KYORITSU KEW 1061 User manual

60
Thank you for purchasing our Model KEW1061, KEW1062 Digital Multimeter.
This user’s manual describes the specifications and handling precaution for this Digital
Multimeter.
Before using this Digital Multimeter, thoroughly read this manual to get a clear understanding
on proper use.
Always observe the following instructions.
Failure to do so may impair the protection provided by the instrument and probes, and may
result in electrical shock or other dangers that may lead to serious injury or the loss of life.
KYORITSU is in no way liable for any damage resulting from the user’s mishandling of the
product.
For safe use of this product, the following safety symbols are used on the product:
nAbout This Manual
• Everyefforthasbeenmadetoensureaccuracyinthepreparationofthismanual.
However, should any errors or omissions come to the attention of the user, please contact
KYORITSU.
• Thecontentsofthismanualaresubjecttochangewithoutpriornoticebecauseof
improvement in performance or function.
• Allrightsreserved.Nopartofthismanualmaybereproducedinanyfromwithout
KYORITSU'S written permission.

61
Regarding Safe Use of This Product
For safe use of this product, the following safety symbols are used on the product and manual:
WARNING
This indicates that the operator must refer to an explanation in the
instruction manual in order to avoid the risk of serious injury or the loss
of life.
CAUTION
This indicates that the operator must refer to an explanation in the
instruction manual in order to avoid the risk of injury or damage to the
product.
Note This indicates information that is essential for handling the instrument
or should be noted in order to familiarize yourself with the instrument’s
operating procedures and/or functions.
Danger! Handle with Care
This symbol indicates that the operator must refer to an explanation in the
instruction manual in order to avoid risk of injury or death of personnel
or damage to the instrument.
Double Insulation
This symbol indicates double insulation or reinforced insulation.
Direct Current
This symbol indicates DC voltage/current.
Alternating Current
This symbol indicates AC voltage/current.
DC/AC
This symbol indicates AC and DC.
Fuse
This symbol indicates a fuse.
Battery
This symbol indicates a battery.
Ground
This symbol indicates ground (earth).

62
WARNING
nAlways observe the following instructions. Failure to do so may result in
electrical shock or other dangers that may lead to serious injury or the loss of
life.
Test leads / Test leads with crocodile clip (optional accessory)
• UsetheprobessuppliedbyKYORITSUwiththisinstrument.
• Donotusetestleads/testleadswithcrocodileclipthathavedeterioratedoraredefective.
Check test leads/test leads with crocodile clip continuity.
• Disconnecttestleads/testleadswithcrocodileclipfromthecircuitundertestbeforeopening
the casing to replace the batteries or for any other reason.
• Disconnecttestleads/testleadswithcrocodileclipfromthecircuitundertestbefore
attaching/detaching the test leads/test leads with crocodile clip to/from the instrument.
• Disconnecttestleads/testleads withcrocodileclipfrom theinstrumentbeforeopening the
casing to replace the batteries or for any other reason.
• Acapisprovidedonthetipofatest lead.
Use a test lead with the cap on for safety (safety standards: IEC 61010-031).
•Do not use the crocodile clip of test leads being loosen or removed conditions.
Casing
• Donotusetheinstrumentifthereisanydamagetothecasingorwhenthecasingis
removed.
Fuses
• Usefusesofthespeciedratingwhenthefuseisreplaced.
Operating Environment
• Do not operate the instrument in an atmosphere where any ammable or explosive gas is
present.
• Avoidusing theinstrument ifithasbeenexposed torain ormoisture orif yourhandsare
wet.
Disassembly
• Noperson,exceptpersonnelfromKYORITSU,isauthorizedtodisassemblethisinstrument.

