manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. KYORITSU
  6. •
  7. Multimeter
  8. •
  9. KYORITSU KEW SNAP Series KEW2040 User manual

KYORITSU KEW SNAP Series KEW2040 User manual

INSTRUCTIONMANUAL
DIGITALCLAMPMETER
KEWSNAP
 
SERIES
KEW2040
 
600AACOnlyType
KEW2055
 
1000AAC/DCType
1.Features
●
Designedtomeetinternationalsafetystandards.
 
IEC61010-1.IEC61010-031:2002&IEC61010-2-032
 
MeasurementCategory(CAT.)IV600V
 
PollutionDegree2
●
Double molded main body provides comfortable
singlehandedgrip
 
●
DataHoldFunction
●
LCD Backlight function to facilitate working at
dimlylitsituations.(KEW2055only)
●
RELfunctiontoindicatemeasurementvariation.
 
(Current,voltage,Resistancemeasurement)
●
MIN/MAX function enables easy reading of min &
maxvalueduringmeasurement.
●
WithContinuity&DiodeCheckFunction
●
NCV
(
Non Contact Voltage
)
Function for wiring
check.
●
600Vinputprotection.
●
Sleepfunctiontoextendbatterylife.
●
WithBarGraph,6039counts
2.SafetyWarnings
This instrument has been designed, manufactured and
testedaccordingtoIEC61010:Safetyrequirementsfor
ElectronicMeasuringapparatus,anddeliveredin
the best condition after passed the inspection. This
instructionmanualcontainswarningsandsafety
rules which must be observed by the user to ensure
safeoperation of the instrument and retainit in safe
condition.
Therefore, read through these operating instructions
beforeusingtheinstrument.
 
WARNING
●
Read through and understand the instructions
containedinthismanualbeforeusingthe
instrument.
●
Keep themanual athandto enablequick
referencewhenevernecessary.
●
Theinstrumentistobeusedonlyinitsintended
applications.
●
Understand andfollowallthesafety
instructionscontainedinthemanual.
●
It is essential that the above instructions are
adheredto.
●
Failure to follow the above instructions may
causeinjury,instrumentdamageand/ordamage
toequipmentundertest.
Thesymbol indicatedontheinstrumentmeansthat
theusermustrefertotherelatedpartsinthemanual
forsafeoperationoftheinstrument.Itisessential to
read the instructions wherever the  symbol appears
inthemanual.

DANGERisreservedforconditionsandactions
thatarelikelytocauseseriousorfatalinjury.

WARNINGisreservedforconditionsandactions
thatcancauseseriousorfatalinjury.

CAUTIONisreservedforconditionsandactions
thatcancauseinjuryorinstrumentdamage.
●
Marks listed in the table below are used on this
instrument.
Usermustrefertothemanual.
Instrumentwithdoubleorreinforcedinsulation
Indicates that this instrument can clamp on
bare conductors when measuring a voltage
corresponding to the applicable measurement
category,whichismarkednexttothissymbol.
AC
 DC
AC&DC
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltageoverAC600Vexists.
●
Do not attempt to make measurement in the
presence of flammable gasses. Otherwise, the
useoftheinstrumentmaycausesparking,
whichcanleadtoanexplosion.
●
Transformerjaw tipsaredesignednotto short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
●
Never attemptto usethe instrumentifits
surfaceoryourhandiswet.
●
Donotexceedthemaximumallowableinputof
anymeasuringrange.
●
Neveropen theBattery coverduringa
measurement.
●

Theinstrumentistobeusedonlyinitsintended
applications or conditions. Otherwise, safety
functions equipped with the instrument doesn

t
work,andinstrumentdamageorseriouspersonal
injurymaybecaused.
 
WARNING
●
Never attempt to make measurement if any
abnormal conditions, such as broken case
and exposed metal parts are found on the
instrument.
 
●
Do not rotate the Function Switch while the
testleadsarebeingconnected.
●
Do not install substitute parts or make any
modification to the instrument. For repair or
re-calibration, return the instrument to your
localdistributorfromwhereitwaspurchased.
●
Donottrytoreplacethebatteriesifthe
surfaceoftheinstrumentiswet.
●
Disconnect all the cords and cables from the
objectunder testandpowerofftheinstrument
before opening the Battery Cover for Battery
replacement.
●
Verify properoperation ona knownsource
before use or taking action as a result of the
indicationoftheinstrument.
●
Use appropriate personal protective equipment
suchasinsulatinggloves,Insulatingboots,and
safetyglasses.

 
CAUTION
●
Set theFunction Switchto anappropriate
positionbeforestartingmeasurement.
 
●
Firmlyinsertthetestleads.
 
●
Disconnect the test leads from the instrument
forcurrentmeasurement.
 
●
Donotexposetheinstrumenttothedirectsun,
hightemperatureandhumidityordewfall.
 
●
Altitude 2000m or less. Appropriate operating
temperatureiswithin0
℃
and40
℃
.
●
Thisinstrumentisn

tdust&waterproofed.Keep
awayfromdustandwater.
●
Be sure to power off the instrument after use.
When the instrument will not be in use for a
longperiod, place it in storageafter removing
thebatteries.
●
Usea clothdipped inwateror neutral
detergent for cleaning the instrument. Do not
useabrasivesorsolvents.
Measurementcategories
(Over-voltagecategories)
To ensure safe operation of measuring instruments,
IEC61010establishessafetystandardsforvarious
electrical environments, categorized as CAT.I to
CAT.IV, and called measurement categories. Higher-
numbered categories correspond to electrical
environments with greater momentary energy,
so a measuring instrument designed for CAT.III
environments can endure greater momentary energy
thanonedesignedforCAT.II.
CA T.Ⅰ
:Secondaryelectrical circuitsconnectedtoan
ACelectricaloutletthroughatransformeror
similardevice.
CAT.Ⅱ
:Primary electrical circuits ofequipment
connected to an AC electrical outlet by a
powercord.
CA T.Ⅲ
:Primary electrical circuits of the equipment
connected directly to the distribution panel,
and feeders from the distribution panel to
outlets.
CAT.Ⅳ
:The circuitfromthe servicedrop tothe
service entrance, and to the power meter
and primary over current protection device
(distributionpanel).
3.Specification
3-1.Measuringrange&accuracy
(accuracyguaranteedat23
℃±
5
℃
.humidity45
〜
85%)
ACCurrent600A,1000AFunction
Function Measuring
Range
Accuracy
KEW2040 KEW2055
600A0-600.0A
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
1000A0-1000A
N/A
DCCurrent600A,1000AFunction
Function Measuring
Range
Accuracy
KEW2040 KEW2055
600A
0-600.0A
N/A
±
1.5%rdg
±
5dgt
1000A
0-1000A
N/A
ACVoltageFunction

