kENDALL SCD SEQUEL tyco Healthcare 6325 Installation and operating instructions

Operation
&
Service
Manual
MODEL
6325
COMPRESSION
SYSTEM
#YCO
/
Healthcare
KENDALL

Parts
Listing
SCD
Sequel
Model
6325
To
order
repair
parts
listed
here,
call
Kendall/Tyco
Healthcare
at
(800)
765-4324.
Contact
Customer
Service
for
availability
of
parts
not
listed
below.
Description
Order
Part
Number
POWER
SUPPLY
PC
BOARD
7774006780
CPU
PC
BOARD
7774006840
TRANSDUCER
-
PRESSURE
7774020504
COMPRESSOR
(THOMAS,
GAST)
7774010283
COMPRESSOR
FILTER
7774031569
COMPRESSOR
MUFFLER
7774031575
COMPRESSOR
CABLE
7774090530
VALVE
ASSEMBLY
(KIP,
SKINNER)
7774010300
TRANSFORMER
7774010301
FRONT
ENCLOSURE
ASSEMBLY
(CASE,
FOAM,
LABELS)
7774109780
FRONT
ENCLOSURE
7774031601
REAR
ENCLOSURE
ASSEMBLY
(CASE,
FOAM,
LABELS)
7774109790
REAR
ENCLOSURE
7774031600
NYLON
HOOK
PAD
7774170580
FAN
GUARD
7774031602
FAN
FILTER
7774031572
HOOK
(BED)
7774031567
CASE
RUBBER
FEET
7774031570
ROCKER
SWITCH
7774070156
BOARD
TO
BOARD
CABLE
7774080385
POWER
CORD
ASSEMBLY
7774090529
STRAIN
RELIEF
7774131980
TUBING,
TRANSDUCER
(1/8
X
1/4)
7774090551
TUBING,5/32
X
9/32
7774090579
TUBING,(COMP-
MUFFLER),
5.50"
7774090540
FUSEHOLDER
7774010289
FUSE,
.75
AMP
7774010290
SHRINK
TUBING
(FUSEHOLDER)
7774090533
FAN
ASSEMBLY
7774109190
MEMBRANE
SWITCH
7774070163
CAUTION
LABEL
7774181923
GROUNDING
LABEL
7774181924
OPERATING
INSTRUCTIONS
LABEL
7774181947
QUIET FOAM,
LOWER
FRONT
7774031576
QUIET
FOAM,
MIDDLE
FRONT
7774031577
QUIET
FOAM,
UPPER FRONT
7774031578
QUIET
FOAM,
BOTTOM
7774031579
QUIET
FOAM,
CHASSIS
7774031580
QUIET
FOAM,
END
7774031599
F181957

Operation
&
Service
Manual
MODEL
6325
Table
of
Contents
M
GENERAL
OPERATING
INSTRUCTIONS
e
Pre-Operation
Check
.......................................
1
*
Use
of
Bed
Hook............................................
1
+
Operation
....................................,..,.,,,,,4.
1
e
Cycle
Monitor
..........................,,..................
1
©
Comfort
Cooling
...................................,........
1
+
Power
Cord
Storage
........................................
1
+
Sleeve
Compatibility
........................................
1
M
FAULT
CONDITIONS
e
Fault
Messages
.................................,..,..,..
24
ㆍ
Watchdog
Circuit
.........................................1
4
M
CONTRAINDICATIONS
...........................,.............
5
M
CAUTIONS
.............................,.,....,..........,.,.
5
E
SERVICE
AND
MAINTENANCE
slntroduction.............................
eee
6
e
Warranty
and
Factory
Service
...............................
6
+
Maintenance
.........................................,....
6-7
e
Ventilation
.................................................
7
efFuses.................................
eee
7
+
Electrical
Safety
............................................
7
e
Cleaning
.................................,...............
7
*
Electrical/Electronics
Description
...........................
7
e
Pneumatic
Description
...............................:
Fer
8
e
Portidentity
............................................
8
»
System
Start-Up
Routine
....................................
8
+
Pressure
Monitoring
........................................
8
(Continued
on
next
page}

Operation
&
Service
Manual
MODEL
6325
Table
of
Contents
(CONTINUED)
M
CALIBRATION/TEST
METHODS
.latroduction..............................
1...
eee
9
*
Calibration
Verification
Procedure
...........................
9
.CalibrationProcedure.....................................
9-10
e
General
Function
Test
......................................
10
+
Procedure
.....................,..........................
10
Ш
DISASSEMBLY/REASSEMBLY
»eCaseRemoval
.............................................
1
eCircuitBoardRemoval......................................
1
+
Circuit
Board
Installation
...................................
1
+
Compressor..........................................,.....
1
+
Solenoid
Valves
...................,...,.,.......,..........
12
«
Case
Reassembly
..........................................
12
画
SPEEIFICATIONS
ων
νε
νο
ων κκ
νε
κκ
18
HE
QUALITY
SPECIFICATIONS
..................................
、.4
E
FIGURE
1:
Diagram
of
Bed
Hook
Use
...........................
15
@
FIGURE
2:
System
—
View
of
Controls
&
Indicators
.............
16
@
FIGURE
3:
Diagram
of
Power
Cord
Storage
.....................
7
M
FIGURE
4:
Pneumatics
Schematic
.............................
18
E
FIGURE
5:
System
—
Exploded
View
...........................
19
Ш
FIGURE
6:
Pneumatics
&
Electrical
Circuits
....................
20
BB
FIGURE
7:
Electrical
Schematic
...............................
2

