manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. KYORITSU
  6. •
  7. Measuring Instrument
  8. •
  9. KYORITSU KEW2040 User manual

KYORITSU KEW2040 User manual

INSTRUCTION MANUAL
DIGITAL CLAMP METER
KEW SNAP
 
SERIES
KEW2040
 
600A AC Only Type
KEW2055
 
1000A AC/DC Type
1.Features
●
Designedtomeetinternationalsafetystandards.
 
IEC61010-1.IEC61010-031:2002&IEC61010-2-032
 
MeasurementCategory(CAT.)IV600V
 
PollutionDegree2
●
Double molded main body provides comfortable

singlehandedgrip
 
●
DataHoldFunction
●
LCD Backlight function to facilitate working at
dimlylitsituations.(KEW2055only)
●
RELfunctiontoindicatemeasurementvariation.
 
(Current,voltage,Resistancemeasurement)
●
MIN/MAX function enables easy reading of min &
maxvalueduringmeasurement.
●
WithContinuity&DiodeCheckFunction
●
NCV
(
Non Contact Voltage
)
Function for wiring
check.
●
600Vinputprotection.
●
Sleepfunctiontoextendbatterylife.
●
WithBarGraph,6039counts
2. Safety Warnings
This instrument has been designed, manufactured and
testedaccordingtoIEC61010:Safetyrequirementsfor
ElectronicMeasuringapparatus,anddeliveredin
the best condition after passed the inspection. This
instructionmanualcontainswarningsandsafety
rules which must be observed by the user to ensure
safeoperation of the instrumentand retain it insafe
condition.
Therefore, read through these operating instructions
beforeusingtheinstrument.
 
WARNING
●
Read through and understand the instructions
containedinthismanualbeforeusingthe
instrument.
●
Keepthemanualathandtoenablequick
referencewhenevernecessary.
●
Theinstrumentistobeusedonlyinitsintended
applications.
●
Understandandfollow allthesafety
instructionscontainedinthemanual.
●
It is essential that the above instructions are
adheredto.
●
Failure to follow the above instructions may
causeinjury,instrumentdamageand/ordamage
toequipmentundertest.
Thesymbol indicatedontheinstrumentmeansthat
theusermustrefertotherelatedpartsinthemanual
forsafeoperationoftheinstrument.It is essential to
read the instructions wherever the  symbol appears
inthemanual.
 
DANGER is reserved for conditions and actions
that are likely to cause serious or fatal injury.
 
WARNING is reserved for conditions and actions
that can cause serious or fatal injury.
 
CAUTION is reserved for conditions and actions
that can cause injury or instrument damage.
●
Marks listed in the table below are used on this
instrument.
Usermustrefertothemanual.
Instrumentwithdoubleorreinforcedinsulation
Indicates that this instrument can clamp on
bare conductors when measuring a voltage
corresponding to the applicable measurement
category,whichismarkednexttothissymbol.
AC
 DC
AC&DC
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltageoverAC600Vexists.
●
Do not attempt to make measurement in the
presence of flammable gasses. Otherwise, the
useoftheinstrumentmaycausesparking,
whichcanleadtoanexplosion.
●
Transformerjawtips aredesignednottoshort
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
●
Never attempttousetheinstrumentifits
surfaceoryourhandiswet.
●
Donotexceedthemaximumallowableinputof
anymeasuringrange.
●
NeveropentheBatterycoverduringa
measurement.
●

Theinstrumentistobeusedonlyinitsintended
applications or conditions. Otherwise, safety
functions equipped with the instrument doesn

t
work,andinstrumentdamageorseriouspersonal
injurymaybecaused.
 
WARNING
●
Never attempt to make measurement if any
abnormalconditions,suchasbrokencase
andexposedmetalpartsarefoundonthe
instrument.
 
●
Do not rotate the Function Switch while the
testleadsarebeingconnected.
●
Do not install substitute parts or make any
modification to the instrument. For repair or
re-calibration, return the instrument to your
localdistributorfromwhereitwaspurchased.
●
Donottrytoreplacethebatteriesifthe
surfaceoftheinstrumentiswet.
●
Disconnect all the cords and cables from the
objectundertestandpowerofftheinstrument
before opening the Battery Cover for Battery
replacement.
●
Verifyproperoperationonaknownsource
before use or taking action as a result of the
indicationoftheinstrument.
●
Use appropriate personal protective equipment
suchasinsulatinggloves,Insulatingboots,and
safetyglasses.

