Kuntze K 100 PR User manual

Manual
K 100 PR
K 100 W PR
Measuringandcontrolinstruments
forpHand ORP measurements
7.02 pH 25.0°C
S1 S2 Auto
7.02 pH 25.0°C
S1 S2 Auto

Dr.A. Kuntze GmbH
Robert-Bosch-Str. 7a
D-40668 Meerbusch
Tel. +49-21 50-70 66-0
Fax +49-21 50-70 66-60
www.kuntze.com
Dr. A. Kuntze
Gutes Wasser mit System

3
K 100 CM/IL & K 100 W CM/IL
Contents
1. Your K 100 .....................................................................................................4
1.1General and Safety instructions ...............................................................5
1.2Application ...............................................................................................6
1.3Intendeduse ............................................................................................6
1.4Features...................................................................................................7
1.5Technical data ..........................................................................................8
2. Instructions for installation and connections .............................................9
2.1Dimensions ............................................................................................ 10
2.2 Installation K 100 ................................................................................... 11
2.3 Installation K 100 W ............................................................................... 12
2.4Connectiondiagram K 100 ..................................................................... 13
2.5 Connection diagram K 100 W................................................................. 14
3. Operation of the instrument....................................................................... 15
3.1How to adjust parameters....................................................................... 16
3.2Where to look forinformation.................................................................. 17
3.3Menuoverview ........................................................................................ 18
4. Code and laguage...................................................................................... 19
5. Adjustments for measurement .................................................................. 20
5.1 Calibration .............................................................................................. 21
5.2Temperaturecompensation .................................................................... 22
5.3ORPmeasurements............................................................................... 23
6. Adjustment of the controller ...................................................................... 24
6.1ON/OFFcontroller.................................................................................. 25
6.2P / PIcontroller as impulse-frequencycontroller..................................... 26
6.3 P / PI controller as pulse-pause controller .............................................. 27
6.4Activation and deactivation ofthe controller............................................. 28
6.5Turn-on delay ......................................................................................... 28
6.6External controller stop(digital input) ..................................................... 28
6.7Manual operation of therelays................................................................ 29
6.8Limit values andAlarm ........................................................................... 30
6.9 Dosage check ........................................................................................ 31
7. Data output.................................................................................................. 32
7.1Currentoutput ........................................................................................ 32
7.2Current output as controlleroutput ......................................................... 32
7.3SerialinterfaceRS485 (option) ............................................................... 32
8. Operation and maintenace........................................................................ 33
9. Service ........................................................................................................ 35
9.1Product info............................................................................................ 35
9.2Analog inputs ......................................................................................... 35
9.3 Erase settings (reset)............................................................................. 35
10.Error messages ......................................................................................... 36
Index ............................................................................................................... 37
Customer settings - for reference! ................................................................. 39

K 100 PR & K 100 W PR
4
1. Your K 100
Is an instrument by Dr. A. Kuntze GmbH which offers high quality and reliability for
years.
It is one of our economy series K 100 with which we are trying to meet the rising
demand for low cost / high end instrumentation.
The K 100 instruments are defined by an excellent value for money. They were
developped to maximize functionability on standard applications.
Operation requirements are reduced to an absolute minimum: For standard
electrodes no adjustments are required. The calibration solutions are automatically
recognised, regardless of the order in which they are used. After every single
calibration step the calibration curve is recalculated, so you can work with a 1-point
calibration, if you like.
For any non-standard application you can alter all parameters connected with
calibration: for restricted or outlying measuring ranges you can adjust the calibration
solutions as well as the zero-point of your measurement which is defined by the inner
buffer of the electrode.
All K 100 instruments have an integrated controller which provides bidirectional PI
control, via two relays or as a steady-state controller via the analog output. You can
define a turn-on delay to prevent incorrect dosage after power failure and operate the
controller by remote control. Connect a level sensor, and the fail-safe will shut down
the controller automatically in a low water situation. Activate the dosage check function
to get an alarm if dosage achieves no results, indicating damages in the feeding lines.
Let´s not forget the alarm function with minimum and maximum limit and turn-on
delay...
You have certainly made a good choice. On the following pages you learn more about
your K 100. If, however, you have further questions or are looking for information not
included in this manual or if you are interested in supplementing products like
sensors or flow cells or in our other instrument series, just give us a call - we will be
delighted to help you!

