
Note)When current is flowing from the upside to the underside of the instrument, reading is
positive(+), on the contrary, reading to be negative(-) when current is flowing from the
undersidetotheupsideoftheinstrument.
6-1-2ACcurrentmeasurement
(1)Setthefunctionselectorswitchto"
〜
A"position.
("
〜
"and"A"markswillbedisplayedontheLCD.)
(2)Place one measured conductor lower than the triangle mark indicated on the fork shaped
sensorandmakeameasurement.(shadedpartinthefigure)
ThenmeasuredvalueisdisplayedontheLCD.
(Whenthecenteroftheconductorisnotlowerthanthetrianglemarkindicatedonthefork
shapedsensor,erroroccurs.)
Note)For the measurement of AC current, zero adjustment, which is required for the
measurementofDCcurrent,isnotnecessary.Currentflowingdirectionhasnorelationto
theindicationpolarity.
6-2Noncontactvoltagedetection(NCV)
This function is to check the presence of voltage without touching wires or electrodes
directly.
AlsocancheckthepresenceofACvoltageincable,outlet,fuseandcircuitbreaker.
[Details]
While voltage is applied to a cable or outlet, the electric field depending on the voltage is
generated. This instrument detects the generated electric field, and verifies the presence of
ACvoltage.Officially,itiscalledasaninstrumentfordetectingelectricalfield.Butitisnota
familiar term, so we call it as "Non contact voltage detection". General detectors detect
voltage by contacting polarized voltage(contacts and terminals). But this instrument is
developedtosatisfythisfunctionandforsafetypurposewithoutcontactingvoltage.
# DANGER
●
To avoid getting an electrical shock, never make measurement on the circuit in which
electricalpotentialoverAC/DC300Vexists.
●
Beforeameasurement,besuretochecktheinstrumentoperationwithwell-knownpower
supply.If"Err"isdisplayedontheLCD,donotmakeameasurement.
●
Donotmakemeasurementwithbatterycoverremoved.
●
Indication on NCV range is a reference value. Make sure to check the voltage with a
preciseequipmentinadvancewhenoperatorwilldirectlytouchorconnectwires.
●
Indication of voltage may be affected by non-grounded metal tube or metal case, the
placewhereaffectedbyothervoltages,handgriporthemeasuringpositionofsensor.
●
Keepyourfingersandhandsbehindthebarrierduringmeasurement.
6-2-1Measurements
(1)Setthefunctionselectorswitchto"NCV"position.
(2)
The sensing mode (100V or 200V) in effect is displayed on the LCD for 1min, and NCV
measurementstarts.
(3)Positionthetippartofforktypedsensoragainstthemeasuredobject.
Whenvoltageisdetected,"Hi"willbedisplayedontheLCD.
(Error could occur depending on the direction, angle and contact surface of the instrument
againstthemeasuredobject.OnNCVrange,dataholdfunctioncannotbeused.
)
Note)
When set the function selector switch to NCV range, self-check function operates and
Indicates"Err",ifthereissomefaultorabnormalcondition.Donotmakeameasurementif
suchindicationdisplayedontheLCD.
5. Preparation
(1)Checkbatteryvoltage
SettheFunctionselectorswitchtothepositionotherthanOFFposition.
Battery Voltage is enough if indications are displayed clearly and "BATT" mark is not
displayed on the LCD. If "BATT" mark is indicated or no indication on the LCD, replace
batterieswithnewoneaccordingtobatteryreplacementproceduresshowninclause8inthis
document.
#
CAUTION
●
IndicationsmaynotbeingdisplayedontheLCDdespitethefunctionselectorswitchisat
thepositionotherthanOFFposition.
Thisisbecausepower-offfunctionoperatedautomaticallyandtheinstrumentturnedoff.
Powerofffunction canbereleasedby turningthefunctionselector switchtoOFF, and
thensetittotherangeonwhichyouwouldliketomakeameasurement.
IfLCDstillblank,batteriesarecompletelyexhausted.Pleasereplacebatteries.
(2)Checkthefunctionselectorswitchissettotheappropriaterange.Andalsocheckdatahold
function is not enabled. If inappropriate range is selected, desired measurement cannot be
made.
6. Measurements
6-1CurrentMeasurement
#
DANGER
●
To avoid getting an electrical shock, never make measurement on the circuit in which
electricalpotentialoverAC/DC300Vexists.
●
Donotmakemeasurementwithbatterycoverremoved.
●
Keepyourfingersandhandsbehindthebarrierduringmeasurement.