(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
6/60/600V1-600.0V
±
1.3%rdg
±
4dgt(50/60Hz)
±
3.0%rdg
±
5dgt(40
〜
400Hz)
DCVoltageFunction

(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
600mV/6/
60/600V0-600.0V
±
1.0%rdg
±
3dgt
Resistance
(
Continuity/DiodeCheck
)
Function
Range Measuring
Range
Accuracy
KEW2040 KEW2055
600
Ω
/6k/
60k/600k
Ω
6M
Ω
0-6M
Ω
±
1.0%rdg
±
5dgt
60M
Ω
6.00M-
60.00M
Ω
±
5%rdg
±
8dgt
ContBuzzer 0-600.0
Ω
Thresholdvalue:60
Ω±
30
Ω

Diode
Testvoltage:0-2V
Frequency/ DUTY Function
(
Auto-ranging for
Frequency
)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
ACA 40Hz
−
400Hz
±
0.5%rdg
±
5dgt
ACV 1Hz
〜
10kHz
0.1-99.9%
(Pulsewidth/Pulseperiod)
±
2.5%rdg
±
5dgt
Note:Measurableinputsare:40Vrms@ACVor
60Arms@AC600A,350A@AC1000ARange
3-2.GeneralSpecification
●
Modeofoperation
 
:
Δ
∑mode
●
Display 
 
:max.6039counts

     
(Frequency:9999)&Bargraph
●

Over-range indication : "OL" displayed when exceeding
the measuring range. (except for AC/DCV and 1000A
Function)
●
Rangeswitching :
 
Auto-ranging/Voltage,ResistanceRange
 
Singlerange/Continuity,Diodecheckand
DUTY
●
Samplerate
  
:threetimespersecond
●
Functionalconstruction:
 
OFF/ACA/ACV/DCV/
Ω
KEW2040
 
OFF/ACA/DCA/ACV/DCV/
Ω
KEW2055
●
Keys  :
 
SELECT(AC/DCswitching&/
Ω
/ / ),REL
Δ
,
 
Hz/DUTY,MIN/MAX,HOLD/BackLight(KEW2055)
●
Powersource :DC3V/R03(UM-4)x2pcs
●
Lowbatterywarning
:
" "markisdisplayedat
 
   
2.4V
±
0.15Vorless.
●

Temperature & humidity : 23
℃±
5
℃
, relative humidity

accuracyguaranteed85%orless(nocondensation)
●
Operatingtemperature:0
〜
40
℃
,relativehumidity
 
&humidityrange85%orless(nocondensation)
●
Storagetemperature :-20
〜
60
℃
,relativehumidity

 
&humidityrange85%orless(nocondensation)
●
Currentconsumption:approx.12mA
●
SleepFunction:Automaticallypoweredoffinabout
 
15minafterthelastFunctionswitchoperation.
 
Press any key or rotate the Function Switch from
OFFtoanypositiontoexitfromtheSleepstate.
●
ApplicableStandards
 
IEC61010-1:2001
 
MeasurementCAT.IV600VPollutiondegree2
 
IEC61010-031:2002,IEC61010-2-032
 
EMC
:
EN61326
 ・
EN55022
 ・
EN61000-4-2(performancecriterionB)
 ・
EN61000-4-3(performancecriterionB)
●
OverloadProtection
 
CurrentRange :720AAC/10sec@KEW2040

 
1200AAC/DC/10sec@KEW2055
 
VoltageRange :720VAC/DC/10sec
 
ResistanceRange :600VAC/DC/10sec
●
WithstandVoltage
 
6880VAC
(
TRMS50/60Hz
)
/5sec
 
(betweenJawsandelectricalcircuit/betweeninternal

 
circuitandenclosure)
●
InsulationResistance:10M
Ω
ormore/1000V
 (
betweenelectricalcircuitandenclosure
)
●
Conductorsize
 
KEW2040:approx.33mm
 
KEW2055:approx.40mm
●
Dimension
 
approx.254(L)
×
82(W)
×
36(D)mm/KEW2055
 
approx.243(L)
×
77(W)
×
36(D)mm/KEW2040
●
Weight:approx.:300g@KEW2040
 
:310g@KEW2055
●
Accessories
 
 
TestLeads Model7066/1set
 
BatteryR03
(
UM-4
)
/2pcs
 
InstructionmanualEnglish,Japanese/1pce
 
CarriyngCaseModel9094/1pcs
3-3.FunctionKeys
The"
●
"markshowsavailablefunctionateachRange.
DCCurrentMeasurement(KEW2055only)
HOLD SELECT ZERO Hz/
DUTY MAX/
MIN
ACA
● ● ● ● ●
ACV
●
-
● ● ●
DCA
● ● ●
-
●
DCV
●
-
●
-
●
Ω ● ● ●
-
●
-
●
- - -
-
●
- - -
4.Preparationformeasurement
4-1.CheckingBatteryVoltage
Set the Function Switch to any position other than
"
OFF
".When the displayisclearwithout "
BATT
" mark,
showing, battery voltage is enough. When the display
is blank or
 
 "
BATT
"  mark is indicated, replace the
batteriesaccordingtoSection7,BatteryReplacement.
 
