.................................................................................................................................................................
2
1

EN-1
Indications
The SCD EXPRESS
TM
Compression System is designed to apply intermittent pneumatic compression to increase venous blood flow in
at-risk patients in order to help prevent deep vein thrombosis and pulmonary embolism. The System consists of the Controller, the Tubing Sets (provided
with the Controller) and single-patient use garments (purchased separately from this Controller). The garments, both leg sleeves and foot cuffs,
compress the limbs to enhance venous blood movement. After the compression, the Controller measures the time it takes for the limbs to refill with
blood and waits that period of time before the next compression is initiated.
Leg Compression
The use of the SCD EXPRESS Compression System with Leg Sleeves is indicated for:
1. Deep vein thrombosis and pulmonary embolism prophylaxis.
Foot
Compression
The use of the SCD EXPRESS Compression System with Foot Cuffs is indicated for:
1. Circulation enhancement. 5.
E
xtremity pain incident to trauma or surgery.
2.
Deep vein thrombosis prophylaxis.
6.
L
eg Ulcers.
3. Edema - Acute. 7.
V
enous stasis / venous insufficiency.
4. Edema - Chronic.
If you need further
information regarding the SC
D EXPRESS Compression System or its clinical benefits, please contact your
Contraindications
Leg Compression
The SCD EXPRESS Compression System
may not
be recommended for use with Leg Sleeve on patients with the following:
1. Any local leg condition in which the sleeves may interfere, such as: (a) dermatitis, (b) vein ligation [immediate postoperative], (c) gangrene, or
(d)
recent skin graft.
2. Severe arteriosclerosis or other ischemic vascular disease.
3. Massive edema of the legs or pulmonary edema from congestive heart failure.
4. Extreme deformity of the leg.
5. Suspected pre-existing deep venous thrombosis.
Foot Compression
The SCD EXPRESS Compression System
may not
be recommended for use with Foot Cuffs on patients with the following:
1. Conditions where an increase of fluid to the heart may be detrimental.
2. Congestive heart failure.
3. Pre-existing deep vein thrombosis, thrombophlebitis or pulmonary embolism.
Use with caution on the infected or insensitive extremity.
Cautions
1. Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
2. Patients with diabetes or vascular disease require frequent skin assessment.
3. Explosion hazard. Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
WARNING:
Do not attempt to repair or replace broken tubing connect
ors as hazardou
s
inflation of
the sleeves may occur
.
K
endall’s SCD EXPRESS Compression System is protected by one or more of the following U.S. patents: 5,022,387; 5,031,604;
5,478,119; 5,876,359;
Des.363,988; Des.364,460; Des.369,664; Des.373,192. Other patents pendi
ng.
Tyco Healthcare
expressly states that no implied license with respect to
use of non-Tyco Healthcare components is grante
d by the sale or lease of this system.
The use of
an intermittent pneumatic compression controller with a compression garment to measure venous refill time in a limb, without permission
from Tyco H
ealthcare Group LP,
may constitute infring
ement
of Tyco Healthcare’s U.S. Patent No. 6,231,532.

EN-2
REF
SN
CUS
L
U
C
L
A
S
S
I
F
I
E
D
®
Explanation of Symbols Used
Refer to Accompanying
Documents
Type BF Protection Against
Electronic Shock
Manufacturing Date Code
Underwriters Laboratories
(UL) Classifi cation mark for
Canada and the United States
Controller Serial Number
Reorder number for the device
located on the carton label
Vascular Refi ll Detection
occurring while illuminated
CE Mark
0123
Front Panel Display
1
4
8
7
6
3
10
11
12
Item Explanation Item Explanation
5
9
13
14
2
1 AC Power Indicator
2 Battery Charging Indicator
3 Battery Status Indicators 1-3
4 Test Mode/Error Code Indicators 1-8
5 Power On Indicator
6 Power On/Standby Button
7 Port B Foot Indicator
8 Port B Garment Confi guration Button
9 Port B Leg Indicator
10 Service Required Error Indicator
11 Vascular Refi ll Assess Indicator
12 Port A Leg Indicator
13 Port A Garment Confi guration Button
14 Port A Foot Indicator

