KERN SFB 10K1HIP User manual

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating instructions
Stainless steel scale
KERN SFB
Version 1.3
02/2011
GB
SFB-BA-e-1113

GB KERN SFB
Version 1.3 02/2011
Operating instructions stainless steel scale
Table of Contents
1Technical Data ................................................................................................... 4
2Appliance overview........................................................................................... 8
2.1 Keyboard overview .......................................................................................................9
2.1.1 Numeric input via navigation keys ................................................................................... 10
2.2 Overview of displays...................................................................................................10
3Basic Information (General) ........................................................................... 11
3.1 Proper use ..................................................................................................................11
3.2 Improper Use..............................................................................................................11
3.3 Warranty .....................................................................................................................11
3.4 Monitoring of Test Resources.....................................................................................12
4Basic Safety Precautions................................................................................ 12
4.1 Pay attention to the instructions in the Operation Manual ..........................................12
4.2 Personnel training.......................................................................................................12
5Transportation & Storage ............................................................................... 12
5.1 Testing upon acceptance ...........................................................................................12
5.2 Packaging / return transport .......................................................................................12
6Unpacking and implantation .......................................................................... 13
6.1 Installation Site, Location of Use ................................................................................13
6.2 Unpacking/implantation ..............................................................................................13
6.2.1 Tripod............................................................................................................................... 15
6.3 Mains connection........................................................................................................18
6.4 Rechargeable battery operation .................................................................................18
6.5 Adjustment..................................................................................................................18
6.6 Linearisation (not verifiable models)...........................................................................21
6.7 Verification..................................................................................................................22
SFB-BA-e-1013
2

SFB-BA-e-1013 3
7Operation ......................................................................................................... 23
7.1 Start-up.......................................................................................................................23
7.2 Switching Off ..............................................................................................................23
7.3 Zeroing .......................................................................................................................23
7.4 Simple weighing .........................................................................................................23
7.5 Switch-over weighing unit (only not verifiable models)...............................................24
7.6 Weighing with tare ......................................................................................................25
7.7 Weighing with tolerance range ...................................................................................25
7.8 Manual totalizing.........................................................................................................28
7.9 Automatic adding-up...................................................................................................30
7.10 Animal weighing .........................................................................................................31
7.11 Lock keyboard ............................................................................................................31
7.12 Display background illumination .................................................................................31
7.13 Automatic switch-off function “AUTO OFF“ ...................................................................32
8Menu................................................................................................................. 33
9Service, maintenance, disposal ..................................................................... 40
9.1 Cleaning .....................................................................................................................40
9.2 Service, maintenance .................................................................................................40
9.3 Disposal......................................................................................................................40
9.4 Error messages ..........................................................................................................40
10 Data output RS 232C (optional) .................................................................. 41
10.1 Technical Data............................................................................................................41
10.2 Printer mode ...............................................................................................................42
10.3 Continuous data output ..............................................................................................42
11 Instant help................................................................................................... 43

SFB-BA-e-1013
4
1 Technical Data
KERN SFB 10K1HIP SFB 15K5HIPM SFB 20K2HIP
Readability (d) 1 g 5 g 2 g
Weighing range (max) 10 kg 15 kg 20 kg
Minimum load (Min) - 100 g -
Verification value (e) - 5 g -
Verification class - III -
Reproducibility 1 g 5 g 2 g
Linearity ± 1 g ± 5 g ± 2 g
Recommended
adjustment weight, not
added (class)
10 kg
(M1)
15 kg
(M1)
20 kg
(M1)
Warm-up time 30 minutes 10 minutes 30 minutes
Stabilization time (typical) 2-3 sec.
Weighing unit kg
Auto Off Available options 5, 15 min.
Ambient temperature 0°C – 40°C -10°C – 40°C 0°C – 40°C
Moist environment 0 % - 95 % (non-condensing)
Input voltage 220 V – 240 V, 50 Hz
Electric Supply
Power pack secondary voltage 9V, 800mA
Service life background light on for 70 h
Service life background light off 90 h
Rechargeable battery
(Standard)
Charge time 12 h
Dimensions display unit
(B x D x H) mm 266 x 165 x 96
Weighing surface mm 300 x 240
IP protection IP 67 as per DIN 60529
Interface RS 232 optional

