manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. KROHNE
  6. •
  7. Measuring Instrument
  8. •
  9. KROHNE OPTIMASS 1400 User manual

KROHNE OPTIMASS 1400 User manual

For initial set up, we strongly recommend that you also refer to the relevant manual!
Quick Start
1
Installation, assembly, start-up and maintenance may only be performed by appropriately
trained personnel. Check the nameplate for correct operating conditions.
For use in hazardous areas, special codes and regulations are applicable. Instruments must
not be connected to power supply before reading instructions described in the supplementary
manual.
This instrument complies with the requirements of the Pressure Equipment Directive. Please
refer to the nameplate for operating condition limits. Instruments must not be pressurised
before reading instructions described in the manual.
This instrument complies with the requirements of the Low Voltage Directive. Instruments must
not be connected to the power supply before reading the instructions described in the manual.
For devices used in SIL applications, additional safety notes apply. For detailed information
refer to the “Safety Manual”.
The responsibility as to the suitability, intended use and corrosion resistance of the used
materials against the measured fluid of this device rests solely with the operator.
For complete documentation (manuals, supplementary manuals, data sheets and certificates)
please refer to www.krohne.com/Downloads.
1 Installation
Quick Start
Hygienic connectionsFlange connections
Handling the meter
Supporting the meter
Mounting positions
Self draining (hygienic)
Special conditions of use to be observed
• The OPTIMASS measuring system and MFC converter must be included in the equipoten-
tial bonding system of any hazardous area installation.
• Allow the electronics to de-energise before opening the electronics enclosure inside a
hazardous area. The time needed to de-energise is 35 minutes for temp. class T6 and 10
minutes for Temp. class T5. There is no waiting time for Temp. class T1
• Use certified cable glands on cable entries. Use a certified plug to seal unused cable
entries.
• Install connecting cables as fixed wiring so that they are protected against damage.
Connecting cables must not exceed the parameters specified in the Supplementary
Instructions for Hazardous Areas.
• For details on the relationship between: maximum permissible ambient temperature, maxi-
mum medium temperature, maximum surface temperature and the relevant temperature
class for each individual meter, please refer to the tables provided in the relevant meter
Handbook and/or the Supplementary Instructions for Hazardous Areas.
If the hygienic installation requires the meter to
be self draining, mount the meter at an angle.
DO NOT mount the meter horizontally!
04/2020 - 4008121701- QS-OPTIMASS 1400-R01- en
OPTIMASS 1400
Twin, straight tube mass flowmeter
1
2
CAUTION!
If there is excessive vibration in the process pipework, you must isolate the meter from its
mounting. It is recommended that you place an insert made from rubber (or similar material)
between the meter body, or pipework, and the mounting clamp or fixing point. For more
information, contact the manufacturer.
Danger:
• Turn offthe power supply before working on electrical connections.
• Check the voltage data on the nameplate
• Observe national regulations for electrical installations
• The device MUST be grounded in accordance with regulations to protect personnel
against electric shock
Vibration from pumps and other equipment in the process line can have an effect on the meter that, in
some situations, can cause damage. Damage is most likely to occur if the meter has hygienic
connections and the level of vibration is excessive.
Excessive vibration
2 Electrical connections
Compact
Procedure
1. Unscrew the housing cover 1
2. Open the touch guard 2
3. Put the electrical cable through the related
cable gland
4. Connect the electrical wires 3
5. Close the touch guard 2
6. Put the signal cable/s through the related cable
gland/s
7. Connect the signal cables 4
8. Replace the housing cover and tighten it with
your hand 5.
Remote
Signal cable preparation
The signal cable is made up of five (twisted) pairs of cables. Each pair also has a drain wire and the
three wires are covered in a foil wrap. The five cable pairs are covered in an outer foil wrap and a braided
screen. The complete cable is protected by a PVC outer sheath
1Outer Sheath
2Braided screen
3Outer foil wrap
4Foil wrap
5Cable pairs
6Drain wires
Strip back all the foil wraps and discard them. Unwrap the braided screen so that it can be twisted to form
a connection underneath the earth clamp. Bend all five drain wires back and twist them together with the
braided screen.
It is important that ALL five drain wires AND the braided screen form part of the shielding.
MFC 400 Remote
Wall mount Field mount
The label (above) on the inside of the terminal
compartment lid, gives information about out-
put connections.
04/2020 - 4008121701- QS-OPTIMASS 1400-R01- en
3
Cable connections (sensor and converter)
Procedure
1 Unscrew the housing cover 1
2 Put the electrical cable/s through the related
cable gland/s 4
3 Connect the conductors 2
4 Run the twisted sink wires and the braided
screen under the earth clamp 3
5 Fix the clamp in place using the two screws
6 Tighten the cable gland/s 4
7 Replace the housing cover and tighten it with
your hand 5
1Signal converter
2Flow sensor
3Cable shielding
4
Shielding
Display
Converter keys
3 Quick setup
1Display (ER1.0.x shown)
2Optical operating keys
Procedure:
Touch the glass directly in front of the optical operating keys to access menus.
If the optical keys are not used for a period of 5 minutes, the display will return to the measuring
mode.
2
1
Zero calibration
ER 1.0.x ER 2.x.x
The menus shown below are only for the quick setup of the meter. For a complete list of menus please
refer to the relevant converter Handbook.
The quick setup menus shown here are for electronic revision (ER)1.x and electronic revision (ER) 2.x.x
If you are not sure which electronics revision is installed on the converter, please check the nameplate.
Details of standard HART menus are shown in the converter Handbook. For information regarding
menu structures for additional communication protocols, please refer to the relevant Supplementary
Instructions.
Key Measuring mode Menu mode Submenu or function mode Parameter and data mode
Switch from measuring mode to menu
mode. Press Optical key for 2.5s display
shows “A0.0.0.0 Quick Setup”
Access to display menu, and
then 1st submenu is displayed.
Access to displayed submenu
or function
For numerical values, move
cursor (highlighted in blue) one
position to the right
Reset display. "Quick Access" functionReturn to measuring mode Press 1 to 3 times, return to
menu mode, data saved
Return to submenu or function,
data saved
Switch between display pages: measured
value 1 + 2, trend page and status page
Select menu Select submenu or function Use cursor to change number,
unit, property and to move the
decimal point