EN-3
Section I - General Operating Instructions
Se
t up
• Place the Co
ntroller on the footboard via the bed hook or place it on a horizontal surface appropriate for the environment, such as on a table, within
reasonable proximity to the point of use. Be sure to allow adequate air flow to the vents located on either end of the controller.
• The controller can operate with one or two garments attached to the patient.
• Plug the tubing set(s) into the back of the controller. Route the tubing toward the patient’s limbs, being careful to leave access ways clear and
eliminate tripping hazards.
• Plug the tubes into garment(s) wrapped onto the patient’s limbs.
•
M
atch the left and right ports with the left and right limbs of the patient. Although the operation of the controller
is not affected, troubleshooting can
be easier. Check tubing set(s) for kinking and secure attachment at the controller and the garment(s).
•
Plug
the
controller power cord into a properly grounded hospital grade receptacle.
Start-up
• Press the Power On/Standby button to begin normal operation. If using leg sleeves, no further user intervention is required unless there is a fault
condition detected or if therapy must be discontinued.
• The Controller will beep and then flash all the LED’s.
LED Indicator Sequence:
1. Port A Leg, Port A Foot, Port B Leg, Port B Foot , Vascular Refill, Battery Status 1-3
➙
GREEN
2. Port A Leg,
Port A Foot,
Port B Leg, Port B Foot, Service Required, Battery Status 1
➙
RED
3
. Error Code/ Test Mode 1-8
➙
RED
•
An automatic valve and pump test will be quickly performed as the system tests microprocessor function and system memory.
• The pump will begin to operate as part of the Garment Detection procedure.
•
Detection of inoperative LED’s and the audible alarm function at start-up is the user
’s responsibility.
Garment Configuration & Detection
After startup, the Garment Configuration procedure allows the user to select when foot compression is required at either of the two Controller ports:
•
On the front display, the Port A Leg and Port B Leg indicators for both ports will be flashing green to indicate the
default garment configuration (leg
compression).
•
Pressing a port configuration button will cause the corresponding port
’
s Leg indicator to turn off and the Foot indicator will be lit to signify foot
’s Leg indicator to turn off and the Foot indicator will be lit to signify foot ’
compression. The port configuration buttons must be pressed for each port that is connected to a foot cuff to turn on the corresponding Foot
indicator(s).
NOTE:
Leg sleeve compression is the default garment configuration when the Controller is first powered on. Therefore, the configuration
buttons do
not have to be pressed to begin compression therapy when leg sleeves are being used.
The configuration buttons need to be pressed only when foot compression is to be used.
Also after startup, the Controller immediately begins conducting the Garment Detection procedure at each port to determine if the garments have been
properly attached to the controller:
•
If necessary, the port configuration buttons may be pressed again to turn the Foot indicator off and revert back to leg compression.
•
During Garment Detection, the compressor and valves are operating and air is delivered out the Controller ports to detect the number and type(s) of
garment(s) connected [Leg Sleeve(s) and/or Foot Cuff(s)].
•
If the Controller senses a properly attached garment and the type of garment detected matches the User-selected garment (or the default)
configuration, then the corresponding Port Status indicator (Leg or Foot, A or B side) on the front display will be lit green.
•
If the Controller senses a properly attached garment but the type of garment detected does not match the User-selected garment (or the default)
configuration, then a Garment Mismatch alarm is triggered. Garment Mismatch errors can be corrected by pushing the corresponding Port
Configuration button to change the User-selected garment type (Leg or Foot).
•
Once the Garment Detection procedure is completed and any garment mismatch errors are addressed, the Port configuration buttons will be disabled
and normal operation begins by starting the compression therapy.
•
If only one Controller port is connected to a garment for single-limb compression, then the User-selected garment (or the default) configuration
setting
(Leg or Foot) for the open port will be ignored and both the leg and foot indicators will be turned off after Garment Detection
.
•
If any garments are not properly detected or if no garments are attached to the Controller, the system will alarm until it is turned off. Check the
garment application and tubing connections then restart the Controller if necessary.
NOTE:
If a garment is attached anytime after the Garment Detection procedure has started, the system must be restarted to ensure that the
proper therapy will be applied to the limb(s).

EN-4
Normal Operation
• Verify that the corresponding Port Status Indicators are green for each disposable garment attached to the controller.
•
After garments have been
successfully detected, the Controller begins the process of applying intermittent compression alternating from on
e port to
the other if two garments are attached. Otherwise, the controller applies compression to only one port when one garment is attac
hed.
•
On successive cycles, the controller automatically adjusts its operating parameters to maintain set pressure.
Pressure Settings
•
The Controller features microprocessor-controlled automatic pressure adj
ustment.
•
The pressure setting depends on the type of garment: 45 mmHg for Leg Sleeves; 130 mmHg for Foot Cuffs.
Vascular Refill Detection
•
The SCD EXPRESS Compression System incorporates Tyco Healthcare’s patented “Vascular Refill Detection” method to customize the therapy for each
patient’s physiology. This system measures the time it takes for the veins in the limb to refill after having been compressed
by the system. The time is
then used in subsequent cycles as the time between compressions.
• The vascular refill detection method is used when first powering on the System after it reaches set pressure and every thirty minutes thereafter.
• The hourglass-shaped Vascular Refill Indicator on the front display will be illuminated during a measurement cycle.
• The method works best when the patient is still, however it will accommodate movement.
• If an error is detected during any measurement or if the compression is not within the System pressure specifications, the refill time measurement
will
be repeated after the next compression cycle.
• The time between compressions on the same limb will never be shorter than twenty seconds or longer than sixty seconds.
• If both Controller ports are being used, then the longer of the two measurements will
be used to adjust the time between cycles.
Garment
Compatibility
The SCD EXPRESS Compression System is designed for use with KENDALL garment Reorder Codes:
•
9545
Leg Sleeve
(Small Thigh Length)
•
9545T
Leg Sleeve
(Small Thigh Length Tear-Away)
• 9530
Leg Sleeve
(Medium Thigh
Length)
•
9530T
Leg Sleeve
(Medium Thigh Length Tear-Away)
•
9780
Leg Sleeve
(Large Thigh Length)
•
9780T
Leg Sleeve
(Large Thigh Length
Tear-Away)
•
9529
Leg Sleeve
(Medium Knee Length)
•
5897 Foot Cuff (Regular Foot)
•
9789
Leg Sleeve
(Large Knee Length)
•
5898 Foot Cuff (Large Foot)
•
9736
Leg Sleeve
(Sterile Thigh Length)
Further instructions for garment application and use are included with the Leg Sleeve and Foot Cuff packaging.
Tubing Set Com
patibility
The garments connect to the Controller via the Tubing Sets provided with the Controller. Additional or replacement Tubing Sets are available as Reorder
Code 9528. The Extension Tubing Sets are also available as Reorder Code 9595.
Portable Controller Configuration
•
The SCD EXPRESS Compression System Controller can be configured without the bed hook for easier portability. The System can also be used
without
the cord cover door so that the power cord can be disconnected at the Controller instead of at the AC power wall outlet.
• An optional Shoulder Strap for the SCD EXPRESS Compression System is available as Reorder Code 9527.
NOTE:
Removal of the bed hook and the cord cover door is described in Section VI – Disassembly and Reassembly.
Section II -
Battery Operation
The SCD EXPRESS Compression System is designed to operate normally on AC line power or DC battery power without interruption. If the Controller
is equipped with the optional battery pack there are three Battery Status Indicator LED’s used to represent the charge level of the battery. Once the
Controller is powered on, it may take the system a few seconds to establish communication with the battery and display the charge level. If the
Controller is not connected to a battery then the indicator lights will remain off.