SFB-BA-e-1013 5
KERN SFB 30K10HIPM SFB 50K5HIP SFB 50K5LHIP
Readability (d) 10 g 5 g 5 g
Weighing range (max) 30 kg 50 kg 50 kg
Minimum load (Min) 200 g - -
Verification value (e) 10 g - -
Verification class III - -
Reproducibility 10 g 5 g 5 g
Linearity ± 10 g ± 5 g ± 5 g
Recommended
adjustment weight, not
added (class)
30 kg
(M1)
50 kg
(M1)
50 kg
(M1)
Warm-up time 10 minutes 30 minutes 30 minutes
Stabilization time (typical) 2-3 sec.
Weighing unit kg
Auto Off Available options 5, 15 min.
Ambient temperature -10°C – 40°C 0°C – 40°C 0°C – 40°C
Moist environment 0 % - 95 % (non-condensing)
Input voltage 220 V – 240 V, 50 Hz
Electric Supply
Power pack secondary voltage 9V, 800mA
Service life background light on for 70 h
Service life background light off 90 h
Rechargeable battery
(Standard)
Charge time 12 h
Dimensions display unit
(B x D x H) mm 266 x 165 x 96
Weighing surface mm 300 x 240 400 x 300
IP protection IP 67 as per DIN 60529
Interface optional RS232

SFB-BA-e-1013
6
KERN SFB 60K20HIPM SFB 60K20LHIPM
Readability (d) 20 g 20 g
Weighing range (max) 60 kg 60 kg
Minimum load (Min) 400 g 400 g
Verification value (e) 20 g 20 g
Verification class III III
Reproducibility 20 g 20 g
Linearity ± 20 g ± 20 g
Recommended
adjustment weight, not
added (class)
60 kg
(M1)
60 kg
(M1)
Warm-up time 10 minutes
Stabilization time (typical) 2-3 sec.
Weighing unit kg
Auto Off Available options 5, 15 min.
Ambient temperature -10°C – 40°C
Moist environment 0 % - 95 % (non-condensing)
Input voltage 220 V – 240 V, 50 Hz
Electric Supply
Power pack secondary voltage 9V, 800mA
Service life background light on for 70 h
Service life background light off 90 h
Rechargeable battery
(Standard)
Charge time 12 h
Dimensions display unit
(B x D x H) mm 266 x 165 x 96
Weighing surface mm 300 x 240 400 x 300
IP protection IP 67 as per DIN 60529
Interface optional RS232

SFB-BA-e-1013 7
KERN SFB 100K10HIP SFB 120K50HIPM
Readability (d) 10 g 50 g
Weighing range (max) 100 kg 120 kg
Minimum load (Min) - 1 kg
Verification value (e) - 50 g
Verification class - III
Reproducibility 10 g 50 g
Linearity ± 10 g ± 50 g
Recommended
adjustment weight, not
added (class)
100 kg
(M1)
120 kg
(M1)
Warm-up time 30 minutes 10 minutes
Stabilization time (typical) 2-3 sec.
Weighing unit kg
Auto Off Available options 5, 15 min.
Ambient temperature 0°C – 40°C -10°C – 40°C
Moist environment 0 % - 95 % (non-condensing)
Input voltage 220 V – 240 V, 50 Hz
Electric Supply
Power pack secondary voltage 9V, 800mA
Service life background light on for 70 h
Service life background light off 90 h
Rechargeable battery
(Standard)
Charge time 12 h
Dimensions display unit
(B x D x H) mm 266 x 165 x 96
Weighing surface mm 400 x 300
IP protection IP 67 as per DIN 60529
Interface optional RS232

2 Appliance overview
1. Battery status display
2. Keyboard
3. Weight display
4. Tolerance tag, see chap. 7.7
5. Weighing unit
6. Levelling screw
7. Spirit level (underneath weighing platform)
SFB-BA-e-1013
8

2.1 Keyboard overview
Button Function
Turn on/off
•Zeroing
Navigation key •Confirm entry
Taring
Navigation key At numeric input increase flashing digit
Scroll forward in menu
•Display sum total
Navigation key •Digit selection to the right
•Add weighing value in summation memory
Navigation key •Digit selection to the left
•Calculate weighing data via interface
C •Delete
•Switch-over gross weight net weight
ESC •Back to menu/weighing mode
•Activate animal weighing function
•Activate weighing with tolerance limits
•Delete total added memory
SFB-BA-e-1013 9