General
Operating
Instructions
The
SCD
SEQUEL
Compression
System
Model
6325
is
designed
to
apply
sequential
gradient
compression
to
the
lower
limbs
to
prevent
deep
vein
thrombosis
in
patients
at
risk.
M
PRE-OPERATION
CHECK
Before
using
the
SCD
SEQUEL
Compression
System
on
the
patient,
verify
that:
©
The
Tubing
Set
is
properly
connected
to
the
System
and
the
Sleeves.
Both
Sleeves
must
be
connected.
If
the
System
can
only
be
used
on
one
leg
(e.g.,
amputee),
connect
the
second
sleeve
but
leave
it
packaged
in
its
plastic
bag.
©
There
are
no
kinks
or
sharp
bends
in
the
Tubing
Set.
©
The
left
and
right
sides
of
the
System
are
unobstructed
for
free
air
flow.
©
The
System
power
cord
is
plugged
into
a
grounded
AC
receptacle
of
the
correct
voltage.
©
No
flammable
anesthetic
gases
are
present.
Ш
USE
OF
BED
HOOK
The
SCD
SEQUEL
Compression
System
features
a
bed
hook.
This
feature
may
be
used
by
placing
the
System
on
the
footboard
of
the
hospital
bed
with
the
System
facing
away
from
the
patient
(see
Figure
1).
NOTE:
After
placing
the
System
on
the
footboard,
confirm
that
the
System
is
securely
held
by
the
footboard,
and
does
not
interfere
with
patient
care.
E
OPERATION
Turn
the
power
switch
on
(located
on
the
lower
right
side
of
the
System).
The
System
performs
a
self-diagnosis
and
LED
check
upon
start-up.
The
LED
segments
illuminate,
starting
at
the
top
of
the
display
panel.
The
System
then
begins
normal
operation
with
ankle
inflation.
See
Figure
2
for
the
location
of
each
control
and
indicator.
NOTE:
When
the
System
is
turned
on, the
Sleeve
Cooling
will
be
off.
To
activate
Cooling,
press
the
Cooling
Button
while
the
machine
is
operating.
The
Cooling
LED
will
illuminate
when
the
option
is
on.
In
the
case
of
a
fault
condition,
the
alarm
will
sound.
The
appropriate
fault
code
will
be
displayed
and
the
System
will
shut
down.
M
CYCLE
MONITOR
The
SCD
SEQUEL
Compression
System
has
a
Cycle
Monitor,
which
continuously
displays
the
status
of
the
System’s
compression
sequence.
The
Monitor
consists
of
two
back-lit
panels
which,
when
illuminated,
read
INFLATE
and
VENT.
These
represent
the
two
major
divi-
sions
of
one
complete
cycle.
During
operation,
the
INFLATE
and
VENT
lights
will
illuminate
to
indicate
which
part
of
the
cycle
the
system
is
in.
If
the
Sleeve
Cooling
is
off,
the
compressor
will
shut
off
during
the
Vent
Cycle.
AUTOMATIC
PRESSURE
ADJUSTMENT
The
SCD
SEQUEL
Compression
System
features
micro-
processor
controlled
automatic
pressure
adjustment.
This
feature
automatically
sets
the
pressure
to
45mmHg
and
maintains
this
set
pressure
even
when
the
patient
moves
or
changes
position.
After
the
start-up
sequence,
the
System
will
set
itself
and
display
45mmHg.
During
the
first
few
inflation
cycles
the
Controller
will
adjust
itself
to
meet
the
45mmHg.
During
the
initial
setting
period
the
decimal
points
on
each
side
of
the
45
will
display
“.45.”
Within
five
cycles
the
System
will
stabilize
between
43
and
47mmHg
and
the
decimal
points
will
turn
off,
leaving
“45”
displayed.
NOTE:
Even
after
the
initial
45mmHg
pressure
is
reached,
the
System
will
continue
to
make
small
adjustments
in
order
to
maintain
the
45mmHg.
MW
POWER
CORD
STORAGE
Power
cord
storage
for
the
SCD
SEQUEL
Compression
System
is
provided
by
wrapping
the
power
cord
around
the
bed
hook.
This
is
shown
in
Figure
3.
E
SLEEVE
COMPATIBILITY
The
SCD
SEQUEL
Compression
System
is
designed
for
use
with
SCD
Sleeve
Reorder
#5330,
5329,
5480,
5345,
and
5336.

Fault
Conditions
Model
6325
Fault
Messages
”
CODE
FAULT
TYPE
DESCRIPTION
SL
System
Low
There
is
less
than
8mmHg
pressure
at
the
end
of
the
first
inflation
cycle
or
at
the
end
of
two
consecutive
cycles
thereafter.
POTENTIAL
CAUSES
*
Tubing
Set
is
not
properly
connected.
*
There
is
a
leak
inside
the
System.
CORRECTIVE
ACTIONS
©
Ensure
proper
Tubing
connections.
*
Turn
the
System
off
and
restart.
If
the
System
displays
the
same
fault,
the
unit
requires
servicing.
Lo
Low
The
System
is
unable
to
bring
the
pressure
up
above
43mmHg
for
five
consecu-
tive
cycles.
The
pressure
in
the
Sleeves
is
consistently
below
the
set
pressure.
POTENTIAL
CAUSES
*
Tubing
is
not
properly
connected.
*
Sleeves
are too
loose.
CORRECTIVE
ACTIONS
Ensure
proper
Tubing
connections.
*
Apply
Sleeves
so
that
only
two
fingers
fit
snugly
between
each
Sleeve
and
the
patients
leg.
*
Turn
the
System
off
and
restart.
If
the
System
displays
the
same
fault,
the
unit
requires
servicing.
SH
System
High
Pressure
has
exceeded
90mmHg
after
1.2
seconds
of
thigh
cycle:
*
On
first
cycle,
System
will
switch
to
vent
mode
and
adjust
flow
control
appropriately.
*
On
subsequent
cycles,
Systern
will
switch
to
the
fault
mode
and
display
this
code.
(Continued
on
next
page)
(1)
When
the
microprocessor
detects
a
fault
condition,
it
interrupts
the
normal
operation
of the
controller,
deactivates
all
valves,
displays
a
fault
code,
and
sounds
an
audible
alarm.
This
alarm
will
remain
active
until
the
System
power
switch
is
turned
off.