#
CAUTION
●
Max.diameterofmeasuredobject(conductor)is
Φ
10mm.
6-1-1DCcurrentmeasurement
(1)Setthefunctionselectorswitchto" A"position.
(" "and"A"markswillbedisplayedontheLCD.)
(2)PressHOLD/0ADJbuttonfor2secormoretoenable0ADJfunctionandadjusttheindication
ontheLCDtobe0.
(Indicationshallbeadjustedto0.Otherwise,erroroccurs.)
(3)Place one measured conductor lower than the triangle mark indicated on the fork shaped
sensorandmakeameasurement.(shadedpartinthefigure)
ThenmeasuredvalueisdisplayedontheLCD.
(Whenthecenteroftheconductorisnotlowerthanthetrianglemarkindicatedonthefork
shapedsensor,erroroccurs.)
1. Safety Warnings
This instrument has been designed and tested according to IEC Publication 61010: Safety
RequirementsforElectronicMeasuring Apparatus. Thisinstructionmanualcontainswarnings
and safety rules which must be observed by the user to ensure safe operation of the
instrument and to retain it in safe condition. Therefore, read through these operating
instructionsbeforestartingusingtheinstrument.
# WARNING
●
Readthroughandunderstandinstructionscontainedinthismanualbeforestartingtouse
theinstrument.
●
Saveandkeepthemanualhandytoenablequickreferencewhenevernecessary.
●
Besuretousetheinstrumentonlyinitsintendedapplications.
●
Besuretounderstandandfollowallsafetyinstructionscontainedinthemanual.
Besuretoobserveaboveinstructions.
Failuretofollowtheaboveinstructionsmaycauseinjury,instrumentdamageand/ordamage
toequipmentundertest.
Thesymbol
#
indicatedontheinstrumentmeansthattheusermustrefertorelatedparts
in the manual for safe operation of the instrument. Be sure to carefully read the
instructionsfollowingeach
#
symbolinthismanual.
#
DANGER
isreservedforconditionsandactionsthatarelikelytocauseseriousorfatal
injury.
#
WARNING
is reserved for conditions and actions that can cause serious or fatal
injury.
#
CAUTION
is reserved for conditions and actions that can cause minor injury or
instrumentdamage.
Followingsymbolsareusedontheinstrumentandintheinstructionmanual.Attentionshouldbe
paid
toeachsymboltoensureyoursafety.
Referto theinstructionsinthemanual.Thissymbolismarked wheretheusermustrefer to
theinstructionmanual
soasnottocausepersonalinjuryorinstrumentdamage.
Indicatesaninstrumentwithdoubleorreinforcedinsulation.
Indicates that this instrument can clamp on bare conductors when measuring
a
voltage
correspondingtotheapplicableMeasurementcategory,whichismarke
dnexttothissymbol.
IndicatesAC(AlternatingCurrent).
IndicatesDC(DirectCurrent).
IndicatesACandDC.
# DANGER
●
NevermakemeasurementonthecircuitaboveAC/DC300V.
●
Donotattempttomakemeasurementinthepresenceofflammablegasses.Otherwise,theuse
oftheinstrumentmaycausesparking,whichcanleadtoanexplosion.
●
Neverattempttousetheinstrumentifitssurfaceoryourhandiswet.
●
Donotexceedthemaximumallowableinputofanymeasurementrange.
●
Donotopenthebatterycoverandtheinstrumentcasewhenmakingmeasurement.
●
Nevertryto makemeasurementifany abnormalconditions,suchas broken
Transformer
jawsorcaseisnoted.
●
The instrument is to be used only in its intended applications or conditions. Otherwise,
safety functions equipped with the instrument doesn
t work, and instrument damage
or
seriouspersonalinjurymaybecaused.
#
WARNING
●
Neverattempttomakeanymeasurement,iftheinstrumenthasanystructuralabnormality
noted,suchascrackedcaseorexposedmetalparts.
●
Do not install substitute parts or make any modification to the instrument. Return the
instrumenttoKyoritsuoryourdistributorforrepairorre-calibration.
●
Donottrytoreplacethebatteriesifthesurfaceoftheinstrumentiswet.
●
Alwaysswitchofftheinstrumentbeforeopeningthebatterycompartmentcoverforbattery
replacement.
#
CAUTION
●
Alwaysmakesure to checkthefunctionselectorswitch is settotheappropriate range
beforestartingmeasurement.
●
Donotexposetheinstrumenttothedirectsun,hightemperatureandhumidityordewfall.
●
Be sure to set the function selector switch to the "OFF" position after use. When the
instrumentwillnotbeinuseforalongperiod,placeitinstorageafterremovingthebatteries.
