KV4000 Kinematic Viscosity Bath
Operation and Technical Manual
KV4000-Manual - 13 -
Withdrawal Bulb. Use to pull sample into
viscometer tube.
Catalog Number: K22090
Viscometer Tube Stopper. Insert in the top
opening of the viscometer tube to prevent
sample from flowing due to gravity.
Catalog Number: K23311 for a pack of 12
Operation near or below ambient
temperatures. A water supply or external
chiller may be used to maintain the
temperature of the bath for cooler
temperature. Hose-barb connections for
water or refrigerated coolant are located at
the back of the bath. There are two
connections: inlet and outlet. The hose
connections can be secured with copper
wire or pull ties to prevent the tubing from
disengaging from the connection.
4.3 Temperature Controller Operation
Figure 4. Temperature Controller
1. Process Temperature Display. The
upper red LED display shows the
process temperature as read from the
RTD probe.
2. Set Point Temperature Display. The
lower green LED display shows the set
point temperature of the controller.
3. Programming Key. Permits scrolling
through controller menu parameters.
One Level Forward
4. Down Key. Used to decrease the set
point temperature and to decrease or
change parameter values when
programming the temperature controller.
5. Up Key. Used to increase the set point
temperature and to increase or change
parameter values when programming
the temperature controller.
6. Exit / Function Key. This key is used to
exit or leave a level. One level backward
IMPORTANT NOTE: The digital
temperature controller for the unit
comes pre-programmed from the
Koehler factory. Please do NOT attempt
to re-program the digital temperature
controller as this will void the product
warranty. If assistance is required,
please do not hesitate to contact the
Koehler technical service department.
Setting the Temperature. Set the desired
operating temperature by adjusting the set
point with the up and down keys. The set
point will be displayed in the lower green Set
Point LED display and the actual
temperature will be displayed in the upper
red Process LED display. Please allow the
instrument to fully equilibrate before
proceeding with any testing.
Temperature Calibration. This routine allows
the digital temperature controller to be
calibrated to a certified thermometer. In the
calibration mode, the display will always give
2 decimal places.
a. Use a certified calibrated measuring
device to acquire the temperature.
Calculate the difference between the
measuring device and the Process
value displayed on the controller.
b. Press the program key two times until
PCt is displayed in the lower green LED
display. Press the DOWN key. CAL will
display on the lower green display. If
there is a value observed in the upper
red LED display, add it to the calculated