Segue
Fault
Conditions
comwe
CODE
FAULT
TYPE
DESCRIPTION
SH
System
High
(continued)
POTENTIAL
CAUSES
©
Tubing
Set
is
kinked.
©
Sleeves
are
too
tight.
CORRECTIVE
ACTIONS
*
Straighten
out
kinked
Tubing
Set.
+
Apply
the
Sleeves
so
that
two
fingers
can
fit
snugly
between
each
Sleeve
and
the
patients
leg.
e
Tum
the
System
off
and
restart.
If
the
System
displays
the
same
fault,
the
unit
requires
servicing.
HI
High
The
System
is
unable
to
bring
the
pressure
down
below
47mmH
for
five
consecutive
cycles.
The
pressure
in
the
Sleeves
is
consistently
above
the
set
pressure.
POTENTIAL
CAUSES
*
Sleeves
are
too
tight.
CORRECTIVE
ACTIONS
|
+
Apply
the
Sleeves
so
that
two
fingers
can
fit
snugly
between
each
Sleeve
and
the
patients
leg.
*
Turn
the
System
off
and
restart.
If
the
System
displays
the
same
fault,
the
unit
reguires
servicing.
SP
System
Pressure
The
System
has
not
controlled
to
35
-
55mmHg
for
12
consecutive
cycles.
POTENTIAL
CAUSES
+
Sleeves
are
too
tight.
CORRECTIVE
ACTIONS
+
Apply
the
Sleeves
so
that
two
fingers
can
fit
snugly
between
each
Sleeve
and
the
patients
leg.
e
Turn
the
System
off
and
restart.
If
the
System
displays
the
same
fault,
the
unit
reguires
servicing.
di
internal
Diagnostic
One
If
the
solenoid
valve
stays
open,
the
microprocessor
will
detect
this
condition,
shut
the
System
off,
and
display
this
code.
The
System
should
be
returned
to
the
manufacturer
for
repair.

Model
6325
Fault
Conditions
commun
CODE
FAULT
TYPE
DESCRIPTION
d2
Internal
Upon
start-up,
the
microprocessor
performs
diagnostic
tests.
If
the
System
fails
Diagnostic
Two
to
pass
all
of
these
tests,
it
will
not
start
and
this
fault
code
will
be
displayed.
The
System
should
be
returned
to
the
manufacturer
for
repair.
d3,
d4
Internal
These
faults
are
indicated
only
during
special
test
conditions
and
should
not
Diagnostic
Three,
occur
during
normal
operation.
Four
The
System
should
be
returned
to
the
manufacturer
for
repair.
ds
Internal
This
fault
is
indicated
if
the
pump
speed
is
not
what
is
expected.
Diagnostic
Five
The
Compressor
Outlet
Filter
should
be
replaced
or
the
Muffler
should
be
cleaned.
If
fault
reoccurs
after
replacement,
the
System
should
be
returned
to
the
manufacturer
for
repair.
de
internal
This
fault
is
detected
if
the
pressure
exceeds
25
mmHg
after
54
seconds
Diagnostic
Six
of
the
vent
cycle.
The
System
should
be
returned
to
the
manufacturer
for
repair.
d7
-
d9
Internal
These codes
have
not
been
implemented.
Diagnostic
Seven,
They
have
been
reserved
for
future
use.
Eight,
Nine
WATCHDOG
CIRCUIT
If
the
Microprocessor
cannot
continue
normal
function,
the
watchdog
circuit
will
trip.
This
causes
the
System
to
go
into
reset
and
restart
normal
operation.
If
the
cause
of
the
disruption
is
still
present,
the
unit
will
continue
to
attempt
to
reset,
which
will
cause
the
alarm
to
sound
once
every
second.
If
the
cause
of
the
disruption
was
transient,
such
as
a
high
energy
RF
pulse,
the
System
will
restart
with
the
COOLING
OFF.

SCD
9
の
Contraindications
The
SCD
SEQUEL
Compression
System
may
not
be
recommended
for
patients
with
the
following:
1.
Any
local
leg
condition
in
which
sleeves
would
interfere
such
as:
a.
Dermatitis
b.
Vein
ligation
(immediate
postoperative)
c.
Gangrene
d.
Recent
skin
graft
Severe
arteriosclerosis
or
other
ischemic
vascular
disease.
Massive
edema
of
legs
or
pulmonary
edema
from
congestive
heart
failure.
Extreme
deformity
of
leg.
Suspected
existing
deep
venous
thrombosis.
Cautions
1.
2.
3.
WARNING:
Do
not
attempt
to
repair
or
replace
broken
tubing
connectors
as
hazardous
inflation
When
this
device
is
used
in
the
Operating
Room,
keep
Comfort
Cooling
OFF
to
maintain
air
quiescence.
Federal
(U.S.A.)
law
restricts
this
device
to
sale by
or
on
the
order
of
a
physician.
Explosion
Hazard.
Do
not
use
in
the
presence
of
flammable
anesthetics.
of
the
sleeves
may
occur.
This
product
is
a
component
of
Kendall’s
SCD
System
that
is
covered
by
various
U.S.
Patents.
The
use
of
non-Kendall
components
in
place
of
and/or
in
combination
with
this
Kendall
component
may
constitute
infringement
of
Kendall
U.S.
Patent(s).
Kendall
expressly
states
that
no
implied
license
with
respect
to
use
of
non-Kendall
components
is
granted
by
the
sale
or
lease
of
this
component.
Other
Patents
Applied
For