5
K 100 PR & K 100 W PR
1.1 General and Safety instructions
This manual applies to the following instruments:
Instrument and type Revision date
K 100 PR 01/08
K 100 W PR 01/08
It contains technical information for the installation, start-up and maintenance. If you
have any questions not included in this manual please contact your supplier or the
official representative of Dr. A. Kuntze GmbH in your country.
We would like to point out that the warranties specified in our general trading
conditions are valid only if
- installation, connections, adjustments, start-up, and maintenance of the
instrument are carried out by authorized personnel with adequate qualification.
- the instrument is used according to the description in this manual.
Please check for damages immediately after receiving the instruments and report any
damages within 24 hours to the delivering company. Never work with a damaged
instrument.
Keep this manual at a safe place where you can always look up the safety instructions
and the informations on handling and usage. According to DIN 61010 the manual is
part of the product and must be maintained as long as the instrument is used, and
given to the next owner if the instrument is sold.
This instrument was designed and built according to the safety measurements for
electronic devices and has left our company in perfect working condition. To preserve
this condition and to ensure safe usage follow all instructions carefully and pay special
attention to all warnings issued in this manual. If the instrument is visibly damaged or
has been stored inappropriately or if there are any doubts concerning safe usage, shut
it down and make sure it cannot be restarted by accident.
You will notice that important safety instructions are highlighted:
WARNING highlights instructions for the protection of people. Disregarding these
instructions may cause accidents and injuries!
ATTENTION higlights instructions for the protection of the instrument and the
equipment. Disregarding these instructions may lead to damage or
destruction of the instrument or equipment!
NOTE is used to highlight interesting details.
1.1 General and safety instructions

K 100 PR & K 100 W PR
6
1.2 Application
The instruments K 100 PR and K 100 W PR can be used to measure and control the
pH or ORP value of water. Both have an integrated controller with two set points. With
this you can control actuators such as dosing pumps or valves to add chemicals until
the desired pH or ORP value is reached and maintained.
Applications are detoxication of industrial waste water, water treatment, and
desinfection.
While the controller is set to Automatic, it controls independently the dosing of possibly
hazardous chemicals, according to the measured values.
For safety measures, both the measurement and the calibration are checked for
failure. Failures are indicated in the display and via the alarm relay, which can set off a
horn or lamp or relate to a central control. If that failure makes control unreliable, the
controller is automatically switched off until the failure has been taken care off.
CAUTION The instrument checks the input signals, the calibration results, and
the water flow, if a flow sensor is connected. It cannot detect
erroneous settings or failures in the treatment system, nor can it
check for plausibility! The safety of the system of which the
instrument is part of, lies within the reach of responsibility of whoever
built the system.
1.3 Intended use
Use these instruments only for the monitoring and control of water.
Use only sensors, fittings, and accessories fit for your application.
Ensure that the required measuring conditions are constantly maintained, such as
flow, pressure, temperature, etc.
Set-up the instrument according to this manual. Carry out all the steps described, and
check all measurements and settings before you activate the controller.
Use all available safety measures such as the alarm relay, the dosage check, and the
lack-of-water indication.
Regularly check that all safety measures are in good working order.
CAUTION The protection built into the instrument is impaired if they are not used
as intended!
1.2Application and intendeduse