EN-5
NOTE:
To prevent a potential shock hazard, the battery compartment should not be left vacant while operating the SCD EXPRESS Compres-
sion System. The battery pack, or a dummy pack (if so equipped), must be installed in the controller to prevent the user from having access
to the battery’s electrical connection.
T
he battery pack
for the SCD EXPRESS
Compression System is optional. To order
a
new or replacement battery
please contact Kendall Customer Service.
1
3
2
Battery State Battery Status 1 Battery Status 2 Battery Status 3
Battery State Battery Status 1 Battery Status 2 Battery Status 3
Battery State Battery Status 1 Battery Status 2 Battery Status 3
Battery State Battery Status 1 Battery Status 2 Battery Status 3
100% charge Green Green Green
100% charge Green Green Green
100% charge Green Green Green
100% charge Green Green Green
67-99% charge Green Green Green (Pulsing)
67-99% charge Green Green Green (Pulsing)
67-99% charge Green Green Green (Pulsing)
67-99% charge Green Green Green (Pulsing)
34-66% charge Green Green (Pulsing) Off
34-66% charge Green Green (Pulsing) Off
34-66% charge Green Green (Pulsing) Off
34-66% charge Green Green (Pulsing) Off
0-33% charge Green (Pulsing) Off Off
0-33% charge Green (Pulsing) Off Off
0-33% charge Green (Pulsing) Off Off
0-33% charge Green (Pulsing) Off Off
Uni
t plugged in and Powered On (Charging)
Battery Indicator Status
:
Battery State
Battery Status 1 Battery Status 2 Battery Status 3
Battery Status 1 Battery Status 2 Battery Status 3
Battery Status 1 Battery Status 2 Battery Status 3
67-100% charge Green Green Green
67-100% charge Green Green Green
67-100% charge Green Green Green
67-100% charge Green Green Green
34-66% charge Green Green Off
34-66% charge Green Green Off
34-66% charge Green Green Off
34-66% charge Green Green Off
< 34% charge Green Off Off
< 34% charge Green Off Off
< 34% charge Green Off Off
< 34% charge Green Off Off
15-40 minutes left* Amber (Flashing) Off Off
15-40 minutes left* Amber (Flashing) Off Off
15-40 minutes left* Amber (Flashing) Off Off
15-40 minutes left* Amber (Flashing) Off Off
< 15 minutes left* Red (Flashing) Off Off
< 15 minutes left* Red (Flashing) Off Off
< 15 minutes left* Red (Flashing) Off Off
< 15 minutes left* Red (Flashing) Off Off
Unit not plugged in and Powered On (Operating on Battery)
Unit
Powered Off (charging when plugged in)
Battery State Battery Status 1 Battery Status 2 Battery Status 3
Battery State Battery Status 1 Battery Status 2 Battery Status 3
Battery State Battery Status 1 Battery Status 2 Battery Status 3
Battery State Battery Status 1 Battery Status 2 Battery Status 3
0-100% charge Off Off Off
0-100% charge Off Off Off
0-100% charge Off Off Off
0-100% charge Off Off Off
Charging the Battery
The battery will begin charging as soon as the unit is plugged into an AC power source. The amount of time required to charge the battery will vary
depending on the battery’s overall condition, age, and the controller’s state during charging. For example, charging a new, fully drained battery will
take approximately 4 hours with the controller on standby and 8 hours with the controller powered on. The Battery Status indicators should always be
used to determine the state of charge for the battery. A fully charged battery will typically provide 6-8 hours of operation time depending on the sleeve
configuration, sleeve application, and the battery condition.
NOTE:
If the operation time on battery power is extremely short the battery should be returned for service or replace
ment.
NOTE:
The battery may sustain permanent damage and become unrecoverable if it is left
unused for extended periods of
time. It is
recommended that
the battery pack be stored with a minimum charge of 50% and kept near 25
° C
(77
°F)
if prolonged storage is necessary.
* With 15-40 minutes of battery charge left, an alarm will sound in a sequence of three beeps once every two minutes. Once
there is less than 15 minutes of battery charge left, the alarm will sound continuously.

EN-6
1
2
3
•
Press the port configuration
button(s) to turn the foot selection
on/off depending on what type
of garment(s) is connected to the
Controller.
• If the proper garment is selected
and the problem persists have the
Controller serviced by a professional.
- Check tubing set for any
kinks.
- C
heck garment
application
(too
loose or tight).
- Check tubing connections.
- Check for leaks in the
garment(s)
and the tubing.
If susp
ect, substitute
a
new tubing set or garment(s).
-
Turn the Controller off an
d
restart
- If the problem persists have the
Controller serviced
by a professional.
•
Garment Mismatch
• System High
• High Pressure
• Low Pressure
• The Garment Detection procedure has
detected a garment confi guration (Leg or
Foot fl ashing green) that does not match
the User-selected confi guration (Leg or
Foot red).
•
System pressure has exceeded
90 mmHg
(Leg sleeve) o
r
180 mmHg (Foot
Cuff
).
• Leg Sleeve pressure is greater
than
47
mmHg after
5 consecutive cycles
;
Foot Cuff
pressure is greater
than
135 mmHg after 5
consecutive
cycles.
• Leg Sleeve pressure is less than
43 mmHg
afte
r
5 consecutive
cycles
;
Foot Cuff
pressure is
less than 125 mmHg after
5
consecutive cycles.
T
his alarm will also be triggered
when no
ga
rments are detected
at either Controller
port (A or B)
during startup.
Battery Warnings
The SCD EXPRESS Compression System battery pack contains Lithium Ion (Li-Ion) battery cells and must be used properly for safety and to maintain
optimal performance.
• Store only between –20
°
C (-4
°
F
)
and 60
°
C (1
4
0
°
F
)
.
•
Do not drop, impact, or immerse in water.
•
Do not touch or ingest any leaking electrolyte. If contact occurs, rinse skin and/or eyes immediately and seek medical attention if irritation develops.
If ingested, contact local poison control center.
•
Do not open battery, dispose of in fire, or short circuit. Doing so may cause the battery to ignite, explode, leak, or become hot and cause personal
injury.
•
Dispose of improperly working or damaged battery packs according to local regulations.
•
Charge only with specified chargers according to Kendall’s instructions.
Section III - Fault Conditions and Troubleshooting
When the microprocessor detects a fault condition, it interrupts the normal operation of the Controller, deactivates all valves to vent the air from the
garment(s), displays a f
ault code,
garment(s), displays a fault code,garment(s), displays a f
and sounds an audible alarm. If a Garment Mismatch alarm is triggered the user may remedy the problem by
pressing the corresponding port configuration button(s). All other alarms will remain active until the Controller is turned off, or the battery runs out of
charge (if operating on battery power).
• The most common Controller alarm is the Garment Pressure alarm, indicated by flashing red Port Status Indicator(s). Garment Pressure alarms
can usually be remedied by checking the disposable garment(s) for proper application and tubing set connection(s) for any leaks or kinks, and then
restarting the System. However, if the alarm condition persists or if the Servi
ce Required Error indicator is illuminated then the Controller should be
returned for service
.
Watchdog Circuit
•
If the microprocessor cannot continue normal function, the watchdog circuit will trip. This causes the Controller to go into reset and restart normal
operation.
• If the cause of the disruption is still present, the unit will continue to attempt to reset, which will cause the alarm to sound once every second.
• If the cause of the disruption was transient, such as a high energy RF pulse, the Controller will restart and continue normal operation.
Alarm Code
Alarm Code
Fault Type
Description
Troubleshooting