Service
and
Maintenance
Model
6325
M
INTRODUCTION
The
service
technician
should
be
familiar
with
the
oper-
ator's
portion
of
this
manual
and
the
operating
principles
of
the
SCD
SEQUEL
Compression
System.
The
fault
codes
dis-
played
by the
System
are
useful
in
diagnosing
service
prob-
lems.
If
a
System
is
to
be
returned
for
service,
a
description
of
the
operating
conditions
and
the
fault
code
displayed
should
accompany
the
unit.
This
manual
describes
service
procedures
to
the
board
level,
with
an
exploded
view
of
the
System
shown
in
Figure
5.
If
a
component
failure
on
a
circuit
board
is
suspected,
the
unit
should
be
returned
for
service.
It
is
recommended
that
the
instrument
be
returned
with
the
circuit
board
in
place,
as
removal
of
the
board(s)
involves
additional
risk
of
mechanical
damage
and
damage
from
electrostatic
dis-
charge
(ESD).
E
WARRANTY
AND
FACTORY
SERVICE
The
Kendall
Healthcare
Products
Company
warrants
that
your
SCD
SEOUEL
Compression
System
is
free
from
defec-
tive
material
and
workmanship.
Our
obligation
under
this
warranty
is
limited
to
the
repair
of
Controllers
returned
to
the
service
address
indicated
at
right,
transportation
charges
prepaid,
within
one
year
of
delivery
to
the
original
pur-
chaser.
Specifically,
we
agree
to
service
and/or
adjust
any
Controller
as
reguired
if
returned
for
that
purpose,
and
to
replace
and
repair
any
part
which,
upon
our
examination,
is
proven
to
have
been
defective.
This
warranty
does
not
apply
to
the
Tubing
Set
or
the
individual
disposable
leg
Sleeves,
or
to
eguipment
damaged
through
shipping
,
tam-
pering,
negligence,
or
misuse,
including
liquid
immersion,
autoclaving,
or
ETO
sterilization.
This
limited
warranty
does
not
cover,
and
is
intended
to
exclude,
any
and
alt
liability
on the
part
of
the
Company,
whether
this
limited
warranty
or
any
warranty
implied
by
law,
for
any
indirect
or
consequential
damages
for
breach
hereaf
or
thereof.
Except
as
expressly
provided
above
in
the
limited
warranty,
the
Company
hereby
negates
and
disclaims
all
express
and
implied
warranties,
including
the
warranties
or
merchantability
and
fitness
for
a
particular
purpose.
Some
states
do
not
allow
the
exclusion
or
limitation
of
inci-
dental
or
consequential
damages,
so
the
foregoing
limita-
tion
or
exclusion
regarding
damages
may
not
apply.
Also,
this
Limited
Warranty
gives
you
specific
legal
rights,
and
you
may
also
have
other
rights
which
may
vary
from
state
to
state.
This
service
manual
is
intended
for
use
as
a
guide
to
technically-qualified
personnel
when
evaluating
instrument
malfunctions.
It
is
not
to
be
construed
as
authorization
to
perform
warranty
repairs.
Unauthorized
service
will
void
the
warranty.
Controllers
requiring
warranty
repairs
must
be
shipped,
pre-
paid
and
insured,
to
the
Service
Center.
The
Kendall
Health-
care
Products
Company
maintains
service
facilities
with
the
capability
to
promptly
repair
the
SCD
SEQUEL
Compression
System.
In
the
event
of
any
service
mainte-
nance
problem,
contact
Customer
Service.
The
toll-free
number
is
(800)
765-4324.
If
the
instrument
is
to
be
returned
for
service,
please
call
the
above
number
to
obtain
an
authorization
return
number
and
send
it
prepaid
and
insured,
in
the
original
carton,
to:
TYCO
HEALTHCARE
GROUP
LP
98.6
Faichney
Drive
Watertown, NY
13601
E
MAINTENANCE
CAUTION:
Unplug
the
System
before
filter
maintenance.
There
are
two
filters
used
in
the
SCD
SEOUEL
Compression
System
—
a
fan
filter
and
a
compressor
outlet
filter.
It
is
very
important
that
the
fan
filter
be
cleaned
frequently
to
ensure
continued
trouble-free
operation.
The
Controller
should
never
be
run
without
the fan
filter
in
place.
To
clean
the
fan
filter,
gently
pry the
fan
cover
off
the
left
side
of
the
System.
Carefully
remove
the
filter
from
the
fan
cover
and
brush
loose
dust
and
lint
from
the
surface.
Wash
the
filter
in
a
mild
detergent
solution,
dry
thoroughly,
and
reinstall.
(Continued
on
page
7)

Service
and
Maintenance
(60000
Mi
MAINTENANCE
(Continued)
Removal
of
the
front
cabinet
exposes
the
compressor
filter.
The
filter
is
mounted
below
the
compressor.
The
filter
is
designed
to
exceed
the
life
of
the
System.
In
the
event
the
filter
becomes
loaded,
flow
will
become
obstructed
and
the
System
will
repeatedly
trigger
a
d5,
LO,
or
SL
alarm.
When
this
occurs
replace
the
filter
as
follows:
Remove
the
inlet
and
outlet
tube
from
the
filter.
Take
the
filter
off
the
Nylon
hook
pad,
replace
with
a
new
filter,
and
then
reattach
the
tubing.
Make
sure
the
tube
exiting
the
compressor
is
connected
to
the
inlet
side
of
the
filter.
Direction
of
flow
is
from
the
compressor
to
filter
to
valves
{see
Figure
4),
and
is
identified
on
the
filter
housing
{avoid
kinking
any
tubing).
Carefully
replace
front
cabinet.
№
VENTILATION
Obstruction
of
the
left
fan
cover and
right
vents
should
be
avoided.
Free
flow
of
air
is
necessary
to
prevent
overheating
and
premature
component
faiture.
圖
FUSES
CAUTION:
Unplug
the
System
before
removing
the
fuse.
A
blown
fuse
should
only
be
replaced
with
a
3AG
Slo
Blo
fuse,
0.75
ampere.
If
a
fuse
blows
a
second
time,
it
should
be
presumed
that
the
instrument
is
defective
and
requires
further
service.
E
ELECTRICAL
SAFETY
CAUTION:
Be
sure
the
System
is
disconnecteď
from
the
power
source
before
removing
the
front
cover.
A
potential
SHOCK
HAZARD
exists
when
the
front
cover
is
removed
even
with
the
unit
turned
off.
Periodic
inspection
should
include
a
resistance
check
from
the
power
plug
grounding
prong
to
the
metal
chassis
inside
the
instrument.
The
chassis
can
be
accessed
by
removing
the
fan
cover.
This
resistance
should
not
exceed
0.1
ohm.
If
ground
resistance
exceeds
this
value
or
the
insulation
integrity
of
the
unit
has
been
compromised
through
mechanical
damage,
the
System
should
be
returned
to
the
Service
Center
for
testing
and
repair.
M
CLEANING
The
System
cabinet
can
be
cleaned
as
necessary
with
a
soft
cloth
dampened
with
water.
If
necessary,
a
mild
disinfectant
and/or
detergent
can
be
used;
excess
fluid
should
be
avoided.
The
System
should
be
wiped
with
a
clean,
dry
cloth
afterward.
Do
not
immerse
in
water.
The
SCD
SEQUEL
Compression
System
cannot
be
effec-
tively
sterilized
by
liquid
immersion,
autoclaving,
or
ETO
sterilization,
as
irreparable
damage
to
the
System
will
occur.
MB
ELECTRICAL/ELECTRONICS
DESCRIPTION
The
electronics
of
the
controller
are
located
on
two
PC
boards
(see
Figures
6
and
7).
The
first
board
can
be
found
beneath
the
membrane
switch
panel
in
the
front
case
cover.
There
is
no
high
voltage
on
the
PC
boards.
The
CPU
board
contains
the
LEDs
and
associated
electronics
which
display
information
about
the
state
of
the
System
during
normal
operation,
as
well
as
in
the
fault
mode
or
the
var-
ious
calibration
and
diagnostic
modes.
This
board
also
contains
the
pressure
transducer,
the
microprocessor,
the
buzzer,
and
the
valve
control.
The
second
board
is
the
Power
board.
It
contains
voltage
regulation
circuits,
analog
conditioning
circuits,
and
motor
control
circuitry.
Operator
input
is
provided
through
the
membrane
switch
panel.
Component
or
System
failure
will
result
in
de-
energizing
the
valves
and
the
compressor.
This
will
release
pressure
from
the
attached
leg
sleeves.