CAUTION
The Sleep feature automatically powers the
instrument off in about 15 min after the last
switch or key operation. Therefore, the display
may be blank even with the Function Switch set
to a position other than  "
OFF
". To operate the
instrument in this case, turn the switch back to
the  "
OFF
"  position, then to any other position,
orpressanykey.Replacethebatteriesifnothing
wasdisplayedafteraboveoperations.
4-2.CheckingSwitchSetting&Operation
Confirm the Function Switch is set to the correct
position, the instrument is set to the correct
measurementmode,
and the Data hold function is disabled. Otherwise,
desiredmeasurementcannotbemade.
5.Measurement
5-1.ACCurrentMeasurement
 
DANGER
●
Never make measurement on a circuit in which
voltage over AC600V exists to avoid getting
electricalshock.
●
Transformer jaw tips are designed not to short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimizethe
possibilityofshorting.
●
DonotmakemeasurementwiththeBattery
Coverremoved.
●
Disconnectthetestleadsfromtheinstrumentfor
currentmeasurement.
(1)Set theFunction Switchto"
600A
" or "
1000A
"
position.(onKEW2040,only"600A"isavailable)
AChasbeenselectedbydefault;presstheSELECT
key, when DC has been selected, to change it to
AC.ACmarkisdisplayedattheupperleftonthe
display.(KEW2055only)
(2)Press the trigger to open the transformer jaws
andclampthemontotheoneconductorundertest,
then take the reading on the display. Pressing the
"
Hz/DUTY
"Keyswitchestheindicationinfollowing
sequence.
ACCurrent

⇨

Hz

⇨

DUTY
Hz/DUTY
 Function requires 60A or more at AC600A
Rangeand350AormoreatAC1000Arange.
 
CAUTION
●
MaxconductorsizeforKEW2040isapproxdia.
33mm and for KEW2055 is approx dia. 40mm.
During current measurement, keep the trans
former jaws fully closed. Other wise, accurate
measurementscannotbetaken.
5-2.DCCurrentMeasurement(KEW2055only)
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
(1)Set theFunction Switchto"
600A
" or "
1000A
"
position. AC has been selected by default; press
the SELECT  key, when AC has been selected, to
changeittoDC.
DC mark is displayed atthe upperleft onthe
Transformer
INSTRUCTIONMANUAL
DIGITALCLAMPMETER
KEWSNAP
 
SERIES
KEW2040
 
600AACOnlyType
KEW2055
 
1000AAC/DCType
1.Features
●
Designedtomeetinternationalsafetystandards.
 
IEC61010-1.IEC61010-031:2002&IEC61010-2-032
 
MeasurementCategory(CAT.)IV600V
 
PollutionDegree2
●
Double molded main body provides comfortable
singlehandedgrip
 
●
DataHoldFunction
●
LCD Backlight function to facilitate working at
dimlylitsituations.(KEW2055only)
●
RELfunctiontoindicatemeasurementvariation.
 
(Current,voltage,Resistancemeasurement)
●
MIN/MAX function enables easy reading of min &
maxvalueduringmeasurement.
●
WithContinuity&DiodeCheckFunction
●
NCV
(
Non Contact Voltage
)
Function for wiring
check.
●
600Vinputprotection.
●
Sleepfunctiontoextendbatterylife.
●
WithBarGraph,6039counts
2.SafetyWarnings
This instrument has been designed, manufactured and
testedaccordingtoIEC61010:Safetyrequirementsfor
Electronic Measuring apparatus, anddeliveredin
the best condition after passed the inspection. This
instruction manual contains warnings and safety
rules which must be observed by the user to ensure
safeoperation of the instrument and retain it in safe
condition.
Therefore, read through these operating instructions
beforeusingtheinstrument.
 
WARNING
●
Read through and understand the instructions
contained in this manual before using the
instrument.
●
Keep themanual athandto enablequick
referencewhenevernecessary.
●
Theinstrumentistobeusedonlyinitsintended
applications.
●
Understand and follow allthesafety
instructionscontainedinthemanual.
●
It is essential that the above instructions are
adheredto.
●
Failure to follow the above instructions may
causeinjury,instrumentdamageand/ordamage
toequipmentundertest.
Thesymbol indicatedontheinstrumentmeansthat
theusermustrefertotherelatedpartsinthemanual
forsafeoperationoftheinstrument.It is essential to
read the instructions wherever the  symbol appears
inthemanual.

DANGERisreservedforconditionsandactions
thatarelikelytocauseseriousorfatalinjury.

WARNINGisreservedforconditionsandactions
thatcancauseseriousorfatalinjury.

CAUTIONisreservedforconditionsandactions
thatcancauseinjuryorinstrumentdamage.
●
Marks listed in the table below are used on this
instrument.
Usermustrefertothemanual.
Instrumentwithdoubleorreinforcedinsulation
Indicates that this instrument can clamp on
bare conductors when measuring a voltage
corresponding to the applicable measurement
category,whichismarkednexttothissymbol.
AC
 DC
AC&DC
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltageoverAC600Vexists.
●
Do not attempt to make measurement in the
presence of flammable gasses. Otherwise, the
useoftheinstrumentmaycausesparking,
whichcanleadtoanexplosion.
●
Transformerjaw tipsaredesignednotto short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
●
Never attemptto usethe instrumentifits
surfaceoryourhandiswet.
●
Donotexceedthemaximumallowableinputof
anymeasuringrange.
●
Neveropen theBattery coverduringa
measurement.
●

Theinstrumentistobeusedonlyinitsintended
applications or conditions. Otherwise, safety
functions equipped with the instrument doesn

t
work,andinstrumentdamageorseriouspersonal
injurymaybecaused.
 
WARNING
●
Never attempt to make measurement if any
abnormal conditions, such asbrokencase
and exposed metal parts are found on the
instrument.
 
●
Do not rotate the Function Switch while the
testleadsarebeingconnected.
●
Do not install substitute parts or make any
modification to the instrument. For repair or
re-calibration, return the instrument to your
localdistributorfromwhereitwaspurchased.
●
Do not try toreplacethebatteriesifthe
surfaceoftheinstrumentiswet.
●
Disconnect all the cords and cables from the
objectunder testandpowerofftheinstrument
before opening the Battery Cover for Battery
replacement.
●
Verifyproper operationon aknown source
before use or taking action as a result of the
indicationoftheinstrument.
●
Use appropriate personal protective equipment
suchasinsulatinggloves,Insulatingboots,and
safetyglasses.

 
CAUTION
●
Setthe FunctionSwitch toanappropriate
positionbeforestartingmeasurement.
 
●
Firmlyinsertthetestleads.
 
●
Disconnect the test leads from the instrument
forcurrentmeasurement.
 
●
Donotexposetheinstrumenttothedirectsun,
hightemperatureandhumidityordewfall.
 