 
CAUTION
●
SettheFunctionSwitchtoanappropriate
positionbeforestartingmeasurement.
 
●
Firmlyinsertthetestleads.
 
●
Disconnect the test leads from the instrument
forcurrentmeasurement.
 
●
Donotexposetheinstrumenttothedirectsun,
hightemperatureandhumidityordewfall.
 
●
Altitude 2000m or less. Appropriate operating
temperatureiswithin0
℃
and40
℃
.
●
Thisinstrumentisn

tdust&waterproofed.Keep
awayfromdustandwater.
●
Be sure to power off the instrument after use.
When the instrument will not be in use for a
longperiod, place it in storage after removing
thebatteries.
●
Usea cloth dipped in water or neutral
detergent for cleaning the instrument. Do not
useabrasivesorsolvents.
Measurement Category
To ensure safe operation of measuring instruments,
IEC 61010 establishes safety standards for various
electrical environments, categorized as O to CAT IV,
and called measurement categories. Higher-numbered
categories correspond to electrical environments with
greatermomentary energy, soameasuringinstrument
designedforCATIII environmentscanendure greater
momentaryenergythanonedesignedforCATII.
O
:Circuits which are not directly connected
tothemainspowersupply.
CAT II
:Electrical circuits of equipment connected
toanACelectricaloutletbyapowercord.
CAT III
:Primary electrical circuits of the
equipment connected directly to the
distribution panel, and feeders from the
distributionpaneltooutlets.
CAT IV
:The circuit from the service drop to the
service entrance, and to the power meter
andprimaryover-currentprotectiondevice
(distributionpanel).
3. Specification
3-1. Measuring range & accuracy
(accuracyguaranteedat23
℃±
5
℃
.humidity45
〜
85%)
ACCurrent600A,1000AFunction
Function Measuring
Range
Accuracy
KEW2040 KEW2055
600A 0-600.0A
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
1000A 0-1000A
N/A
DCCurrent600A,1000AFunction
Function Measuring
Range
Accuracy
KEW2040 KEW2055
600A
0-600.0A
N/A
±
1.5%rdg
±
5dgt
1000A
0-1000A
N/A
ACVoltageFunction

(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
6/60/600V 1-600.0V
±
1.3%rdg
±
4dgt(50/60Hz)
±
3.0%rdg
±
5dgt(40
〜
400Hz)
DCVoltageFunction

(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
600mV/6/
60/600V 0-600.0V
±
1.0%rdg
±
3dgt
Resistance
(
Continuity/DiodeCheck
)
Function
Range Measuring
Range
Accuracy
KEW2040 KEW2055
600
Ω
/6k/
60k/600k
Ω
6M
Ω
0-6M
Ω
±
1.0%rdg
±
5dgt
60M
Ω
6.00M-
60.00M
Ω
±
5%rdg
±
8dgt
ContBuzzer 0-600.0
Ω
Thresholdvalue:60
Ω±
30
Ω

Diode
Testvoltage:0-2V
Frequency/DUTYFunction
(
Auto-rangingfor
Frequency
)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
ACA 40Hz
−
400Hz
±
0.5%rdg
±
5dgt
ACV 1Hz
〜
10kHz
0.1-99.9%
(Pulsewidth/Pulseperiod)
±
2.5%rdg
±
5dgt
Note:Measurableinputsare:40Vrms@ACVor
60Arms@AC600A,350A@AC1000ARange
3-2. General Specification
●
Modeofoperation
 
:
Δ
∑mode
●
Display 
 
:max.6039counts

     
(Frequency:9999)&Bargraph
●

Over-range indication : "OL" displayed when exceeding
t h e  m e a s u r i n g  r a n g e .  ( e x c e p t  f o r  A C / D C V  a n d  1 0 0 0 A 
Function)
●
Rangeswitching :
 