Model
6325
Service
and
Maintenance
comue
Mi
PNEUMATIC
DESCRIPTION
The
rotary
vane
compressor,
driven
by
a
low-noise
brush-
less
DC
motor,
runs
only
when
needed.
Air
can
be
directed
from
the
compressor
to
the
sleeve
cooling
chambers
during
the
60-second
vent
phase
by
pressing
the
Sleeve
Cooling
button
on
the
membrane
switch
panel.
This
causes
the
compressor
to
turn
on
and
the
solenoid
valve
V
to
be
energized
during
the
VENT
phase
(See
Figure
4).
Air
flow
through
the
solenoid
manifold
assembly
is
con-
trolled
by the
motor
control
circuitry
working
with
the
brushless
DC
motor.
From
the
manifold
assembly,
the
air
is
gated
to
the
sleeve
chambers
during
the
11-second
com-
pression
phase
by
energizing
solenoid
valves
A,C,
and
T
in
sequence.
At
the
end
of
this
period,
all
three
solenoid
valves
are
de-energized
simultaneously
which
disconnects
the
compressor
from
the
sleeves
and
allows
the
valves
to
release
sleeve
pressure
to
the
atmosphere.
The
ankle
pres-
sure
transducer
monitors
the
pressure
in
the
ankle
portion
of
the circuit
and
provides
data
input
to
the
micro-
processor
for
feedback
control
(see
Figure
6).
E
PORT
IDENTITY
The
Tubing
Set
connecting
the
Sleeves
to
the
System
is
attached
to
the
fitting
on
the
rear
of
the
System
in
the
upper
left
hand
corner
(as
viewed
from
behind
the
System).
Inside
this fitting
are
four
ports
which
are,
from
left
to
right:
ANKLE,
VENT
(COOLING),
CALF,
and
THIGH
(see
Figure
4).
M
SYSTEM
START-UP
ROUTINE
@
PHASE
I:
When
the
System
is
turned
on,
a
series
of
tests
is
com-
menced
lasting
three
seconds'2):
1.
All
LEDs
illuminate
and
the
beeper
sounds
for
0.5
seconds.
2.
The two
seven-segment
LEDs
illuminate
(displaying
“.88.")
and
then they
go
off.
3.
The
COOLING
LED
illuminates.
The
beeper
sounds
briefly
and
the
COOLING
LED
goes
off.
4.
The
Cycle
Monitor
LEDs
(INFLATE
and
VENT)
illuminate
then
go
off.
+
PHASE
Il:
The
45mmHg
set
pressure
is
displayed.
During
this
phase,
the
compressor
flow
is
adjusted
to
a
predeter-
mined
start
value.
®
PHASE
Ill:
The
normal
operating
cycle
begins
with
ankle
inflation.
M
PRESSURE
MONITORING
The
LED
display
that
is
used
to
display
the
45
mmHg
set
pressure
and
the
fault
codes
can
also
be
used
to
display
the
actual
pressure.
Press
the
left
end
of
the
white
horizontal
line
of
the
contro!
panel
for
two
seconds.
This
will
cause
the
display
to
change
to
Pressure
Monitor
Mode.
The
System
will
stay
in
this
mode
throughout
the
compression
and
deflation
portions
of
the
current
cycle
(or
the
next
compression
cycle
if
the
button
is
pressed
during
the
VENT
portion
of
the
cycle).
The
last
reading
of
the
compression
portion
of
the
cycle
will
be
displayed
for
30
seconds.
At
the
beginning
of
the
next
inflation
cycle
the
display
will
revert
to
displaying
45
mmHg
pressure.
When
the
System
is
in
the
Pressure
Monitor
Mode,
a
decimal
point
will
flash
on
either
side
of
the
7-segment
display,
indicating
that
the
actual
pressure
is
being
displayed.
NOTE:
While
in
the
Pressure
Monitor
Mode,
it
is
normal
for
the
displayed
pressure
to
change
rapidly
during
the
compression
portion
of
the
cycle.
(2)
Detection
of
inoperative
LEDs
and
the
audible
alarm
function
is
the
user's
responsibility.
The
start-up
routine
also
tests
microprocessor
function
and
system
memory.
If
a
fault
condition
is
detected,
either
a
fault
code
will
be
displayed
or
the
unit
will
repeatedly
go
into
reset
which
will
result
in
a
long
beep
at
a
1
Hz
rate.

