Koehler K4304X Guide

Operation and Instruction Manual
REV B
Koehler Instrument Company, Inc.
1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA
Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815
http://www.koehlerinstrument.com • e-mail: info@koehlerinstrument.com
Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing
K4304X
High Temperature 'Sequence IV'
Liquid Foam Test Bath


CERTIFICATE OF CONFORMANCE
Liquid Foaming Characteristics Apparatus
K4304X
This certificate verifies that part number K4304X, Liquid Foaming Characteristics
Apparatus, was manufactured in conformance with the applicable standards set
forth in this certification.
Specifications: ASTM D6082
IP 146
DIN 51566
FTM 791-3211
FTM 791-3213
NF T 60-129
This unit is tested before it leaves the factory, to ensure total functionality and
compliance to the above specifications and ASTM standards. Test and inspection
records are on file for verification.
Jesse Kelly
Application Engineer
Koehler Instrument Company


K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -1-
Table of Contents
1Introduction ...................................................................................................................3
1.1 Koehler’s Commitment to Our Customers .............................................................................................3
1.2 Recommended Resources and Publications.........................................................................................3
1.3 Instrument Specifications.......................................................................................................................3
1.4 Software Specifications..........................................................................................................................3
2Safety Information and Warnings ................................................................................4
3Getting Started...............................................................................................................4
3.1 Packing List............................................................................................................................................4
3.2 Unpacking ..............................................................................................................................................4
3.3 Setup......................................................................................................................................................4
4Descriptions...................................................................................................................5
4.1 Main Instrument Controls.......................................................................................................................5
4.2 Top Plate Descriptions...........................................................................................................................6
4.3 Flowmeter Setup and Operation ............................................................................................................6
4.4 Flowmeter Calibration Data....................................................................................................................6
4.5 Temperature Controller Operation.........................................................................................................6
5Operation .......................................................................................................................8
5.1 Bath........................................................................................................................................................8
5.2 Power .....................................................................................................................................................8
5.3 Preparation of Glassware and Tubing ...................................................................................................8
6Safety Features..............................................................................................................8
6.1 Over-Temperature Protection ................................................................................................................8
6.2 Over-Power Protection...........................................................................................................................8
7Maintenance...................................................................................................................8
7.1 Routine Maintenance .............................................................................................................................9
7.2 Replacement Part...................................................................................................................................9
8Wiring Diagrams..........................................................................................................10
8.1 Unit Wiring............................................................................................................................................10
9Troubleshooting..........................................................................................................11
9.1 Unit does not power up ........................................................................................................................11
9.2 Unit is on and keeps resetting into start up routine..............................................................................11
9.3 Unit is on but bath does not heat up ....................................................................................................11
9.4 Bath heats up but temperature does not stabilize................................................................................11
10 Service..........................................................................................................................11
11 Storage.........................................................................................................................11

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -2-
12 Warranty.......................................................................................................................11
13 Returned Goods Policy...............................................................................................12
Notes..................................................................................................................................13
Notes..................................................................................................................................14