SFB-BA-e-1013
10
2.1.1 Numeri
Press
c input via navigation keys
current setting appears. The first digit is flashing and can be
changed.
If the first digit is not to be changed, press and the second digit will start
flashing. Each time you press , the display unit jumps to the subsequent
digit, returning to the first digit after the last digit has been pressed.
To change the selected (flashing) digit, press repeatedly until the desired
value appears. Then select by using additional digits and change these by using
Finish entry with .
2.2 Overview of displays
Display Significance
Rechargeable battery very low
STABLE Stability display
ZERO Zero display
GROSS Gross weight
NET Net weight
AUTO Automatic add-up enabled
Kg Weighing unit
M+ Adding
LED
+ / / - Indicators for weighing with tolerance limits

SFB-BA-e-1013 11
3.1 Proper use
f material
, i.e. the
placed in the centre of the weighing
plate. As soon as a stable weighing value is reached the weighing value can be read.
quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly
system.
maximum load (max) of the weighing
plate, minus a possibly existing tare load, must be strictly avoided. Balance may be
xplosive environment. The serial version is not explosion
rotected.
he structure of the balance may not be modified. This may lead to incorrect
d faults and destruction of the balance.
d according to the described conditions. Other areas of
se must be released by KERN in writing.
3.3 nt
Warranty claims shall be voided in case
•dit ion manual are ignored
•plia tside the described uses
•lia or opened
•liquids, natural wear and tear
•
•The measuring system is overloaded
3 Basic Information (General)
The balance you purchased is intended to determine the weighing value o
to be weighed. It is intended to be used as a “non-automatic balance”
material to be weighed is manually and carefully
3.2 Improper Use
Do not use balance for dynamic weighing. In the event that small
draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring
Impacts and overloading exceeding the stated
damage by this.
Never operate balance in e
p
T
weighing results, safety-relate
The balance may only be use
u
Warra y
Our con ions in the operat
The ap nce is used ou
The app nce is modified
Mechanical damage or damage by media,
The appliance is improperly set up or incorrectly electrically connected

3.4 Monitoring of Test Resources
SFB-BA-e-1013
12
In the framework of quality assurance the measuring-related properties of the
plicable, the testing weight, must be checked regularly. Thebalance and, if ap
responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com with regard to
the monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
alibrated (return to the national standard) fast and at moderate cost.c
4 Basic Safety Precautions
onnel training
tion & Storage
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you
are already familiar with KERN balances.
4.2 Pers
The appliance may only be operated and maintained by trained personnel.
5 Transporta
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
.2 Packaging / return transport5
Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning.
plied transport securing devices.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly sup
Secure all parts such as glass wind screen, weighing
platform, power unit etc. against shifting and damage.

SFB-BA-e-1013 13
vel surface;
raughts due to open windows and doors;
eighing;
strong humidity for extended periods. Non-
permitted condensation (condensation of air humidity on the appliance) may
taken to a considerably warmer environment. In
be weighed or weighing container.
•Being of protection type IP 67 as per DIN EN 60529, the weighing scale is
Major display deviations (incorrect wei xperienced should
electro ipment), static
electric ion or remove
source
6.2 U
Scope of delivery / serial accessories:
•
•Transit
•Mains
•Rechargeable battery
•Instruction Manual
arefully remove the balance from the packaging, remove plastic cover, assemble
6.2.1) and setup balance at the intended
workstation.
6 Unpacking and implantation
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
On the installation site observe the following:
•Place the balance on a firm, le
•Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
•Protect the balance against direct d
•Avoid jarring during w
•Protect the balance against high humidity, vapors and dust;
•Do not expose the balance to
occur if a cold appliance is
this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
•Avoid static charge of goods to
suitable for short-term use in wet conditions.
ghing results) may be e
magnetic fields (e.g. due to mobile phones or radio equ
ity accumulations or instable power supply occur. Change locat
of interference.
npacking/implantation
Balance, see chap. 2
Securing
adapter
C
the tripod and the display unit (see chap.