Calibration/Test
Methods
See
Disassembly/Reassembly
procedure
before
testing.
M
INTRODUCTION
The
SCD
SEQUEL
Compression
System
has
a
special
test
mode
for
pressure
transducer
Calibration
Verification,
pres-
sure
transducer
Calibration,
and
a
general
function
test
that
can
verify
microprocessor,
solenoid
valve,
and
pneumatic
functionality.
1,
Factory
calibration
certification
is
void
if
the
unit
is
opened.
2.
The
unit
should
only
be
opened
by
persons
trained
in
Electrostatic
Discharge
prevention
methods
and
equipment.
@
CALIBRATION
VERIFICATION
PROCEDURE
Equipment
Required:
A
regulated
air
source
with
a
con-
stant
output
pressure
set
at
45.0
mmHg
+
0.2
mmHg.
NOTE:
The
pressure
transducers
used
in
the
SCD
SEQUEL
Compression
System
are
state-of-the-art,
highly
precise,
and
virtually
drift-free
devices.
Complete
the
following
steps
to
verify
System
Calibration:
1.
Unplug
the
System.
2.
Turn
the
System
on
its
face
and
remove
the
five
screws
located
on
the
rear
cabinet.
3.
Carefully
turn the
System
right
side
up
so
the front
cab-
inet
is
facing
you.
4.
Gently
lower
the front
cabinet
toward
you
until
the
components
inside
are
accessible.
(Make
sure
the
var-
ious
electrical
and
pneumatic
connections
between
the
CPU
Board
and
the
rest
of
the
unit
are
not
dis-
lodged.)
5.
Carefully
disconnect
the
Tubing
that
is
attached
to
the
Pressure
Transducer
on
the
Control
Board.
6.
Plug
the
System
into
a
wall
outlet.
CAUTION:
ELECTRICAL
SHOCK
HAZARD.
Do
not
touch
the
electrical
components
on
the
chassis.
The
PC
Boards
are
not
at
line
voltage.
7.
Place
the
System
in
the
Calibration
Verification
Mode
by
first
pressing
the
Hidden
and
Cooling
Buttons
simultaneously
while
turning
on
the
System.
(The
Hidden
Button
is
located
on
the
control
panel
at
the
very
left
end
of
the
horizontal
white
line.)
This
places
the
System
in
the
Test
Access
Mode.
The
display
will
show
“01.”
Press
the
Hidden
Button
until
“04”
appears
in
the
System
display.
NOTE:
Test
Access
Modes
“01,”
“02,”
“05,”
and
“07”
require
dedicated
manufacturing
test
equipment
and
are
not
intended
for
customer
use.
Do
not
use.
8.
Press
the
Cooling
Button
to
initiate
the
Calibration
Verification
Mode.
(The
System
will
continuously
dis-
play
the
pressure
sensed
at
the
pressure
transducer.)
9.
Connect
the
calibrated
air
pressure
source
to
the
Pressure
Transducer
using
a
piece
of
Tubing.
10.
Apply
45.0
+
0.2
mmHg.
Read
the
pressure
on
the
dis-
play.
11.
Disconnect
Tubing.
With
no
pressure
applied
read
the
display.
12.
The
display
readings
should
be
compared
to
the
pres-
sure
applied
to
verify
proper
calibration.
13.
To
exit
Calibration
Verification
Mode,
tum
off
the
System.
14.
If
the
System
requires
calibration,
proceed
to
the
Calibration
Procedure
step
7.
15.
Unplug
the
Controller.
16.
Reconnect
the
Tubing.
Particular
care
should
be
taken
to
avoid
stressing
or
kinking
the
tubing.
17.
Carefully
replace
the front
cabinet
and
turn
the
Controller
on
its
face.
Check
for
proper
placement
of
feet,
foam,
and
tubing.
18.
Refasten
the
five
screws.
(Do
not
over-tighten.)
@
CALIBRATION
PROCEDURE
Equipment
Required:
A
regulated
air
source
with
a
con-
stant
output
pressure
set
at
45.0
mmHg
+
0.2
mmHg.
NOTE:
The
pressure
transducers
used
in
the
SCD
SEQUEL
Compression
System
are
state-of-the-art,
highly
precise,
and
virtually
drift-free
devices.
It
is
not
expected
to
require
recalibration
for
several
years
or
more
of
continuous
use.
Complete
the
following
steps
to
Calibrate
the
System:
1.
Unplug
the
System.
2.
Turn
the
System
on
its
face
and
remove
the
five
screws
located
on
the
rear
cabinet.
3.
Carefully
turn the
System
right
side
up
so
the
front
cab-
inet
is
facing
you.
4.
Gently
rotate
the
front
cabinet
toward
you
until
the
components
inside
are
accessible.
(Make
sure
the
var-
ious
electrical
and
pneumatic
connections
between
the
CPU
Board
and
the
rest
of
the
unit
are
not
dis-
lodged.)
5.
Carefully
disconnect
the
Tubing
that
is
attached
to
the
Pressure
Transducer
on
the
Control
Board.
(Continued
on
page
10)

Calibration/Test
Methods
соли
Model
6325
Ἡ
CALIBRATION
PROCEDURE
(Continued)
6.
11.
12.
13.
14.
15.
16.
17.
Plug
the
System
into
a
wall
outlet.
CAUTION:
ELECTRICAL
SHOCK
HAZARD.
Do
not
touch
the
electrical
components
on
the
chassis.
The
PC
Boards
are
not
at
line
voltage.
Place
the
System
in
the
Calibration
Mode
by
pressing
the
Hidden
and
Cooling
Buttons
simultaneously
while
turning
on
the
System.
(The
Hidden
Button
is
located
on
the
control
panel
at
the
very
left
end
of
the
hori-
zontal
white
line.)
This
places
the
System
in
the
Test
Access
Mode.
The
display
will
show
“01.”
Press
the
Hidden
Button
until
“03”
appears
in
the
System
display.
NOTE:
Test
Access
Modes
“01,”
“02,”
“05,”
and
“07”
require
dedicated
manufacturing
test
equipment
and
are
not
intended
for
customer
use.
Do
not
use.
.
Press
the
Cooling
Button
to
initiate
the
Calibration
Mode.
The
System
will
now
display
the
pressure
in
the
line.
.
The
INFLATE
LED
will
be
illuminated.
10.
Connect
the
calibrated
air
pressure
source
to
the
Pressure
Transducer
using
a
piece
of
tubing.
With
45.0
+
0.2mmHg
applied,
press
the
Cooling
Button.
(This
will
record
the
45mmHg
set
point
in
System
software.)
The
Vent
LED
will
be
illuminated.
Disconnect
the
tubing.
With
no
pressure
applied,
press
the
Cooling
Button.
(This
will
record
the
OmmHg
set
point
in
System
software.)
Turn
the
System
off.
Reconnect
the
tubing.
Particular
care
should
be
taken
to
avoid
stressing
or
kinking
the
tubing.
Carefully
replace
the
front
cabinet
and
turn the
System
on
its
face.
Check
for
proper
placement
of feet,
foam,
and
tubing.
Refasten
the
five
screws.
(Do
not
over-tighten.)
CAUTION:
If
the
Cooling
Button
is
pressed
inadver-
tently
during
the
Calibration
other
than
when
specified
in
steps
8,
11,
and
13,
the
System
must
be
recalibrated.
IB
GENERAL
FUNCTION
TEST
The
following
describes
a
General
Function
Test
which
can
be
used
to
exercise
the
System.
This
test
is
not
part
of
the
normal
testing
and
maintenance
routine,
but
can
be
used
in
the
event
of
disassembly
to
verify:
The
correct
wiring
and
connection
of
the
solenoid
valves.
The
correct
orientation
and
connection
of
the
pneu-
matic
circuit.
There
is
no
blockage
or
leaking
in
the
pneumatic
circuit.
M
PROCEDURE
1.
Press
the
Hidden
and
Cooling
Buttons
simultaneously
while
turning
on
the
System.
(The
Hidden
Button
is
located
on
the
control
panel
at
the
very
left
end
of
the
white
horizontal
line.)
This
places
the
System
in
the
Test
Access
Mode.
The
display
will
show
“01.”
Press
the
Hidden
Button
until
“06”
is
on
the
System
display.
NOTE:
Test
Access
Modes
“01,”
“02,”
“05,”
and
“07”
reguire
dedicated
manufacturing
test
eguipment
and
are
not
intended
for
customer
use.
Do
not
use.
During
the
procedure
you
will
gently
hold
a
finger
near
each
air
outlet
port
on
the
back
of
the
System
to
verify
air
flow
is
present
for
each
cycle
at
the
correct
port.
The
order
of
the
ports
from
left
to
right,
looking
at
the
back
of
the
System,
is
Ankle,
Vent
(Cooling),
Calf,
and
Thigh
(see
Figure
4).
The
flow
will
gradually
increase
for
the
Ankle,
Calf,
and
Thigh.
The
flow
will
start
a
maximum
and
stop
abruptly
for
the
Vent.
To
initiate
the
test,
press
the
Cooling
Button.
*
The
ankle
solenoid
valve
will
open,
the
System
will
beep,
and
the
compressor
will
run
at
minimum
flow
for
five
seconds.
*
The
compressor
will
ramp
up
to
maximum
flow
in
two
seconds.
4
The
System
will
beep
and
remain
at
full
speed
for
five
seconds.
4
The
ankle
valve
will
close.
*
The
sequence
will
be
repeated
for
the
calf
and
thigh
portions.
9
At
the
end
of
the
thigh
portion,
the
thigh
valve
will
shut
and
the
vent
valve
will
open.
The
vent
valve
will
go
immediately
into
maximum
flow.
This
will
last for
five
seconds.
+
When
the
vent
circuit
has
completed
its
cycle,
the
System
will
return
to
the
Test
Access
Mode.