Auto-ranging/Voltage,ResistanceRange
 
Singlerange/Continuity,Diodecheckand
DUTY
●
Samplerate
  
:threetimespersecond
●
Functionalconstruction:
 
OFF/ACA/ACV/DCV/
Ω
KEW2040
 
OFF/ACA/DCA/ACV/DCV/
Ω
KEW2055
●
Keys  :
 
SELECT(AC/DCswitching&/
Ω
/ / ),REL
Δ
,
 
Hz/DUTY,MIN/MAX,HOLD/BackLight(KEW2055)
●
Powersource :DC3V/R03(UM-4)x2pcs
●
Lowbatterywarning
:
" "markisdisplayedat

   
2.4V
±
0.15Vorless.
●

Temperature & humidity : 23
℃±
5
℃
, relative humidity

accuracyguaranteed85%orless(nocondensation)
●
Operatingtemperature:0
〜
40
℃
,relativehumidity
 
&humidityrange85%orless(nocondensation)
●
Storagetemperature :-20
〜
60
℃
,relativehumidity

 
&humidityrange85%orless(nocondensation)
●
Currentconsumption:approx.12mA
●
SleepFunction:Automaticallypoweredoffinabout
 
15minafterthelastFunctionswitchoperation.
 
Press any key or rotate the Function Switch from
OFFtoanypositiontoexitfromtheSleepstate.
●
Locationforuse
Outdooruse,Altitudeupto2000m
●
ApplicableStandards
 
IEC61010-1,61010-2-033
 
MeasurementCAT.IV600VPollutiondegree2
 
IEC61010-031,IEC61010-2-032
 
EMC
:
EN61326-1
 ・
EN55022
 ・
EN61000-4-2(performancecriterionB)
 ・
EN61000-4-3(performancecriterionB)
●
RoHS
:
EN50581
●
OverloadProtection
 
CurrentRange :720AAC/10sec@KEW2040

 
1200AAC/DC/10sec@KEW2055
 
VoltageRange :720VAC/DC/10sec
 
ResistanceRange :600VAC/DC/10sec
●
WithstandVoltage
 
6720VAC
(
TRMS50/60Hz
)
/5sec
 
(betweenJawsandelectricalcircuit/betweeninternal

 
circuitandenclosure)
●
InsulationResistance:10M
Ω
ormore/1000V
 (
betweenelectricalcircuitandenclosure
)
●
Conductorsize
 
KEW2040:approx.33mm
 
KEW2055:approx.40mm
●
Dimension
 
approx.254(L)
×
82(W)
×
36(D)mm/KEW2055
 
approx.243(L)
×
77(W)
×
36(D)mm/KEW2040
●
Weight:approx.:300g@KEW2040
 
:310g@KEW2055
●
Accessories
 
 
TestLeads Model7066A/1set
 
BatteryR03
(
UM-4
)
/2pcs
 
InstructionmanualEnglish,Japanese/1pce
 
CarriyngCaseModel9094/1pcs
3-3. Function Keys
The"
●
"markshowsavailablefunctionateachRange.
DCCurrentMeasurement(KEW2055only)
HOLD SELECT ZERO Hz/
DUTY MAX/
MIN
ACA
●●●●●
ACV
●
-
●●●
DCA
●●●
-
●
DCV
●
-
●
-
●
Ω●●●
-
●
-
●
---
-
●
---
4. Preparation for measurement
4-1. Checking Battery Voltage
Set the Function Switch to any position other than
"
OFF
".Whenthedisplay isclearwithout "
BATT
"mark,
showing, battery voltage is enough. When the display
is blank or
 
"
BATT
"  mark is indicated, replace the
batteriesaccordingtoSection7,BatteryReplacement.
 
