Chapter 2: Prerequisites
Product Service 1-800-522-7658
Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a blower system. Refer to Appendix G: VAV Open Area
Guidelines to properly assess the fume hood area opening, allowing the variable
air volume (VAV) mechanical system to be programmed at various sash
positions.
Electrical Requirements
The Protector Hood models feature internal wiring for the LED light assembly
and light switch. All internal wiring is terminated at the single point wiring
junction box for hook-up by a qualified electrician. The blower switch, and light
switch wires are also terminated at the single point wiring junction box for hook-
up by a qualified electrician. Refer to Chapter 3: Getting Started and Appendix
C: Protector Premier Specifications for the wiring diagram for proper electrical
installation.
Projected Airflows and Static Pressure
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at 62.5% Static Pressure (inches of water) Static Pressure (inches of water)
Full open open 4' 5' 6' 8' 4' 5' 6' 8'
at 28" at 18" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 160 725, 0.22" 955, 0.31" 1180, 0.41" 1640, 0.28" 760, 0.23" 1000, 0.33" 1235, 0.43" 1715, 0.30"
80 128 580, 0.14" 765, 0.20" 945, 0.26" 1310, 0.18" 605, 0.15" 800, 0.21" 990, 0.28" 1370, 0.19"
60 96 435, 0.08" 575, 0.11" 710, 0.15" 985, 0.10" 455, 0.08" 600, 0.12" 745, 0.16" 1030, 0.11"
50 80 365, 0.06" 480, 0.08" 590, 0.10" 820, 0.07" 380, 0.06" 500, 0.08" 620, 0.11" 860, 0.08"
40 64 290, 0.04" 380, 0.05" 470, 0.07" 655, 0.05" 305, 0.04" 400, 0.05" 495, 0.07" 685, 0.05"
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at Static Pressure (inches of water) Static Pressure (inches of water)
62.50% Full open 4' 5' 6' 8' 4' 5' 6' 8'
open at 18" at 28" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 62 450, 0.09" 595, 0.12" 735, 0.16" 1025, 0.11" 485, 0.09" 640, 0.13" 790, 0.17" 1100, 0.12"
80 50 365, 0.06" 480, 0.08" 590, 0.10" 820, 0.07" 390, 0.06" 515, 0.08" 635, 0.11" 880, 0.08"
60 38 270, 0.03" 360, 0.04" 440, 0.06" 615, 0.04" 290, 0.03" 385, 0.05" 475, 0.06" 660, 0.04"
Projected Airflows and Static Pressure
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at 62.5% Static Pressure (inches of water) Static Pressure (inches of water)
Full open open 4' 5' 6' 8' 4' 5' 6' 8'
at 28" at 18" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 160 705, 0.21" 930, 0.29" 1150, 0.39" 1600, 0.27" 740, 0.22" 975, 0.31" 1205, 0.42" 1675, 0.29"
W/Bypass 80 128 565, 0.13" 745, 0.19" 920, 0.25" 1280, 0.17" 590, 0.14" 780, 0.20" 965, 0.27" 1340, 0.19"
Block 60 96 425, 0.08" 560, 0.10" 690, 0.14" 960, 0.10" 445, 0.08" 585, 0.11" 725, 0.15" 1005, 0.10"
50 80 350, 0.05" 465, 0.07" 575, 0.10" 800, 0.07" 370, 0.06" 490, 0.08" 605, 0.11" 840, 0.07"
40 64 280, 0.03" 370, 0.05" 460, 0.06" 640, 0.04" 295, 0.04" 390, 0.05" 480, 0.07" 670, 0.05"
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at Static Pressure (inches of water) Static Pressure (inches of water)
62.50% Full open 4' 5' 6' 8' 4' 5' 6' 8'
open at 18" at 28" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 62 440, 0.08" 580, 0.11" 720, 0.15" 1000, 0.10" 475, 0.09" 625, 0.12" 775, 0.16" 1075, 0.11"
W/Bypass 80 50 350, 0.05" 465, 0.07" 575, 0.10" 800, 0.07" 380, 0.06" 500, 0.08" 620, 0.11" 860, 0.07"
Block 60 38 265, 0.03" 350, 0.04" 430, 0.06" 600, 0.04" 285, 0.03" 375, 0.05" 465, 0.06" 645, 0.04"