Koehler K44100 Guide

Operation and Instruction Manual
REV A
Koehler Instrument Company, Inc.
1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA
Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815
Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing
K441XX
Automatic Non-Woods Metal Noack
Evaporative Apparatus

- 1 -

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 2 -
Table of Contents
1Introduction ............................................................................................- 3 -
1.1 Koehler’s Commitment to Our Customers..................................................................- 3 -
1.2 Recommended Resources and Publications..............................................................- 3 -
1.3 Instrument Specifications............................................................................................- 4 -
2Safety Information and Warnings .........................................................- 4 -
3Getting Started........................................................................................- 5 -
3.1 Packing List.................................................................................................................- 5 -
3.2 Unpacking ...................................................................................................................- 5 -
3.3 Installation...................................................................................................................- 5 -
3.4 Additional Accessories Required for Running Tests...................................................- 5 -
3.5 Optional Accessories Available...................................................................................- 5 -
4Descriptions............................................................................................- 6 -
4.1 Piping Diagram............................................................................................................- 6 -
4.2 Instrument Controls and Connections.........................................................................- 7 -
6Operation ..............................................................................................- 10 -
6.1 Summary of Test Method..........................................................................................- 10 -
6.2 Analysis Setup Procedure.........................................................................................- 10 -
6.3 Test Procedure..........................................................................................................- 11 -
6.4 Analysis of the Results..............................................................................................- 12 -
6.5 The Results Browser................................................................................................. - 13 -
7Maintenance..........................................................................................- 15 -
7.1 Routine Maintenance ................................................................................................- 15 -
7.2 Replacement Parts....................................................................................................- 15 -
8Service...................................................................................................- 16 -
9Storage..................................................................................................- 16 -
10 Warranty................................................................................................- 16 -
11 Returned Goods Policy........................................................................- 16 -
Notes...........................................................................................................- 17 -

- 3 -
1 Introduction
The Automatic Non-Woods Metal Noack
Evaporative Apparatus tests for the evaporation loss
tendencies of lubricating oils at temperatures of up to
275°C. The newly designed electrically heated
aluminum block allows for testing without the use of
hazardous Woods Metal. The Noack tester is
equipped with an Electronic regulator allowing for
automatic control of temperature and differential
pressure. The system is managed by an integrated
6.5” Touch Screen Panel PC by means of the Noack
Evaluation Software run by a Windows® based
operating System. The Evaluation Software is
capable of recording all analytical parameters,
allowing for user customizable parameters, methods
and result reports as well as printing graphs and test
results.
• Conforms to ASTM D5800, Procedure B
• 6.5” Integrated Touch Screen Panel PC
• Integrated Vacuum Pump with automatic electronic
control system
• Direct sample temperature measurement via
PT100 probe
• Equipped with high resistant Kalrez valve, inlet filter
to remove product residuals
• USB port for connection to an external printer
and/or external PC
• Storage capacity for more then 60,000 analysis
• CE Marked
This manual provides important information
regarding safety, technical references, installation
requirements, operating condition specifications,
user facility resource requirements, and operating
instructions for the Automatic Non-Woods Metal
Noack Evaporative Apparatus. This manual should
also be used in conjunction with applicable published
laboratory procedures. Information on these
procedures is given in section 1.2.
1.1 Koehler’s Commitment to Our
Customers
Providing quality testing instrumentation and
technical support services for research and testing
laboratories has been our specialty for more than
50 years. At Koehler, the primary focus of our
business is providing you with the full support of
your laboratory testing needs. Our products are
backed by our staff of technically knowledgeable,
trained specialists who are experienced in both
petroleum products testing and instrument service
to better understand your requirements and
provide you with the best solutions. You can
depend on Koehler for a full range of accurate and
reliable instrumentation as well as support for your
laboratory testing programs. Please do not
hesitate to contact us at any time with your
inquiries about equipment, tests, or technical
support.
Toll Free: 1-800-878-9070 (US only)
Tel: +1 631 589 3800 • Fax: +1 631 589 3815
Email: info@koehlerinstrument.com
http://www.koehlerinstrument.com
1.2 Recommended Resources and
Publications
1. American Society for Testing and Materials
(ASTM)
100 Barr Harbor Drive
West Conshohocken, Pennsylvania 19428-
2959, USA
Tel: +1 610 832 9500 • Fax: +1 610 832 9555
http://www.astm.org • email: service@astm.org
ASTM Publication:
ASTM D5800: Standard Test Method
Evaporation Loss of Lubricating Oils by
the Noack Method
2. Energy Institute (IP)
61 New Cavendish Street
London, WIM 8AR, United Kingdom
Tel: 44 (0)20 7467 7100
Fax: 44 (0)20 7255 1472
http://www.energyinstpubs.org.uk/
IP Publication:
IP 421