EN-7
4
5
6
7
8
9
(Battery
Indicator 1)
- Check garment application
(too loose or tight).
-
Turn the Controller off and
restart
.
-
Verify that the valve assembly
is properly connecte
d.
-
Tur
n the Controller off
and restart.
-
If the problem persists have the
Controller
serviced by a professional.
-
T
urn the Controller off
and restart.
-
If the problem persists have the
Controller
serviced by a professional.
-
Verify that the pump is
properly
connected
.
-
Turn the Controller off and restart.
-
If the problem persists have the
Controller
serviced by a professional.
-
Check tubing for kinks.
-
Check garment application
(too loose or tight).
-
Check to make sure there
are no kinks in the internal tubing.
-
Turn the Controller off
and restart
.
-
If the problem persists
h
ave the
Controller serviced by a professional.
-
Plug the Controller into
an AC power
outlet.
-
If the problem persists recalibrate or
replace the battery pack. The battery
can be calibrated by using test
m
ode 01 Burn-In (Sect
io
n V).
System Pressure
Valve Error
Software Error
Pump Error
Vent Error
Low Battery
Alarm
Leg Sleeve pressure is not
bet
ween
35 and
55 mmHg for
12 consecutive cycles; Foot Cuff
pressure
is not between
110 and
150 mmHg
f
or
for f
12 consecutive
cycles.
If a valve electrically
malfunctions,
this error
will be
displayed.
Upon startup, the
microprocessor performs
diagnostic tests. If the Controller fails to pass
these
tests this error will be displayed.
I
f the pump electrically
malfunctions this
error
will be displayed.
The pressure in a garment
is greater than 20
mmHg at
the end of any vent period.
T
he pressure detected during
an inflation
cycle does not ri
se
above 5mmHg.
T
here is less than 10 minutes of
battery
charge remaining. The
pump and valves
will
continue to operate for
as long as there is
enough power.
Alarm Code
Fault Type
Description
Troubleshooting

EN-8
Ala
rm Code
10
Battery Error
(Battery Indicator
1
)
11
(Error Indicators 5 & 7)
Fault Type
Battery Error
Temperature
Error
Description
If a battery calibration error or
battery cell
failure is detected
this error will be displayed.
If the internal case temperature of
the
Controller
drops
below 5° C
(41° F) or exceeds
55
°
C (131
° F)
.
Troubleshooting
-
Turn the Controller off
, plug it into an
Turn the Controller off, plug it into an Turn the Controller off
AC po
wer outlet,
and restart.
- If the problem persists
recalibrate or
replace the battery pack.
-
Make sure the controller
is not covered and that
the fan port
is not
obstructed.
- Turn the Controller off, allow it to
cool, and restart.
- Verify that the cooling fan is
properly connected.
- Have the Controller serviced by a
professional if the problem persists.
Section IV - Service and Maintenance
This service manual is intended for use as a guide to technically-qualified personnel when evaluating System malfunctions. It is not to be construed as
authorization to perform warranty repairs. Unauthorized service will void the warranty.
Introduction
The SCD EXPRESS Controller contains no user serviceable parts. User maintenance is covered in the sections that follow. All other maintenance must be
performed by appropriately qualified service technical personnel.
Service technicians should be familiar with the operator’s portion of this manual and the operating principles of the SCD EXPRESS Compression System.
If a Controller is to be returned to Kendall for service, a description of the operating conditions and the fault code displayed should accompany the unit.
The fault codes displayed by the Controller are useful in diagnosing service problems.
This manual describes service procedures to the board level, with an exploded view of the Controller shown in
Figure1. If a component failure on a circuit board is suspected, the unit should be returned for service. It is recommended that the system be returned
with the circuit board in place, as removal of the board(s) involves additio
nal risk of mechanical damage and damage from electrostatic discharge (ESD
).
Warranty and Factory Service
Tyco Healthcare Group LP warrants that your SCD EXPRESS Compression System is free from defective material and workmanship. Our obligation
under this warranty is limited to the repair of Controllers returned to a service center,
transportation charges prepaid, within one year of delivery to the
original purchaser. Specifically, we agree to se
rvice and/or adjust any Controller as required if returned for that purpose, and to replace and repair any
part which, upon our examination, is proven to have been defective. This warranty does not apply to the Tubing Set or the disposable garments, or
to equipment damaged through shipping, tampering,
negligence, or misuse, including liquid immersion, autoclaving, ETO sterilization,
or the use
of
unapproved cleaning solutions.
T
o the extent permitted by applicable law, this limited warranty does not cover, and is intended to exclude, any and all liability on the part of the
To the extent permitted by applicable law, this limited warranty does not cover, and is intended to exclude, any and all liability on the part of the T
Company, whether under this limited warranty or any warranty implied by law, for any indirect or consequential damages for breach hereof or thereof.
Except as expressly provided above in the limited warranty, to the extent permitted by applicable law, the Company hereby negates and disclaims
all express and to the extent permitted by applicable law, implied warranties, including the warranties or merchantability and fitness for a particular
purpose.
Controllers requiring repairs should be sent to a service center. Call one of the service centers listed. Obtain a return material authorization number and
ship the controller, prepaid and insured in the original carton.