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -3-
1 Introduction
The Koehler K4304X High Temperature
'Sequence IV' Liquid Foam Test Bath is the latest
design for determining foaming characteristics of
lubricating oils according to the ASTM D6082 test
method and related test specifications.
This manual provides important information
regarding safety, technical reference, installation
requirements, operating condition specifications,
user facility resource requirements, and operating
instructions for the High Temperature 'Sequence
IV' Liquid Foam Test Bath. This manual should
also be used in conjunction with applicable
published laboratory procedures. Information on
these procedures is given in section 1.2.
1.1 Koehler’s Commitment to Our
Customers
Providing quality testing instrumentation and
technical support services for research and testing
laboratories has been our specialty for more than
50 years. At Koehler, the primary focus of our
business is providing you with the full support of
your laboratory testing needs. Our products are
backed by our staff of technically knowledgeable,
trained specialists who are experienced in both
petroleum products testing and instrument service
to better understand your requirements and
provide you with the best solutions. You can
depend on Koehler for a full range of accurate and
reliable instrumentation as well as support for your
laboratory testing programs. Please do not
hesitate to contact us at any time with your
inquiries about equipment, tests, or technical
support.
Toll Free: 1-800-878-9070 (US only)
Tel: +1 631 589 3800
Fax: +1 631 589 3815
Email: info@koehlerinstrument.com
http://www.koehlerinstrument.com
1.2 Recommended Resources and
Publications
1. American Society for Testing and Materials
(ASTM)
100 Barr Harbor Drive
West Conshohocken, Pennsylvania 19428-
2959, USA
Tel: +1 610 832 9500
Fax: +1 610 832 9555
http://www.astm.org
email: service@astm.org
ASTM Publication:
ASTM D6082: Standard Test Method for
High Temperature Foaming
Characteristics of Lubricating Oils
1.3 Instrument Specifications
Models:
K43041
K43049
Electrical
Requirements:
115V, 60Hz
230V, 50/60Hz
Temperature Range:
Ambient to 150°C
(302°F)
Temperature Control
Stability:
Exceeds ASTM
requirements
Digital Setpoint and
Displays °C/°F
Built-in Over-temperature
Cut-off Protection
Capacity:
2 Test Cylinders
Bath Medium:
Suitable Heat Transfer
Fluid
(see section 5.1)
Bath Medium
Capacity:
38.5L (9 gal) each
1.4 Software Specifications
PC System
Requirements:
Intel® Pentium II Processor
or similar (minimum)
Operating
System:
Windows® 98 SE, 2000, NT,
XP
Memory
Requirements:
256Mb RAM (512Mb RAM
recommended) 15 Mb Hard
Disk Space (minimum)
Other:
Two RS-232 communication
ports
Microsoft® Excel (97 or
above)

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -4-
2 Safety Information and Warnings
Safety Considerations. The use of this
equipment may involve hazardous materials and
operations. This manual does not purport to
address all of the safety problems associated with
the use of this equipment. It is the responsibility of
any user of this equipment to investigate,
research, and establish appropriate safety and
health practices and determine the applicability of
regulatory limitations prior to use.
Equipment Modifications and Replacement
Parts. Any modification or alteration of this
equipment from that of factory specifications is not
recommended voids the manufacturer warranty,
product safety, performance specifications, and/or
certifications whether specified or implied, and
may result in personal injury and/or property loss.
Replacement parts must be O.E.M. exact
replacement equipment.
Unit Design. This equipment is specifically
designed for use in accordance with the applicable
standard test methods listed in section 1.2 of this
manual. The use of this equipment in accordance
with any other test procedures, or for any other
purpose, is not recommended and may be
extremely hazardous.
Chemical Reagents Information. Chemicals and
reagents used in performing the test may exhibit
potential hazards. Any user must be familiarized
with the possible dangers before use. We also
recommend consulting the Material Data and
Safety Sheet (MSDS) on each chemical reagent
for additional information. MSDS information can
be easily located on the internet at
http://siri.uvm.edu or http://www.sigma-
aldrich.com.
3 Getting Started
The instructions for preparing the equipment
assume that the user is aware of the contents of
this document, which lists the warranty conditions
and important precautions.
3.1 Packing List
K4304X High Temperature 'Sequence IV' Liquid
Foam Test Bath including:
- (2) 1000mL Test Cylinders (K43012)
- (2) Calibrated and Certified Stainless Steel
'Mott' Diffuser Stones (344-005-01C)
- (2) Air Delivery Tube Assemblies (K430-3A)
- (2) Air Outlet Tubes (K430-0-13)
- (2) Rubber Stoppers (K430-0-8)
- (1) 12 x 18 Pyrex Bath Jar (332-001-003)
- (1) 18” Safety Shield (K23425-00000)
- (1) Test Tube Rack (K43002-0-13)
K4304X-Manual High Temperature 'Sequence
IV' Liquid Foam Test Bath Operation and
Instruction Manual
3.2 Unpacking
Carefully unpack and place the instrument and
accessories in a secure location. Use extra care
while unpacking the Borosilicate Glass bath jar.
Ensure that all parts listed on the packing list are
present. Inspect the unit and all accessories for
damage. If any damage is found, keep all packing
materials and immediately report the damage to
the carrier. We will assist you with your claim, if
requested. When submitting a claim for shipping
damage, request that the carrier inspect the
shipping container and equipment. Do not return
goods to Koehler without written authorization.
3.3 Setup
Equipment Placement. Place the base on a firm,
level table in an area with adequate ventilation or
in a hood. The unit may be leveled by making
minor turning adjustments to the feet located at
the base of the unit. Please note that Koehler does
not supply a level with this equipment.
Install the borosilicate glass test jar. Fill the jar with
appropriate bath medium as per ASTM D6082.
Please see section 5.1 for further details on the
bath medium.
Install the bath top assembly on the jar and plug
the control cables into the receptacle on the base
unit.