Return to menu mode without
acceptance of data
Return to submenu or function
without acceptance of data
or
Esc ][ +
Key Screen Description and setting
A Quick Setup Press and hold for 2.5 s,
then release key
2 x C Setup
3 x C1.1.1 Zero Calibration
Calibrate Zero?
Automatic
Please wait
Countdown from 40s
Calibrate zero
Passed
Zero Calibration
+XX.XXX%
Display of measured zero
calibration in %
5 x Save Configuration?
Yes
Display page
Key Screen Description and setting
A0.0.0 Quick Setup Press and hold for 2.5 s,
then release key
2 x C0.0.0 Setup
3 x C1.1.1 Zero Calibration
Calibrate Zero?
Automatic
Please wait
Countdown from 40s
Calibrate zero
Passed
Zero Calibration
+XX.XXX%
Display of measured zero
calibration in %
5 x Save Configuration?
Yes
Display page
04/2020 - 4008121701- QS-OPTIMASS 1400-R01- en
4
Contact
To view all KROHNE locations and contact
details visit:
www.krohne.com
Then “Select your country“ from the drop down
list at the top left of the page
Quick setup menu
Measuring mode Select Menu Select menu and/or submenu Select function & set data
Press for 2.5 s
A Quick Setup A1 Language
A2 Tag
A3 Reset A3.1 Reset Errors
A3.2 All Totalisers
A3.3 Totaliser 1
A3.4 Totaliser 2
A3.5 Totaliser 3
A4 Analogue Output A4.1 Measurement
A4.2 Units
A4.3 Range
A4.4 Low Flow Cutoff
A4.5 Time Constant
A5 Digital Outputs A5.1 Measurement
A5.2 Pulse Value Unit
A5.3 Value p. Pulse
A5.4 Low Flow Cutoff
A6 GDC IR Interface
A7 Flow Direction
A8 Zero Calibration
A9 Operation Mode
ER 1.0.x
Measuring mode Select Menu Select menu and/or submenu Select function & set data
Press for 2.5 s
A Quick Setup

A1.0.0 Language
A2 Reset A2.1.0 Reset Errors
A2.2.0 Stop all Simulations
A2.3.0 All Totalisers
A2.4.0 Totaliser 1
A2.5.0 Totaliser 2
A2.6.0 Totaliser 3
A3 Configuration A3.1.0 Tag
A3.2.0 Measurement
A3.3.0 Range
A3.4.0 Alarm Code
A3.5.0 Low Flow Cutoff
A3.6.0 Damping
A3.7.0 Terminals C Type
A3.8.0 Flow Direction
A 4.0.0 Safety Mode
A 5.0.0 SIL Verification
A 6.0.0 Unlock Device
A 7.0.0 Calibrate Zero
A 8.0.0 Operation Mode
ER 2.x.x
04/2020 - 4008121701- QS-OPTIMASS 1400-R01- en