7
K 100 PR & K 100 W PR
1.4 Features
Meter
Measuring ranges 0.00 - 14.00 pH
-1500 - 1500 mV
-30.0 - 140.0 °C
Display Measured value with dimension
Temperature with dimension
Status display sensor, calibration, controller & alarm
Temperature compensation manual or automatic with Pt100
Calibration automatic recognition of the calibration solutions
in arbitrary order
Controller
Set points 2 set points with adjustable direction
Controller types ON/OFF controller with hysteresis
P controller as Pulse-Pause-, Impulse-Frequency- or steady controller
PI controller as Pulse-Pause-, Impulse-Frequency- or steady controller
Hysteresis ajustable within the measuring range
P range XPadjustable within the measuring range
Integral time TN0 - 2000 sec.
Least pulse 0.1 - 9.9 sec.
Pulse+Pause time 02 - 99 sec.
Impulse frequency 100 - 7200 pulses/h
Turn-on delay 0 - 200 sec.
Dosage check 0 - 90 min
Alarm function min. and max. limit and onset delay
Connections
Relays 3 potential-free contacts (2x controller, 1x alarm)
6A, 250 V, max. 550 VA
Analog output 0/4-20 mA galvanically isolated, max. loading 500 Ohm
Analog inputs 1 measuring input for pH or ORP
1 measuring input for temperature sensor Pt100
Digital input external controller stop or low water indication
NC or NO selectable via menu
Serial interface RS485, Baud rate 9600, data formate 8Bit,
(Option) 1start and 1stop bit, no parity
1.4Features

K 100 PR & K 100 W PR
8
1.5Technical data
Installation panel-type housing wall-mounting housing
Dimensions 96 x 96 x 135 mm (WxHxD) 165 x 160 x 85 mm
Weight 0.8 kg 1.0 kg
Terminals screw terminals spring-loaded terminals
for cables up to 1.5mm2for cables up to 1.5mm2
Protection class Front IP54 IP65
Power supply 230 V +6/-10%, 40..60 Hz, optionally 117 V or 24 V
internal fuse none 230V: 63mA HRC
117V: 125mV HRC
24V: 800mA NRC
Power consumption 10VA
Kontact rating 6A/ 250 V, max. 550 VAresistive load
(with RC protective circuit)
Operation temperature 0 - 50°C
Storage temperature -20 - +65°C
Humidity max. 90% at 40°C non condensing
view
Feature K 100 K 100 W
1.5 Technical data

9
K 100 PR & K 100 W PR
Current up to Capacitor C Resistance R
60 mA 10 nF 260 V 390 Ohm 2 Watt
70 mA 47 nF 260 V 22 Ohm 2 Watt
150 mA 100 nF 260 V 47 Ohm 2 Watt
1,0A 220 nF 260 V 47 Ohm 2 Watt
R
C
2.Instructions for installationand connections
2. Instructions for installation and
connections
Installation:
On the next pages you will find detailed instructions for the installation.
For panel-type meters you have to prepare an opening of 92x92mm. Install the
instrument and fix it with the two mounting clips which were part of the delivery.
You can install instruments in wall-mounting housings either by hanging them upon
the center slot or by sliding the slot under a screw, which is an alternative for limited
space. Either way you have to fix it additionally with two screws.
ATTENTION Install the instrument in a place where it is not put under mechanical
or chemical strain!
Mind the protection class:
K 100: Front IP54
K 100 W: IP65 (closed terminal cover)
Connections:
You will find detailed connection diagrams on the following pages.
Before connecting the power supply check the information on the instrument label!
ATTENTION Input, output and control lines must be installed separate from each
other and separate from power lines!
For inputs and outputs use screened lines and connect the screen on one side only.
The measurements are interference-sensitive. Special screened cables with a very
high insulation over short distances, and an impedance converter for longer
connections.
For the connection of temperature sensors use a low-resistance cable with a large
diameter.
When using the relays, mind that with inductive loads, interference must be
suppressed. If that is not possible, the relay must be protected at the terminal block in
the K 100 by a resistance-capacitance filter or, in case of direct current, by a free-
wheeling diode.