●
Altitude 2000m or less. Appropriate operating
temperatureiswithin0
℃
and40
℃
.
●
Thisinstrumentisn

tdust&waterproofed.Keep
awayfromdustandwater.
●
Be sure to power off the instrument after use.
When the instrument will not be in use for a
longperiod, place it in storageafter removing
thebatteries.
●
Usea clothdipped inwateror neutral
detergent for cleaning the instrument. Do not
useabrasivesorsolvents.
Measurementcategories
(Over-voltagecategories)
To ensure safe operation of measuring instruments,
IEC61010establishessafetystandardsforvarious
electrical environments, categorized as CAT.I to
CAT.IV, and called measurement categories. Higher-
numbered categories correspond toelectrical
environments with greater momentaryenergy,
so a measuring instrumentdesignedforCAT.III
environments can endure greater momentary energy
thanonedesignedforCAT.II.
CA T.Ⅰ
:Secondaryelectrical circuitsconnectedtoan
ACelectricaloutletthroughatransformeror
similardevice.
CAT.Ⅱ
:Primary electricalcircuitsofequipment
connected to an AC electrical outlet by a
powercord.
CA T.Ⅲ
:Primary electrical circuits of the equipment
connected directly to the distribution panel,
and feeders from the distribution panel to
outlets.
CAT.Ⅳ
:The circuitfromthe servicedrop tothe
service entrance, and to the power meter
and primary over current protection device
(distributionpanel).
3.Specification
3-1.Measuringrange&accuracy
(accuracyguaranteedat23
℃±
5
℃
.humidity45
〜
85%)
ACCurrent600A,1000AFunction
Function Measuring
Range
Accuracy
KEW2040 KEW2055
600A0-600.0A
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
1000A0-1000A
N/A
DCCurrent600A,1000AFunction
Function Measuring
Range
Accuracy
KEW2040 KEW2055
600A
0-600.0A
N/A
±
1.5%rdg
±
5dgt
1000A
0-1000A
N/A
ACVoltageFunction

(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
6/60/600V1-600.0V
±
1.3%rdg
±
4dgt(50/60Hz)
±
3.0%rdg
±
5dgt(40
〜
400Hz)
DCVoltageFunction

(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
600mV/6/
60/600V0-600.0V
±
1.0%rdg
±
3dgt
Resistance
(
Continuity/DiodeCheck
)
Function
Range Measuring
Range
Accuracy
KEW2040 KEW2055
600
Ω
/6k/
60k/600k
Ω
6M
Ω
0-6M
Ω
±
1.0%rdg
±
5dgt
60M
Ω
6.00M-
60.00M
Ω
±
5%rdg
±
8dgt
ContBuzzer 0-600.0
Ω
Thresholdvalue:60
Ω±
30
Ω

Diode
Testvoltage:0-2V
Frequency/ DUTY Function
(
Auto-ranging for
Frequency
)
Range Measuring
Range
Accuracy
KEW2040 KEW2055
ACA 40Hz
−
400Hz
±
0.5%rdg
±
5dgt
ACV 1Hz
〜
10kHz
0.1-99.9%
(Pulsewidth/Pulseperiod)
±
2.5%rdg
±
5dgt
Note:Measurableinputsare:40Vrms@ACVor
60Arms@AC600A,350A@AC1000ARange
3-2.GeneralSpecification
●
Modeofoperation
 
:
Δ
∑mode
●
Display 
 
:max.6039counts

     
(Frequency:9999)&Bargraph
●

Over-range indication : "OL" displayed when exceeding
the measuring range. (except for AC/DCV and 1000A
Function)
●
Rangeswitching :
 
Auto-ranging/Voltage,ResistanceRange
 
Singlerange/Continuity,Diodecheckand
DUTY
●
Samplerate
  
:threetimespersecond
●
Functionalconstruction:
 
OFF/ACA/ACV/DCV/
Ω
KEW2040
 
OFF/ACA/DCA/ACV/DCV/
Ω
KEW2055
●
Keys  :
 
SELECT(AC/DCswitching&/
Ω
/ / ),REL
Δ
,
 
Hz/DUTY,MIN/MAX,HOLD/BackLight(KEW2055)
●
Powersource :DC3V/R03(UM-4)x2pcs
●
Lowbatterywarning
:
" "markisdisplayedat
 
   
2.4V
±
0.15Vorless.
●

Temperature & humidity : 23
℃±
5
℃
, relative humidity

accuracyguaranteed85%orless(nocondensation)
●
Operatingtemperature:0
〜
40
℃
,relativehumidity
 
&humidityrange85%orless(nocondensation)
●
Storagetemperature :-20
〜
60
℃
,relativehumidity

 
&humidityrange85%orless(nocondensation)
●
Currentconsumption:approx.12mA
●
SleepFunction:Automaticallypoweredoffinabout
 
15minafterthelastFunctionswitchoperation.
 
Press any key or rotate the Function Switch from
OFFtoanypositiontoexitfromtheSleepstate.
●
ApplicableStandards
 
IEC61010-1:2001
 
MeasurementCAT.IV600VPollutiondegree2
 
IEC61010-031:2002,IEC61010-2-032
 
EMC
:
EN61326
 ・
EN55022
 ・
EN61000-4-2(performancecriterionB)
 ・
EN61000-4-3(performancecriterionB)
●
OverloadProtection
 
CurrentRange :720AAC/10sec@KEW2040

 
1200AAC/DC/10sec@KEW2055
 
VoltageRange :720VAC/DC/10sec
 
ResistanceRange :600VAC/DC/10sec
●
WithstandVoltage
 
6880VAC
(
TRMS50/60Hz
)
/5sec
 
(betweenJawsandelectricalcircuit/betweeninternal

 
circuitandenclosure)
●
InsulationResistance:10M
Ω
ormore/1000V
 (
betweenelectricalcircuitandenclosure
)
●
Conductorsize
 
KEW2040:approx.33mm
 
KEW2055:approx.40mm
●
Dimension
 
approx.254(L)
×
82(W)
×
36(D)mm/KEW2055
 
approx.243(L)
×
77(W)
×
36(D)mm/KEW2040
●
Weight:approx.:300g@KEW2040
 
:310g@KEW2055
●
Accessories
 
 
TestLeads Model7066/1set
 
BatteryR03
(
UM-4
)
/2pcs
 
InstructionmanualEnglish,Japanese/1pce
 
CarriyngCaseModel9094/1pcs
3-3.FunctionKeys
The"
●
"markshowsavailablefunctionateachRange.
DCCurrentMeasurement(KEW2055only)
HOLD SELECT ZERO Hz/
DUTY MAX/
MIN
ACA
● ● ● ● ●
ACV
●
-
● ● ●
DCA
● ● ●
-
●
DCV
●
-
●
-
●
Ω ● ● ●
-
●
-
●
- - -
-
●
- - -
4.Preparationformeasurement
4-1.CheckingBatteryVoltage
Set the Function Switch to any position other than
"
OFF
".Whenthedisplayisclear without "
BATT
"mark,
showing, battery voltage is enough. When the display
is blank or
 
 "
BATT
"  mark is indicated, replace the
batteriesaccordingtoSection7,BatteryReplacement.
 