Other KROHNE Measuring Instrument manuals

KROHNE BM 70 A User manual

KROHNE

KROHNE BM 70 A User manual

KROHNE OPTIFLEX 1300 C User manual

KROHNE

KROHNE OPTIFLEX 1300 C User manual

KROHNE ifc 300 Wiring diagram

KROHNE

KROHNE ifc 300 Wiring diagram

KROHNE ALTOSONIC V12 User manual

KROHNE

KROHNE ALTOSONIC V12 User manual

KROHNE UFC 300 Wiring diagram

KROHNE

KROHNE UFC 300 Wiring diagram

KROHNE OPTIWAVE 7500 C User manual

KROHNE

KROHNE OPTIWAVE 7500 C User manual

KROHNE OPTISWIRL 5080 Wiring diagram

KROHNE

KROHNE OPTISWIRL 5080 Wiring diagram

KROHNE OPTISONIC 3400 User manual

KROHNE

KROHNE OPTISONIC 3400 User manual

KROHNE GA24 Wiring diagram

KROHNE

KROHNE GA24 Wiring diagram

KROHNE OPTIMASS 2000 User manual

KROHNE

KROHNE OPTIMASS 2000 User manual

KROHNE MagCheck User manual

KROHNE

KROHNE MagCheck User manual

KROHNE OPTISWIRL 4200 User manual

KROHNE

KROHNE OPTISWIRL 4200 User manual

KROHNE IFC 020 K User manual

KROHNE

KROHNE IFC 020 K User manual

KROHNE OPTIPROBE User manual

KROHNE

KROHNE OPTIPROBE User manual

KROHNE OPTIFLUX 5000 User manual

KROHNE

KROHNE OPTIFLUX 5000 User manual

KROHNE SD 200 W/R User manual

KROHNE

KROHNE SD 200 W/R User manual

KROHNE OPTISONIC 7300 Wiring diagram

KROHNE

KROHNE OPTISONIC 7300 Wiring diagram

KROHNE OPTISWIRL 4070 User manual

KROHNE

KROHNE OPTISWIRL 4070 User manual

KROHNE ALTOSONIC V User manual

KROHNE

KROHNE ALTOSONIC V User manual

KROHNE OPTIWAVE 7300 C User manual

KROHNE

KROHNE OPTIWAVE 7300 C User manual

KROHNE OPTIMASS 3400 User manual

KROHNE

KROHNE OPTIMASS 3400 User manual

KROHNE OPTIMASS Series User manual

KROHNE

KROHNE OPTIMASS Series User manual

KROHNE DK 46 Wiring diagram

KROHNE

KROHNE DK 46 Wiring diagram

KROHNE OPTISWIRL 5080 Wiring diagram

KROHNE

KROHNE OPTISWIRL 5080 Wiring diagram

Popular Measuring Instrument manuals by other brands

YOKOGAWA ScopeCorder DL750 user manual

YOKOGAWA

YOKOGAWA ScopeCorder DL750 user manual

Stoneridge OPTAC user manual

Stoneridge

Stoneridge OPTAC user manual

Geokon 4050 instruction manual

Geokon

Geokon 4050 instruction manual

Omron K3MA-J Series manual

Omron

Omron K3MA-J Series manual

Ludlum Measurements 44-172 manual

Ludlum Measurements

Ludlum Measurements 44-172 manual

Greenlee CMT-80 instruction manual

Greenlee

Greenlee CMT-80 instruction manual

Kamstrup MULTICAL 602 Technical description

Kamstrup

Kamstrup MULTICAL 602 Technical description

Blue-White ProSeries Sonic-Pro S1 Installation and operating manual

Blue-White

Blue-White ProSeries Sonic-Pro S1 Installation and operating manual

Keller CellaTemp PA 4 Series quick guide

Keller

Keller CellaTemp PA 4 Series quick guide

Rohde & Schwarz EFL 110 Setting up

Rohde & Schwarz

Rohde & Schwarz EFL 110 Setting up

Envent 330S user manual

Envent

Envent 330S user manual

EZY SWITCH SMS-T4 installation manual

EZY SWITCH

EZY SWITCH SMS-T4 installation manual

Task Force Tips EF1 RC MONITOR Instructions for installation, safe operation and maintenance

Task Force Tips

Task Force Tips EF1 RC MONITOR Instructions for installation, safe operation and maintenance

WiT WT901C user manual

WiT

WiT WT901C user manual

PRÜFTECHNIK VIBSCANNER 2 Getting started

PRÜFTECHNIK

PRÜFTECHNIK VIBSCANNER 2 Getting started

AEMC F09 user manual

AEMC

AEMC F09 user manual

Digga TORQUE LOGIC Installation, operating and troubleshooting instructions

Digga

Digga TORQUE LOGIC Installation, operating and troubleshooting instructions

HOPETECH HT3530 user manual

HOPETECH

HOPETECH HT3530 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.