Kinetik IET2 User manual

Inner Ear Thermometer IET2
User Instructions
The ear thermometer is a devicecapable of achievinginfra-red temperature measurement when placed in
the auditory canal of a subject.It is asafe means ofmeasuring human bodytemperature through theear.
Specification:
Range of displayed temperature : 34 ~44 (93.2 ~111.2 )
Display L (L ) when thetemperature is under34 (93.2 )
Display H (H ) when thetemperature is over44 (111.2 )
Operating ambient:
Temperature:16 ~35 (60.8 ~95.0 )
Humidity: 80%rh
Storage and transport condition:
Temperature: -10 to 55 (14 to 131 )
Humidity: 30%rh to 90%rh
Display resolution temperature range: 0.1 (0.1 )
Accuracy: 0.2 (from 35.5 to 42.0 )
0.4 (from 96.0 to 107.6 )
Power consumption: 30 mW
Battery: two 1.5V alkalescence batteries (AAA)
Battery life: 4000 takes or 6 months with3-min.usege per day
Dimension: 134mm by 37mm by 67mm
Net weight: 74g
Beeper sign: on/off, measuring finishand low voltagewarning, etc.
Self-testing sequence: Press the 'on/recall' button to turn on the thermometer and all of symbol
(See Fig A) should be displayed on the LCD in one second.
Calibration: This thermometer is calibrated at the time of manufacture. If the thermometer is used
according to the instructions, periodic re-calibration is not required.
Manufacturing Batch code: see the label within the battery compartment.
Probe membrane
Disposable probe cover
Probe cap
On/recall button
Batterry cover
Scan button
LCD
Display: liquid crystal display, 4 digits
temperature value: display the maximum temperature
in measuring process
temperature unit: centigrade or fahrenheit
display of memory: last ten memories
low battery warning: the LCD display and then abeep sound isheard
Please read the instructions carefully before using.
Cautions
1. Condition the thermometer at room temperature for 30 minutes before taking a reading.
2. Use a new probe cover with each and every measurement to prevent inaccurate readings.
3. Please keep your ear canalclear, ifnot, the measuringresult shall beinaccurate.
4. The
5. Do not scratch the probe membrane as this will affect the accuracy of measurements.
6. If the LCD does not change or respond when you press any of the buttons, remove the batteries
and then replace to reset.
12. This device must always bekept in aclean, dry area.
13. : Degree of protection against electric shock is Type BF applied part.
' ' symbol on the display indicates the device is ready for use.
14. Battery disposal should be completed in accordance with local regulations.:
7. Please don't use the thermometer if your ear canal has become inflamed. Do not force the thermometer
into the ear canal. If pain occurs, stop using immediately.
8. Do notexpose this thermometer to electric shock.
9. Do notexpose the thermometerto sunlight orto water.
10. Do notuse near strongelectromagnetic fields, i.e. Keep it awayfrom any radiosystems and mobile phones.
11. Donot modify thisdevice without authorizationof the manufacturer.
IET2 IB UK 20180525
The standard we adopted is EN12470-5:2003 Clinical thermometers -- Part 5: Performance of infrared
ear thermometers (with maximum device).
This device complies with the requirements of IEC 60601-1,IEC 60601-1-11 and IEC 60601-1-2.
Guidance and manufacturer's declaration - electromagnetic immunity
The IET2 Device is intended for use in the electromagnetic environment specified below.
The customer or the user of the IET2 Device should assure that it is used in such an environment.
IMMUNITY test IEC 60601
test level
Compliance
level Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 Ghz 3 A/m
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
Portable and mobile RF communications equipment should
be used no closer to any part of the IET2 Device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance :
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,ashould be
less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked
with the following symbol:
d1.2 P
d1.2 P
d2.3 P
d2.3 P
80 MHz to 800 MHz
800 MHz to 2,5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
A. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted the
oretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
IET2 Device is used exceeds the applicable RF compliance level above, the IET2 Device should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the IET2 Device.
B. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The IET2 Device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the IET2 Device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the IET2 Device as recommended below, according to
the maximum output power of the communications equipment.
Recommended separation distances between
portable and mobile RF communications equipment and the IET2 Device
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
150 kHz to 80 Mhz
d1.2 P
80 MHz to 800 Mhz 800 MHz to 2,5 Ghz

Replacing the batteries
2. The thermometeris supplied withtwo 1.5VALKALESCENCE BATTERIES(AAA). Insert new batteries when
the low battery symbol appears on the LCD.
3. Remove the battery cover andtake out theold batteries.
4. Place twonew batteries according to the '+' or ' -'
5. Please take out the batteriesto avoid batteryleaking if unitnot used forover six months.
6. The disposal of the probe cover, battery and device shall comply with the local environment requirements.
The lithium battery or fuel cell may lead to excessive temperatures, fire or explosion.