Sige]
Disassembly
/Reassembly
M
CASE
REMOVAL
CAUTION:
SHOCK
HAZARD.
Disconnect
line
cord
from
the
power
supply.
CAUTION:
Disassembly
of
the
case
and
particularly
the
disconnection
of
pneumatic
and/or
electronic
components
can
disrupt
the
function
and
calibration
of
the
instrument.
It
is
strongly
recommended
that
the
Calibration
and
General
Valve
Function
Tests
be
performed
upon
reassembly.
°
ESD
WARNING:
Unit
should
only
be
opened
by
person
trained
in
ESD
methods
and
equipment.
Remove
the
case
components
in
the
following
sequence:
(see
Figures
5
and
6
located
in
the
Appendix).
1.
Remove
front
cover.
To
remove
the front
cover,
place
the
System
on
its
face
and
remove
the
five
screws
located
on
the
back
of
the
unit.
2.
With
the
System
standing
upright
and
facing
you,
care-
fully
rotate
the
front
cover
toward
you.
3.
Note
the
location
and
orientation
of
the
connectors
from
the
rear
case
components
on
the
CPU
Board.
Remove
the
connectors
by
gently
pulling
on
each
indi-
vidual
connector.
Some
connectors
have
locking
clips
that
must
be
gently
squeezed
to
disengage.
4.
Very
gently
disconnect
the
Tubing
at
the
pressure
trans-
ducers
on
the
CPU
Board.
BE
CIRCUIT
BOARD
REMOVAL
The
CPU
Board
is
located
on
the
front
case.
The
Power
Board
is
located
on
the
rear
case
directly
above
the
com-
pressor.
CAUTION:
Use
a
grounded
strap
when
handling
any
elec-
tronic
components.
1.
Note
the
location
and
orientation
of
all
the
pneumatic
and
electrical
connections
to
the
PC
Boards
(see
Figure
7).
(Label
cables
and
tubing
if
necessary.)
2.
Disconnect
all
the
pneumatic
and
electronic
leads
to
the
PC
Boards.
3.
Remove
the
six
mounting
screws
located
on
the
CPU
Board.
4.
Carefully
slide
the
Power
Board
out
of
the
System
from
between
the
compressor
and
top
of
rear
cabinet.
CAUTION:
Components
on
the
PC
Boards
should
not
be
removed
or
replaced.
If
a
circuit
board
is
suspected
of
being
faulty
the
unit
should
be
returned
for
repair
by
calling
the
toll-free
service
number
(800)
765-4324.
M
CIRCUIT
BOARD
INSTALLATION
CAUTION:
Use
a
grounded
strap
when
handling
any
elec-
tronic
components.
1.
Carefully
align
the
CPU
Board
with
the
mounting
bosses
and
fasten
the
six
mounting
screws.
2.
Carefully
slide
the
Power
Board
into
the
System
between
the
compressor
and
top
of
rear
cabinet.
E
COMPRESSOR
The
Rotary
Vane
Compressor
is
driven
by
a
state-of-the-art
brushless
DC
motor.
Flow
is
controlled
by
adjusting
the
rotational
speed
of
the
motor.
The
higher
the
speed,
the
greater
the
flow.
The
motor
control
circuit
which
controls
the
motor
speed
is
located
on
the
Power
Board
and
is
not
integral
to
the
motor
on
the
Compressor.
NOTE:
The
Compressor
used
in
the
instrument
is
not
a
user
serviceable
component.
Special
jigs
and
fixtures
are
required
to
ensure
proper
alignment
during
reassembly.
Do
not
disassemble.
Do
not
oil.

