
employ this method of measurement.
The allowable range of the resistance measuring current I that the calibrator receives from a resistance
measuring device under calibration is rated as 0.1 mA to 3 mA. To ensure accuracy, the resistance
measuring current I from the device under calibration shall be strictly confined within the range . For
further details, see Chapter 17 , “ Specification ” .
Any resistance signal being sourced does not include the resistance component of the lead cables for
source. The whole resistance, when measured at the ends of the lead cables for source, is given by
adding the resistance of the lead cables (approximately 0.1 Ωon a round-trip basis) to the sourced
resistance signal. For source of precise resistance signals, use three-wire or four-wire connection.
If capacitance between the terminals of a device under calibration is greater than 0.1 ų F, the calibrator
may fail to source correct resistance signals.
Step
Step
Step
Step
1:
1:
1:
1:
Using
the
〔OHM
OHM
OHM
OHM
〕
key to select Ohm function. Using the 〔RANGE
RANGE
RANGE
RANGE
〕
key, select the desired range. The
selected function and the default range source value and unit shall be shown in the main districts part of the
LCD.
Step
Step
Step
Step
2
2
2
2
:
:
:
:
Set the output value digit by digit using each pair of 〔〕/〔〕keys.
Each pair of 〔〕/〔〕keys corresponds to each digit of the LCD reading. Each press of the 〔〕/〔〕
key increases or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to
overflow or underflow, allowing you to set the output value without interruption. Holding down the 〔〕/〔
〕
key continuously changes the digit in question. And the value won ’ t change if it is increased or decreased to
the Maxim or Minimum value. Pressing the 〔ZERO
ZERO
ZERO
ZERO
〕key initializes the output set point to the default value(0).
Step
Step
Step
Step
3
3
3
3
:
:
:
:
The calibrator sources the preset resistance value between the output terminals.