K 100 PR & K 100 W PR
10
2.1 Dimensions
K 100 W
K 100
2.1 Dimensions

11
K 100 PR & K 100 W PR
2.2 Installation K 100
Prepare an
opening of 92 x
92 mm.
Screw tight until
the instrument is
fixedperfectly.
Install the
instrument from
the front side
and fix it with the
two mounting
clips which were
part of the
delivery.
2.2 Installation K 100

K 100 PR & K 100 W PR
12
2.3 Installation K 100 W
2.3 Installation K 100 W
Remove the terminal
cover.
Drill three holes (max.
M5) according to the
drawing. Mind that there
are two ways for
installation: (1) You can
hang the instrument upon
the upper screw. In that
case drill the upper hole
120mm above the lower
two. (2) Or you can slip
the fixture on the back of
the isntrument under the
upper screw. In that case
the upper hole has to be
another 15mm higher.
Mount the instrument and
fix it with the two lower
screws. Close the
terminal cover or start
with the connections.

13
K 100 PR & K 100 W PR
PE
Sensor
Pt100
0/4-20 mA
Dig. IN
Rel 1
Rel 2
Rel 3
L
N
1
2
3
4
5
6
7
8
9 +
10 -
11 +
12 -
13
14
5
4
3
2
1
9
8
7
6
RS 485
-+
15
16
17
18
19
20
21
22
23
24
25
26
K 100
M
K 100K 100
M
2.4Connectiondiagram K 100
2.4 Connection diagram K 100
Connection Terminals Note
pH sensor 1 + 2 1 = Reference electrode = screen
2 = Measuring electrode = core
with impedance 1 - 4 Measurement = brown & white
converter 1 = brown, 2 = white
Supplied voltage = yellow & green
3 = yellow = - 6 V
4 = green = + 6 V
ORP sensor 1 + 2 1 = Measuring electrode = core
2 = Reference electrode = screen
Pt100 5 + 6
Analog output 9 + 10 9 = +, 10 = -, max. load 500 Ohm
Digital input 11 + 12 11 = +, 12 = -, external controller stop
and / or low water indication
Relay 1 18 + 19
Relay 2 20 + 21
Relay 3 22 + 23 Alarm relay
Power supply 24 - 26 Check information on instrument label!
RS485 (Option) Sub-D 3 = +, 8 = -
4/7 bridged activates terminating resistance

K 100 PR & K 100 W PR
14
- + + -
1 2 3 4 5 6 7 10 11 12 13 14 15 16 17 18 19 20 21 22
23 24 25 26 27 28
Sensor Pt100 0/4-20 mA Rel. 1 Rel. 2 Rel. 3 L N PE
RS 485 Dig. IN
A
Display
+ -
M
K 100
M
K 100
2.5 Connection diagram K 100 W
2.5 Connection diagram K 100 W
Connection terminals Note
pH sensor 1 + 2 1 = reference electrode = screen
2 = measuring electrode = core
with impedance 1 - 4 Measurement = brown & white
converter 1 = brown, 2 = white
Supplied voltage = yellow & green
3 = yellow = -6V
4 = green = +6V
ORP sensor 1 + 2 1 = measuring electrode = core
2 = reference electrode = screen
Pt100 6 + 7
Display contrast Display Potentiometer to adjust brightness
Analog output 11 + 12 11 = +, 12 = -, max. load 500 Ohm
Relay 1 14 + 15
Relay 2 16 + 17
Relay 3 18 + 19 Alarm relay
Power supply 20 - 22 Check information on instrument label!
RS485 (Option) 23 + 24 23 = -, 24 = +
Jumper A activates terminating resistance
Digital input 26 + 27 26 = +, 27 = -, external controller stop

15
K 100 PR & K 100 W PR
1 2 3 4 5 6
5.49 pH 25.0°C
S1
S2 Man
7 8 9 10 11
1 Measured value
2 Status relay 1
3 Status relay 2
relay OFF
relay ON
4 Temperature
5 Controller
AUTO: controller ON
MAN : controller OFF
(manual operation
of the relays)
6 Orientation aids
7 Key left ()
8 Key up ()
9 Key down ()
10 Key right ()
3.Operation of the instrument
3. Operation of the instrument
When turned on the instrument shows the measured value and temperature together
with the controller mode (Man) and the status of the relays S1 and S2 (both OFF).
With five membrane keys you can move within the menu:
With key you enter the main menu.
With keys and you move up and down in the menu.
With key you adress a menu or parameter.
With key you leave a menu or store a change.
For your convenience triangles in the display indicate the directions you can take from
your position in the menu.
With key „M“ you get back to the display of the measured values from any point in the
menu. Changes that have not been stored will be lost.
NOTE The instruments K 100 and K 100 W differ only in the enclosures.
Operation and menus are the same.