Guarantee
1. When voltage ofthe batteries arelow, theLCD will display ' 'symbol, please replace two new
batteries in unit. Thethermometer cannot workaccurately under thecondition of lowvoltage.
This product is guaranteed for a period of one year from the date of purchase against mechanical and electrical manufacturing defects. There
are no serviceable parts inside this device. Any attempted repair by unauthorised persons invalidates the warranty. In the unlikely event that you
experience a problem, please return it to the retailer where you made the purchase, along with your receipt. This does not affect your statutory
rights.
Guidance and manufacturer's declaration - electromagnetic immunity
Guidance and manufacturer's declaration - electromagnetic emissions
The IET2 Device is intended for use in the electromagnetic environment specified below.
The customer or the user of the IET2 Device should assure that it is used in such an environment.
The IET2 Device is intended for use in the electromagnetic environment specified below.
The customer or the user of the IET2 Device should assure that it is used in such an environment.
IMMUNITY test
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Electrostatic discharge
(ESD) IEC 61000-4-2
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
IEC 60601
test level
6 kV contact
8 kV air
3 A/m
Compliance
Compliance
Group 1
Class B
Not applicable
Not applicable
Electromagnetic environment - guidance
Electromagnetic environment - guidance
6 kV contact
8 kV air
3 A/m
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
The IET2 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The IET2 is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
While pulling the outer ear, insert the probe snugly into the ear canal (insert as deep as is comfortable, do
not force, the probe into the ear canal). Press and immediately release the ‘SCAN’ button. Do not remove
the thermometer from the ear until you hear the beep sound – this beep means that the measurement is
complete. Remove the thermometer from the ear. The display shows the measured temperature.
OOO
6.
The LCD will displays 'Er 1' as Fig E and cannot displays tem
when the ambient temperature is out of the range of 16 Cto 35 C.
perature
5. Memory seach:
Press the 'on/recall'button, the last ten
memories (NO.9-NO.0) will take turns to display on the LCD
as Fig C and Fig D.
7. Shut off:
The thermometer shall automatically shut off without manipulating in one minute.
3.
4. Repeat measurements: wait at least 1 minute between measurements. Ensure the Ear symbol
is visible in the LCD.
1. Press the 'on/recall' button to turn on the thermometer. After the
beep the thermometer will be ready to read when the LCD
display matches Fig B.
How to use ( C displays for example )
Measuring human body temperature in the Ear Canal
O
Fig A Fig B
Fig E
Note:
In order to avoid the spreading of germs, use a new probe
cover for each measurement
Check to ensure probe cover isfitted on firmlybefore use (pleasesee
the diagrams right).
If the probe cover is broken,discard and applya new oneimmediately.
2. Straighten the ear canal by pulling the outer ear up and back to give
a clear view of the eardrum.
For children under 1 year, Pull the ear straight back.
For children ages 1 year to adult, Pull the ear up and back.
(Caution: Many pediatricians suggest use of ear thermometers for
children older than 6 months.)
Newborn-1 year
1 year-adult
Fig C Fig D
Incorrect Correct
Explanation of symbols
O
Cleaning and Storage
1.
Store thermometer and the Disposable probe covers in a dry location free from dust
and contamination and away from direct sunlight. The ambient temperature at the
storage location should remain fairly constant and within the range of -10 C to 55 C.
2. Use an alcohol swab or cotton swab moistened with alcohol (70% Isopropy) to clean the
thermometer casing and the measuring probe . Ensure that no liquid enters the interior of
the thermometer.
3. Never use abrasive cleaning agents, thinners or gasoline for cleaning and never immerse the instrument
in water or other cleaning liquids. Take care not to scratch the surface of the probe membrance or display.
before each use
O
battery is empty
Product disposal instructions
for electronic devices
The battery in this product complies with the
requirements stated in European Directives
2006/66/EEC
Type BF equipment
Memory
L C or L F
H C or H F
subject temperature isunder 34 C
L
O
T
subject temperature is above 44 C
Classification according to the degree of protection
against ingress of water as detailed in IEC 529
Lot Number
Authorised representative in the european community
Read IFU carefully
or
Note: Before measuring, the thermometer shall be stabilized at the operating
ambient condition for a minimum of 30 min.
Changing from Celsius to Fahrenheit
To switch the display between C and F, turn the unit on. Press and hold the on/recall
button for 3 seconds, the display will show as Fig. F , the unit will prepare for a measurement.
OO
Fig F
Product disposal
Please ensure environmental protection. Batteries do not belong in the domestic waste. Please
hand them in at collection point or the municipal recycle material centre as special waste.
This symbol on products and/or accompanying documents means that consumed electronic
products must not be mixed with conventional domestic waste. Take these products to the
corresponding collection points for correct treatment and reycling, where they will be accepted free
of charge. For more information on the closeset collection point, Please enquire with your local
authorities.
Other Kinetik Thermometer manuals