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 4 -
1.3 Instrument Specifications
Model: K44100
K44190
Electrical
Requirements:
115V ±15%, 60Hz
220V ±15%, 50/60Hz
Test Temperature
Range:
225°C to 275°C
Measurable
Temperature Range:
0°C to 320°C
Temperature
Resolution:
0.01°C
Temperature
Accuracy:
± 0.2°C
Maximum Ambient
Temperature:
35°C
Relative Humidity: 80%
Heating Power: 420W
Repeatability /
Reproducibility:
According to ASTM
D5800 or better
Dimensions:
(L x W x H, in.(cm)) 15.75 x 17.72 x 17.72
(40 x 45 x 45)
Net Weight: 48.5 lbs (22kg)
2 Safety Information and Warnings
Safety Considerations. The use of this
equipment may involve hazardous materials and
operations. This manual does not purport to address
all of the safety problems associated with the use of
this equipment. It is the responsibility of any user of
this equipment to investigate, research, and
establish appropriate safety and health practices and
determine the applicability of regulatory limitations
prior to use.
DO NOT allow unauthorized persons to access
the analyzer.
STAY ALERT! Do not operate this unit while
under the influence of medication, alcohol, or
drugs.
Equipment Modifications and Replacement
Parts: Any modification or alteration of this
equipment from that of factory specifications is not
recommended and voids the manufacturer warranty,
product safety, performance specifications, and/or
certifications whether specified or implied, and may
result in personal injury and/or property loss.
Replacement parts must be O.E.M. exact
replacement equipment.
Chemical Reagents Information: Chemicals and
reagents used in performing the test may exhibit
potential hazards. Any user must be familiarized
with the possible dangers before use. We also
recommend consulting the Material Data and
Safety Sheet (MSDS) on each chemical reagent
for additional information. MSDS information can
be easily located on the internet at
http://siri.uvm.edu or http://www.sigma-
aldrich.com.
Disposal of the Instrument
Please note the crossed out symbol of a trash can
on the instrument (please see symbol below). This
indicates that the product must be disposed of
separate from waste at the end of its operational
life. The user that intends to dispose of this
equipment must contact the manufacturer to
ensure proper disposal.

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 5 -
3 Getting Started
3.1 Packing List
Automatic Non-Woods Metal Noack Evaporative
Apparatus
Evaporation Crucible for Procedure B
PT100 Probe for Crucible
Test Ball (10)
Nozzle Cleaner
Crucible Holder
Protective Gloves
Hook Wrench
Pliers
Operation Software
3.2 Unpacking
Carefully unpack and place the instrument and
accessories in a secure location. Inspect the unit for
damage. If any damage is found, keep all packing
materials and immediately report the damage to the
carrier. We will assist you with your claim, if
requested. When submitting a claim for shipping
damage, request that the carrier inspect the shipping
container and equipment. Do not return goods to
Koehler without written authorization.
3.3 Installation
Equipment Placement: Place the instrument on a
firm surface in a secure area.
1. Connect the cord cable from the analyzer to
a network that conforms to the electric safety
specifications
2. Set up glassware as described in the
procedure section
3. Turn ON the analyzer
3.4 Additional Accessories Required for
Running Tests
Part Number Description
K44100-1 Glassware Accessory Set
Includes:
2L Glass Bottle (2)
Rubber Stopper (4)
Glass Delivery Tubes
Silicon Connection Tubing
K44100-2 Stand for glass bottles w/
inclined manometer, 0-50
mmH2O
K44100-3 Noack Reference Oil, RL223,1L
KLA-DB-KIT Set of Connectors and Cables
KLA-PT100-CAL
Calibration Decade Box –
PT100 Simulator
3.5 Optional Accessories Available
Part Number Description
K44100-SFW Noack Evaluation Software
K44100-14 Noack Reference Oil, RL208, 1L