SFB-BA-e-1013
14
Remove the transportation lock:
m
Remove the marked screws.
1. Models platform size 300 x 240 m
2. Models platform size 400 x 300 mm
Remove the screw marked by the label .
Attention: The sealed screws must not be unscrewed.
Accurate weighing results require a weighing bridge with perfect horizonta
During initial installation and after each change of work area it is necessary to level
l alignment.
the weighing bridge.
As the air bubble is located under
is in the prescribed circle.
the weighing plate, remove it.
Level balance with foot screws until
the air bubble of the water balance

6.2.1 Tripod
SFB-BA-e-1013 15
s rm size 300 x 240 mm:A semble example models platfo
tach the tripod to the platform acc. to fig. u rewAt sing the 4 sc s [1], securing disks
nd washers. Ensure that the cable is not damaged nor squeezed. Screw-in supporta
screw [2] till it is safely fixed.

SFB-BA-e-1013
16
rom holder, for that remove the turning knobs [3] on the side.Remove display unit f
Attach the tripod with the four raised counter-sunk head screws [4] and the nuts on
the holder of the display unit.
Re-attach and position display unit using the turning knobs [3].

SFB-BA-e-1013 17
cope of delivery models platform size 400 x 300 mm:S
Tripod tube
Adapter display unit
Tripod foot

SFB-BA-e-1013
18
6.3 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage. Only use original KERN mains adapters. Using other
makes requires consent by KERN.
6.4 Rechargeable battery operation
Before the first use, the battery should be charged by connecting it to the mains
power supply for at least 12 hours.
The symbol appearing on the weight display indicates that the battery is
getting low. Approximately 10 h of instrument usage are left; afterwards it will shut
off automatically. Use the supplied battery charger for charging the battery.
Charge status of rechargeable battery is indicated by the LED display.
red: Voltage has dropped below prescribed minimum.
green: Rechargeable battery is completely charged
yellow: Charging storage battery
To save battery life, you can enable the automatic switch-off function "AUTO OFF",
see chap. 7.13.
6.5 Adjustment
s the acceleration value due to gravity is not the same at every location on earth,
the balance must be coordinated in compliance with the underlying physical weighing
principle to the existing acceleration due to gravity at its place of location (only if the
balance has not already been adjusted to the location in the factory). This adjustment
process must be c ed issioning, after each change of location
as well as in case of fluctuating environment temperature. To receive accurate
measuring values it is also recommended to adjust the balance periodically in
weighing operation.
A
arri out for the first comm
•In balances with a resolution of < 15 000 dividing steps an adjustment is
recommended.
In balances with a resolution of > 15 000 dividing steps a linearisation is
recommended (see chap. 6.6).
•The weight to be used depends on the capacity of the scale. Carry out
adjustment as near as possible to the scale’s maximum weight. Info
about test weights can be found on the Internet at: http://www.kern-
sohn.com.
•Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.

SFB-BA-e-1013 19
erified models:
V
access, both contacts of the printed circuit board must be short-circuited by
a jumper (see chap. 6.7).
Attention:
In verified balances the access to menu block „P2 CAL“ is locked. For menu
weighing system must be re-verified by an
rification wire/seal mark fitted before it can
After destruction of the seal the
authorised agency and a new ve
be reused for applications subject to verification.
Switch-on balance and during the selftest press .
Press subsequently , , , the first menu block
„PO CHK“ will be displayed.
Press repeatedly until „P2 CAL“ will be displayed.
Press , the first menu item „COUNT“ will be displayed
Press repeatedly until „CAL“ will be displayed.
Acknowledge with . Ensure that there are no objects on
the weighing plate.
Wait for stability display, then press .
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with .
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press .
during selftest, balance will return
After the adjustment the balance will carry out a self-test.
Remove adjusting weight
into weighing mode automatically. An adjusting error or
incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.
Prior to a new verification the access to the menu block „P2 CAL“ must be
locked again by removing the jumper.

SFB-BA-e-1013
20
els:
C ll-
Not verifiable mod
a up menu item P3 CALCalnoLin, see chap. 8
Acknowledge with . Ensure that there are no objects on
eighing plate.the w
Wait for stability display, then press .
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with
.
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press .
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
ight will be
stment
An adjusting error or incorrect adjusting we
indicated by the error message; repeat adju
procedure.