●
Useaclothdippedinwaterorneutraldetergentforcleaningtheinstrument.Donotuse
abrasivesorsolvents
Fork Jaws
Barrier
Function Selector Switch
Data Hold / Zero Adjustment Button
Hand Strap
Display
Low Battery Warning
AC/DC
NCV Range
Data Hold Mode
Unit
Fork Jaws
Barrier
Function Selector Switch
Data Hold / Zero Adjustment Button
Hand Strap
Display
Low Battery Warning
AC/DC
NCV Range
Data Hold Mode
Unit
Place the center of the
measured conductor lower
than the triangle mark indicated
on the fork shaped sensor.
(shaded part in the figure)
IEC61326(EMCstandard)
EN50581(RoHS)
●Indication
LCDMax.1049units,symbols
●Over range display
"OL"symbolisdisplayedontheLCD.
(Onlyoncurrentrange)
●Response time
Approx.2sec.
●Sampling rate
Approx.twicepersecond
●Location for use
Indooruse,Altitudeupto2000m
●Temperature &
Humidity range
(guaranteed accuracy)
23
℃±
5
℃
Relativehumidity:75%orless(nocondensation)
●Operating Temperature &
Humidity range
0
〜
40
℃
Relativehumidity:85%orless(nocondensation)
●Storage Temperature &
Humidity range
-20
〜
60
℃
Relativehumidity:85%orless(nocondensation)
●Power source
DC3V:R03(UM-4)x2pcs
●Current consumption
Approx.12mAorless
Todecreasecurrentconsumption,detectingcircuitisononlyfor
0.1/0.5sec.
●Power off function
Powerofffunctionoperatesautomaticallyafteraswitchremains
for10min.
●Overload Protection
AC/DCcurrent:AC/DC120A/10sec.
ACvoltage(NCV):AC360V/10sec.
●Withstand Voltage
AC3470VforFivesec.
(Betweenelectricalcircuitandenclosures.)
●Insulation Resistance
10M
Ω
/1000V
(Betweenelectricalcircuitandenclosures.)
●Max. diameter of
measured object
Max.10mm
●Dimensions
161.3(L)x40.2(W)x30.3(D)mm
●Weight
110g(includingbatteries)
●Accessories
BatteryR03---------------------2
Instructionmanual------------1
Carryingcase-------------------1
Reference
*Effective Value (RMS)
Mostalternatingcurrentsandvoltagesareexpressedineffectivevalues,whicharealsoreferred
toasRMS(Root-Mean-Square)values. Theeffectivevalue isthesquarerootoftheaverageof
squareofalternatingcurrentorvoltagevalues.Manyclampmetersusingaconventional
rectifying circuit have "RMS" scales for AC measurement. The scales are, however, actually
calibratedintermsoftheeffectivevalueofasinewavethoughtheclampmeterisrespondingto
theaveragevalue.Thecalibrationisdonewithaconversionfactorof1.111forsinewave,which
isfoundbydividingtheeffectivevaluebytheaveragevalue.Theseinstrumentsarethereforein
erroriftheinputvoltageorcurrenthassomeothershapethansinewave.
*CF(CrestFactor)isfoundbydividingthepeakvaluebytheeffectivevalue.
Examples:
Sinewave:CF=1.414
Squarewavewitha1:4dutyratio:CF=2
3. Specification
ACcurrent
〜
A
DCcurrent A
ACvoltage
〜
V
Note)NCVrangeiscalibratedtodetectthevoltage,wherenon-groundedsinglewire,AC80Vor
more.However,detectingsensitivitymaybeaffectedby theabsenceofgroundedornon-
groundedmetaltubeormetalcase.Alsoitmaybeaffectedintheplacewhereinfluenced
byothervoltages,howyougriptheinstrumentorthemeasuringpositionofsensor.
● CF(Crest Factor)
CF=2.5orless
● Standards
IEC61010-1
MeasurementCATIII300V,pollutiondegree2
IEC61010-2-032
InstructionManual
6-2-2Sensingmode
●
Therearetwotypesofsensingmode:100Vmodeand200Vmode.
●
Abovetwomodescanbechangedoverbypressingthedataholdbutton2secormore.
(The selected sensing mode is stored even if switching off the instrument. When setting the
functionswitchto"NCV"again,measurementcanbedoneonthesamemode.)
●
Factorysetting:200Vmode
(1)100Vmode
Sensitivityonthismodeissetsharp,therefore,thepresenceofACvoltagecanbechecked
onlyby placingtheinstrumentclosertothemeasuredobject,such asanoutlet,aplugand
parallelcords,asshowninfigure.