63
Contents
1. Overview........................................................................................................... 65
2. Measurement Category ..................................................................................... 66
3. Specications .................................................................................................... 67
3.1 GeneralSpecications ............................................................................ 67
3.2 Accuracy ................................................................................................. 70
4. Operation .......................................................................................................... 78
4.1 Precautions Before Measurement ........................................................... 78
4.2 Components ............................................................................................ 79
4.3 Measuring Instructions............................................................................ 84
4.3.1 AC Voltage Measurement ................................................................ 84
4.3.2 DC Voltage Measurement ................................................................ 84
4.3.3 DC+AC Voltage Measurement ........................................................ 84
4.3.4 DC, AC Voltage Dual Display ......................................................... 85
4.3.5 Resistance Measurement.................................................................. 85
4.3.6 LowPower-Ω................................................................................... 86
4.3.7 Continuity Check ............................................................................. 86
4.3.8 Diode Test ........................................................................................ 87
4.3.9 Temperature Measurement............................................................... 88
4.3.10 Current Measurement ..................................................................... 89
4.3.11 DC+AC Current Measurement....................................................... 90
4.3.12 DC, AC Current Dual Display........................................................ 91
4.3.13 Capacitor Measurement.................................................................. 91
4.3.14 Frequency Measurement, Duty cycle ratio..................................... 92
4.3.15
FunctiontochangeRMSdetectionto/fromMEANdetection
mode ........................................................................................................93
4.3.16Functiontoturnthelteron/off ..................................................... 93
4.3.17 AUTO HOLD Function.................................................................. 94
4.3.18 PEAK HOLD Function .................................................................. 95
4.3.19 Relative and percentage calculation ............................................... 95
4.3.20 Decibel calculation ......................................................................... 96
4.3.21 MIN/MAX/AVGFunction ............................................................. 97
4.4 Memory Function.................................................................................... 98
4.5 AUTO POWER OFF Function............................................................. 101
4.6 Set-up Function..................................................................................... 102
4.7 AdditionalfunctionssimplysetwhenPOWERON............................. 105
4.8 Averaging Function............................................................................... 105
4.9 5000 display mode ................................................................................ 106
4.10 LCD Check ......................................................................................... 106

64
5. User Calibration Function............................................................................... 107
6. Battery and Fuse Replacement ....................................................................... 112
6.1 Battery Replacement............................................................................. 112
6.2 Fuse Replacement ................................................................................. 113
7. Calibration and Maintenance .......................................................................... 114
8 Disposing the Product..................................................................................... 114

65
1. Overview
lFast and more accurate measurement
The KEW1061 and KEW1062 digital multimeters use Σmodulation for A/D conversion,
which enable fast and more accurate measurement.
lDisplay
5-digit (LCD)
Maximum Reading: 50000
Bar graph indicator
lSupports a variety of measurement function
Measurement function
DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature,
Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test,
LowPower-Ω*
Other functions
DataHold(D•H),AutoHold(A•H),PeakHold*(P•H),RangeHold(R•H),
Maximum value (MAX), Minimum value (MIN), Average value (AVG),
Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory,
LCD backlight.
LowPower-Ω:Measuresresistanceunderlowmeasurementcurrent.
*:FormodelKEW1062only
lSwitching detection modes
Effective value (root mean square value) detection (RMS) and mean value detection (MEAN)
can be switched during AC voltage or AC current measurement (KEW1062 only).
l Low-passlter
Thelow-pass lter canbe switched on/offduringAC voltage orACcurrent measurement
(KEW1062 only).
lCommunication: optional communication package is required
• MeasurementdatacanbetransferredtoaPCusinganoptionalUSBcommunicationset.
The data can be read by certain applications to make trend graphs or can be converted
intoExcelles.
• Thedatacanalsobeoutputfromanoptionalprinterviaanoptionalprintercommunication
set.
lSafety design
Complied standards: CE standards
Usesacurrent-inputterminalshutterforpreventingwronginput.
Useshigh-performanceUL-standardfuses.