- 6 -
4 Descriptions
4.1 Piping Diagram
Figure 1: Piping Diagram

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 7 -
4.2 Instrument Controls and Connections
Figure 2: Top / Front / Left Side View

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 8 -
Figure 3: Top / Left / Back Side View
Figure 4: Analyzer Complete with Assembled Glassware Set

- 9 -
5 Instrument Calibration
1. From the main menu screen (See Figure 5
below) select Diagnostic Maintenance.
Figure 5: Diagnostic Maintenance Panel
2. In the Diagnostic Maintenance Panel (See
Figure 6 below) select Calibration. This will
open the calibration panel.
Figure 6: Diagnostic Maintenance Panel
3. At the top of the Calibration Screen (See Figure
7) three buttons will allow the user to choose
which probe/ sensor will be displayed. The
Calibrated in label directly below these buttons
shows the selected probe’s last calibration date.
Figure 7: Calibration Screen
4. On the left side of the screen, a graph shows
the calibration curve along with the selected
probe’s value in both °C and volts.
5. On the right side of the screen, the Offset field
allows for fine tuning of the calibration.
NOTE:If an offset has been entered, the
original calibration curves will be shown as a
green line. The black line will move according
to the offset value entered in real time.
6. The Perform Calibration button starts the
guided calibration procedure: to perform a
calibration of a temperature probe it is
necessary to use a RTD/PT100 Simulator. The
recommended RTD/PT100 simulator is the
KLA-PT100-CAL Calibrated Decade Box with
KLA-DB-KIT Connectors and Cables.

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 10 -
6 Operation
6.1 Summary of Test Method
A quantity of 65 grams of sample is heated in a
crucible at 250°C with a constant flow of air drawn
through it for 60 minutes. At the end of the test, the
sample is cooled and reweighed. The loss in mass
of the oil is then determined.
6.2 Analysis Setup Procedure
WARNING:DO NOT INSERT THE CRUCIBLE
INTO THE BATH UNTIL THE SOFTWARE
PROMPTS THE USER AT THE START OF
OPERATION.
1. Press the Start Analysis button on the main
menu screen (Figure 5) to open the analysis
startup panel. See Figure 8 below:
Figure 8: Analysis Startup Screen
2. Touch the Analysis Name field to identify the
sample. An on-screen keyboard will pop up
(See Figure 9 below). The analysis name
entered will automatically be listed as the test
title in the results browser. This will allow the
operator to easily recognize the analysis when
using the results browser for review.
Figure 9: On-Screen Keyboard Pop-Up
NOTE:If the analysis name is left blank the
analysis will not start.
3. Select the method by touching the Method
field. Method A is shown in Figure 8 above.
Method A: Temperature Control via PT100
probe in the Heating Block (Bath)
Method B: Temperature Control via PT100
probe in the Crucible.
4. Select the set-point temperature for the test
based on the selected method. Switching the
method will automatically change the
temperature to that prescribed by the method.
5. Weigh the empty crucible without the cover to
the nearest 0.01g and enter the measured
weight.
6. Pour 65g ± 0.1g of sample into the crucible and
enter the measured weight to the nearest 0.01g
into the dedicated fields.
7. When all of the above parameters are entered,
start the analysis by pressing the START
button.
WARNING:DO NOT INSERT THE CRUCIBLE
INTO THE BATH UNTIL THE SOFTWARE
PROMPTS THE USER AT THE START OF
OPERATION.
8. According to the method and temperature
selected, the bath will heat to the prescribed
set-point.
NOTE: The initial heating phase will show the
bath’s temperature even if method B was
selected. During this phase the graph will not
display data and the aspiration will remain idle.
9. While the bath is heating, ensure the glassware
is correctly set-up as shown on Page Figure 1,
Page 6. Verify that the manometer is level and
fill with water as needed so that the inclined
scale reads zero. Make sure the bleed valve
is fully open.
10. Once the bath has reached the set temperature
an audible alarm will sound and the Crucible
Prompt will appear on the screen. See Figure
10:

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 11 -
Figure 10: Crucible / Bath Ready Prompt
11. The shut off chime button will appear at the
top right corner of the screen when the acoustic
alarm is sounded. The volume of the chime is
very loud so the operator may want to shut off
the chime for the current alarm.
NOTE:Shutting off the chime will only silence
the current crucible alarm. Therefore, future
alarms or prompts will still give off an audible
chime.
12. As per the on-screen instructions, ensure that
the crucible’s cover is correctly screwed on the
crucible’s body.
13. Insert the crucible into the bath and connect the
extraction tube to the Y shaped pipe on the
Woulff bottle as illustrated on Page 6, Figure 1.
NOTE:Ensure that the sample PT100
temperature probe is inserted into the crucible
and connected to the analyzer even if Method A
has been selected.
14. Once the crucible is inserted and the extraction
tube is connected, you may touch the screen to
start the vacuum and the analysis.
Figure 11: Aspiration Activation Screen
15. The aspiration / vacuum will now be activated.
See Figure 11 above. Regulate the aspiration /
vacuum to -2 mbar by slowly closing the bleed
valve halfway and regulating the vacuum
regulator on the analyzer’s side in order to
reach 20 ± 0.2mm on the inclined manometer (-
2mbar on screen).
16. The Aspiration label at the right side of the
screen will silently flash during this phase until
the aspiration reaches its desired setting and is
stabilized.
NOTE:Although the aspiration level should
automatically be maintained at -2 mbar it may
need to be readjusted at times due to the
heating vapor tensions. An acoustic alarm will
sound to alert the operator when the aspiration
level derives from the prescribed tolerance
range. It is imperative the aspiration is
readjusted in the shortest time possible.
17. Once the aspiration has been regulated, touch
the screen to start the analysis. The analyzer
will check if the aspiration is stable, if not, the
instructions for its regulation will pop up again
to allow the operator to repeat the procedure.
6.3 Test Procedure
1. The Method icon is disabled while the analysis
is running. If Method B was chosen, the graph
will show the Sample PT100 probe
temperature. The Start Analysis button will
become a STOP button, allowing the operator
to interrupt/abort the analysis.
2. If the analysis is interrupted within 5 minutes
from the start, any collected data is discarded
and the analysis will not be recorded,

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 12 -
otherwise, even an incomplete analysis running
longer than 5 minutes will be recorded.
3. The status bar will describe the progress of the
analysis. In the bottom right hand side of the
screen the graph will display the temperature
data.
4. An alarm will sound along with a full screen
notice three minutes before the end of the
analysis. See Figure 12 below. Touch the
screen to shut off the alarm and close the full
screen notice. The status bar will display the
countdown to the end of the analysis.
Figure 12: End Analysis Screen
5. As the analysis is approaching the end, prepare
3 cm deep water bath at ambient temperature.
This will be used to cool down the crucible;
therefore it should be large enough to hold a
sufficient amount of water. Sufficient cool down
of the crucible will take approximately 30
minutes so accommodate a water bath where
the crucible can safely rest without interfering
with regular laboratory activities. (A laboratory
sink would be sufficient so long as no
interference occurs).
6. When the analysis time reaches 60 minutes
(the countdown reaches 0), a new alarm will
sound. Touch the screen to stop the alarm,
then CAREFULLY remove the crucible and
place it in the above mentioned water bath. Let
the crucible rest in the bath for 30 minutes.
NOTE: WHILE REMOVING THE CRUCIBLE
AND PLACING IT IN THE WATER BATH
MAKE SURE THE SAMPLE DOES NOT
SPLASH THE CRUCIBLE’S COVER.
7. After 30 minutes, carefully dry the outside of the
crucible then, unscrew the cover.
NOTE: This step should be performed without
tilting the crucible. THE SAMPLE SHOULD
NOT SPLASH THE CRUCIBLE’S COVER.
8. Reweigh the crucible with the remaining (un-
evaporated) part of the sample to the nearest
0.01g.
9. In the Results Browser enter the final weight.
The analyzer will calculate the final result.
6.4 Analysis of the Results
Subtract the empty crucible’s weight from the end
result in order to maintain the weight of the
remaining sample. The result of the following
formulas will provide the value of Noack
Evaporation Loss (Test Method D5800)
Method A:
(B-A) – (C-A)
Evaporation Loss = ――――――――
B-A
Where:
A= Empty Crucible weight
B= Crucible plus Sample weight
C= Crucible plus Sample after 1
hour of heating
Method B:
[(M1-M2)/M1] x100
Where:
M1= sample weight before the test
M
2= sample weight after the test
The final result is given to the nearest 0.1%
on both methods
NOTE:The analyzer is able to perform all
calculations on the above mentioned formulas, the
operator is only required to:
Note and introduce the empty
crucible’s weight
Note and introduce the sample’s
weight before the test
Note and introduce the crucible’s
plus sample’s weight after the test
The analyzer automatically gives the result
correlation between the two procedures as
prescribed in the ASTM D5800 method:
Result on B= 1.030 x Result on A
Result on A= 0.970 x Result on B