CAUTION
The Sleep feature automatically powers the
instrument off in about 15 min after the last
switch or key operation. Therefore, the display
may be blank even with the Function Switch set
to a position other than  "
OFF
". To operate the
instrument in this case, turn the switch back to
the  "
OFF
"  position, then to any other position,
orpressanykey.Replacethebatteriesifnothing
wasdisplayedafteraboveoperations.
4-2.CheckingSwitchSetting&Operation
Confirm the Function Switch is set to the correct
position, the instrument issettothecorrect
measurementmode,
and the Data hold function is disabled. Otherwise,
desiredmeasurementcannotbemade.
5.Measurement
5-1.ACCurrentMeasurement
 
DANGER
●
Never make measurement on a circuit in which
voltage over AC600V exists to avoid getting
electricalshock.
●
Transformer jaw tips are designed not to short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
●
DonotmakemeasurementwiththeBattery
Coverremoved.
●
Disconnectthetestleadsfromtheinstrumentfor
currentmeasurement.
(1)SettheFunction Switchto "
600A
" or "
1000A
"
position.(onKEW2040,only"600A"isavailable)
AChasbeenselectedbydefault;presstheSELECT
key, when DC has been selected, to change it to
AC.ACmarkisdisplayedattheupperleftonthe
display.(KEW2055only)
(2)Press the trigger to open the transformer jaws
andclampthemontotheoneconductorundertest,
then take the reading on the display. Pressing the
"
Hz/DUTY
"Keyswitchestheindicationinfollowing
sequence.
ACCurrent

⇨

Hz

⇨

DUTY
Hz/DUTY
 Function requires 60A or more at AC600A
Rangeand350AormoreatAC1000Arange.
 
CAUTION
●
MaxconductorsizeforKEW2040isapproxdia.
33mm and for KEW2055 is approx dia. 40mm.
During current measurement, keep the trans
former jaws fully closed. Other wise, accurate
measurementscannotbetaken.
5-2.DCCurrentMeasurement(KEW2055only)
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
(1)SettheFunction Switchto "
600A
" or "
1000A
"
position. AC has been selected by default; press
the SELECT  key, when AC has been selected, to
changeittoDC.
DCmark isdisplayed atthe upperleft onthe
Transformer
5-4.DCVoltageMeasurement
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
●

Keep your fingers behind the barrier on the
instrumentduringmeasurement.
(1)SettheFunctionSwitchto"
DCV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connecttheredandblacktestleadstothe
positive (+)and negative (-) sides of the circuit
under test respectively. Take the reading on the
display. If the connection is reversed, the display
indicatesthe"-"mark.
5-5.Resistance/Cont/DiodeMeasurement
 
DANGER
●
Neverusetheinstrumentonanenergizedcircuit.
●
Do not make measurement with the Battery
Coverremoved.
Resistance
(1)Set the Function Switchto  "
Ω/ Cont/ Diode
"
position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "OL" is
indicatedonthedisplay,andthenshort-circuitthe
tipsoftestleadstomaketheindicationzero.
(3)Connect the test leads to the both ends of the
resistorundertest.
(4)Takethereadingonthedisplay.
     
 
CAUTION
●

Evenifshortthetestleadtips,indicatedvalue
may not be zero. But this is because of the
resistanceoftestleadsandnotafailure.
●
Whentestleadsareopen,"OL"isindicatedon
thedisplay.
Continuity
(1)Set the Function Switchto  "
Ω/ Cont/ Diode
"
position. "
Ω
"  has been selected by default; press
theSELECTkeytochangeitto"
Continuity
".
Resistance⇨Cont⇨Diode
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "
OL
" is
indicated on the display and short-circuit the tips
of test leads. Indication should become zero and
buzzersounds.
(3)Connect the test leads to the both ends of the
conductor under test. The buzzer sounds, if the
resistanceundertestis100
Ω
orless.
Diode
(1)Setthe FunctionSwitch to"
Ω/ Cont/ Diode
"
position. "
Ω
" has been selected by default; press
theSELECTkeytochangeitto"
Diode
"
Resistance⇨Cont⇨Diode
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)ConnecttheredandblacktestleadstotheAnode
and Cathode of the diode under test respectively.
Takethereadingonthedisplay.Iftheconnection
isreversed,thedisplayindicates"
OL
".
 
CAUTION
●
Someofdiodescannotbetested.Indicationon
thedisplaywillbe"OL".
 

(Zenerdiode,LEDand
soon)
6.Otherfunctions
6-1.SleepFunction
(1)This is a function to prevent the instrument from
beingleftpoweredoninordertoconservebattery
life.
Thisfunction causesthe instrumentto enter
Sleep mode about 15 minutes after the last
key operation. To exit the Sleep mode, turn the
Functionswitchto"
OFF
",thentoanyotherposition,
orpressanyKey.
(2)SleepFunctionisdisabledwhen;MIN/MAX
Function is selected. Continuous measurement is
made with the Sleep Function being disabled. To
activate Sleep Function again, disable the MIN/
MAXFunction.
 
CAUTION
●
Theinstrument consumessmallamount of
battery power in the Sleep mode.Set the
FunctionSwitchtotheOFFpositionafteruse.
6-2.HOLDKey
(1)DataHoldFunction
This is a function to freeze the measured value
onthedisplay.Pressthe"HOLD"keytofreeze the
reading.
The reading will be held regardless of subsequent
variationininput."
H
"isindicatedontheupperleft
cornerofthedisplaywhiletheinstrumentisinthe
DataHoldmode.ToexitDataHoldmode,pressthe
"HOLD"keyagain.
    