66
2. Measurement Category
WARNING
nMeasurement Category (CAT.)
The restrictions on the maximum voltage level for which the KEW1061, KEW1062 can
beused,dependonthemeasurementcategoriesspeciedbythesafetystandards.
Do not apply any input level higher than maximum allowable input.
AC/DC 1000V CAT. III
AC/DC 600V CAT. Ⅳ
Measurement
category Description Remarks
ⅠCAT. I For measurements performed on circuits not
directly connected to MAINS.
ⅡCAT. II For measurements performed on circuits directly
connected to the low voltage installation.
Appliances, portable
equipment, ect.
ⅢCAT.III For measurements performed in the building
installation.
Distribution board,
circuit breaker, ect.
ⅣCAT. ⅣFor measurements performed all the source of the
low-voltage installation.
Overhead wire,
cable systems, ect.
Internal Wiring
Outlet
Entrance
Cable
Fixed Equipment,
etc.
Distribution
Board
CAT. II
Equipment
CAT.I
CAT.IV
CAT. III
Note
Radiation immunity affects the accuracy of the KEW1061, KEW1062 under the
conditionsspeciedinIEC61326-1.
Useofthisinstrumentislimitedtodomestic,commercial,andlightindustry
applications. If equipment generating strong electromagnetic interference is located
nearly, the instrument may malfunction.

67
3. Specications
3.1 GeneralSpecications
Measurement function:
DC Voltage, AC Voltage, DC Current, AC Current, Resistance,
Frequency, Temperature, Capacitor, Duty cycle ratio,
Decibel(dBV,dBm),ContinuityCheck,DiodeTest,LowPower-Ω*
Otherfunctions: DataHold(D•H),AutoHold(A•H),PeakHold*(P•H),
RangeHold(R•H),Maximumvalue(MAX),Minimumvalue(MIN),
Average value (AVG), Zero Adjustment (Capacitor, Resistance),
Relative values, Save to Memory, LCD backlight.
LowPower-Ω:Measuresresistanceunderlowmeasurementcurrent.
*:FormodelKEW1062only
Measuring method: Σmodulation
Display: 5-digit (LCD)/7-segment
Maximum Reading: 50000
Polarity Indicator: “–” Appears automatically when the polarity is
negative
Overrange Indicator: “ OL ”
Low-battery Indicator: “ ” Appears when the batteries become
low
Measurement cycle: 6 times per second
(except frequency measurement: one times per second, Resistance
measurement : four times per second, capacitor measurement (50mF):
max.0.03timepersecond)
Bar graph display 15 times per second
Operating temperature and humidity ranges:
-20 to 55ºC, 80%RH or less (no condensation)
70%RH or less at 40 to 55ºC.
Storage temperature and humidity ranges:
-40 to 70ºC, 70%RH or less (no condensation)
Temperaturecoefcient:
(Accuracyat23±5ºC)×0.05/ºCorless
At -20 to 18ºC and 28 to 55ºC
When continuously DCV and DCA measuring, add 1digit / ºC
(exceptfor50mV,5A,10Arange,add3digits/ºC).
Power supply: AA-size (R6) 1.5V batteries: 4

68
Battery life: Approximately 120 hours
(Operating hours of alkaline batteries when in DC voltage-mode.)
Note: The battery life varies depending on the operating conditions.
Insulationresistance:1000VDC,100MΩormore
Withstandvoltage: 6.88kVrmsACforveseconds
(across input terminals and casing)
Externaldimensions: Approximately192(L)×90(W)×49(D)mm
Weight: Approximately 560g (including batteries)
Complied standards: Safety standards
IEC61010-1,IEC61010-031
CAT.III (Max. input voltage: AC/DC1000V)
CAT.IV (Max. input voltage: AC/DC600V)
Pollution degree 2, indoor use,
2000m max. above sea level
EMC standards
IEC61326-1ClassB
Effect of radiation immunity:
Intheradio-frequencyelectromagneticeldof3V/m,accuracyis
withinvetimestheratedaccuracy.