EN-9
Fan Screen and Ventilation
CAUTION
:
Unplug the Controller and remove the battery pack (see Section VI) before fan screen maintenance.
It is very important that the fan screen be cleaned as required to ensure continued trouble-free operation. The Controller should never be run without
the fan screen in place. To clean the fan screen, follow the instructions later in this manual in the “Fan and Screen (Removal/Installation)” Section to
first remove the fan screen. Brush any loose dirt and lint from the surface of the screen. Reinstall the fan and screen, making sure the gasket m
aterial is
oriented so as to prevent the screen from touching the fan directly.
During system use, obstruction of the fan cover and vents should b
e avoided. Free flow of air is necessary to prevent overheating and premature
component failure.
Fuses
CAUTION:
Unplug the Controller
and remove the battery pack
(see Section VI)
before replacing the fuse(s).
Blown fuses should only be replaced by those indicated on the power supply board near the location of the fuses at the AC inlet. Use only 1.6 A,
250 VAC, 5x20mm Slo Blo fuses. The use of fuses that have the Semko and/or VDE marking is preferred. If a fuse blows a second time, it should be
presumed that the Controller is defective and requires further service. Please contact your service center. Fuses are not accessible from the outside of the
controller. Refer to the Disassembly/Reassembly pr
ocedures later in the manual. The fuses are located on the power supply board as part of the
power
inlet module under the fuse cover.
Electrical Safety
CAUTION:
Be sure the Controller is disconnected from the power source (AC line power and DC battery power) before any disassembly. A potential
SHOCK HAZARD exists when the front cover is removed even with the unit turned off.
NOTE:
T
he power supply cord/plug serves as the electrical supply mains disconnect device.
The system should be periodically inspected to ensure its electrical safety. Check the resistance of the power cord ground from the plug to the IEC
60320 connector. This resistance should not exceed 0.1 ohm. If ground resistance exceeds this value or the insulation integrity of the unit has been
compromised through mechanical damage, the Controller should be returned to a service center for testing and repair
.
compromised through mechanical damage, the Controller should be returned to a service center for testing and repair.compromised through mechanical damage, the Controller should be returned to a service center for testing and repair
* The SCD EXPRESS was designed to provide sufficient isolation inside the controller between all exposed metal components and the electrical
supply mains. As a result there is no externally accessible ground on the device. The recommended electrical safely test for proper grounding is a
resistance check of the power cord ground.
After Any Repair Once Per Year
After Any Repair Once Per Year
Inspect and Clean Fan Screen
As Required
Verify Tr
ansducer Calibration
(Test Modes 03 and 04)
X X
X X
Electrical Safety Tests
X
X
Leak Test (Test Mode 06) X
Leak Test (Test Mode 06) X
General Function Test (Test Mode 07)
X
Suggested Preventative Maintenance Schedule
Outside U.S. and
Canada
Tyco Healthcare Group LP
20 Garryduff Road
Ballymoney, BT53 7AP
00-44-(0)-2827-661719 Service Center.
00-44-(0)-1329-224226 Customer Care
CANADA
Tyco Healthcare Group Canada
7300 Trans Canada Highway
Pointe-Claire, Qc H9R 1C7
877-664-TYCO (8926)
UNITED STATES
Tyco Healthcare Group LP
98.6 Faichney Drive
Watertown, NY 13601
(800) 448-0190 or (315) 788-5246

EN-10
Cleaning
The Controller cabinet can be cleaned with a soft cloth dampened with water. If necessary, a mild disinfectant and/or detergent can be used; excess
fluid should be avoided. The Controller should be wiped with a clean, dry cloth afterward. Do not immerse in any liquid. Do not use products
containing ammonium chloride, acetone or other aromatic solvents, as those chemicals will degrade the integrity of the case and cause it to become
embrittled and possibly crack.
The SCD EXPRESS Compression System cannot be effectively sterilized by liquid immersion, autoclaving, or ETO sterilization, as irreparable damage to
the System will occur.
Electrical/Electronics Description
Line voltage is fed into the controller through the power cord to the power supply mounted on the Power Supply PC Board in the rear case of the
controller. It is important to disconnect the power cord at the outlet before opening the controller case. Exposure to high voltage on the Power Supply
PC Board is likely to occur if it is electrically live.
The power supply converts AC line voltage, 100 to 240 VAC, to DC voltage to power the controller components, including the main Controller PC Board
that is mounted onto the front case. Alternately, the main Controller PC Board may be directly powered by the optional battery pack. The Controller PC
Board controls all functionality of the system and includes the transducer and buzzer. It does not contain any high voltage. The buttons and indicator
LED’s on the front display of the controller are integrated into the membrane panel which connects to the Controller PC Board.
Kendall does not recommend any attempt to repair printed circuit boards. In manufacturing, extensive testing is performed that cannot be duplicated in
the field without specialized equipment. Improper repair could result in patient or user hazards.
Pneumatic Operation Description
When the controller is turned on, the compressor operates and the valves are cycled to determine which type of garment is connected. After garment
detection has completed, an inflation cycle is initiated, releasing air through the set of valves, mounted to a manifold. A transducer monitors the
pressure in the garments. The reading from the transducer assists the controller in adjusting the pump’s motor speed to deliver the proper pressure to
the garments in the appropriate amount of time.
Section V - Test Methods and Calibration
The SCD EXPRESS Compression System has various test modes that can be accessed by the service technician. They are intended for use by qualified
personnel. To activate the test modes follow these steps for entering “Test Access Mode”.
1. Plug the controller into an outlet supplying the appropriate line voltage. Do not activate the test modes while operating on battery power.
2
. Press the Port A and Port B configuration buttons at the same time while turning the controller on.
3. The buzzer will beep and the Test Mode 01 indicator will illuminate signifying “Test Mode 01”.
NOTE:
The Error Code Indicators 1-8 double as Test Mode Indicators 1-8.
4. The user can cycle through the test modes by pressing the Port A button. Each test mode is indicated by the illumination of the corresponding
test
mode indicator number. Pressing the Port A button with the last test mode number illuminated (Test Mode 08) will cycle the test mode back to Test
Mode 01.
5. After selecting the desired test mode, the Port B button can be pressed to initiate the test.
6. If test access is entered but no test mode is selected within 2 minutes, it is assumed that the test access mode was entered inadvertently and a Low
Pressure alarm will be triggered.
7. If a test mode is entered and left idle for 5 minutes the unit will revert back to test access mode selection.
8. To exit Test Access Mode, turn the controller off.
Test Mode 01 - Burn-In
NOTE:
Burn-In mode is used in manufacturing to ensure proper assembly, to identify premature failures and to calibrate the battery pack.
This mode is
not generally used outside of the manufacturing environment.
1
.
Verify nothing is plugged into the ports on the back of the controller and enter test access mode. Select Test Access Mode 01.
2
.
Press
the
Port B button to begin Burn-In. The compressor will operate and the valves will actuate, releasing air out of the ports. The process
repeats
continuously until the Burn-In period is complete (approximately 16 hours).
3
.
The battery, if equipped, will be discharged then charged for calibration during Burn-in.
4
.
When 16 hours of Burn-In is completed the controller will go into alarm mode, flashing Test Access Mode 01 red, and the Port A Leg , Port A Foot,
Port B Leg, Port B Foot indicators green. The buzzer will not pip during this alarm.