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -5-
Install the RTD probes in the bath top with the cork
/ thermometer probe holder provided
Ventilation. A fume hood or exhaust system is
required when operating the unit. Flammable
vapors and/or steam are generated during
operation and must not be permitted to
accumulate. A canopy-style hood may be used if
the height from the top of the unit to the canopy is
5 feet or less. The exhaust blower should have a
rating of 1000 C.F.M. or greater.
Power. Connect the line cords to properly fused
and grounded receptacles with the correct voltage
as indicated in section 1.3 or on the back of the
unit.
WARNING: For safety, disconnect the power
when performing any maintenance and/or
cleaning. Do NOT turn the power on unless the
bath is filled with the proper medium; otherwise,
damage may occur to the unit and the warranty
will be void.
4 Descriptions
4.1 Main Instrument Controls
Figure 1: Main Instrument Descriptions
1. Line Switch. This switch controls the power to
motor and heater / temperature controller.
When the power switch is in the ON position,
the digital temperature controller and the
stirrer are powered on.
2. Temperature Controller. The temperature
controller regulates the bath temperature its
corresponding bath. Refer to Section 4.3 for
full operational details.
3. Input Switch. This switch allows the user to
control the heater power capacity. Pressing
the switch to HI will utilize 100% of the power
available resulting in a faster heating rate.
This is recommended when initially heating
the bath from ambient temperature to the
150°C test temperature. Pressing the switch to
LO will cut the heater capacity to 50% for
more accurate control of the bath temperature
for stabilization during testing.
4. Flowmeters. To regulate the flow of air into
each air delivery tube assembly. Each
flowmeter is configured to deliver air at a rate
of 200 mL/min. Detailed flowmeter calibration
data is given is section 4.4 of this document.
5. Stirrer Motor. The stirrer constantly circulates
bath medium to prevent temperature gradients
and ensures temperature stability. When
cleaning and/or servicing, please be sure to
disconnect unit power to avoid possible injury.
6. Temperature Probe. Detects the bath
temperature and communicates with the
temperature controller to maintain desired
temperature for each bath.
7. Air Delivery Tube. Connect to air conditioning
coil and flowmeter for passing air through gas
diffuser located at the bottom of the Test
Cylinder.
8. Heat Shield. An 18” safety shield is provided
to protect the user from accidentally touching
the Bath while operating the instrument.
9. Bath Reservoir. For maintaining the sample
at proper test temperature

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -6-
4.2 Top Plate Descriptions
Figure 3: Hot Bath (Right) Top Plate Descriptions
10. Double Heater. Provides high heating to
attain 150°C as required by sequence IV to
the bath.
11. Conditioning Coil for Sample 1. Connect to
flowmeter. Air then flows through coil located
within bath set at test temperature. Connect
outlet, located at the corresponding position
on the opposite side of the plate, to the air
delivery tube for Sample 1.
12. Conditioning Coil. Not used for this model.
4.3 Flowmeter Setup and Operation
1. Connect the inlet on the flowmeter manifold
fitting to a source of clean dry air. Instrument
grade compressed air is recommended for this
application. Tygon or heavy wall rubber tubing
is suitable for connection to the manifold.
2. Adjust the inlet air pressure to the regulator to
fifteen (15) psi for normal operation.
3. Connect the outlets on the flowmeters to the
air tube assemblies using new ¼” tygon
tubing.
4. Set the proper airflow by first consulting the
flowmeter calibration data in section 4.5. Find
the millimeter line that will yield a flow of 200
ml/min for the sequence IV procedure and set
both flowmeters to this value.
5. Adjust the valve on the flowmeter, so that the
ball float in the tube is centered on the
millimeter value specified in the data.
6. Permit the air to flow for the duration specified
in the test procedure and disconnect the tygon
tubing from the flowmeter outlet when the
aeration period as elapsed.
4.4 Flowmeter Calibration Data
K43002-4-12 Flow Rate 200 mL/min
Float Material:
Carboloy
Float Density:
14.98 g/ml
Metered Fluid:
Air
Minimum Flow:
22.0 ml/min
Maximum Flow:
246.2 ml/min
Tolerance:
± 2% F.S
Scale Reading (mm)
Flowrate (ml/min)
65
246.2
60
221.1
55
199.7
50
175.5
45
154.0
40
130.3
35
110.2
30
91.5
25
75.6
20
60.4
15
45.0
10
32.3
5
22.0
4.5 Temperature Controller Operation