69
Standard accessories: Batteries : 4
Test leads: 1set (M-7220A)
Fuse (included): 440mA/1000V (M-8926), 10A/1000V (M-8927)
Instruction manual: 1
Optional accessories: Carrying case M-9150
(for the main unit with test leads
and communication cable)
Test leads (1set) M-7220A
Test leads with crocodile clip (1set)
M-7234
Fuse
440mA/1000V M-8926
10A/1000V M-8927
Temperature probes M-8405, 8406, 8407, 8408
USBCommunicationset M-8241(Software,USBadapterand
cable)
PrinterAdapterandCable M-8243
Printer M-8246
AC adapter (for printer, Europe)
M-8248
Thermal paper for printer (10 rolls)
M-8247

70
3.2 Accuracy
Test conditions:
Temperatureandhumidity:23±5ºCat80%RHorless
Accuracy:±(%ofreading+digits)
Note: Each response time is a value to rated accuracy within selected range.
DC Voltage Measurement V
Range Resolution Accuracy Input Resistance Maximum
Input Voltage
KEW1061, KEW1062
50mV 0.001mV 0.05+10
Approx.100MΩ 1000V DC500mV 0.01mV 0.02+2
2400mV 0.1mV
5V 0.0001V 0.025+5
10MΩ 1000V rms AC
50V 0.001V
0.03+2500V 0.01V
1000V 0.1V
NMRR: 80dBormore50/60Hz±0.1%
(70dBormore50/60Hz±0.1%when50mVRange)
CMRR: 100dBormore50/60Hz(Rs=1kΩ)
Responsetime:0.3secmax.
AC Voltage Measurement [RMS] V
KEW1061
ACCoupling,Rms-valuedetection,Crestfactor*:<3
Range Resolution
Accuracy Input
Impe-
dance
Maximum
Input
Voltage
10 to 20Hz 20Hz to
1kHz
1k to
10kHz
10k to
20kHz
500mV 0.01mV
1.5+30*1 0.7+30*1 2+50*2
11MΩ
<50pF 1000V
rms AC
1000V DC
5V 0.0001V
50V 0.001V 10MΩ
<50pF
500V 0.01V
1000V* 0.1V *2 *2 3+30*2 -
*: Crestfactor<1.5at1000Vrange
Accuracy*1:At5to100%ofrange,*2:At10to100%ofrange
CMRR:80dBormoreDCto60Hz(Rs=1kΩ)
Response time: 1 sec max.

71
AC Voltage Measurement [RMS] V
KEW1062
ACCoupling,Rms-valuedetection,Crestfactor*:<3
Range Reso-
lution
Accuracy
Input
Impedance
Maximum
Input
Voltage
10 to
20Hz
20Hz to
1kHz
1k to
10kHz
10k to
20kHz
20k to
50kHz
50k to
100kHz
50mV
0.001mV
2+80
*2
0.4+40
*2
5+40
*2
5.5+40
*2
15+40
*2 11MΩ
<50pF 1000V
rmsAC
1000V
DC
500mV 0.01mV
1+30
*1
0.4+30
*1
1+40
*1
2+70
*2
5+200
*2
5V 0.0001V
50V 0.001V
10MΩ
<50pF
500V 0.01V
1000V* 0.1V *2 *2 3+30
*2 -
*: Crestfactor<1.5at1000Vrange
Accuracy*1:At5to100%ofrange,*2:At10to100%ofrange
CMRR:80dBormoreDCto60Hz(Rs=1kΩ)
Response time: 1 sec max.
AC Voltage Measurement [MEAN] V
KEW1062
AC Coupling, MEAN value detection, RMS value calibration (sine wave)
Range Reso-
lution
Accuracy
Input
Impedance
Maximum
Input
Voltage
10 to 20 Hz 20 to 500 Hz 500Hz to 1 kHz
50mV
0.001mV
4+80
*2
1.5+30
*2
5+30
*2 11MΩ
<50pF
1000V
rmsAC
1000V
DC
500mV 0.01mV
2+30
*1
1+30
*1
3+30
*1
5V 0.0001V
50V 0.001V 10MΩ
<50pF
500V 0.01V
1000V 0.1V *2 *2 *2
Accuracy*1:At5to100%ofrange,*2:At10to100%ofrange
CMRR:80dBormoreDCto60Hz(Rs=1kΩ)
Response time: 1 sec max.

