Installing the 6×6 cm chamber
1Hold the chamber in position, with the latch springs up
toward the handle, and tighten the two captive cham-
ber screws (146-15344).
Caution: Use the captive chamber screws (part number
146-15344) to install the chamber. Other screws will
cause leaks and problems with the mixing fan.
Make them snug—turn each screw until it stops, and
then just a little bit more (up to ¼ turn). If you have a
torque screwdriver, tighten them to 2.5 N·m (22 inch
lbs).
2Install the latch screws.
With the latch open, tighten until they are snug. If you
have a torque screwdriver, tighten them to 1.1 N·m (9.4
inch lbs).
3If you unplugged the light sensor cable, install the
light sensor connector to the connection labeled PAR.
4Install the leaf temperature thermocouple connectors
to the connections labeled T1 and T2.
5Adjust the thermocouple height.
6Install the connector cover.
After swapping chambers, it is a good idea to verify that the
chamber is sealed and that the software recognizes it. Run
the leak test under Start up > System Tests > Chamber Leak.
Check to see that the chamber is recognized under Constants
> System Constants > ChType.
Copyright © 2020 LI-COR, Inc. All rights reserved.
984-16473•01/2020
LI-COR Biosciences
4647 Superior Street
Lincoln, Nebraska 68504
Phone: +1-402-467-3576
Toll free: 800-447-3576 (U.S. and Canada)
envsales@licor.com
LI-COR Distributor Network:
www.licor.com/env/distributors
Regional Offices
LI-COR Biosciences GmbH
Siemensstraße 25A
61352 Bad Homburg
Germany
Phone: +49 (0) 6172 17 17 771
envsales-gmbh@licor.com
LI-COR Biosciences UK Ltd.
St. John’s Innovation Centre
Cowley Road
Cambridge
CB4 0WS
United Kingdom
Phone: +44 (0) 1223 422102
envsales-UK@licor.com