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 13 -
6.5 The Results Browser
Every analysis performed is automatically
recorded. The results browser grants access
to the analysis archive.
Use the two arrows on the top of the screen
to choose the month/year. See Figure 13
below:
oThe Month’s corresponding days
are shown in gray if no analysis
has been performed.
oThe day is shown in black if the day
holds more than one analysis
oWhen a day is selected, a green
box will appear around that day
From top to bottom the analysis are enlisted
in order of time:
oFor example in the picture below
the two analysis where performed
at 01:38:14 and 02:18:48 on June
25, 2010.
A - sign (as shown in sample 1)
means a final weight has not
been entered
A ± sign (as shown in sample
2) means the sample has been
fully completed.
oThe two arrows located on the side
allow the operator to select the
sample.
oPress the Load Analysis Button (in
between the two arrows) to view
the analysis.
Figure 13: Results Browser Main Screen
If the final weight has not been introduced a
question mark will appear in the Crucible+ Sample
Final Weight Field as shown in Figure 14:
Figure 14: Incomplete Final Weight Screen
To enter the final result, or review any previous
parameters, press the Review Analysis button (top
button on the right hand side). See Figure 15 below:
Figure 15: Review Analysis Screen
Once the final weight is entered the result will
automatically be calculated according to the method
chosen. It is also possible to review/change the
other parameters by selecting the Change button
located on the right hand side of the shown values.
Once the final weight has been introduced press
the ACCEPT button to confirm, or CANCEL to
discard the changes and close the panel. Refer to
Figure 15 above.
Once the analysis has been reviewed and
completed with the final weight, the result browser
will automatically show the result and the equivalent
result for the other available method. See Figure
16:

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 14 -
Figure 16: Procedure Result Equivalent Screen
If a USB Flash Drive is connected, the Export to
USB button will become available. This allows the
analysis to be exported for review on an external
USB. If a printer is installed the Add to Print
Queue and Print Queue will be available.
Any USB Windows XP compatible printer can be
connected to this analyzer. For detailed
instructions on how to install a printer please
contact a Koehler Representative as each printer
has different characteristics and specific drivers.
Touching the Show Details button causes the
analysis to reload and the data to be displayed
second by second on top of the grid. See Figure 17.
Use the buttons on the side to scroll through the
data and the 60’, 10’ and 1’ buttons to change the
zoom factor on the graph located at the bottom.
Figure 17: Temperature Graph Result Screen
If a USB Flash is connected, the Camera Icon is
shown. Use this button to save a screenshot as a
.jpg image.
NOTE:The first panel in the Result Browser will
also be updated when a final weight is entered.
The minus sign between the first time and sample
name is now a ±. See Figure 18 below.
Figure 18: Result Browser Update Screen