K 100 PR & K 100 W PR
16
Temp. Comp.
Manual Comp.
Temp. Comp.
Automat. Comp.
Temp. Comp.
Manual Comp.
Enter code
058 Code
Enter code
058 Code
Enter code
062 Code
3.1How to adjust parameters
3.1 How to adjust parameters
1) When you adress a parameter the actual setting is
displayed.
2) Switch to the next alternative setting with key.
3) When you have come to the last alternative, pressing the
key once more will bring you back to the start.
Selection between alternatives
For many parameters you have the choice between two or more alternatives, e. g.
between manual and automatic temperature compensation. For these parameters you
need only key. Switch from one alternative to the next until you either come back to
where you started or until you reach the alternative you were looking for.
With these parameters any changes are immediately valid - there is no need to store
the change.
1) Adress the parameter with key.
2) A double triangle appears behind the number indicating
that the number can be changed now with keysand .
3) Store the new value with key. The double triangle
disappears - the new value is stored.
Adjustment of numerical parameters
Numerical parameters can only be altered when a double triangle is visible behind the
number. This double triangle appears when you adress the parameter with key.
Adjust the parameter with keys and . A short pressure on the key changes the last
decimal by 1. If you keep the key pressed, the value will continue changing until the
pressure is released.
Store the changes with key. The double triangle disappears.
NOTE If you do not want to store the change, press key „M“ instead of
key
.

17
K 100 PR & K 100 W PR
Calibrate
Temp. Comp.
Enter password
Set points
Limit values
Basic settings
Service
5.48 pH 25.0°C
S1S2 Auto
Main menu
Cal. Pt
Contr. settings
Start delay
Analog output
Language
Bus adress
pH / ORP
Dig. input
Basic settings
3.2 Where to look
3.2 Where to look for information
Main display
Enter the menu with key .
Information is in chapter:
Adjustment of the meter - calibration
Adjustment of the meter - temperature compensation
Code and language
Adjustment of the controller - ON/OFF, P/ PI controller
Adjustment of the controller - alarm
Information is in chapter:
Adjustment of the meter - temperature compensation
Adjustment of the controller - ON/OFF, P/PI controller
Adjustment of the controller - activation and deactivation
Data output
Code and language
Data output
Adjustment of the meter - ORP measurement
Adjustment of the controller - activation and deactivation
Main menu and basic settings
The parameters are sorted into two menus: In the main menu you will find all functions
which are used regularly, such as calibration. The menu basic settings contains all
parameters which are set just once during start-up.
On the following pages you will find information on how to adjust parameters and
which parameters you need for which application, in the following order:
1) General adjustments: password and language
2) Adjustments for measurement: calibration and temperature compensation for
pH or selection of ORP measurement
3) Adjustments of the controller: selection of the controller version and
corresponding parameters
4) Adjustments to read out data: analog, digitally and/or as alarm