72
DCV+ACV +
KEW1061
MaximumReading50000,Crestfactor*:<3
Range Resolution
Accuracy Input
Impe-
dance
Maximum
Input
Voltage
DC, 10 to
20Hz
DC, 20Hz
to 1kHz
DC, 1k to
10kHz
DC, 10k
to 20kHz
5V 0.0001V 1.5+10
*1
1+10
*1
2+10
*2
11MΩ
<50pF 1000V
rms AC
1000V DC
50V 0.001V 10MΩ
<50pF
500V 0.01V
1000V* 0.1V *2 *2 -
DCV+ACV +
KEW1062
MaximumReading50000,Crestfactor*:<3
Range Reso-
lution
Accuracy Input
Impe-
dance
Maxi-
mum
Input
Voltage
DC, 10
to 20Hz
DC,
20Hz to
1kHz
DC,
1k to
10kHz
DC,
10k to
20kHz
DC,
20k to
50kHz
DC,
50k to
100kHz
5V 0.0001V 1.5+10
*1
0.5+10
*1
1+10
*1
2+10
*2
5+20
*2
11MΩ
<50pF
1000V
rms AC
1000V
DC
50V 0.001V 10MΩ
<50pF
500V 0.01V
1000V* 0.1V *2 *2 -
*: Crestfactor<1.5at1000Vrange
Accuracy*1:At5to100%ofrange,*2:At10to100%ofrange
CMRR:80dBormoreDCto60Hz(Rs=1kΩ)
Response time: Approx. 2 sec
DC Current Measurement A
Range Resolution Accuracy Voltage Drop Maximum Input
Current
KEW1061, KEW1062
500μA 0.01μA
0.2+5
<0.11mV/μA 440mA Protected by a
440mA/1000V fuse.
5000μA 0.1μA
50mA 0.001mA <4mV/mA
500mA*3 0.01mA
5A 0.0001A 0.6+10 <0.1V/A 10A Protected by a
10A/1000V fuse.
10A 0.001A 0.6+5
*3:Maximummeasurementcurrent:440mAat500mArange
Responsetime:0.3secmax.
Note : After measuring over 500mA DC (especially 10A DC), “Zero error” occurs for a
while. In that case, please wait for a while at zero input until the value stabilizes before
measuring again.

73
AC Current Measurement [RMS] A
KEW1061
Rms-valuedetection,Crestfactor:<3
Range Resolution Accuracy Voltage Drop Maximum Input
Current
10 to 20Hz 20Hz to 1kHz
500μA 0.01μA
1.5+20 1+20
<0.11mV/μA 440mA Protected
by a 440mA/
1000V fuse.
5000μA 0.1μA
50mA 0.001mA <4mV/mA
500mA*3 0.01mA
5A 0.0001A <0.1V/A 10A Protected by a
10A/1000V fuse.
10A 0.001A
AC Current Measurement [RMS] A
KEW1062
Rms-value detection, Crestfactor:<3
Range Resolution Accuracy Voltage
Drop
Maximum Input
Current
10 to 20Hz
20Hz to 1kHz
1k to 5kHz
500μA 0.01μA
1+20 0.75+20 1+30
<0.11
mV/μA 440mA Protected
by a 440mA/
1000V fuse.
5000μA 0.1μA
50mA 0.001mA <4mV/mA
500mA*3 0.01mA
5A 0.0001A 1.5+20 1+20 2+30 <0.1V/A
10A Protected by
a 10A/1000V fuse.
10A 0.001A
Model KEW1061/1062
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3:Maximummeasurementcurrent:440mAat500mArange
Response time: 1 sec max.