CAUTION
TheSleepfeatureautomaticallypowersthe
instrument off in about 15 min after the last
switch or key operation. Therefore, the display
may be blank even with the Function Switch set
to a position other than  "
OFF
". To operate the
instrument in this case, turn the switch back to
the  "
OFF
"  position, then to any other position,
orpressanykey.Replacethebatteriesifnothing
wasdisplayedafteraboveoperations.
4-2. Checking Switch Setting & Operation
Confirm the Function Switch is set to the correct
position,theinstrumentissettothecorrect
measurementmode,
and the Data hold function is disabled. Otherwise,
desiredmeasurementcannotbemade.
5. Measurement
5-1. AC Current Measurement
 
DANGER
●
Never make measurement on a circuit in which
voltage over AC600V exists to avoid getting
electricalshock.
●
Transformer jaw tips are designed not to short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precautionshouldbetakentominimizethe
possibilityofshorting.
●
Donotmake measurement with the Battery
Coverremoved.
●
Disconnectthetestleadsfromtheinstrumentfor
currentmeasurement.
(1)SettheFunctionSwitchto"
600A
" or "
1000A
"
position.(onKEW2040,only"600A"isavailable)
AChasbeenselectedbydefault;presstheSELECT
key, when DC has been selected, to change it to
AC.ACmarkisdisplayedattheupperleftonthe
display.(KEW2055only)
(2)Pressthetriggertoopenthetransformerjawsand
clampthemontotheoneconductorundertest,then
take the reading on the display. Pressing the "
Hz/
DUTY
" Key switches the indication in following
sequence.
AC Current
⇨
Hz
⇨
DUTY
Hz/DUTY
 Function requires 60A or more at AC600A
Rangeand350AormoreatAC1000Arange.
 
CAUTION
●
MaxconductorsizeforKEW2040isapproxdia.
33mm and for KEW2055 is approx dia. 40mm.
During current measurement, keep the trans
former jaws fully closed. Other wise, accurate
measurementscannotbetaken.
●
Keepyourfingersandhandsbehindthebarrier
duringmeasurement.
5-2. DC Current Measurement (KEW2055 only)
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
(1)SettheFunctionSwitchto"
600A
" or "
1000A
"
position. AC has been selected by default; press
the SELECT  key, when AC has been selected, to
changeittoDC.
DCmarkisdisplayedattheupperleftonthe
Barrier
O: Device which is
not directly
connected to the
mains power supply
Barrier
:
It isapartprovidingprotectionagainst
electricalshockand ensuring the minimum
requiredairandcreepagedistances.
5-4. DC Voltage Measurement
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
●

Keep your fingers behind the barrier on the
instrumentduringmeasurement.
(1)SettheFunctionSwitchto"
DCV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connectthe red and black test leads to the
positive (+)and negative (-) sides of the circuit
under test respectively. Take the reading on the
display. If the connection is reversed, the display
indicatesthe"-"mark.
5-5. Resistance/ Cont/ Diode Measurement
 
DANGER
●
Neverusetheinstrumentonanenergizedcircuit.
●
Do not make measurement with the Battery
Coverremoved.
Resistance
(1)Set the Function Switch to  "
Ω/ Cont/ Diode
" 
position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "OL" is
indicatedonthedisplay,andthenshort-circuitthe
tipsoftestleadstomaketheindicationzero.
(3)Connect the test leads to the both ends of the
resistorundertest.
(4)Takethereadingonthedisplay.
     
 
CAUTION
●

Evenifshortthetestleadtips,indicatedvalue
may not be zero. But this is because of the
resistanceoftestleadsandnotafailure.
●
Whentestleadsareopen,"OL"isindicatedon
thedisplay.
●
Keepyourfingersandhandsbehindthebarrier
duringmeasurement.
Continuity
(1)Set the Function Switch to  "
Ω/ Cont/ Diode
" 
position. "
Ω
"  has been selected by default; press
theSELECTkeytochangeitto"
Continuity
".
Resistance ⇨ Cont ⇨ Diode
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "
OL
" is
indicated on the display and short-circuit the tips
of test leads. Indication should become zero and
buzzersounds.
(3)Connect the test leads to the both ends of the
conductor under test. The buzzer sounds, if the
resistanceundertestis100
Ω
orless.
Diode
(1)SettheFunctionSwitchto"
Ω/ Cont/ Diode
"
position. "
Ω
" has been selected by default; press
theSELECTkeytochangeitto"
Diode
"
Resistance ⇨ Cont ⇨ Diode
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)ConnecttheredandblacktestleadstotheAnode
and Cathode of the diode under test respectively.
Takethereading on the display. If the connection
isreversed,thedisplayindicates"
OL
".
 