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -7-
Figure 4. Temperature Controller
1. Process Temperature Display. The
upper red LED display shows the process
temperature as read from the RTD probe.
2. Set Point Temperature Display. The
lower green LED display shows the set
point temperature of the controller.
3. Programming Key. Permits scrolling
through controller menu parameters. One
Level Forward
4. Down Key. Used to decrease the set
point temperature and to decrease or
change parameter values when
programming the temperature controller.
5. Up Key. Used to increase the set point
temperature and to increase or change
parameter values when programming the
temperature controller.
6. Exit / Function Key. This key is used to
exit or leave a level. One level backward
IMPORTANT NOTE: The digital
temperature controller for the unit comes
pre-programmed from the Koehler factory.
Please do NOT attempt to re-program the
digital temperature controller as this will
void the product warranty. If assistance is
required, please do not hesitate to contact
the Koehler technical service department.
Setting the Temperature. Set the desired
operating temperature by adjusting the set
point with the up and down keys. The set point
will be displayed in the lower green Set Point
LED display and the actual temperature will be
displayed in the upper red Process LED
display. Please allow the instrument to fully
equilibrate before proceeding with any testing.
Temperature Calibration. This routine allows the
digital temperature controller to be calibrated
to a certified thermometer.
a. Use a certified calibrated measuring
device to acquire the temperature.
Calculate the difference between the
measuring device and the Process value
displayed on the controller.
b. Press the program key two times until PCt
is displayed in the lower green LED
display. Press the DOWN key. CAL will
display on the lower green display. If there
is a value observed in the upper red LED
display, add it to the calculated difference
obtained in the previous step. This is the
offset value.
c. Press the Program Key. The lower green
display will flash. Use the up or down keys
to adjust to the new calibration offset
value on the upper red display calculated
in the previous step. When the value has
been entered, the controller will
automatically store the value. The lower
green display will stop flashing. If further
adjustments are necessary, press the
Program Key again. Resume regular
operations by pressing the Exit / Function
key two times. Verify if the new calibration
is correct by observing the upper red
display and comparing the value with the
calibrated reference device.
Auto Tune. This routine allows the digital
temperature control to learn the heating
parameters needed for any particular set point
temperature. This operation should be done
when installing a new unit, after replacing or
changing the bath medium type, or utilizing a
different temperature set point 20% different
from the previously used set point
temperature.
a. Set the operating temperature to the
desired setting.