74
AC Current Measurement [MEAN] A
KEW1062
MEAN value detection, RMS value calibration (sine wave)
Range Resolution
Accuracy Voltage
Drop
Maximum Input
Current
10 to 20Hz 20 to 500Hz 500Hz to
1kHz
500μA 0.01μA
2+20 1.5+20 2+30
<0.11
mV/μA 440mA Protected
by a 440mA/
1000V fuse.
5000μA 0.1μA
50mA 0.001mA <4mV/mA
500mA*3 0.01mA
5A 0.0001A
3+20 2+20 4+30 <0.1V/A
10A Protected
by a 10A/
1000V fuse.
10A 0.001A
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3:Maximummeasurementcurrent:440mAat500mArange
Response time: 1 sec max.
DCA+ACA +
KEW1061
MaximumReading50000,Crestfactor:<3
Range Resolution
Accuracy
Voltage Drop Maximum Input
Current
DC, 10 to
20Hz
DC, 20Hz to
1kHz
500μA 0.01μA
2+10 1.5+10
<0.11mV/μA 440mA Protected by a
440mA/1000V fuse.
5000μA 0.1μA
50mA 0.001mA <4mV/mA
500mA*3 0.01mA
5A 0.0001A <0.1V/A 10A Protected by a
10A/1000V fuse.
10A 0.001A
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3:Maximummeasurementcurrent:440mAat500mArange
Response time: 2 sec max.

75
DCA+ACA +
KEW1062
MaximumReading50000,Crestfactor:<3
Range Resolution
Accuracy
Voltage Drop Maximum Input
Current
DC, 10 to
20Hz
DC, 20Hz
to 1kHz
DC, 1k to
5kHz
500μA 0.01μA
1.5+10 1+10 1.5+10
<0.11mV/μA 440mA Protected
by a 440mA/
1000V fuse.
5000μA 0.1μA
50mA 0.001mA <4mV/mA
500mA*3 0.01mA
5A 0.0001A 2+10 1.5+10 3+10 <0.1V/A 10A Protected by a
10A/1000V fuse.
10A 0.001A
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3:Maximummeasurementcurrent:440mAat500mArange
Response time: Approx. 2 sec
ResistanceMeasurementΩ
Range Resolution Accuracy Maximum
Measuring
Current
Open-loop
Voltage
Input Protective
Voltage
KEW1061 KEW1062
500Ω 0.01Ω
0.1+2
*1
0.05+2
*1
<1mA
<2.5V 1000V rms
5kΩ 0.0001kΩ <0.25mA
50kΩ 0.001kΩ <25μA
500kΩ 0.01kΩ <2.5μA
5MΩ 0.0001MΩ 0.5+2 <1.5μA
50MΩ 0.001MΩ 1+2 <0.13μA
*1:Accuracyisspeciedafterzeroadjustment(resistance).
Responsetime:1secmax.at500Ωto500kΩ
5secmax.at5MΩto50MΩ
LowPower-ΩLP-Ω
Maximum Reading 5000
Range Resolution Accuracy Maximum
Measuring
Current
Open-loop
Voltage
Input Protective
Voltage
KEW1062 only
5kΩ 0.001kΩ
0.2+3
<10μA
<0.7V 1000V rms
50kΩ 0.01kΩ <1.0μA
500kΩ 0.1kΩ <0.6μA
5MΩ 0.001MΩ 1+3 <0.05μA
LowPower-Ω: Measures resistance under lowmeasurement current.