CAUTION
●
Someof diodescannotbetested.Indication on
thedisplaywillbe"OL".
 

(Zenerdiode,LEDand
soon)
6. Other functions
6-1. Sleep Function
(1)This is a function to prevent the instrument from
beingleftpoweredoninordertoconservebattery
life.
Thisfunctioncausestheinstrumenttoenter
Sleepmodeabout15minutesafterthelast
key operation. To exit the Sleep mode, turn the
Functionswitchto"
OFF
",thentoanyotherposition,
orpressanyKey.
(2)SleepFunctionisdisabledwhen;MIN/MAX
Function is selected. Continuous measurement is
made with the Sleep Function being disabled. To
activate Sleep Function again, disable the MIN/
MAXFunction.
 
CAUTION
●
The instrument consumes small amount of
batterypowerintheSleepmode. Set the
FunctionSwitchtotheOFFpositionafteruse.
6-2. HOLD Key
(1)DataHoldFunction
This is a function to freeze the measured value
onthedisplay.Pressthe"HOLD" key to freeze the
reading.
The reading will be held regardless of subsequent
variationininput."
H
"isindicatedontheupperleft
cornerofthedisplaywhiletheinstrumentisinthe
DataHoldmode.ToexitDataHoldmode,pressthe
"HOLD"keyagain.
    
 
CAUTION
●
Held readings are released when Sleep
Functionisactivatedwhiletheinstrumentisin
theDataHoldmode.
(2)BacklightON/OFF(KEW2055only)
PressingtheHOLDkey2secormorelightsupthe
Backlight. Pressing the HOLD key 2 sec or more
againturnsofftheBacklight.
6-3. NCV Function
Red LED on the upper area on the Panel lights up
at all functions except for OFF when electric field
exceeding 100V is detected by the sensor installed in
theJaws.
Itindicatesapresenceof
voltageinanelectricalcircuit
or equipment without touching
them.
NCVSensorcandetect
electrical field only from the
directionindicatedintheright
figure.
Putthefixedelement(left
side) closer to the conductor
undertest.
Detectionagainstin-wall
outletisimpossible.
 
DANGER
●

The LED may not light up due to installation
condition of electrical circuit or equipment.
Never touch the circuit under test to avoid
possibledangereveniftheLEDforNCV
doesn

tlightup.
●

CheckthefunctionalityofLEDonawell-known
power supply prior to measurement. When the
LEDdoesn

tlightup,donotmakemeasurement.
●
NCVindicationisaffectedbyexternalvoltage,
howtoholdorplacetheinstrument.
6-4. MIN/MAX Function
 
CAUTION
●
SELECT, ZERO, Hz/DUTY keys are disabled
whileMIN/MAXFunctionisbeingactivated.
(1)
AC/DC Current Range
(AC600AonlyonKEW2040)
PressingtheMIN/MAXKey at 600A&1000A
Function enables min or max value measurement.
PresstheMIN/MAXKeytoselectMAXorMIN.The
maxorminvaluewithinmeasuringrange isbeing
held until this function is disabled. "MIN" or "MAX"
is indicated on the display while this function is
being activated. To disable this function, press
down the MIN/MAX Key at least 2 sec or change
functions.
(2)
AC/DC Voltage Range

 
CAUTION
●
Pressing the MIN/MAX Key without applying
voltagedisablestheAuto-rangingfunctionand
fixes the Range to 6V. Connect the test leads
tothecircuitundertestandpresstheMIN/MAX
Key after an appropriate range is selected by
Auto-rangingfunction.
Pressing the MIN/MAX Key enables min or max value
measurement. Press the MIN/MAX Key to select MAX
orMIN.Themaxorminvaluewithinmeasuringrange
is being held until this function is disabled. "MIN" or
"MAX"isindicatedonthedisplaywhilethisfunctionis
beingactivated.
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-5. ZERO Function
 