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -8-
b. Press the up and down arrow buttons
simultaneously for about 5 seconds. When
Auto Tune is active, the lower green LED
display will blink TUNE. Auto Tune will
automatically toggle off when the set point
temperature is reached. Auto tune can be
terminated by pressing the up & down
buttons simultaneously again.
5 Operation
5.1 Bath
Fill the baths with the appropriate heat transfer
fluid based upon the testing temperature. Fill the
bath with the medium to 2" (5 cm) from the top of
the bath to allow room for fluid expansion. This will
provide the proper depth for immersing the test
cylinders and allow for thermal expansion.
Testing
Temperature
Recommended Fluid
Above 50°C
Koehler supplies highly refined white
technical oil (part # 355-001-001) that
contains an oxidation inhibitor to limit
clouding at higher temperatures.
Above 110°C
Koehler supplies a clear silicone heat
transfer fluid (part # 355-001-002) with
high oxidation resistance and low
volatility.
5.2 Power
1. Turn on the main power switches to the unit.
2. The temperature controllers will illuminate and
display the actual and set point temperatures.
3. Set the bath temperature to desired test
temperatures as per D6092 using the
operating instructions in section 4.6.
WARNING: Do NOT turn the power on unless the
bath is filled with the proper medium; otherwise,
damage may occur to the unit and the warranty
would be void.
5.3 Preparation of Glassware and Tubing
1. Clean the graduated cylinders as per ASTM
D6082 instructions and pour the test samples
into the cylinders to the recommended level as
per the aforementioned test methods.
2. Clean and flush the air delivery tube assembly
with diffuser stone as per ASTM D6082.
3. Place the two 1000 ml test cylinders into the
tops & lock in place with the four clamps
provided on the top plate.
4. Proceed with the test and record the test
results in accordance to ASTM D6082.
6 Safety Features
The Koehler K4304X High Temperature
'Sequence IV' Liquid Foam Test Bath is equipped
with several safety and protection features, which
are described in the following sections.
6.1 Over-Temperature Protection
The Koehler K4304X High Temperature
'Sequence IV' Liquid Foam Test Bath is equipped
with Over-temperature Protection (OTP) circuitry,
which prevents the unit from exceeding unsafe
operating temperatures. If the unit cannot maintain
the set point temperature and begins to decline,
the OTP circuitry may have been activated. Please
follow these steps.
1. Turn off the unit power switch and disconnect
the line cord.
2. Determine the source of the problem and
correct the situation.
3. Restart the unit. Monitor the operations to
ensure that the unit is operating properly. If
you are still experiencing trouble, please
contact Koehler technical service for
assistance.
6.2 Over-Power Protection
The Koehler K4304X High Temperature
'Sequence IV' Liquid Foam Test Bath is equipped
with Over-power Protection circuitry, which
prevents the unit from unsafe electrical conditions.
If power to the unit is lost, then turn off the main
power and turn it back on again. The main power
switch also functions as a circuit breaker.
7 Maintenance

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -9-
WARNING. Disconnect power to the unit before
servicing to avoid exposure to high voltages
and/or temperatures which may result in personal
injury or death. If you have any questions about
maintaining your equipment, then please do not
hesitate to contact the Koehler technical service
department.
7.1 Routine Maintenance
The K4304X High Temperature 'Sequence IV'
Liquid Foam Test Bath requires little routine
maintenance to provide many years of continuous
service. However, over the course of time, some
instrument parts may need to be replaced. When
ordering replacement part(s), please provide the
model number, serial number, and product
shipment date of your equipment so that we can
ensure you will receive the proper replacement
part(s).
7.2 Replacement Part
Part Number
Description
091-240-002
Solid State Relay, 90 - 240V, 25A
165
Thermometer / Probe Holder
265-400-002
RTD Temperature Probe
275-103-044
Temperature Controller, 100-240V,
50/60Hz 7A
278-001-002
Fuse, 1A, Slo-Blo, 5 X 20 mm
332-001-003
12 x 18 Borosilicate Bath Jar
K23425-03000
18” Safety Shield
K23700-03013A
Motor 115V 60HZ 1/20 HP
K23700-03014A
Motor 230V 50/60HZ 1/15 HP
K430-0-13
Air Outlet Elbow Tube
K430-0-8
Rubber Stopper
K430-3A
Air Inlet (Delivery) Tube without diffuser
K43002-4-12
Flowmeter
K43002-0-13
Test Tube Rack
K43002-0-15
Locking Post Assembly
K43012
1000ml Cylinder w/ Ring
K43002-0-9
Hot Bath Outer Heater, 750W, 115V
K43002-0-11
Hot Bath Inner Heater, 750W, 115V
K43092-0-9
Hot Bath Outer Heater, 750W, 230V
K43092-0-11
Hot Bath Inner Heater, 750W, 230V

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -10-
8 Wiring Diagrams
8.1 Unit Wiring