(2)200Vmode
Sensitivityonthismodeissetdull,sotheearthsideandnon-earthsideof100Vcableway
candeverified.(Wherecablesarecrowded,suchasinadistributionboard,earthsidecould
notbeverified.)
AlsocancheckthepresenceofACvoltagein200Vcableway,plug,outlet,fuseandcircuit
breaker.
7. Other functions
7-1Autopowerofffunction
Thisfunctioncausestheinstrument toautomaticallyenter thepower-offmode about10min
afterthelastfunctionselectorswitchoperation.
Toreleasethepower-offfunction,turnofftheinstrumentandthenturnonagain.
7-2Dataholdfunction(OnlyonACA/DCArange)
Thisisafunctiontohold the measured value on theLCD."H"markisshownonthe LCD
whiletheinstrumentisinthedataholdmode.Toexitthedataholdmode,pressthedatahold
buttonagain.
Note)
The measured value being held will be released when auto power-off function operates
whiledataholdfunctionisoperating.
8. Battery replacement
#
WARNING
●
To avoid getting electrical shock, be sure to set the function selector switch to "OFF"
positionbeforetryingtoreplacethebatteries.
#
CAUTION
●
Donotmixnewandoldbatteries.
●
Makesuretoinstallbatteryincorrectpolarityasindicatedinsidethebatterycover.
When"BATT"markisshownontheupperleftcorneroftheLCD,replacethebatteries.
Notethatthebatteryiscompletelyexhausted,theLCDblankswithout"BATT"markshown.
(1)Setthefunctionselectorswitchto"OFF"position.
(2)
Unscrewthebattery coverfixingscrewsand removethebatterycover on thebottomof the
instrument.Thenreplacenewbatteries.(R03x2pcs)
Range
ACA
Measuringrange
0〜
100A
Accuracy
±
2.0
%
rdg
±5
dgt
(
50
/
60Hz
)
±
3.0
%
rdg
±5
dgt
(
50
/
60Hz
)
CF(Crestfactor)
CF
≦2
2<
CF
≦
2.5
Range
DCA
Measuringrange
0〜±
100A
Accuracy
±
2.0
%
rdg
±5
dgt
Range
NCV
Measuringrange
AC300Vorless
Action
Normalcondition:Lo
Atvoltagedetecting(singlewireAC80Vormore):Hi
92-1556E8-16
4. INSTRUMENT LAYOUT
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or
designsdescribedinthismanualwithoutnoticeandwithout
obligations.
AC/DCFORKCURRENTTESTER
KEW
2300R
KEW FORK
Co
rre
ct Wrong
2. Features
●
Thisinstrument,Forkcurrenttester,canmeasureAC/DCcurrentupto100Awithoutopening
andclosingtheJaws.
●
TrueRMSreadingforACcurrent
●
Forkshapedsensorforeasymeasurementattightplacesandcrowdedcableareas.
●
NCVfunction(NonContactVoltage)enableslivewirecheck
●
Autopowerofffunction
●
Dataholdfunction
●
Pocketsizehandytester,adoptedover-moldingforabetterfit
●
Carryingcasefurnishedasastandardaccessory.
●
Designedtointernationalsafetystandards.
IEC61010-2-032measurementCAT.III300VPollutiondegree2
○
MeasurementCategory
Toensuresafeoperationofmeasuringinstruments,IEC61010establishessafetystandards
for various electrical environments, categorized as O to CAT IV, and called measurement
categories. Higher-numbered categories correspond to electrical environments with greater
momentaryenergy,soameasuringinstrumentdesignedforCATIIIenvironmentscanendure
greatermomentaryenergythanonedesignedforCATII.
O :Circuitswhicharenotdirectlyconnectedtothemainspowersupply.
CATII :ElectricalcircuitsofequipmentconnectedtoanACelectricaloutletbyapower
cord.
CATIII :Primaryelectricalcircuitsoftheequipmentconnecteddirectlytothedistribution
panel,andfeedersfromthedistributionpaneltooutlets.
CATIV :Thecircuitfromtheservicedroptotheserviceentrance,andtothepowermeter
andprimaryover-currentprotectiondevice(distributionpanel).
O: Device which is
not directly
connected to the
mains power supply
Barrier:It is a part providing protection against electrical shock and ensuring the minimum
requiredairandcreepagedistances.
KYORITSU ELECTRICAL INSTRUMENTS
WORKS, LTD.
#