CAUTION
●

MIN/MAX,PEAK keys are disabled while ZERO
Functionisbeingactivated.
ZeroAdjustment FunctionatCurrentRange "
Δ
"mark
istobeindicatedattheupperrightonthedisplaywhile
ZEROfunctionisbeingoperated.Indicationofrelative
value at Current, Voltage, Resistance: Pressing the
ZERO Key indicates REL (relative value) Press the
ZERO Key to save the initial value at the start of
measurementasareferencevalue.Thenthedifference
betweenthelatermeasured values and the reference
valueisindicatedonthedisplay.
TheAuto-rangingfunction is disabled, while this
functionisbeingactivated, andtheRange isfixedto
theRangeselectedatthestartofmeasurement.
Relativevalueisindicatedwithinfollowingranges.
(
Measuringrange
)
=(Full-scalevalueatthefixedRange)‒(Initialvalue)
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-6. Over-flow indication
When the input exceeds the measuring range at each
FunctionotherthanVoltageand1000ARange,"OL"or
"-OL"isindicatedonthedisplay.
7. Battery Replacement
 
WARNING
●

MIN/MAX,PEAKkeysare disabledwhileZERO
Functionisbeingactivated.
 
CAUTION
●

Donotmixoldandnewbatteries.
●

Installbatteriesincorrectpolarityasindicated
intheBatteryCompartment.
ReplacethebatterieswhenaLow Battery
Voltage warning "BATT" mark is indicated on the
display. Note that when the battery is completely
exhausted,thedisplayblankswithout"
BATT
"mark
shown.
(1)SettheFunctionSwitchto"OFF"position.
(2)Unscrew and remove the Battery Compartment
Coveronthebottomoftheinstrument.
(3)Replace the batteries observing correct polarity.
UsenewR03(AAA)orLR03/1.5Vbatteries.
(4)Install the Battery Compartment and tighten the
screws.
display.
(2)Withthetransformerjawsclosedandwithout
clampingthemontotheconductor,pressthe
"
ZERO
" key to zero adjust the display. (mark is
displayedattheupperrightonthedisplay.)
(3)Pressthetriggertoopenthetransformerjawsand
clampthemontotheoneconductorundertest,the
conductorshouldbeatthecenterofthejaws,then
takethereadingonthedisplay.
(4)Set theFunctionSwitchto an appropriateposition
accordingtocurrentundertest.
(5)Pressing the  "
ZERO
"  key again releases  "
ZERO
" 
function. (mark at the upper right on the display
disappears.)
 
CAUTION
●
When the current flows from the upside (the
displayside)totheundersideoftheinstrument,
thepolarityofthereadingispositiveandvice
versa.
●
Keepyourfingersandhandsbehindthebarrier
duringmeasurement.
5-3. AC Voltage Measurement
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over AC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
●
Keep your fingers behind the barrier on the
instrumentduringmeasurement.
(1)SettheFunctionSwitchto"
ACV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connect the test leads to the circuit under test.
Takethereadingonthedisplay.Pressingthe
"
Hz/DUTY
"keywhilereadingisindicatedon
the display switches the indication in following
sequence.
 
AC Voltage ⇨ Hz ⇨ DUTY
 
CAUTION
●
Hz/DUTYFunctionrequiresAC40Vorhigher.
●
To measure a frequency, measure the voltage
ontheelectricalcircuitinadvance.
●
Then press the Hz/DUTY key to enter into
frequencymeasurement.
●
Readingsof frequency mayfluctuate orbe
influencedundernoisyenvironment.
Anode
Cathode
92-1801C8-16
8. Maintenance
●
Cleaning
Use a cloth dipped in water or neutral detergent for
cleaningtheinstrument.
Donotuseabrasivesorsolvents.Otherwise,instrument
getdamaged,deformedordiscolored.