 
CAUTION
●
Heldreadings arereleased whenSleep
Functionisactivatedwhiletheinstrumentisin
theDataHoldmode.
(2)BacklightON/OFF(KEW2055only)
PressingtheHOLDkey2secormorelightsupthe
Backlight. Pressing the HOLD key 2 sec or more
againturnsofftheBacklight.
6-3.NCVFunction
Red LED on the upper area on the Panel lights up
at all functions except for OFF when electric field
exceeding 100V is detected by the sensor installed in
theJaws.
It indicates a presence of
voltageinanelectricalcircuit
or equipment without touching
them.
NCV Sensor can detect
electrical field only from the
directionindicatedintheright
figure.
Put the fixed element (left
side) closer to the conductor
undertest.
Detection against in-wall
outletisimpossible.
 
DANGER
●

The LED may not light up due to installation
condition of electrical circuit or equipment.
Never touch the circuit under test to avoid
possible danger even if the LED for NCV
doesn

tlightup.
●

CheckthefunctionalityofLEDonawell-known
power supply prior to measurement. When the
LEDdoesn

tlightup,donotmakemeasurement.
●
NCVindicationisaffectedbyexternalvoltage,
howtoholdorplacetheinstrument.
6-4.MIN/MAXFunction
 
CAUTION
●
SELECT, ZERO, Hz/DUTY keys are disabled
whileMIN/MAXFunctionisbeingactivated.
(1)
AC/DCCurrentRange
(AC600AonlyonKEW2040)
Pressingthe MIN/MAX Key at 600A & 1000A
Function enables min or max value measurement.
PresstheMIN/MAXKeytoselectMAXorMIN.The
maxorminvalue withinmeasuringrangeisbeing
held until this function is disabled. "MIN" or "MAX"
is indicated on the display while this function is
being activated. To disable this function, press
down the MIN/MAX Key at least 2 sec or change
functions.
(2)
AC/DCVoltageRange

 
CAUTION
●
Pressing the MIN/MAX Key without applying
voltagedisablestheAuto-rangingfunctionand
fixes the Range to 6V. Connect the test leads
tothecircuitundertestandpresstheMIN/MAX
Key after an appropriate range is selected by
Auto-rangingfunction.
Pressing the MIN/MAX Key enables min or max value
measurement. Press the MIN/MAX Key to select MAX
orMIN.Themaxorminvaluewithinmeasuringrange
is being held until this function is disabled. "MIN" or
"MAX"isindicatedonthedisplaywhilethisfunctionis
beingactivated.
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-5.ZEROFunction
 
CAUTION
●

MIN/MAX,PEAKkeysaredisabledwhileZERO
Functionisbeingactivated.
ZeroAdjustmentFunctionatCurrent Range "
Δ
"mark
istobeindicatedattheupperrightonthedisplaywhile
ZEROfunctionisbeingoperated.Indicationofrelative
value at Current, Voltage, Resistance: Pressing the
ZERO Key indicates REL (relative value) Press the
ZERO Key to save the initial value at the start of
measurementasareferencevalue.Thenthedifference
betweenthelatermeasuredvalues and the reference
valueisindicatedonthedisplay.
The Auto-ranging function is disabled, while this
functionisbeingactivated, andtheRangeisfixedto
theRangeselectedatthestartofmeasurement.
Relativevalueisindicatedwithinfollowingranges.
(
Measuringrange
)
=(Full-scalevalueatthefixedRange)‒(Initialvalue)
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-6.Over-flowindication
When the input exceeds the measuring range at each
FunctionotherthanVoltageand1000ARange,"OL"or
"-OL"isindicatedonthedisplay.
7.BatteryReplacement
 
WARNING
●

MIN/MAX,PEAKkeysare disabledwhileZERO
Functionisbeingactivated.
 
CAUTION
●

Donotmixoldandnewbatteries.
●

Installbatteriesincorrectpolarityasindicated
intheBatteryCompartment.
Replace the batteries when a LowBattery
Voltage warning "BATT" mark is indicated on the
display. Note that when the battery is completely
exhausted,thedisplayblankswithout"
BATT
"mark
shown.
(1)SettheFunctionSwitchto"OFF"position.
(2)Unscrew and remove the Battery Compartment
Coveronthebottomoftheinstrument.
(3)Replace the batteries observing correct polarity.
UsenewR03(AAA)orLR03/1.5Vbatteries.
(4)Install the Battery Compartment and tighten the
screws.
display.
(2)Withthetransformerjawsclosedandwithout
clamping them onto the conductor, pressthe
"
ZERO
" key to zero adjust the display. (mark is
displayedattheupperrightonthedisplay.)
(3)Pressthetriggertoopenthetransformerjawsand
clampthemontotheoneconductor undertest,the
conductorshouldbeatthecenterofthejaws,then
takethereadingonthedisplay.
(4)Setthe Function Switch toanappropriateposition
accordingtocurrentundertest.
(5)Pressing the  "
ZERO
"  key again releases  "
ZERO
"
function. (mark at the upper right on the display
disappears.)
 
CAUTION
●
When the current flows from the upside (the
displayside)totheundersideoftheinstrument,
thepolarityofthereading ispositiveandvice
versa.
5-3.ACVoltageMeasurement
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over AC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
●
Keep your fingers behind the barrier on the
instrumentduringmeasurement.
(1)SettheFunctionSwitchto"
ACV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connect the test leads to the circuit under test.
Take the reading on the display. Pressing the
"
Hz/DUTY
" keywhilereadingisindicatedon
the display switches the indication in following
sequence.
 
ACVoltage⇨Hz⇨DUTY
 
CAUTION
●
Hz/DUTYFunctionrequiresAC40Vorhigher.
●
To measure a frequency, measure the voltage
ontheelectricalcircuitinadvance.
●
Thenpress theHz/DUTYkey toenter into
frequencymeasurement.
●
Readingsof frequencymay fluctuateor be
influencedundernoisyenvironment.
Anode
Cathode
92-1801B07-01
8.Maintenance
●
Cleaning
Use a cloth dipped in water or neutral detergent for
cleaningtheinstrument.
Donotuseabrasivesorsolvents.Otherwise,instrument
getdamaged,deformedordiscolored.
5-4.DCVoltageMeasurement
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
●

Keep your fingers behind the barrier on the
instrumentduringmeasurement.
(1)SettheFunctionSwitchto"
DCV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connecttheredandblacktestleadstothe
positive (+)and negative (-) sides of the circuit
under test respectively. Take the reading on the
display. If the connection is reversed, the display
indicatesthe"-"mark.
5-5.Resistance/Cont/DiodeMeasurement
 
DANGER
●
Neverusetheinstrumentonanenergizedcircuit.
●
Do not make measurement with the Battery
Coverremoved.
Resistance
(1)Set theFunction Switchto  "
Ω/ Cont/ Diode
"
position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "OL" is
indicatedonthedisplay,andthenshort-circuitthe
tipsoftestleadstomaketheindicationzero.
(3)Connect the test leads to the both ends of the
resistorundertest.
(4)Takethereadingonthedisplay.
     
