
EN FR NL DE SL SV
ABPM-100 EN Instruction manual 9/74
Technical alarm – Description
If the (systolic or diastolic) blood pressure taken is
outside the range indicated in the “Technical Fea-
tures” paragraph, the technical alarm displays the
message “Hi” or “Lo” on the screen. In this case,
consultadoctor or check that youare usingthe device
properly.The technical alarm’slimit valuesarevalues
set by the factory and can neither be changed nor
deactivated. Under the IEC 60601-1-8, these limit
values of the alarm have a lower priority. The tech-
nical alarmis notalocked alarm anddoesnotneedto
be reset. Thesignal displayed on the screen switches
o automatically after approximately 8 seconds.
MAINTENANCE AND STORAGE
•
Clean thedevice and the cucarefully,using only
a slightly damp cloth.
• Do not use detergent or solvent.
• Never put either the device or the cu in water
as this might seep into the device or cu and
damage them.
•
Do not put heavy objects on the device or on
the cu when they are being stored. Remove
the batteries.
• No component can be maintained by user inthe
monitor. The circuit diagrams, component part
lists, descriptions, calibration instructions, or
other information which will assist the user’s
appropriately qualied technical personnel to
repair those parts of equipment which are des-
ignated repairably can be supplied.
• It is recommended the performance should be
checked every 2 years or after repair. Please
contact the service center.
•
The monitorcan maintain thesafety and perfor-
mance characteristics for a minimum of 10,000
measurements or three years.
•
The cu integrity is maintained after 1,000
open–closecycles of the closure.
• It is recommended the cu should be disinfect-
ed 2 times every week if needed (For example,
in hospital or in clinique). Wipe the inner side
(the side contacts skin) of the cu by a soft cloth
squeezed after moistened with Ethyl alcohol
(75-90%), then dry the cu by airing.
•
Clean thecuafterthe usage of every 200 times
is recommended.
TECHNICAL SPECIFICATIONS
OF THE PRODUCT
Model number LA090206
Measuring
method Measuring blood pressure on the arm,
oscillometric and non-invasive
Measuring
range Pressure in the cu 0 – 300 mmHg
Systolic pressure 60 – 260 mmHg
Diastolic pressure 40 – 199 mmHg
Pulse 40 – 180 beats/minute
Accurac y of
the indicator systolic ±3 mmHg
diastolic ±3 mmHg
pulse ±5% of the value displayed
Memory 4 x 30 memory locations
Dimensions +/- 107 x 80 x 52mm
Weight +/- 287g (without any batteries)
Cu size 22 - 30cm
Admin. operat-
ing conditions + 10°C a + 40°C, ≤85% relative air
humidity (non-condensing)
Permissible
storage
conditions
-20°C a +50°C, ≤85% relative air humidity,
800 – 1 050 hPa of ambient pressure
Power supply 4 x 1.5 V AAA batteries
Battery lifespan Approximately 200 measurements,
depending on the increased blood
pressure and ination pressure
Accessories Instructions for use, storage box
Ranking Internal power supply, IP20, no AP
or APG, continuous use, BF device
Pressure meas-
uring system
components
Pump,Valve, LCD, Cu, Sensor
Changes could be made to the technical features
without any prior warning for updating purposes.
•
This device complies with European standard
EN60601-1-2 and meets the special safety re-
quirementsconcerningelectromagneticcompat-
ibility. Please note that HF portable and mobile
communication devices are likely to aect this
device. For further details, please contact the
after-salesservice at the address stated or consult
the end of the instructions for use.