This manual suits for next models

1

Other KYORITSU Measuring Instrument manuals

KYORITSU KEW2046R User manual

KYORITSU

KYORITSU KEW2046R User manual

KYORITSU KEW 3128 User manual

KYORITSU

KYORITSU KEW 3128 User manual

KYORITSU KEW SNAP Series User manual

KYORITSU

KYORITSU KEW SNAP Series User manual

KYORITSU KEW 2413F User manual

KYORITSU

KYORITSU KEW 2413F User manual

KYORITSU KEW 6305 User manual

KYORITSU

KYORITSU KEW 6305 User manual

KYORITSU KEW3121B User manual

KYORITSU

KYORITSU KEW3121B User manual

KYORITSU KEW6205 User manual

KYORITSU

KYORITSU KEW6205 User manual

KYORITSU KEW 2062 User manual

KYORITSU

KYORITSU KEW 2062 User manual

KYORITSU 6300 Installation guide

KYORITSU

KYORITSU 6300 Installation guide

KYORITSU KEW SNAP 2002PA User manual

KYORITSU

KYORITSU KEW SNAP 2002PA User manual

KYORITSU KEW2200R User manual

KYORITSU

KYORITSU KEW2200R User manual

KYORITSU 2062 User manual

KYORITSU

KYORITSU 2062 User manual

KYORITSU KEW FORK 2300R User manual

KYORITSU

KYORITSU KEW FORK 2300R User manual

KYORITSU KEW 2007R User manual

KYORITSU

KYORITSU KEW 2007R User manual

KYORITSU 8115 User manual

KYORITSU

KYORITSU 8115 User manual

KYORITSU KEW Smart User manual

KYORITSU

KYORITSU KEW Smart User manual

KYORITSU KEW 2413R User manual

KYORITSU

KYORITSU KEW 2413R User manual

KYORITSU KEW 6300 Assembly instructions

KYORITSU

KYORITSU KEW 6300 Assembly instructions

KYORITSU KEW 6310 Installation guide

KYORITSU

KYORITSU KEW 6310 Installation guide

KYORITSU KEW 2200R User manual

KYORITSU

KYORITSU KEW 2200R User manual

KYORITSU 3005A User manual

KYORITSU

KYORITSU 3005A User manual

KYORITSU KEW SNAP 2009R User manual

KYORITSU

KYORITSU KEW SNAP 2009R User manual

KYORITSU KEW3551 User manual

KYORITSU

KYORITSU KEW3551 User manual

KYORITSU KEW SNAP Series User manual

KYORITSU

KYORITSU KEW SNAP Series User manual

Popular Measuring Instrument manuals by other brands

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Oakton pH 310 operating instructions

Oakton

Oakton pH 310 operating instructions

Raycus RFL-A4000D user guide

Raycus

Raycus RFL-A4000D user guide

METREL MD 9273 instruction manual

METREL

METREL MD 9273 instruction manual

Agreto HFM II user manual

Agreto

Agreto HFM II user manual

Endress+Hauser Stamolys CA71SI Series operating instructions

Endress+Hauser

Endress+Hauser Stamolys CA71SI Series operating instructions

RADWAG PUE CY10 operating manual

RADWAG

RADWAG PUE CY10 operating manual

Vynco VMC51 quick start guide

Vynco

Vynco VMC51 quick start guide

Mecmesin Compact Force Gauge operating manual

Mecmesin

Mecmesin Compact Force Gauge operating manual

Horiba Scientific LAQUA-PD210 instruction manual

Horiba Scientific

Horiba Scientific LAQUA-PD210 instruction manual

Asclepion QuadroStar 532/980 user manual

Asclepion

Asclepion QuadroStar 532/980 user manual

C2Max XTREME HD user manual

C2Max

C2Max XTREME HD user manual

CNC CNCWL- 5136-XN Operator's manual

CNC

CNC CNCWL- 5136-XN Operator's manual

Etatron eSelectM 2 PH-CL Operating	instructions and maintenance

Etatron

Etatron eSelectM 2 PH-CL Operating instructions and maintenance

DAKOTA ULTRASONICS UMX-2 Operation manual

DAKOTA ULTRASONICS

DAKOTA ULTRASONICS UMX-2 Operation manual

Ele 45-9300 operating instructions

Ele

Ele 45-9300 operating instructions

Leviton Series 2000 installation instructions

Leviton

Leviton Series 2000 installation instructions

Elster ALPHA A1800 Technical manual

Elster

Elster ALPHA A1800 Technical manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.