K 100 PR & K 100 W PR
18
Dig.input
5.48 pH 25.0°C
S1S2 Auto
Main menu
3.3Menuoverview
3.3 Menu overview
Press key to enter the menu.
Meas. val.
6.99pH
Cal. press&
Slope
58.8 mV/pH
Sensor zero
01.2 mV
Temp. comp.
Manualcomp.
Temperature
25.0 °C
Enter password
056 Code
Set point S1
7.50 pH
P range S1
0.20 pH
Integraltime S1
000 sec.
Set point S2
6.50 pH
P range S2
0.20 pH
Integraltime S2
000 sec.
LimitS1
8.00 pH
LimitS2
6.00 pH
start delay
005 sec.
Calibrate
Temp.Comp.
Enter password
Set points
Limit values
Basic settings
Basic settings
Cal.Pt
Contr. settings
Start delay
Analog output
Cal.Pt
0.0 °C
Pulse-freq. S1
00*100/h
Pulse-freq. S2
00*100/h
Direction S1
raise
Direction S2
lower
Hysteresis
0.01 pH
Pulse+Pause
10 sek.
Min. pulse
0.1 sec.
Start delay
ON
Start delay
180 sec.
Select output
0 - 20 mA
Begin 0/4mA
0.00 pH
End 20 mA
14.00 pH
Analog output
Measuring value
Language
Bus adress
pH/ORP
We speak
English
Bus adress
Nr. 00
pH/ORP
pHmeasurement
Buffers used
Product info
Buffer 1
7.00 pH
Buffer 2
4.00 pH
Sensor buffer
7.00 pH
Analog inputs
Clearmemory
Service
Unitno.
Nr. 041
Software date
M/Y 1.00
Product. date
M/Y 1.00
Clearmemory
Press&
Service
Analog inputs
input 1 003 mV
Analog inputs
input 2 25.0°C
Dig.input
NO

19
K 100 PR & K 100 W PR
.
Basic settings .
.
Language
Enter Password
056 Code
Main menu
Basic settings
We speak
english
Language
.
.
Enter password
4.Codeandlaguage
4. Code and laguage
Enter password
To get access to the various parameters you have to enter the correct password:
Code 11 gives access to the parameters „calibration“, „temperature
compensation“, and „set points“.
Code 86 gives access to all parameters and functions.
With any other number it is impossible to select, view or change any parameter.
.
Main menu
Enter password
Language
For the communication with the instrument you can choose from a variety of
languages.
Since choosing a language is part of the basic settings, it requires code 86. If a
different code is set, you will be asked to enter the correct password.

K 100 PR & K 100 W PR
20
Calibrate
Temp. Comp.
.
.
.
Basic settings Cal. Pt100
Buffers used
.
.
pH / ORP
Main menu
Basic settings
5.Adjustmentsfor measurement
5. Adjustments for measurement
Calibration of the pH measurement, display of slope and zero-point.
Selection of manual or automatic temperature compensation,
adjustment of temperature for manual compensation.
Calibration of the temperature measurement
Adjustment of the calibration solutions and the
inner buffer of the electrode.
Selection between pH and ORP measurement.
The instruments K 100 PR and K 100 W PR can be used for pH or ORP
measurements.
pH measurement
pH measurement requires a calibration.
By means of calibration the electrode´s characteristics are determined and stored. The
procedure is a subsequent measurement of two solutions with a known pH. The pH
values of the buffers used asd calibration solutions are stored in the menu basic
settings.
You have to adjust the calibration data only if you want to use special calibration
solutions or electrodes with a special buffer solution.
The electrode´s slope is influenced by temperature. This influence can be
compensated by manual or automatic compensation. Manual compensation means
that the temperature is entered manually. Automatic compensation requires
measuring the temperature with a Pt100.
ORP measurement
At works the instruments are set for pH measurement. For ORP measurements you
have to switch to ORP measurement.
This is the only adjustment. ORP measurements do not require calibration or
temperature compensation. Therefor the parameters „calibration“ and „temperature
compensation“ do not appear in the menu if ORP measurement has been selected.
This manual suits for next models
1
Table of contents
Other Kuntze Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Fluidwell
Fluidwell F010-P manual

Endress+Hauser
Endress+Hauser Cerabar M PMC41 operating instructions

Spectrum
Spectrum 2103AL product manual

Endress+Hauser
Endress+Hauser dosimass operating instructions

PCB Piezotronics
PCB Piezotronics M353B16 Installation and operating manual

Leader Electronics Corp.
Leader Electronics Corp. LMV-181A instruction manual