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 15 -
7 Maintenance
WARNING:Disconnect power to the unit before
servicing to avoid exposure to high voltages and/or
temperatures which may result in personal injury or
death. If you have any questions about maintaining
your equipment, then please do not hesitate to
contact the Koehler technical service department.
7.1 Routine Maintenance
Each time a test is run clean the crucible and
cover assembly
Each time a test is run clean the Woullf
bottles and pipes
Each time a test is run clean the heating
jacket
Periodically clean the vacuum filter.
7.2 Replacement Parts
Some instrument parts may need to be replaced.
When ordering replacement part(s), please provide
the model number, serial number, and product
shipment date of your equipment so that we can
ensure you will receive the proper replacement
part(s).
Part Number Description
K44100-4 Evaporation Crucible
K44100-5
PT100 Sample Temperature
Probe
K44100-6 Silicon Tubing, 2m
K44100-7 Air Filter (for vacuum pump)
K44100-8
Main electronic board for Noack
Tester
K44100-9 Hook Wrench
K44100-10 Pliers
K44100-11 Crucible Holder
K44100-12 Test Balls, pk of 10
K44100-13 Nozzle Cleaner

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 16 -
8 Service
Under normal operating conditions and with
routine maintenance, the K44100 Noack
Evaporation Tester should not require service. Any
service problem can be quickly resolved by
contacting Koehler’s technical service department
either by letter, phone, fax, or email. In order to
assure the fastest possible service, please provide
us with the following information.
Model Number:
Serial Number:
Date of Shipment:
9 Storage
This laboratory test instrument is equipped with
electrical components. Storage facilities should be
consistent with an indoor laboratory environment.
This testing equipment should not be subjected to
extremes of temperature and/or moisture.
This equipment was shipped from the factory in a
corrugated cardboard container. If long term
storage is anticipated, re-packing the instrument in
a water-resistant container is recommended to
ensure equipment safety and longevity.
10 Warranty
We at Koehler would like to thank you for your
equipment purchase, which is protected by the
following warranty. If within one (1) year from the
date of receipt, but no longer than fifteen (15)
months from the date of shipment, Koehler
equipment fails to perform properly because of
defects in materials or workmanship, Koehler
Instrument Company, Inc. will repair or, at its sole
discretion, replace the equipment without charge
F.O.B. its plant, provided the equipment has been
properly installed, operated, and maintained.
Koehler Instrument Company must be advised in
writing of the malfunction and authorize the return
of the product to the factory. The sole responsibility
of Koehler Instrument Company and the
purchaser’s exclusive remedy for any claim arising
out of the purchase of any product is the repair or
replacement of the product. In no event shall the
cost of the purchaser’s remedy exceed the
purchase price, nor shall Koehler Instrument
Company be liable for any special, indirect,
incidental, consequential, or exemplary damages.
KOEHLER INSTRUMENT COMPANY, INC.
DISCLAIMS ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTIES OF FITNESS FOR A
PARTICULAR PURPOSE. Please save the
shipping carton in the event the equipment needs to
be returned to the factory for warranty repair. If the
carton is discarded, it will be the purchaser’s
responsibility to provide an appropriate shipping
carton.
11 Returned Goods Policy
To return products for credit or replacement, please
contact Koehler Customer Service with your
purchase order number, our packing list/invoice
number, the item(s) to be returned and the reason
for the return. You will be issued a Returned
Authorization (RA) number, which must be
prominently displayed on the shipping container
when you return the material to our plant. Shipping
containers without an RA number prominently
displayed with be returned to the sender. Goods
must be returned freight prepaid. Returns will be
subject to a restocking charge, the application of
which will depend upon the circumstances
necessitating the return. Some returns cannot be
authorized, including certain products purchased
from outside vendors for the convenience of the
customer, products manufactured on special order,
products shipped from the factory past ninety (90)
days, and products which have been used or
modified in such a way that they cannot be returned
to stock for future sale.

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 17 -
Notes

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 18 -
Notes

K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus
Operation and Instruction Manual
K441XX-Manual - 19 -
Notes
This manual suits for next models
1
Table of contents
Other Koehler Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Beckman Coulter
Beckman Coulter Cytomics FC 500 Getting started

Thermo Scientific
Thermo Scientific 146i instruction manual

Thermo Scientific
Thermo Scientific VF-F20-A operating manual

Biohit
Biohit Proline 710520 instruction manual

THORLABS
THORLABS Telesto Series user manual

Light Progress
Light Progress UV-BOX-E3/40H-NX-R manual