Disassembly/Reassembly
como)
Model
6325
EE
SOLENOID
VALVES
+
All
four
solenoid
valves
are
arranged
in
a
2-by-2
single
manifold.
The
solenoid
vaive
that
connects
the
pump
to
the
Sleeve
Cooling
(Vent)
Port
is
a
2-way,
normally
closed
type.
When
energized,
it
passes
air
to
the
port;
when
de-energized,
flow
is
blocked.
The
solenoid
valve
connecting
air
flow
from
the
flow
control
valve
to
the
ANKLE,
CALF,
and
THIGH
ports
are
of
the
3-way,
normally
closed
type.
When
energized,
these
valves
pass
air
to
their
respective
Sleeve
ports.
When
de-energized,
pump
air
is
blocked
and
air
pres-
sure
in
the
Sleeves
is
released
through
the
port
on
the
top
of
that
particular
valve.
The
coils
on
all
four
valves
are
removable
by
removing
the
nut
at
the top
of
the
valve.
Any
debris
in
the
air
lines
could
cause
mechanical
mal-
function
of
the
solenoid
valves.
If
this
is
suspected,
the
valve
tubing
should
be
disconnected
and
the
valve
flushed
with
filtered
compressed
air
through
the
port
while
energizing
the
valves
respectively
from
a
24-volt
supply.
Filtration
of
incoming
air
guards
against
debris
entering
the
compressor.
If
debris
is
present,
it
may
indicate:
a.
Failure
to
clean
the
filters
when
needed.
b.
Physical
breakdown
of
some
component
of
the
pneumatic
circuit,
for
example,
degradation
of
a
section
of
tubing.
E
UNIT
REASSEMBLYG)
1.
Place
the
rear
case
with
the
complete
chassis
installed,
on
the
bench
in
front
of
you
with
the
front
of
the
System
towards
you.
2.
Place
the
front
case
on
its
side
with
the
bottom
edge
of
the
front
case
against
the
rear
base.
The
front
case
should
be
toward
you
with
the
rear
case
behind.
@
Reconnect
all
of
the
electrical
connections
in
the
proper
location
and
orientation
of
the
leads
from
the
System
components.
Replace
the
connectors
by
gently
pushing
straight
in.
@
Reconnect
the
Tubing
on
the
pressure
transducers.
Particular
care
should
be
taken
to
avoid
stressing
the
fittings
on
the
pressure
transducer.
3.
Rotate
the
front
case
and
place
it
over
the
rear
case
base.
Carefully
align
the
two
halves,
making
sure
the
rubber
feet
are
in
place
and aligned
properly.
Turn
the
closed
System
on
its
face
and
reinstall
the
five
screws.
Turn
the
unit
right
side
up
and
perform
the
General
Function
Test.
(3)
NOTE:
It
is
recommended,
before
case
reassembly,
that
any
disassembly
which
involves
the
electrical
or
pneumatic
circuit
requires
a
recalibration.
(See
Calibration
Procedure.)

Sequel
Specifications
M
MODEL
#6325
UL
Listing
....,...........
Built
to
UL
544
Standard
Listing
KFBQ
File
#E116754
Compression
Type
.......
Sequential,
Gradient
Pressure
Compression
Cycle
.........
11
Seconds
Compression,
60
Seconds
Decompression
Set
Pressure
....................,.....
45
mmHg
Bed
Hook
.................................
Yes
Power
Cord
Storage
.................,.......
Yes
Audible/Visual
Alarms
...
.Low
Pressure,
High
Pressure,
internal
Electronics
Malfunction
Sleeve
Cooling
Function
......................
Yes
Cycle
Monitor
.............................
Yes
Power
Cord
...................
Hospital
Grade
Plug
Controller
Height
.......................
6
inches
Controller
Width
.....................
103/4
inches
Controller
Depth
.....................
41/2
inches
Weight
...............................4
8.9
lbs.
Power
Requirements
.........
115
VAC,
86
VA,
60
Hz
ShippinggUnit
.............................
Each
Shipping
Case
Dimensions
.......
Length
-
13
3/8
inches
Width
-
11
inches
Depth
-
103/8
inches
Shipping
Weight
........................
12.2
lbs.
Tubing
Set
............................
Included
Operating
Instructions
..............
Attached
to
unit
under handle
Operation
&
Service
Manual
..............
Included

Quality
Specifications
Model
6325
M
MODEL
#6325
The
following
information
is
a
brief
description
of
the
tests
performed
on
all
SCD
SEQUEL
Compression
Systems
to
ensure
compliance
to
the
prescribed
System
quality
specifications
and
does
not
include
clinical
performance.
1.0
2.0
Electrical
Tests
1.1
Power
and Control
Printed
Wiring
Board
Assemblies
The
boards
are
tested
on
a
board
test
fixture
for
the
following
functions:
1.1.1.
Individual
component
integrity
1.1.2
Component
installation
1.1.3
Board
calibration
1.2
Controller
1.2.1
Correct
sequential
activation
of
Inflate
and
Vent
LEDs
Cooling
Button
and
LED
function
Dielectric
Voltage
Withstand
Test
(Hi-Pot)
Power
Switch
function
Indicator
lights
and
button
functions
1.2.2
1.2.3
1.2.4
1.2.5
Pneumatic
Functional
Tests
2.1
Controller
calibration
at
45
+
2mmHg
2.2
Airflow
supporting
compression
in
the
ankle,
calf,
and
thigh
System
high
pressure
fault
detection
System
low
pressure
fault
detection
2.3
2.4
3.0
4.0
5.0
Burn-In
Requirements
While
operating
in
the
Burn-In
Mode,
the
System
is
run
for
48
hours
to
enhance
unit
reliability.
Visual
Attributes
Inspection
4.1
Case
parts
appearance
and
alignment
4.2
Labeling
Final
Product
Inspection
Final
release
criteria utilizes
100%
testing
of
Systems
for
the
following:
5.1
Calibration
5.2
Air
flow
levels
5.3
Cosmetic
defects

Sequel
15
Figure
1
Diagram
of
Bed
Hook
Use

16)
Figure
2
Model
6325
System
—
View
of
Controls
&
Indicators
SET
PRESSURE/
FAULT
CODE
INDICATOR
INFLATE
LED
ASSESS
LED
HIDDEN
BUTTON
COOLING
LED
COOLING
ON/OFF
BUTTON
POWER
SWITCH
Table of contents
Other kENDALL Medical Equipment manuals
Popular Medical Equipment manuals by other brands

EASTMAN
EASTMAN Kodak X-Omat 1000 installation instructions

OSCILLA
OSCILLA TSM-300 Instructions for use

Shuttle
Shuttle 2000-1 manual

Integra
Integra MAYFIELD Head Clamp A2079 Instructions for use

Cardinal Health
Cardinal Health Kangaroo Joey user guide

Sunoptic Surgical
Sunoptic Surgical Titan X300 Operator's manual