PMP
PMP K03/0218 page 9
9.2 Analogue output [Out]
Menu item [4nA]: display value at 4 mA
factory setting 0…500,0 mmH20
factory setting: [000.0]
Menu item [20nA]: display value at 20 mA
factory setting 0…500,0 mmH20
factory setting: [500.0]
The 4 mA and 20 mA – value of the analogue output can be related to any
differential pressures.
Condition: [4nA] < [20nA]
Factory setting: 4mA = 000,0; 20 mA = 500,0
9.3 Simulation [SINU]
The main menu Simulation is used to test the switching of the relays and the
analogue output 4-20mA is adjusted manually to the current. During installation,
this simplifies the testing of the connection to the downstream analysing unit in
the control cabinet or master display.
Menu item [Alr1]: simulation Alarm 1
Menu item [Alr2]: simulation Alarm 2
Value setting: [OFF] or [ON] (OFF/ON changeable with )
Press ESC-button to finish.
Basic setting: [OFF]
Menu item [Out]: simulation analogue output
Value setting: [4, 5, 6, 7 …20] mA changeable with )
Press ESC-button to finish.
Basic setting: [4] mA
When leaving the main menu item Simulation [SINU] the menu item
Alarm 1 [ALr1], Alarm 2 [ALr2] and Analogue Output [OUt] are
resetted to the basic setting and the outputs behave according the
display.