 
CAUTION
●

Evenifshortthetestleadtips,indicatedvalue
may not be zero. But this is because of the
resistanceoftestleadsandnotafailure.
●
Whentestleadsareopen,"OL"isindicatedon
thedisplay.
Continuity
(1)Set theFunction Switchto  "
Ω/ Cont/ Diode
"
position. "
Ω
"  has been selected by default; press
theSELECTkeytochangeitto"
Continuity
".
Resistance⇨Cont⇨Diode
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "
OL
" is
indicated on the display and short-circuit the tips
of test leads. Indication should become zero and
buzzersounds.
(3)Connect the test leads to the both ends of the
conductor under test. The buzzer sounds, if the
resistanceundertestis100
Ω
orless.
Diode
(1)Setthe FunctionSwitch to"
Ω/ Cont/ Diode
"
position. "
Ω
" has been selected by default; press
theSELECTkeytochangeitto"
Diode
"
Resistance⇨Cont⇨Diode
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)ConnecttheredandblacktestleadstotheAnode
and Cathode of the diode under test respectively.
Takethereadingonthedisplay.Iftheconnection
isreversed,thedisplayindicates"
OL
".
 
CAUTION
●
Someofdiodescannotbetested.Indicationon
thedisplaywillbe"OL".
 

(Zenerdiode,LEDand
soon)
6.Otherfunctions
6-1.SleepFunction
(1)This is a function to prevent the instrument from
beingleftpoweredoninordertoconservebattery
life.
Thisfunction causesthe instrumentto enter
Sleep mode about 15 minutes after the last
key operation. To exit the Sleep mode, turn the
Functionswitchto"
OFF
",thentoanyotherposition,
orpressanyKey.
(2)SleepFunctionisdisabledwhen;MIN/MAX
Function is selected. Continuous measurement is
made with the Sleep Function being disabled. To
activate Sleep Function again, disable the MIN/
MAXFunction.
 
CAUTION
●
Theinstrument consumessmallamount of
battery power in the Sleep mode. Setthe
FunctionSwitchtotheOFFpositionafteruse.
6-2.HOLDKey
(1)DataHoldFunction
This is a function to freeze the measured value
onthedisplay.Pressthe"HOLD"key to freeze the
reading.
The reading will be held regardless of subsequent
variationininput."
H
"isindicatedontheupperleft
cornerofthedisplaywhiletheinstrumentisinthe
DataHoldmode.ToexitDataHoldmode,pressthe
"HOLD"keyagain.
    
 
CAUTION
●
Heldreadings arereleased whenSleep
Functionisactivatedwhiletheinstrumentisin
theDataHoldmode.
(2)BacklightON/OFF(KEW2055only)
PressingtheHOLDkey2secormorelightsupthe
Backlight. Pressing the HOLD key 2 sec or more
againturnsofftheBacklight.
6-3.NCVFunction
Red LED on the upper area on the Panel lights up
at all functions except for OFF when electric field
exceeding 100V is detected by the sensor installed in
theJaws.
It indicates a presenceof
voltageinanelectricalcircuit
or equipment without touching
them.
NCV Sensor can detect
electrical field only from the
directionindicatedintheright
figure.
Put the fixed element(left
side) closer to the conductor
undertest.
Detection against in-wall
outletisimpossible.
 
DANGER
●

The LED may not light up due to installation
condition of electrical circuit or equipment.
Never touch the circuit under test to avoid
possible danger even iftheLEDforNCV
doesn

tlightup.
●

CheckthefunctionalityofLEDonawell-known
power supply prior to measurement. When the
LEDdoesn

tlightup,donotmakemeasurement.
●
NCVindicationisaffectedbyexternalvoltage,
howtoholdorplacetheinstrument.
6-4.MIN/MAXFunction
 
CAUTION
●
SELECT, ZERO, Hz/DUTY keys are disabled
whileMIN/MAXFunctionisbeingactivated.
(1)
AC/DCCurrentRange
(AC600AonlyonKEW2040)
Pressingthe MIN/MAX Key at 600A & 1000A
Function enables min or max value measurement.
Pressthe MIN/MAXKeytoselectMAXorMIN. The
maxorminvalue within measuringrangeisbeing
held until this function is disabled. "MIN" or "MAX"
is indicated on the display while this function is
being activated. To disable this function, press
down the MIN/MAX Key at least 2 sec or change
functions.
(2)
AC/DCVoltageRange

 
CAUTION
●
Pressing the MIN/MAX Key without applying
voltagedisablestheAuto-rangingfunctionand
fixes the Range to 6V. Connect the test leads
tothecircuitundertestandpresstheMIN/MAX
Key after an appropriate range is selected by
Auto-rangingfunction.
Pressing the MIN/MAX Key enables min or max value
measurement. Press the MIN/MAX Key to select MAX
orMIN.Themaxorminvaluewithinmeasuringrange
is being held until this function is disabled. "MIN" or
"MAX"isindicatedonthedisplaywhilethisfunctionis
beingactivated.
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-5.ZEROFunction
 
CAUTION
●

MIN/MAX,PEAKkeysaredisabledwhileZERO
Functionisbeingactivated.
ZeroAdjustmentFunctionatCurrent Range "
Δ
"mark
istobeindicatedattheupperrightonthedisplaywhile
ZEROfunctionisbeingoperated.Indicationofrelative
value at Current, Voltage, Resistance: Pressing the
ZERO Key indicates REL (relative value) Press the
ZERO Key to save the initial value at the start of
measurementasareferencevalue.Thenthedifference
betweenthelatermeasured values and the reference
valueisindicatedonthedisplay.
The Auto-ranging function is disabled, while this
functionisbeingactivated, and theRangeisfixedto
theRangeselectedatthestartofmeasurement.
Relativevalueisindicatedwithinfollowingranges.
(
Measuringrange
)
=(Full-scalevalueatthefixedRange)‒(Initialvalue)
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-6.Over-flowindication
When the input exceeds the measuring range at each
FunctionotherthanVoltageand1000ARange,"OL"or
"-OL"isindicatedonthedisplay.
7.BatteryReplacement
 
WARNING
●

MIN/MAX,PEAKkeysare disabledwhileZERO
Functionisbeingactivated.
 