76
Continuity Check
Maximum Reading 5000
Range Resolution Range of Operation Measuring
Current
Open-loop
Voltage
Input Protective
Voltage
500Ω 0.1Ω
The buzzer turns on
for resistances lower
than100±50Ω.
Approx.
0.5mA <5V 1000V rms
Diode Test
Range Resolution Accuracy Measuring Current
(Vf=0.6V)
Open-loop
Voltage
Input Protective
Voltage
2.4V 0.0001V 1+2 Approx. 0.5mA <5V 1000V rms
Temperature Measurement TEMP
Range Resolution Accuracy Input Protective
Voltage
-200to1372°C 0.1°C 1%+1.5°C 1000V rms
-328to2501.6°F 0.1°F 1%+2.7°F
UseoptionalTemperatureProbe:ThermocoupleTypeK
Capacitor Measurement
Maximum Reading 5000
Range Resolution Accuracy Input Protective Voltage
5nF 0.001nF
1+5
*1
1000V rms
50nF 0.01nF
500nF 0.1nF
5μF 0.001μF
50μF 0.01μF
500μF 0.1μF 2+5
5mF 0.001mF 3+5
50mF 0.01mF
*1:Accuracyisspeciedafterzeroadjustment(capacitor).

77
Frequency Measurement Hz
AC Coupling, Maximum Reading 9999
Range(AUTO) Resolution Accuracy
2.000 to 9.999Hz 0.001Hz
0.02+1
*1
9.00 to 99.99Hz 0.01Hz
90.0 to 999.9Hz 0.1Hz
0.900 to 9.999kHz 0.001kHz
9.00 to 99.99kHz 0.01kHz *2
Accuracy
*1:At10to100%ofinputvoltageorcurrentrange
*2:At40to100%ofinputvoltageorcurrentrange
Duty cycle ratio %
Range Resolution Accuracy
10 to 90% 1% ±1%*1
Accuracy
*1:At10.00Hzto500.0Hz,squarewave
At 40 to 100% of input voltage or current range
PeakHoldP•H
Model KEW1062 only
Maximum Reading 5000
Range Accuracy Response Time Maximum
DCV, DCA ±100digit >250μS

78
4. Operation
4.1 Precautions Before Measurement
n
Examining Items Contained in the Package
After opening the package, be sure to examine the product as instructed below before use.
Should the delivered product be the wrong model, lack any item, or show any aw in its
appearance, contact the vendor from which you purchased the product.
n
Precautions of Operation and Storage
CAUTION
• Insertthebatteriesintheinstrumentbyreferringto“6.1BatteryReplacement.”
• ABlankcoverisprovidedontheupperpartofbackcasing.
Don’tremovetheBlankcoverexceptwhentheUSBadapterorPrinteradapteris
connected.
• Donotusetheinstrumentnearnoise-emittingequipmentor
where there may be a sudden change of temperature. Otherwise, the instrument may
give an unstable reading or errors.
Removal of Dirt
Do not wipe the instrument using any solvent (chemicals) such as benzine or paint
thinner, as this may damage or discolor the front panel.
Useadryclothtocleantheinstrument.
Storage Conditions
• Donot leavetheinstrument exposedtodirect sunlightor inahot andhumid location
such as the inside of a vehicle, for any prolonged
length of time.
• Iftheinstrumentwillnotbeusedforaprolongedperiod,removethebatteries.
This manual suits for next models
1
Table of contents
Other KYORITSU Multimeter manuals

KYORITSU
KYORITSU KEW1011 User manual

KYORITSU
KYORITSU KEW MATE 2012R User manual

KYORITSU
KYORITSU New SNAP 7 User manual

KYORITSU
KYORITSU KEW MATE 2000A User manual

KYORITSU
KYORITSU KEW 4106 User manual

KYORITSU
KYORITSU KEW1020R User manual

KYORITSU
KYORITSU KEW SNAP Series KEW2040 User manual

KYORITSU
KYORITSU KEW 1018 User manual

KYORITSU
KYORITSU KEW1021R User manual

KYORITSU
KYORITSU 1009 User manual