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -11-
9 Troubleshooting
WARNING: Troubleshooting procedures involve
working with high voltages and/or temperatures
which may result in personal injury or death, and
should only be performed by trained personnel.
Please do not hesitate to contact Koehler for
assistance.
9.1 Unit does not power up
1. Establish that the socket outlet is providing
proper and adequate voltage.
2. Check if Overpower Protection circuitry
located directly behind the temperature
controller inside the front tray has been
activated.
3. Check if line switch is in the ON position.
4. If problem persists, please call the Koehler
technical service department for assistance.
9.2 Unit is on and keeps resetting into
start up routine
For 220V units, make sure that the socket
outlet is greater than 215V.
Check if there is a steady and reliable power
source.
Make sure the connector plug on the rear
panel is firmly plugged in.
9.3 Unit is on but bath does not heat up
Make sure that the actual temperature
reading is not higher than the set-point
temperature.
Determine if the temperature controller is
properly calibrated by comparison to an
ASTM standard thermometer.
Determine if the Overtemperature Protection
(OTP) circuitry as been activated.
9.4 Bath heats up but temperature does
not stabilize
Make sure there are no drafts from open
doors, windows, or environmental control
vents in the vicinity of the bath.
If the set temperature is close to ambient,
then you may need to circulate cold water
through the coils in order to achieve proper
temperature stability.
10 Service
Under normal operating conditions and with
routine maintenance, the K4304X High
Temperature Sequence IV Liquid Foam Test Bath
should not require service. Any service problem
can be quickly resolved by contacting Koehler’s
technical service department either by letter,
phone, fax, or email. In order to assure the fastest
possible service, please provide us with the
following information.
Model Number:
Serial Number:
Date of Shipment:
11 Storage
This laboratory test instrument is equipped with
electrical components. Storage facilities should be
consistent with an indoor laboratory environment.
This testing equipment should not be subjected to
extremes of temperature and/or moisture.
This equipment was shipped from the factory in
a corrugated cardboard container. If long term
storage is anticipated, re-packing the instrument in
a water-resistant container is recommended to
ensure equipment safety and longevity.
12 Warranty
We, at Koehler, would like to thank you for your
equipment purchase, which is protected by the
following warranty. If within one (1) year from the
date of receipt, but no longer than fifteen (15)
months from the date of shipment, Koehler
equipment fails to perform properly because of
defects in materials or workmanship, Koehler
Instrument Company, Inc. will repair or, at its sole
discretion, replace the equipment without charge
F.O.B. its plant, provided the equipment has been
properly installed, operated, and maintained.
Koehler Instrument Company must be advised in
writing of the malfunction and authorize the return
of the product to the factory. The sole
responsibility of Koehler Instrument Company and
the purchaser’s exclusive remedy for any claim
arising out of the purchase of any product is the
repair or replacement of the product. In no event
shall the cost of the purchaser’s remedy exceed
the purchase price, nor shall Koehler Instrument
Company be liable for any special, indirect,
incidental, consequential, or exemplary damages.

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -12-
KOEHLER INSTRUMENT COMPANY, INC.
DISCLAIMS ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTIES OF FITNESS FOR A
PARTICULAR PURPOSE. Please save the
shipping carton in the event the equipment needs
to be returned to the factory for warranty repair. If
the carton is discarded, it will be the purchaser’s
responsibility to provide an appropriate shipping
carton.
13 Returned Goods Policy
To return products for credit or replacement,
please contact Koehler Customer Service with
your purchase order number, our packing
list/invoice number, the item(s) to be returned and
the reason for the return. You will be issued a
Returned Authorization (RA) number, which must
be prominently displayed on the shipping
container when you return the material to our
plant. Shipping containers without an RA number
prominently displayed with will be returned to the
sender. Goods must be returned freight prepaid.
Returns will be subject to a restocking charge, the
application of which will depend upon the
circumstances necessitating the return. Some
returns cannot be authorized, including certain
products purchased from outside vendors for the
convenience of the customer, products
manufactured on special order, products shipped
from the factory past ninety (90) days, and
products which have been used or modified in
such a way that they cannot be returned to stock
for future sale.

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -13-
Notes

K4304X High Temperature 'Sequence IV' Liquid Foam Test Bath
Operation and Instruction Manual
K4304X-Manual -14-
Notes
Table of contents
Other Koehler Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Fisher Scientific
Fisher Scientific HPS RT2 Basic Operation manual

K&S
K&S ARISMD installation guide

IKA
IKA RCT basic safety control operating instructions

Okolab
Okolab 158206049 installation guide

Healthmark
Healthmark Flexible Inspection Scope FIS-007H Instructions for use

Isotech
Isotech TWIST user manual

Thermo Scientific
Thermo Scientific Thermo Scientific SL16R instruction manual

Memmert
Memmert ICH eco Series operating instructions

Hettich
Hettich MIKRO 120 manual

THORLABS
THORLABS BMS001 instruction manual

Heidolph
Heidolph Hei-MIX Unimax 2010 operating instructions

Selecta
Selecta CENCOM 2 instruction manual