EN-11
Test Mode 02 - General Function Test
1. With nothing plugged into the ports on the back of the controller, enter test access mode. Select
Test Access Mode
02.
2.
Press the Port B button to begin the test.
3.
Pressing
the
P
ort A button during this test will cause each one the of the LED’s to illuminate one at a time in succession and the alarm to pip.
4.
Pressing and holding the
P
ort B button
will increase the pump speed to its maximum in 4-5 seconds.
5.
Releasing
the
P
ort B button
will allow the pump to decrease its speed.
6.
The valves will actuate in succession (valve #1 through valve #6) for two seconds each.
Test Mode 03 - Pressure Transducer Calibration
NOTE:
The transducer used in the SCD EXPRESS Compression System is a state-of -the-art, highly precise and virtually drift free device.
Factory
calibration certification is void if the case is opened.
Recalibration is rarely required and
should be done only when necessary.
Always perform test 04 before test 03 to verify the pressure transducer calibration.
REQUIRED EQUIPMENT:
A regulated, precision air source accurate to ±0.1mmHg over a range of 0 to 130 mmHg.
1. With nothing plugged into the ports on the back of the controller, enter test access mode. Select
Test Access Mode
03.
2. Press the P
ort B button
to begin the test.
3.
Te
st Mode
03 indicator will flash red until the calibration procedure is completed or an error condition
occurs.
4. Valve #1 will be energized throughout the procedure, so that the user can verify the calibration of the pressure transducer with the controller case
open or closed. The pressure s
tandard can either be directly connected to the
transducer with the case open, or it can be attached to the Bladder #1
location at Port A with the case closed. The Bla
dder #1 location is the left-most fitting within Port A (as viewed from the back of the Controller).
5.
Test Mode
04 indicator is lit red to signal that the unit is ready for the 0 mmHg calibration. After all air
pressure is disconnected to zero the
transducer, press
the P
ort B button
to complete calibration for this range.
6.
Test Mode
05 indicator is lit red to indicate that the unit is ready for the 18 mmHg calibration. After 18 mmHg (± 0.1 mmHg) of air pressure is
applied to the transducer, press the P
ort B button
to complete the calibration for this range.
7.
Test Mode
06
indicator is lit
red
to indicate that the unit is ready for the 45 mmHg calibration. After 45 mmHg (± 0.1 mmHg) of air pressure is
applied to the transducer, press
the P
ort B button
to complete
the
calibration for this range.
8. Test Mode 07 indicator is lit red to indicate that the unit is ready for the 130 mmHg calibr
atio
n.
Afte
r
130 mmHg (± 0.1 mmHg) of air pressure is
applied to the transducer, press
the P
ort B button
to co
mpl
e
te the
calibration for this range.
9. Upon completion, the new calibration values are recorded into memory and the unit beeps and reverts ba
ck to
Test Access Mode.
10. If the calibration test mode is exited before the process is completed, the previous calibration values remain unchanged.
Test Mode 04 - Pressure Transducer Calibration Verification
Note:
The transducer used in the SCD EXPRESS Compression System is a state-of-the-art, highly precise and virtually drift free device.
Factory
calibration certification is void if the case is opened.
Recalibration is rarely required and should be done only when necessary.
Always perform test 04 before test 03 to verify the pressure transducer calibration.
REQUIRED EQUIPMENT:
A regulated, precision air source accurate to ±0.1mmHg
over a range of 0 to 130 mmHg.
1
.
With nothing plugged into the ports on the back of the controller, enter test access mode. Select Test Access Mode 04.
2
.
Press the
P
ort B button
to begin the test.
3
.
Test Mode 04 indicator will flash red until the calibration procedure is completed or an error condition occurs.
4
.
Valve #1 will be energized throughout the procedure, so that the user can verify the calibration of the pressure transducer with the controller case
closed. The pressure standard can be directly connected to the Bladder #1 location at Port A with the case closed.
The Bladder #1 location is the left
most fitting within Port A (as viewed from the back of the Controller).
NOTE:
For each of the calibration verification ranges, if the supplied pressure is less than the specific calibration range then the Port A
L
eg
indicator will
light red. If the supplied pressure is greater than the specific calibration range then the Port B Leg indicator will light red.
When the supplied pressure is within the calibration range, then both the Port A
L
eg
and Port B
L
eg
indicators will light green.
Test Mode Look up Char
t
0
1
– Burn-In Feature
0
5
– Self Test
0
2
– Functional Test
0
6
– Leak Test
0
3
– Pressure Transducer Calibration
0
7
– Performance Test
0
4
– Pressure Transducer Calibration Verification
0
8
–
Manufacturing
Test