CAUTION
●

Donotmixoldandnewbatteries.
●

Installbatteriesincorrectpolarityasindicated
intheBatteryCompartment.
Replace the batteries when a Low Battery
Voltage warning "BATT" mark is indicated on the
display. Note that when the battery is completely
exhausted,thedisplayblankswithout"
BATT
"mark
shown.
(1)SettheFunctionSwitchto"OFF"position.
(2)Unscrew and remove the Battery Compartment
Coveronthebottomoftheinstrument.
(3)Replace the batteries observing correct polarity.
UsenewR03(AAA)orLR03/1.5Vbatteries.
(4)Install the Battery Compartment and tighten the
screws.
display.
(2)Withthetransformerjawsclosedand without
clamping them onto the conductor, press the
"
ZERO
" key to zero adjust the display. (mark is
displayedattheupperrightonthedisplay.)
(3)Pressthetriggertoopenthetransformerjawsand
clampthemonto theoneconductor undertest,the
conductorshouldbeatthecenterofthejaws,then
takethereadingonthedisplay.
(4)Setthe Function Switch toanappropriateposition
accordingtocurrentundertest.
(5)Pressing the  "
ZERO
"  key again releases  "
ZERO
"
function. (mark at the upper right on the display
disappears.)
 
CAUTION
●
When the current flows from the upside (the
displayside)totheundersideoftheinstrument,
thepolarityofthereading ispositiveandvice
versa.
5-3.ACVoltageMeasurement
 
DANGER
●
Nevermakemeasurementonacircuitinwhich
voltage over AC600V exists to avoid getting
electricalshock.
●
Do not make measurement with the Battery
Coverremoved.
●
Keep your fingers behind the barrier on the
instrumentduringmeasurement.
(1)SettheFunctionSwitchto"
ACV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connect the test leads to the circuit under test.
Take the reading onthedisplay.Pressingthe
"
Hz/DUTY
" keywhilereadingisindicatedon
the display switches the indication in following
sequence.
 
ACVoltage⇨Hz⇨DUTY
 
CAUTION
●
Hz/DUTYFunctionrequiresAC40Vorhigher.
●
To measure a frequency, measure the voltage
ontheelectricalcircuitinadvance.
●
Thenpress theHz/DUTYkey toenter into
frequencymeasurement.
●
Readingsoffrequency mayfluctuate orbe
influencedundernoisyenvironment.
Anode
Cathode
92-1801B07-01
8.Maintenance
●
Cleaning
Use a cloth dipped in water or neutral detergent for
cleaningtheinstrument.
Donotuseabrasivesorsolvents.Otherwise,instrument
getdamaged,deformedordiscolored.

This manual suits for next models

1

Other KYORITSU Multimeter manuals

KYORITSU New SNAP 7 User manual

KYORITSU

KYORITSU New SNAP 7 User manual

KYORITSU KEW 1018 User manual

KYORITSU

KYORITSU KEW 1018 User manual

KYORITSU KM2012RA User manual

KYORITSU

KYORITSU KM2012RA User manual

KYORITSU KEW1011 User manual

KYORITSU

KYORITSU KEW1011 User manual

KYORITSU KEW 1061 User manual

KYORITSU

KYORITSU KEW 1061 User manual

KYORITSU KEW 1018 User manual

KYORITSU

KYORITSU KEW 1018 User manual

KYORITSU 1110 User manual

KYORITSU

KYORITSU 1110 User manual

KYORITSU KEW 1109 User manual

KYORITSU

KYORITSU KEW 1109 User manual

KYORITSU KEW MATE 2000A User manual

KYORITSU

KYORITSU KEW MATE 2000A User manual

KYORITSU KEW MATE 2012RA User manual

KYORITSU

KYORITSU KEW MATE 2012RA User manual

KYORITSU KEW1021R User manual

KYORITSU

KYORITSU KEW1021R User manual

KYORITSU 1009 User manual

KYORITSU

KYORITSU 1009 User manual

KYORITSU KEW 4106 User manual

KYORITSU

KYORITSU KEW 4106 User manual

KYORITSU KEW 1052 User manual

KYORITSU

KYORITSU KEW 1052 User manual

KYORITSU KEW1020R User manual

KYORITSU

KYORITSU KEW1020R User manual

KYORITSU KEW MATE 2012R User manual

KYORITSU

KYORITSU KEW MATE 2012R User manual

Popular Multimeter manuals by other brands

VOLTCRAFT VC135 Safety instructions

VOLTCRAFT

VOLTCRAFT VC135 Safety instructions

Analog Technologies VC480C manual

Analog Technologies

Analog Technologies VC480C manual

Tektronix cdm250 user manual

Tektronix

Tektronix cdm250 user manual

Triplett 9007 instruction manual

Triplett

Triplett 9007 instruction manual

LTR MT 07 manual

LTR

LTR MT 07 manual

NEO 94-002 instruction manual

NEO

NEO 94-002 instruction manual

HoldPeak 770G Operator's manual

HoldPeak

HoldPeak 770G Operator's manual

Simpson 260 Series 5P Operator's manual

Simpson

Simpson 260 Series 5P Operator's manual

Kaise 7050 Operator's manual

Kaise

Kaise 7050 Operator's manual

Solight V15 manual

Solight

Solight V15 manual

PCE Instruments PCE-IT 100 user manual

PCE Instruments

PCE Instruments PCE-IT 100 user manual

Tektronix THM500 Series instruction manual

Tektronix

Tektronix THM500 Series instruction manual

HP 3457A Service manual

HP

HP 3457A Service manual

FLIR Extech EX510 user guide

FLIR

FLIR Extech EX510 user guide

VOLTCRAFT VC151 operating instructions

VOLTCRAFT

VOLTCRAFT VC151 operating instructions

Truper MUT-202 manual

Truper

Truper MUT-202 manual

Fnirsi FNIRSI-S1 instruction manual

Fnirsi

Fnirsi FNIRSI-S1 instruction manual

LOVATO ELECTRIC DMG800RGCA installation manual

LOVATO ELECTRIC

LOVATO ELECTRIC DMG800RGCA installation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.