EN-12
5
.
Test Mode 05 indicator is lit red to indicate that the unit is ready for the 0 mmHg calibration verification. With no pressure applied to the transducer,
press
the P
ort B button
to complete the verification for this range.
6
.
Test Mode
06 indicator is lit red to indicate that the unit is ready for the 18 mmHg calibration verification. Apply 18 mmHg (± 0.1 mmHg) then press
the P
ort B button
to complete the verification for this range.
7.
Test Mode
07 indicator is lit red to indicate that the unit is ready for the 45 mmHg calibration verification. Apply 45 mmHg (± 0.1 mmHg) then press
the P
ort B button
to complete the verification for this range.
8.
Test Mode
08 indicator is lit red to indicate that the unit is ready for the 130 mmHg calibration verification. Apply 130 mmHg (± 0.1 mmHg) then
press
the P
ort B button
to complete the verification for this range
9. Upon completion the unit beeps and reverts back to Test Access Mode.
Test Mode 05 -
Self Test
1.
Enter Test Access Mode and press the Port A button until Test Access Mode 05 is selected.
2.
Press the Port B button to begin the self test.
3.
The Test Mode 05 indicator will flash until the test is completed.
4.
The alarm will pip a
nd
the unit will perform the full array of tests performed during Start-up
.
Test Mode 06 - Leak Test
The pressure decay leak test will determine if there are any significant leaks in the system (including the pneumatic circuitry inside the controller and the
connections with the tubing set).
1. Attach a tubing set connected to leg sleeves wrapped around appropriately sized leg forms.
2.
Enter Test Access Mode and press the Port A button until Test Access Mode 06 is selected.
3. Press the Port B button to begin the Leak Test.
4. The Test Mode 06 indicator will flash until the test is completed.
5. The controller will inflate all the bladders for port A (Valves #1-3) at the same time to 45mmHg and then vent down to 5-7 mmHg.
6. The unit will then hold the pressure for 40 seconds, while reading the pressure at 20 seconds and 40 seconds. (The alarm will pip at the start of the
leak test and at the 20-second and 40-second intervals.)
7. If there is a significant pressure drop over that 20-second time period, a Low Pressure alarm will be triggered.
8. If no leak is detected, the unit will repeat the leak test for Port B (Valves #4-6).
9. If no leaks are detected at either port then the system will simply beep and return to Test Access Mode.
Test Mode 07 - Performance Test
When in this mode, the user can verify the pump and valve performance, pressure delivery, and the airflow through the pneumatic circuit. In
manufacturing, this test is conducted with known volumes connected to the sleeve ports. Then the inflation cycles run during the test at the low and
high pump speeds create backpressures in the volumes that are measured and used to verify system performance .
1. Attach a tubing set connected to leg sleeves wrapped around appropriately sized leg forms
.
2.
Enter Test Access Mode and press the Port A button until Test Access Mode 07 is selected.
3.
Press the Port B button to begin the Performance Test.
4.
The Test Mode 07 indicator will flash until the test is completed.
5.
After initiating the performance test, the Port A Leg indicator will flash green in sync with an audible alarm.
6.
After the Port B button is pressed the Port A Leg indicator will be lit green, the alarm will stop, and the controller will then go through a normal
inflation cycle at Port A with the pump operating at a low speed throughout the cycle.
7.
Next, the Port B Leg indicator will flash green in sync with an audible alarm.
8.
After the Port B button is pressed, the Port B Leg indicator will be lit green, the alarm will stop, and the controller will then go through a normal
inflation cycle at Port B with the pump operating at a high speed throughout the cycle.
9.
Upon completion the unit beeps and reverts back to Test Access Mode.
Test Mode 08
-
Manufacturing Test
NOTE:
Manufacturing Test mode is used in manufacturing with specialized test equipment to ensure proper assembly and performance.
This mode is not designed for use outside of the manufacturing environment.

EN-13
Section VI - General Disassembly / Reassembly
WARNING:
Always make sure the power cord is unplugged and the battery is removed before attempting to perform any installation or
remo
val
procedures.
• Follow ESD (Electrostatic Discharge) safety procedures to protect the electronics located within the controller.
•
Remove the power cord cover by first removing the retaining screw on the cord cover door.
•
Slide the cord cover door off by pressing the door to the right.
•
Remove the power cord by rocking back and forth until the cord is loose.
•
Remove the five (5) screws that hold the front cover to the rear cover with a Phillips head screwdriver.
•
The front cover may now be carefully pulled away. To separate the front and rear covers, reach in and remove the
transducer tube from the transducer
on the front cover. (Diagram 01)
•
Observe and note the locations of all tubing and wiring harnesses for ease in reassembly.
•
If required, disconnect the electrical connectors and tubes so the two case halves can be separated completely.
•
Reassembly is the reverse of disassembly.
Battery Pack (Removal / Installation)
• Remove the battery pack screw located on the bottom center of the rear case of the controller.
• Slide the battery pack out by pulling on the tab on the bottom, right side of the unit.
• Installation is the reverse of removal.
Compressor (Removal / Installation)
•
The compressor is not a user serviceable component. Do not disassemble. Do not oil. The compressor is mounted to its own separate bracket
that, in
turn, mounts through the inside of the battery compartment in the rear case of the controller.
• Disconnect the pump wiring harness from the Controller Board on the front case.
•
D
isconnect the compressor output tube at the check valve. (Diagram 02)
•
Remove the compressor intake tubing from the muffler port. (Diagram 02)
•
Remove the two (2) screws from the compressor mounting bracket. (Screws are accessible through the right side battery compartment once the
battery pack is removed.) (Diagram 01)
•
Slide the entire compressor assembly out of the rear case housing.
•
Installation is the reverse of removal.
Valve Manifold and Bed Hook (Removal / Installation)
•
The valve manifold is located in the center of the controller on the rear case. It is a plastic manifold block with six solenoid valves. No attempt
should
be made to repair a damaged manifold or valve. Return the entire assembly for repair or replacement.
•
Inspect tubes that lead to the manifold for kinks and proper attachment before performing any work. Detach all tubing from the manifold fittings.
Note the location of connections and the tubing routing for ease of reassembly.
• Disconnect the valve wiring harness from the Controller Board on the front case.
• Remove the two side screws from the valve manifold assembly and then pull the bed hook away from the rear case. (Diagram 03)
• Remove the center screw from the manifold assembly and remove the valve manifold.
• Some wires may be secured to the manifold with wire ties. Be prepared to reattach these wires after reinstalling the assembly. New wire ties will
be
required.
• Installation is the reverse of removal.
Muffler (Removal / Installation)
• The muffler is a custom plastic part used to keep the Kendall SCD EXPRESS Compression System running quietly.
• To remove the muffler detach the compressor intake tubing and slide the muffler out from under the power board.
• For reinsertion of the muffler be sure to route the intake tubing properly. (Diagram 02) Tubing must be run between the muffler and the right side
of
the rear case to avoid kinking.

EN-14
Power Supply Board (Removal / Installation)
CAUTION:
Use a grounded strap when handling any electronic components
.
• The power supply has no user serviceable parts except for the fuses. No attempt should be made to repair a damaged supply. Return to the factory
for
repair or replacement.
• Disconnect the 4-pin controller board wiring harness from the power board.
• Remove the muffler as described in “Muffler (Removal/Installation).”
• The power supply board is held in place by channels on the side of the rear case as well as retaining brackets on the front case.
• To remove the power board slide the board out from the rear case.
• Installation is the reverse of removal.
Fan and Screen (Removal / Installation
)
• Disconnect the 2-pin fan connector from the power board.
• Observe the angle and orientation of the fan and screen prior to removal to ease reassembly. (Diagram 04)
• Slide the fan and screen out of the case by pulling straight out.
• Installation is the reverse of removal. For optimum cooling and quietness, use only Kendall replacement fans.
Main Controller Board (Removal/Installation)
CAUTION:
Use a grounded strap when handling any electronic components
.
• The main CPU board has no user serviceable parts. No attempt should be made to repair a damaged board. Return to the factory for repair or
replacement.
• Disconnect the 4 pin controller board wiring harness from the power board.
• The main CPU board is located on the front cover and is held in place with two screws.
• Disconnect the two membrane ribbon cables by sliding them out of the connectors on the controller board.
• Remove the controller board.
• Installation is the reverse of the removal.

EN-15
Section VII - Specifi cations
SCD EXPRESS Compression System
Safety Standards
Built to UL60601-1, CSA-C22.2 No. 601.1-M90, EN60601-1, and IEC 60601-1-2 Standards
UL Classified File # E189131
UL Device Classification
Class I Equipment
Internally Powered
, Portabl
e
Type BF Applied Parts
Not AP or APG Equipments
Mode of Operation
Continuous
Ingress of Water Protection
Ordinary,
IPX0 (IEC 529)
Compression Type
Leg Sleeves: Sequential, Gradient
Foot Cuffs: Uniform
Compression Cycle
Leg Sleeves: 11 Seconds Compression
Foot Cuffs: 5 Seconds Compression
Decompression time based upon Vascular Refill Detection measurement
Set Pressure
Leg Sleeves: 45 mmHg
Foot Cuffs: 130 mmHg
Bed Hook
Yes (Optional)
Bed Hook Yes (Optional)Bed Hook
Power Cord Storage
Yes
Audible/Visual Alarms
Low Pressure, High Pressure, Internal Electronics Malfunction
Power Cord
Hospital Grade Plug
Controller
Dimensions
:
Height:
6.2 inches (15.8 cm)
Width:
7.0 inches (17.8 cm)
Depth:
4.5 inches (11.4 cm)
Controller Weight
3.5 lbs. (1.6 kg)
Power Requirements
100-240 VAC, 50VA, 50/60 Hz
Battery
16.4 V, 2000mAhr, Lithium Ion (Optional)
Run Time: 6-8 hours
Charge Time: 4 hours (charging only)
Shoulder Strap
Yes (Optional)
Shipping Unit
Each
Shipping Case Dimensions
Height:
11
.6 inches (29.4 cm)
Width:
9.
0
inches (22.8 cm
)
Depth:
11.
4
inches (28.9 cm)
Shipping Weight
7 lbs. 6 oz. (3.45 kg)
Tubing Set
Included, set of two individual assemblies
Operation & Service Manual
Included
Transport & Storage
0°C (32
°F)
to 50
°
C
(
122
°F)
If the user suspects that the environment conditions
for transp
ort and storage have been exceeded, return
the unit for service.

EN-16
Section VIII - Schematics
Figure 1 -
Parts Assembly Diagram -
Exploded View
1. Front Membrane Panel
2. Front Enclosure
3. Main Controller Board
4. Valve/Manifold Assembly
5. Wire Barrier
6. Compressor Assembly
7. Power Supply Board
8. Fuses (Set of 2)
9. Muffl er
10. Fan
11. Fan Screen
12. Rear Enclosure
13. Rubber Feet (Set of 2)
14. Battery Pack
15. Power Cord
16. Cord Capture Door
17. Bed Hook
Controller Parts List
17
16
15
13
3

EN-17
14
12
11
10
9
8
7
5
4
6
2
1
Section VIII - Schematics
Figure 1 -
Parts Assembly Diagram -
Exploded View
Table of contents
Other kENDALL Medical Equipment manuals